Search Results
Found 1 results
510(k) Data Aggregation
(463 days)
The F&P 950 Respiratory Humidifier is intended to provide heat and humidity to respiratory gases delivered to patients. It is for use in a hospital or long-term care facility by a health professional.
F&P 950 Breathing Circuit Kits: 950A80J, 950A81J, 950A82J, 950A61J, 950A61J, 950A40J, AA451SU
The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to adult, adolescent and child patients, within the limits of its stated technical specifications.
F&P 950 Breathing Circuit Kits: 950P81J, 950P40J
The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to neonatal, infant, child and adolescent patients of its stated technical specifications.
F&P 950 Breathing Circuit Kits: 950N80J, 950N81J, 950N61J, 950N40J
The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to neonatal, infant and child patients, within the limits of its stated technical specifications.
The F&P 950 Respiratory Humidifier is designed to warm and humidify air after it leaves a commercially available flow source and to maintain the air in a warmed and humidified condition to the point of delivery into the patient via heated tubing.
The F&P 950 Respiratory Humidifier is made up of a heaterbase, sensor cartridge and expiratory heater wire adapter with breathing circuit kits to support the use with adult, pediatric and neonatal patients. Breathing circuit kits enable the use of the F&P 950 Respiratory Humidifier with clinical procedures such as invasive ventilation and nasal high flow.
The device is a programmable electrical medical device that uses sensors throughout the system to humidify the gas. Gas travels through the chamber where it is humidity is requlated through monitoring the gas temperature exiting the chamber to achieve the target humidity level. The gas is then transported to the patient through a heated delivery tube. The gas temperature exiting the heated delivery tube is monitored to minimize the formation of condensate.
The document is a 510(k) Summary for the F&P 950 Respiratory Humidifier and its accessory breathing circuit kits. It outlines the device's indications for use and compares its technological characteristics to a primary predicate device (F&P MR850 Respiratory Humidifier) and several secondary predicate devices for the various breathing circuit kits. The summary also details the non-clinical tests performed to demonstrate substantial equivalence.
Here's the requested information based on the provided text, focusing on acceptance criteria and supporting studies:
-
Table of acceptance criteria and the reported device performance:
Acceptance Criteria Reported Device Performance Delivered humidity level (Bypassed airway) >33mg/L for flow 12mg/L for flow
Ask a specific question about this device