Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232432
    Manufacturer
    Date Cleared
    2024-01-30

    (172 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K201614, K222561, K220782

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    E3D™ -A Static Interbody Devices are thoraco-lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12- L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

    E3D™ -A Integrated Interbody Devices are integrated lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with three screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g., facet screws or posterior fixation). Hyperlordotic implants (≥20° lordosis) are intended for use with supplemental fixation (e.g., facet screws or posterior fixation). When used without screws or anchors, these devices are intended for use with supplemental fixation (e.g., facet screws or posterior fixation).

    When used for multilevel degenerative scoliosis or sagittal deformity, supplemental fixation that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems) must be used.

    Device Description

    The E3D™ -A Interbody System provides interbody fusion devices designed to provide structural stability during spinal fusion. The E3D™-A Interbody System consists of interbodies offered in various sizes to accommodate surgical needs and anatomic requirements. The E3D-A Interbodies were designed to be placed via an anterior approach. All interbodies in the system are additively manufactured from titanium alloy powder, per ASTM F3001. The System offers both Static and Integrated Interbodies. The Integrated version is to be used in conjunction with three (3) screws that are subtractively manufactured from titanium alloy, per ASTM F136.

    The E3D-A Interbody System's implants are available with and without HA®®º Surface®, a 20-40 nanometer thin hydroxyapatite (HA) surface treatment. The surface treatment presents nano-scale topography on the entirety of the implant surface.

    AI/ML Overview

    The provided text describes a medical device, the E3D™-A Interbody System, and its clearance by the FDA based on substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria through clinical performance or AI/software analysis.

    Therefore, the information required to answer the prompt (acceptance criteria for device performance, sample sizes, expert involvement, ground truth, etc.) is not present in the provided text. The document focuses on regulatory clearance for an interbody fusion device, primarily detailing its physical characteristics, indications for use, and comparison to existing similar devices.

    Specifically:

    • No acceptance criteria or device performance report is provided in the context of a study for clinical efficacy or AI performance. The "Performance Data" section solely refers to physical and material testing (ASTM F2077 for dynamic compression shear testing and Promimic's protocols for the HA surface treatment), not clinical outcomes or AI model performance.
    • No mention of a test set, training set, or data provenance for any AI/software component. The device is an interbody fusion system, a physical implant.
    • No information on experts, ground truth, or adjudication methods as these are not relevant to the type of regulatory submission described (510(k) for a physical implant based on substantial equivalence).
    • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned, as this device does not involve human readers interpreting images with or without AI assistance.
    • No standalone algorithm performance is mentioned, as there is no algorithm described.

    In summary, the provided document is a 510(k) clearance letter and summary for a physical medical device (an interbody fusion system), not a document detailing the performance evaluation of an AI or software device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IdentiTi Porous Ti Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the IdentiTi Porous Ti System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The IdentiTi Porous Ti Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with IdentiTi LIF interbody spacers to provide integrated fixation. IdentiTi LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. IdentiTi ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.

    The IdentiTi Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the IdentiTi NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The IdentiTi NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with IdentiTi NanoTec LIF interbody spacers to provide integrated fixation. IdentiTi NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.

    The Transcend PEEK Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Transcend PEEK Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Transcend PEEK Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with Transcend LIF interbody spacers to provide integrated fixation. Transcend LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. Transcend ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.

    The Transcend PEEK Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Transcend NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Transcend NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with Transcend NanoTec LIF interbody spacers to provide integrated fixation. Transcend NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.

