Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K990159
    Date Cleared
    1999-04-19

    (90 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K914790, K902481, K973291

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DYNAMIC Hollow Fiber Dialyzer DH-Series are indicated for use whenever a patient is in acute or chronic renal failure and hemodialysis is prescribed by a physician. Therefore, use of this device should be only on the direction of a physician who has evaluated all of the aspects of the patient's illness.

    Device Description

    Blood enters a blood inlet port where it is distributed to hollow fibers. Each hollow fiber has an inner diameter of 200 microns and a wall thickness of 8 microns. The fibers used in this device which is substantially equivalent to the fibers utilized in the Focus R-Series Hollow Fiber Dialyzers (K914790) and Gambro GFS plus Series Hemodialyzer (K902481), which have been previously approved under a 510(k) Notification. The wall thickness of the hollow fibers in Focus R-Series Hollow Fiber Dialyzers and the proposed device is 8 microns. The inner diameter of hollow fibers in both Focus R-Series Hollow Fiber Dialyzers, Gambro GFS plus Series Hemodialyzer and the proposed device is 200 microns. The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the DYNAMIC Hollow Fiber Dialyzer DH-Series. The primary purpose of a 510(k) is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device. This typically involves showing that the device performs at least as well as the predicate device in relevant performance metrics.

    Based on the provided text, here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as numerical thresholds that must be met. Instead, the approach is to demonstrate substantial equivalence to predicate devices. The "performance" is reported by stating that the device is substantially equivalent to these predicates.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    Urea ClearanceSubstantially equivalent to predicate devices"proposed device is substantially equivalent to Focus R-Series Hollow Fiber Dialyzers 120H-R(K914790) , Gambro GFS plus 20 Hemodialyzer (K902481), and DYNAMIC Hollow Fiber Dialyzer DC-Series (K973291) for in vitro performance."
    Creatinine ClearanceSubstantially equivalent to predicate devicesSame as above.
    Phosphate ClearanceSubstantially equivalent to predicate devicesSame as above.
    Vitamin B12 ClearanceSubstantially equivalent to predicate devicesSame as above.
    Ultrafiltration CoefficientSubstantially equivalent to predicate devicesSame as above.
    Construction, Design, Intended Use, FunctionSubstantially equivalent to predicate devices"substantially equivalent in construction, design, intended use, and function to other hemodialyzers currently marketed in the United States."
    MaterialsSimilar to predicate devices"similarities are noted in the design and materials employed to accomplish the same intended use."
    Sterilization MethodSubstantially equivalent to predicate devices"sterilized by ethylene oxide gas," which is also used by predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "In vitro testing was performed on the proposed device".

    • Sample Size: The document does not specify the sample size for the in vitro testing. It only mentions that testing was performed on "the proposed device."
    • Data Provenance: The data is from in vitro testing. The country of origin for the data is implied to be Taiwan, R.O.C., where the submitter (DYNAMIC TECHNOLOGY CORP.) is located. It is prospective data as it was generated to support this 510(k) submission for the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" for hollow fiber dialyzers in this context refers to the measured performance characteristics (clearances, ultrafiltration) of the device itself and its comparison to predicate devices, not subjective expert assessment of images or clinical outcomes. The performance is measured directly in a lab setting.

    4. Adjudication Method for the Test Set

    This is not applicable. As stated above, the performance metrics are objective laboratory measurements, not subjective evaluations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. MRMC studies are typically used for diagnostic imaging devices where human readers interpret medical images. This document describes a hollow fiber dialyzer, which is a therapeutic device, not an imaging device.

    6. Standalone (Algorithm Only) Performance Study

    This is not applicable. This device is a physical medical device (dialyzer), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. Type of Ground Truth Used

    The "ground truth" for the performance claims are the objective, measurable results of in vitro laboratory tests (urea, creatinine, phosphate, vitamin B12 clearances, and ultrafiltration coefficient). The comparison is then made against the established performance characteristics of the predicate devices.

    8. Sample Size for the Training Set

    This is not applicable. The device is a physical medical device, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1