Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K240647
    Date Cleared
    2024-04-05

    (29 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K133444, K212245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Virage Navigation System instruments are to be used during the preparation and placement of Virage® OCT polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in open procedures. When used with the Medtronic NavLock arrays, the Virage Navigation instruments can be used with the Medtronic StealthStation ® S7. When used with the ZimVie reference arrays, the Virage Navigation instruments can be used with the Brainlab Navigation System.

    Device Description

    The Virage® Navigation System is comprised of nonsterile, reusable instruments including drills, taps, and drivers that can be operated manually and/or under a power surgical technique to prepare for and insert Virage OCT polyaxial screws. The Virage Navigation System instruments are designed for use with either the Medtronic StealthStation S7 or the Brainlab Spine & Trauma Navigation software (version 2.0 and higher). The ZimVie reference arrays can only be used with the Brainlab Spine & Trauma Navigation System. The Virage Navigation System is also compatible with the Zimmer Biomet Universal Power System.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA regarding the Virage® Navigation System. It primarily focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed information about a comprehensive clinical study in the format requested.

    Here's a breakdown of the information that can be extracted or inferred from the provided text, along with what is not available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Positional Accuracy Met"positional accuracy...all met the acceptance criteria when the Virage Navigation instruments were attached to the ZimVie reference arrays."
    Usability Met"usability...all met the acceptance criteria when the Virage Navigation instruments were attached to the ZimVie reference arrays."
    Tolerance Analysis Met"tolerance analysis...all met the acceptance criteria when the Virage Navigation instruments were attached to the ZimVie reference arrays."
    Packaging Effectiveness"packaging...evaluated to determine no risks were introduced to the system."
    Sterilization Effectiveness"sterilization...evaluated to determine no risks were introduced to the system."
    Automated Cleaning Safety"Automated cleaning instructions were reviewed and adopted by the Virage Navigation System, but no risks were introduced to the system."
    Biocompatibility (new arrays)"The new reference arrays are non-patient contacting and are composed of stainless-steel (array body) and aluminum (array holder), and both materials were evaluated per ISO 10993-1. Since the device materials have neither direct nor indirect contact with the body, additional biocompatibility information is not necessary." (Implies acceptance criteria met for material safety.)
    Risk Assessment (overall)"A risk assessment was conducted that found risks have been reduced as far as possible and concluded that the benefits associated with spine surgery and the use of the Brainlab Navigation System with the Virage® Navigation System outweigh the risks related to polyaxial screw placement."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document mentions "verification and validation activities" but does not detail the sample sizes for these tests, nor the data provenance. These are typically included in the full 510(k) submission, but not in this summary letter.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The document mentions "usability" but doesn't specify if experts were involved in establishing ground truth for any performance metrics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. The device in question is a navigation system for surgical procedures, not an AI diagnostic tool that human readers would interpret. The goal is to assist surgeons, not necessarily "improve human readers." The study described is a performance assessment of the navigation system itself.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The studies mentioned ("positional accuracy, usability, and tolerance analysis") appear to be evaluating the device performance, which is inherently a "standalone" evaluation of its mechanical and computational capabilities. However, these are not directly comparable to "algorithm only" performance for AI diagnostic tools. The device's function is to assist in surgical navigation, which implies it is always with a human in the loop during actual use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document mentions "positional accuracy." For this, the ground truth would likely be established through precise measurements using calibrated instruments or imaging techniques to determine the true position relative to the navigated object. However, the specific method of establishing this ground truth is not detailed. For "usability," the ground truth would be user feedback and observation against predefined tasks.

    8. The sample size for the training set

    This information is not provided in the document. The Virage® Navigation System, as described, is a mechanical and software-based navigation tool, not an AI/ML-driven diagnostic device that typically undergoes a separate "training set" for model development. The "training" for such systems would involve software development and testing rather than data-driven machine learning model training.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" in the context of an AI/ML model, this information is not applicable as per the provided document. If there are underlying algorithms that use data to improve, those details are not present here.

    In summary:

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device for regulatory clearance. It confirms that the device met specified acceptance criteria for various performance aspects (positional accuracy, usability, tolerance analysis, packaging, sterilization, cleaning, and material biocompatibility). However, it does not provide the detailed study design, sample sizes, expert qualifications, or ground truth establishment methods typically found in a clinical study report for an AI/ML diagnostic device. The device itself is a surgical navigation system, not a diagnostic AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233443
    Date Cleared
    2023-11-16

    (28 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K133444, K212245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vital Navigation System instruments are used during the preparation and placement of the Vital vy System screws during spinal surgery to precisely locate anatomical structures in either open or minimally invasive procedures. The Vital Navigation System instruments are designed for use with either the Medtronic StealthStation S7 or the Brainlab Spine & Trauma Navigation software. The ZimVie reference arrays can only be used with the Brainlab Spine & Trauma Navigation System.

