Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
Mfinity femoral stems are indicated in patients requiring hip arthroplasty.
Total or partial hip arthroplasty is indicated in the following cases:
- Severely painful and/or disabled joint as a result of osteoarthritis, post-traumatic arthritis, inflammatory arthritis or hip dysplasia.
- Avascular necrosis of the femoral head.
- Acute fracture of the femoral head.
- Acute fracture of the proximal femur, suitable to be treated by means of hip arthroplasty.
- Non-union of proximal femur fracture, suitable to be treated by means of hip arthroplasty.
- Primary pathology involving the femoral head but with a non-deformed acetabulum.
- Failure of previous hip surgery:
- Conservative hip surgery.
- Internal fixation.
- Arthrodesis.
- Partial or total hip arthroplasty.
- Hip resurfacing replacement.
The Mfinity Femoral System includes implantable devices provided individually packed, sterile and single-use intended for cementless use in total or partial hip arthroplasty to replace the native femoral neck for primary or revision surgery. The product range is composed of three different versions (Mfinity collarless, Mfinity collared and Mfinity L) available in standard and lateral offset.
The Mfinity femoral stem can be combined with the CoCr ball head (K072857, K080885 and K103721), Endo Head (K111145), the MectaCer BIOLOX® forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
The subject devices are made of titanium alloy according to ISO 5832-11 and coated with Titanium plasma spray according to ASTM F1580 and Hydroxyapatite according with ASTM F1185.
The provided text is an FDA 510(k) clearance letter for the Mfinity Femoral System, which is a hip prosthesis, not a diagnostic AI device. Therefore, the information requested regarding acceptance criteria and studies that prove a device meets those criteria, specifically concerning AI performance metrics, expert adjudication, training set, and test set details, is not applicable to this document.
The document discusses the substantial equivalence of the Mfinity Femoral System to predicate devices based on non-clinical performance data (fatigue tests, ROM evaluation, surface analyses, biocompatibility, shelf-life, pyrogenicity) and states that no clinical studies were conducted.
The acceptance criteria for this type of medical device (hip prosthesis) would typically involve demonstrating mechanical strength, durability, biocompatibility, and sterilization effectiveness based on recognized standards (e.g., ISO, ASTM, European Pharmacopoeia, USP). The "reported device performance" would consist of the results of these non-clinical tests meeting the specified limits of those standards.
Below is a general interpretation of what "acceptance criteria" and "reported device performance" would entail for a hip prosthesis based on the provided document, acknowledging that specific numerical details of these criteria and results are not explicitly stated in this letter but would be in the full 510(k) submission.
Based on the provided FDA 510(k) Clearance Letter for the Mfinity Femoral System:
This document describes the clearance of a hip prosthesis, not an AI-powered diagnostic device. Therefore, the typical elements of acceptance criteria and study designs relevant to AI (such as sensitivity, specificity, AUC, expert consensus, training/test set sizes, etc.) are not applicable to this submission.
The "acceptance criteria" for this device would generally refer to meeting established mechanical, material, and biological performance standards for medical implants. The "study that proves the device meets the acceptance criteria" refers to the non-clinical performance testing conducted.
Here's an attempt to populate the requested table and answer the questions based on the information available, while highlighting the irrelevance of AI-specific questions:
1. Table of Acceptance Criteria and the Reported Device Performance
Acceptance Criteria Category | Specific Criteria (General based on device type) | Reported Device Performance (Summary from letter) |
---|---|---|
Mechanical Performance | - Fatigue strength (ISO 7206-4, ISO 7206-6) |
- Pull-off strength of femoral head
- Range of Motion (ISO 21535) | - Mfinity Fatigue tests performed according to ISO 7206-4 and ISO 7206-6 (Results are assumed to have met specified limits)
- Pull-Off Test On CoCr Femoral Head performed (Results are assumed to have met specified limits)
- Evaluation of ROM according to ISO 21535 performed (Results are assumed to have met specified limits) |
| Material Performance | - Coating characteristics (cross-section, SEM, XRD) | - Cross sectioned area evaluation of double coated implant surfaces performed - Scanning Electron Microscopy pictures of double coated implant surfaces taken
- XRD analyses comparing HA coating on Mfinity stem and planar samples performed (Results are assumed to demonstrate acceptable coating properties) |
| Biocompatibility | - Material safety - Cytotoxicity, sensitization, irritation, genotoxicity, chronic toxicity, etc. | - Biocompatibility assessment performed (Results are assumed to demonstrate compliance with standards) |
| Sterilization & Endotoxin| - Bacterial endotoxin levels (European Pharmacopoeia §2.6.14 / USP ) - Pyrogenicity (USP ) | - Bacterial endotoxin test (LAL test) performed
- Pyrogen test performed (Results are assumed to meet requirements, though note indicates not labeled as non-pyrogenic or pyrogen free) |
| Shelf-Life | - Maintain integrity and performance over specified shelf-life | - Shelf-life evaluation performed (Results are assumed to demonstrate acceptable shelf-life) |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified in the FDA letter. For non-clinical tests (like fatigue or material analysis), sample sizes would be determined by the specific ISO/ASTM standards referenced, but these are not diagnostic "test sets" in the AI context.
- Data Provenance: The data comes from the "Non-Clinical Studies" and "Performance Testing" conducted by Medacta International S.A. No country of origin is explicitly stated for the testing labs or data generation, but the company is based in Switzerland. The studies are by nature prospective in that they are conducted specifically to test the device's performance against standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This is a hip prosthesis, not an AI diagnostic device. There is no concept of "ground truth" established by experts in the context of interpreting images or medical conditions for this device. The "ground truth" for the performance tests would be the established scientific and engineering principles and the limits defined by the relevant ISO/ASTM standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As this is not an AI diagnostic device, no adjudication method was used. Performance is assessed directly against objective engineering and material standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a hip prosthesis, not an AI-assisted diagnostic tool. No MRMC study was performed, and there is no AI assistance involved. The letter explicitly states: "No clinical studies were conducted."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an algorithm or AI device. The device (a physical implant) is evaluated standalone based on its inherent properties and performance against physical and material standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the non-clinical performance tests is based on established engineering and material science standards and specifications (e.g., ISO 7206, ISO 21535, ASTM F1580, ASTM F1185, European Pharmacopoeia, USP). The device's performance is compared against the requirements and acceptance limits set forth in these standards.
8. The sample size for the training set
- Not Applicable. This device did not involve machine learning or a "training set."
9. How the ground truth for the training set was established
- Not Applicable. This device did not involve machine learning or a "training set."
Ask a specific question about this device
Page 1 of 1