    The IdentiTi ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. Additionally, the IdentiTi ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate. The IdentiTi ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    The IdentiTi ALIF Standalone Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. Additionally, the IdentiTi NanoTec ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi Nanotec ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi NanoTec ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate. The IdentiTi NanoTec ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi NanoTec ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    Device Description

    The IdentiTi and Transcend Interbody Systems are thoracolumbar intervertebral body fusion systems designed to be inserted through anterior and posterior surgical approaches. The interbody spacers are manufactured from PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, tantalum per ASTM F560, titanium alloy (Ti-6Al-4V ELI), and commercially pure titanium (CPTi Grade 2) per ASTM F67. The interbody spacers are available in the following material options: (1) PEEK (polyetheretherketone) with tantalum and titanium alloy markers, or (2) commercially pure porous titanium (PTi), or (3) a combination of commercially pure porous titanium (CPT) Grade 2) per ASTM F67 and titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The subject IdentiTi and Transcend Interbody Systems implants consist of various lengths, widths, heights and lordotic options to accommodate individual patient anatomy. To mitigate risk of expulsion, the interbody endplates feature teeth. All interbody spacers feature an internal graft aperture for placement of graft material to promote fusion through the cage. Additionally, the IdentiTi implants are offered with a microstructure due to the layering of material that forms the porous architecture. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. The IdentiTi and Transcend NanoTec Interbody Systems implant surfaces have been treated with a 20-40 nanometer thin hydroxyapatite (HA) surface treatment. The surface treatment presents nano-scale topography on the entirety of the implant surface, in addition to macro-/micro-scale topography existing from prior to treatment. The purpose of this Traditional 510(k) is to receive clearance for expanded indications for the treatment of degenerative spondylolisthesis, multilevel degenerative scoliosis, spinal stenosis, and sagittal deformity, and use of allogeneic bone consisting of cortical bone, as well as allogeneic bone consisting of demineralized allograft bone with bone marrow aspirate.

    AI/ML Overview

    This document pertains to the Alphatec Spine interbody systems (IdentiTi™ Porous Ti, IdentiTi™ NanoTec™, Transcend™ PEEK, Transcend™ NanoTec™, IdentiTi™ ALIF Standalone, and IdentiTi™ NanoTec™ ALIF Standalone Interbody Systems) and their expanded indications for use.

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state specific acceptance criteria values (e.g., "device must withstand 1000N axial compression"). Instead, it lists the types of non-clinical tests performed, implying that the devices met the requirements of these standard tests. The reported "performance" is implicitly that the devices passed these tests.

    Test CategorySpecific Test / StandardReported Device Performance
    Mechanical TestingASTM F2077Static and dynamic axial compression, compression-shear and torsion tests performed. (Implicitly passed)
    Subsidence TestingASTM F2267Static subsidence test performed. (Implicitly passed)
    Expulsion ResistanceASTM Draft F-04.25.02.02Static push-out test performed. (Implicitly passed)
    Material CharacterizationF1714Gravimetric analysis performed. (Implicitly passed)
    Particulate AnalysisF1877Particulate analysis performed. (Implicitly passed)
    Integrated FixationScrew push-outScrew push-out test performed. (Implicitly passed)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each non-clinical test (e.g., number of interbody devices tested for axial compression). It refers to the tests generally.

    Regarding data provenance, the tests are non-clinical (laboratory tests) and therefore do not involve patient data or data from specific countries of origin. The data is prospective as it was generated specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable to this submission. The "ground truth" for the non-clinical performance tests is based on established engineering standards (ASTM) and the physical properties of the devices themselves, not on expert medical opinion or consensus.

    4. Adjudication Method for the Test Set

    This question is not applicable to this submission. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters, where discrepancies in interpretation need to be resolved. The non-clinical tests outlined here are objective measurements against predefined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable to this submission. The device is an intervertebral body fusion system, a physical implant, not an AI diagnostic or assistance tool. Therefore, MRMC studies involving human readers or AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to this submission. The device is a physical interbody fusion system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" is based on engineering standards and material science specifications. The tests are designed to assess the physical and mechanical properties of the interbody systems against recognized benchmarks for safety and effectiveness in their intended use.

    8. The sample size for the training set

    This question is not applicable to this submission. As the device is a physical implant and not an AI/ML algorithm, there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable to this submission for the same reason as above (no training set).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1