    Device Description

    The Vital™ Navigation System is comprised of nonsterile, reusable instruments including awls, probes, taps, and drivers that can be operated manually. The Vital Navigation System instruments are designed for use with either the Medtronic StealthStation S7 or the Brainlab Spine & Trauma Navigation software. The ZimVie reference arrays can only be used with the Brainlab Spine & Trauma Navigation System. Both combinations are used to assist surgeons in precisely locating anatomical structures in either open or minimally invasive procedures for preparation and placement of Vital and Vitality Screws. This surgical imaging technology provides surgeons visualization for complex and MIS procedures and confirms the accuracy of advanced surgical procedures. The use of these navigation systems provides the surgeon access to real-time multi-plane 3D images) providing confirmation of hardware placement.

    AI/ML Overview

    The provided text states that the Vital™ Navigation System underwent performance testing, including positional accuracy, usability, and tolerance analysis, and that all met the acceptance criteria when the Vital Navigation instruments were attached to the ZimVie reference arrays.

    However, the document does not provide a specific table of acceptance criteria or the reported device performance values for these tests. It also does not detail the study design elements like sample sizes, data provenance, expert qualifications, or the use of MRMC studies.

    Therefore, for aspects not explicitly stated in the document, I will indicate "Not specified in the document."

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Positional accuracy"met the acceptance criteria"
    Usability"met the acceptance criteria"
    Tolerance analysis"met the acceptance criteria"

    Note: The document states that these tests were performed and met the criteria, but the specific metrics or values for the acceptance criteria and the performance are not provided.

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified in the document.
    • Data provenance (country of origin, retrospective/prospective): Not specified in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified in the document. The testing described (positional accuracy, usability, tolerance analysis) seems to be related to the device's technical specifications rather than the generation of clinical ground truth by experts for classification tasks.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified in the document.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study is not mentioned as having been performed. The device is a "Stereotaxic Instrument" and "Navigation System," which aids surgeons in locating anatomical structures and placing screws. It is not an AI-assisted diagnostic imaging device that would typically involve human readers interpreting images.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The device is a "Navigation System" used with human surgeons. It's an instrument assisting a human operator, not a standalone algorithm making autonomous decisions or interpretations that would have an "algorithm only" performance metric in the traditional sense of AI diagnostics. The "positional accuracy" testing would be the closest analogue to a standalone performance metric for a navigation system, indicating its inherent precision.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The nature of the "ground truth" for a navigation system's performance would primarily be related to its engineering precision and accuracy (e.g., actual vs. measured positions, successful and precise placement in simulated or cadaveric models during testing). The document refers to "positional accuracy," "usability," and "tolerance analysis," which imply engineering and usability testing rather than a clinical ground truth established by medical outcomes or expert consensus on a diagnosis.

    8. The sample size for the training set:

    • Not applicable/Not specified. This device is not described as an AI/ML-based system that undergoes a "training" phase with a dataset. Its performance is validated through engineering and usability testing.

    9. How the ground truth for the training set was established:

    • Not applicable. As noted above, this device does not appear to be an AI/ML-based system with a training phase.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223108
    Date Cleared
    2023-01-19

    (111 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    TELIGEN System (K213978), Brainlab Spine & Trauma Navigation System (K212245), Medtronic StealthStation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.

    The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:

    · the use of stereotactic surgery may be appropriate, and

    · reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

    These procedures include but are not limited to spinal fusion. TELIGEN Clear, when used with UNAS, can be precalibrated with the Brainlab Navigation System. TELIGEN Access Probe, when used with UNAS, can be pre-calibrated with the Brainlab Navigation System, where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.

    Device Description

    The TELIGEN Procedure Kit and the TELIGEN Procedure Kit Pro are sterile, single use kits intended for use in surgical spinal procedures allowing for access, visualization, discectomy, graft delivery and navigation.

    The TELIGEN Procedure Kit and the TELIGEN Procedure Kit Pro include a camera, ports and port holder, TELIGEN Clear, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments. Additionally, the TELIGEN Procedure Kit Pro includes an Access Probe as well as the instruments included in the TELIGEN Procedure Kit.

    TELIGEN Access Probe and TELIGEN Clear are part of the DePuy Synthes Navigation Ready Instruments Portfolio and are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS). For further details on UNAS, refer to the UNAS labeling.

    AI/ML Overview

    The provided document, K223108 for the TELIGEN System, primarily focuses on the substantial equivalence of physical medical devices (TELIGEN Access Probe and TELIGEN Clear) which are "Navigation Ready Instruments" used in spinal procedures. It does not describe a study involving an AI/Machine Learning device or software, nor does it provide the detailed performance data, acceptance criteria, ground truth establishment, or study methodologies typically associated with such systems.

    Therefore, I cannot extract the information required to populate the fields related to an AI/ML device study, such as acceptance criteria, sample sizes for test/training sets, expert qualifications, or MRMC studies.

    The document discusses performance data related to the physical instruments:

    • Accuracy Verification: Fulfillment of navigation systems instrument accuracy requirements, Instrument Length Comparison to Predicate Device, Array Characteristics Comparison to Predicate Device.
    • Rigidity of Connections and Instrument during Use
    • CAD Model Evaluation
    • Simulated Use Evaluation

    These are engineering and mechanical performance tests for physical instruments, not an AI/ML algorithm's diagnostic or assistive performance.

