Search Results
Found 1 results
510(k) Data Aggregation
(11 days)
The 3MTM Attest ™ 293 Auto-reader is designed to incubate and automatically read the SM - Allest - - TEO Parfinal negative fluorescence reading at 1 hour for 1291 or 3 hours for 1292.
The 3M™ Attest ™ 293 Auto-reader is also designed to allow for further The SM - Attest 1291 and/or the Attest 1292 Rapid Readout Biological Indicators for a final negative, visual pH color change of the growth media at 24 hours for 1291 and 48 hours for 1292.
The 3M™ Attest ™ 293G Auto-reader is designed to incubate and automatically The SM - Attest - 2300 Auto Todor 16 Cougical Indicators (RRBI) for EO at 37℃ for a final negative fluorescence reading at 4 hours.
The own Attoot 2007 late reas (BI), which are incubated at 60 °C. The initial Bls in this series to receive market clearance were the 1291 (K900771) and 1292 (K926364).
The purpose of these Bls is to assess whether a potential failure of a steam sterilization cycle has occurred. Please note that both of these initial premarket sterfilizations (K900771) and (K926364) listed the 3M Attest 190 Auto-reader as houncations (13007 17 and (102000 1) 1102000 1) 1100 Auto-reader was cleared (K004009) for use with these Bls and serves as the predicate device for the new Model 293 Auto-reader.
The Attest 293 is comprised of three (3) Attest 290 units within a single case.
The 3M Attest 293G Auto-reader is an accessory to the 3M Attest 1294 Rapid Readout EO Biological Indicator (RRBI) which is to be incubated at 37 °C.
The purpose of these Bls is to assess whether a potential failure of an ethylene oxide sterilization cycle has occurred. The model 290G was cleared (K031012) with the Attest 1294 Bl and serves as the predicate for the 293G.
The Attest 293G is comprised of three (3) Attest 290G units within a single case.
The provided document is a 510(k) Premarket Notification summary for the 3M™ Attest™ 293/293G Auto-reader. It states that the new device is "substantially equivalent" to predicate devices (Attest 290 and 290G Auto-readers) because they use the same technology, software, and have the same indications for use and intended use. The new device is essentially a "three units within one case" configuration of the predicate.
Therefore, the submission does not contain information about a new study with acceptance criteria and device performance evaluation as would be typical for a device with a new technological approach or significantly different intended use. Instead, it relies on the established performance of its predicate devices due to its substantial equivalence in design and function.
As such, many of the requested details about a study for acceptance criteria cannot be extracted from this document, as a new, standalone performance study demonstrating new acceptance criteria was not conducted or reported in this 510(k) summary. The "acceptance criteria" here is implicitly demonstrating substantial equivalence to the predicate.
Here's an attempt to answer based on the provided text, acknowledging the limitations:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria or a performance table for a new study. The acceptance criteria for the 510(k) itself is demonstrating "substantial equivalence" to the predicate devices. The "reported device performance" is implicitly considered to be the same as the predicate devices:
Acceptance Criteria (Implied) | Reported Device Performance (Implied) |
---|---|
Same product technology | Uses the same product technology as predicate. |
Same software | Uses the same "look and feel" software as predicate. |
Same indications for use | Has the same indications for use as predicate. |
Same intended use | Has the same intended use as predicate. |
Configuration as "three units within one case" of predicates | Device is essentially three individual predicate units enclosed in one case. |
2. Sample size used for the test set and the data provenance
Not applicable. No new test set data is provided as the submission relies on substantial equivalence to predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No new test set requiring expert ground truth establishment is reported.
4. Adjudication method for the test set
Not applicable. No new test set requiring adjudication is reported.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an auto-reader for biological indicators, not an AI-assisted diagnostic tool that would involve human readers or MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The device itself is a standalone auto-reader. However, no new standalone performance study (beyond demonstrating it functions as three predicate units) is mentioned. Its performance is considered equivalent to the predicate devices which have established standalone performance.
7. The type of ground truth used
For the predicate devices, the ground truth for evaluating biological indicators (BI) would typically involve microbiological confirmation of sterility (or failure to achieve sterility) in a controlled laboratory setting, which could be considered a form of "pathology" or definitive outcome data regarding the sterilization process. However, this document does not detail how the ground truth for the predicates was established, only that the new device is equivalent.
8. The sample size for the training set
Not applicable. This document does not describe a machine learning algorithm or a new training set. The device's operation is based on established auto-reading technology for biological indicators.
9. How the ground truth for the training set was established
Not applicable. No training set is described in this document.
Ask a specific question about this device
Page 1 of 1