    Intended Use

    STERI-OSS Implant System is intended for use in partially or fully edentulous mandibles and maxilla, in support of single of multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. STERI-OSS Implant System is for single stage surgical procedures. This system is intended for delayed loading.

    Device Description

    STERI-OSS Implant System, Fixture is a medical device made of titanium that is inserted into the maxillary or mandibular alveolar bone to support prosthesis such as artificial teeth for the patient's recovery of masticatory function.
    STERI-OSS Implant System. Abutment is inserted to support prosthesis such as artificial teeth, and is a medical device made of titanium alloy. It connects artificial teeth with a fixture implanted in the maxillary or mandibular alveolar bone where teeth are lost.

    AI/ML Overview

    The provided document is a 510(k) summary for the STERI-OSS Implant System, which is a dental implant system. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing a study where the device meets pre-defined acceptance criteria for performance in a clinical setting. For medical devices undergoing 510(k) clearance, the primary goal is to show that the new device is as safe and effective as a legally marketed predicate device, not necessarily to meet specific novel performance acceptance criteria through clinical studies.

    Therefore, the requested information elements such as "acceptance criteria and the reported device performance," "sample sizes used for the test set," "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for the training set," and "how ground truth for the training set was established" are typically not applicable in the context of a 510(k) substantial equivalence submission for a device like a dental implant. These elements are more commonly associated with clinical trials or performance studies for novel devices or software with AI/ML components requiring new clinical evidence.

    The document describes non-clinical testing to demonstrate that the differences between the subject device and predicate devices do not affect substantial equivalence.

    Here's a breakdown of the available and non-applicable information based on your request:


    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not applicable in the format typically used for performance studies with defined acceptance criteria for clinical outcomes. The document instead presents a comparison of technological characteristics with predicate devices and reports on non-clinical engineering and materials testing to support substantial equivalence.

    ItemAcceptance Criteria (Not explicitly defined for clinical performance)Reported Device Performance (as per non-clinical testing)
    Fixture- Material: Pure Titanium Grade 4 (ASTM F67)Matches predicate/reference devices
    - Connection type: Internal HexMatches predicate/reference devices
    - Surface Treatment: SLAMatches predicate/reference devices
    - Sterilization: Gamma SterilizationMatches predicate/reference devices
    - Dimensions: Within similar dimensional range of predicatesDemonstrated to be similar or within range
    Abutment- Material: Ti6Al4V ELI (ASTM F136)Matches reference devices
    (Straight Abutment)- Principle of operation: Cement retained restorationMatches reference devices
    - Surface Treatment: None (machined surface)Matches some reference devices; difference with TiN-coated predicate deemed minor
    - Dimensions: Similar design and dimension rangeDemonstrated to be similar or within range
    Abutment- Material: Ti6Al4V ELI (ASTM F136)Matches reference devices
    (Solid Abutment)- Principle of operation: Cement retained restorationMatches reference devices
    - Surface Treatment: None (machined surface)Matches some reference devices; difference with TiN-coated predicate deemed minor
    - Dimensions: Similar design and dimension rangeDemonstrated to be similar or within range
    Abutment- Material: Ti6Al4V ELI (ASTM F136)Matches reference devices
    (Angled Abutment)- Principle of operation: Cement retained restorationMatches reference devices
    - Surface Treatment: None (machined surface)Matches some reference devices; difference with TiN-coated predicate deemed minor
    - Dimensions: Similar design and dimension rangeDemonstrated to be similar or within range, supported by mechanical bench testing for differences
    Screw- Material: Ti6Al4V ELI (ASTM F136)Matches reference devices; difference with cp-Ti predicate deemed minor
    (Abutment Screw)- Principle of operation: Connect abutment to fixtureMatches reference devices
    - Surface Treatment: NoneMatches reference devices
    - Dimensions: Within reference device dimensionsDemonstrated to be within range
    Screw- Material: Ti6Al4V ELI (ASTM F136)Matches some reference devices; difference with cp-Ti predicate deemed minor
    (Cover Screw)- Principle of operation: Protect internal portion of implantMatches reference devices
    - Surface Treatment: NoneMatches some reference devices; difference with anodized predicate deemed minor
    - Dimensions: Within range of reference device dimensionsDemonstrated to be within range
    Abutment- Material: Ti6Al4V ELI (ASTM F136)Matches some reference devices; difference with cp-Ti predicate deemed minor
    (Healing Abutment)- Principle of operation: Help soft tissue of gum naturally formedMatches reference devices
    - Surface Treatment: NoneMatches some reference devices; difference with anodized predicate deemed minor
    - Dimensions: Similar design and dimension rangeDemonstrated to be similar or within range
    General Non-Clinical- ISO 14801:2016 (Fatigue)Successfully tested for worst-case scenario
    - ISO 11137-1/2 (Gamma Sterilization)Validated
    - ANSI/AAMI ST79, ISO 17665-1/2, ISO 11737-1/2, ISO 11138-1/3 (End User Sterilization)Validated
    - ASTM F88, F1140, F1929, F2096 (Shelf Life)Tested (5 years)
    - ISO 10993-1/5/12 (Biocompatibility)Evaluated and tested
    - USP (Pyrogen and Endotoxin)Endotoxin testing conducted per batch (limit