    In summary, the provided text does not contain the information requested in the prompt regarding acceptance criteria and study details for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212756
    Date Cleared
    2022-05-05

    (247 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K122211,K212245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Discectomy Navigation Ready Instruments:
    The Discectorny Navigation Ready Instruments when used with the compatible Universal Navigation Adaptor Set are intended to assist the surgeon in locating anatomical structures to facilitate disc space preparation, including discectorny or bony resection. These are indicated for use in surgical spinal procedures, in which:
    · the use of stereotactic surgery may be appropriate, and
    · reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT. MR. 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated navigation arrays.
    These procedures include but are not limited to spinal fusion. The Discectomy Navigation Ready instruments can be precalibrated with the Brainlab Navigation System.

    Universal Navigation Adaptor Set:
    The Universal Navigation Adaptor Set (UNAS) is intended for use with the compatible DePuy Synthes Navigation Ready Instruments to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which:
    · the use of stereotactic surgery may be appropriate, and
    · reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.
    These procedures include but are not limited to spinal fusion. The DePuy Synthes Navigation Ready Instrument, when used with UNAS, can be pre-calibrated and/or manually calibrated with the Brainlab Navigation System, where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.

    Device Description

    Discectomy Navigation Ready Instruments:
    The Discectomy Navigation Ready Instruments are reusable instruments used for disc space preparation, including discectomy or bony resection. These instruments are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS) and associated navigation arrays. For further details on UNAS, refer to the UNAS labeling.
    The Discectomy Navigation Ready Instruments are part of the DePuy Synthes Navigation Ready Instruments Portfolio. The instruments include Cobb Elevators, Curettes and a Bone Graft Delivery Device.

    Universal Navigation Adaptor Set:
    The Universal Navigation Adaptor Set (UNAS) contains reusable spine surgical instruments used to aid in determining the correct location and trajectory of spinal instruments and implants. The UNAS has an interface between third-party navigation systems and the DePuy Synthes Navigation Ready Instruments. The UNAS can only be used with Brainlab and Medtronic StealthStation® navigation systems. The UNAS includes:

    • Brainlab compatible UNAS Navigation Arrays, ●
    • Brainlab compatible Navigation Rings and ●
    • Medtronic compatible Navigation Ring ST. ●
      The Navigation Rings and Navigation Ring ST mate with compatible DePuy Synthes Navigation Ready Instruments. These instruments include implant site preparation and implant insertion instruments as well as access and discectomy instruments.
      When the Brainlab compatible Navigation Ring is attached to the Navigation Ready Instrument, a UNAS Navigation Array can be attached and the instrument can be used only with the Brainlab Navigation System as either a manually calibrated and/or pre-calibrated instrument.
      When the Navigation Ring ST is attached to the Navigation Ready Instrument, a Medtronic SureTrak® II Universal Tracker Fighter array (SureTrak II array) can be attached, and the instrument can be manually calibrated only with the Medtronic StealthStation navigation system.
    AI/ML Overview

    The provided text is a 510(k) summary for the Discectomy Navigation Ready Instruments and Universal Navigation Adaptor Set. It details the device description, indications for use, and performance data used to demonstrate substantial equivalence to predicate devices. However, the document does NOT contain information about acceptance criteria or a study proving that the device meets those criteria, specifically concerning AI/ML performance metrics, ground truth establishment, or human reader studies.

    The performance data listed does not relate to an AI/ML algorithm. Instead, it focuses on the physical and functional aspects of the surgical instruments and their compatibility with navigation systems. Therefore, I cannot extract the information required by your prompt, as the provided document does not describe a study involving AI/ML.

    Here's why the prompt's specific requirements cannot be met by the provided text:

    • AI/ML Performance Metrics (e.g., Sensitivity, Specificity, AUC): Not present. The "Performance Data" section discusses "Accuracy Verification" but this refers to the accuracy of the navigation system's ability to track the physical instruments, not the accuracy of an AI algorithm in diagnosing or classifying medical images.
    • Sample size for test set, data provenance: Not present, as there's no AI test set.
    • Number of experts, qualifications, adjudication method: Not present, as there's no expert review for AI-generated output.
    • Multi-Reader Multi-Case (MRMC) study: Not present, as there is no AI assistance for human readers mentioned.
    • Standalone (algorithm-only) performance: Not present, as no AI algorithm is detailed.
    • Type of ground truth (pathology, outcomes data, etc.): Not present, as there's no AI model producing diagnostic or prognostic output.
    • Training set size and ground truth for training set: Not present, as there's no AI model that requires training data.

    The document's "Performance Data" section outlines the following evaluations:

    • Accuracy Verification:
      • Fulfillment of navigation systems instrument accuracy requirements
      • Instrument Length Comparison to Predicate Device
      • Array Characteristics Comparison to Predicate Device
    • Navigation Connection Repeatability for Pre-Calibrated Instruments
    • CAD Model Evaluation
    • Simulated Use Evaluation

    These are standard engineering and functional performance tests for medical devices, particularly surgical instruments used with navigation systems, and do not involve artificial intelligence or machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1