Search Results
Found 3 results
510(k) Data Aggregation
(95 days)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Transesophageal (Cardiac), Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Peripheral vessel and Lung.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals (includes emergency room), private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
The HM70 EVO is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode. CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HM70 EVO also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HM70 EVO has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
The provided text is a 510(k) Premarket Notification for the HM70 EVO Diagnostic Ultrasound System. It describes the device, its intended use, and comparative information to predicates to demonstrate substantial equivalence. However, it explicitly states that clinical studies were not required to support substantial equivalence for this device. Therefore, the document does not contain information about acceptance criteria, a study proving the device meets those criteria, sample sizes, expert involvement, or MRMC studies for AI-assisted performance.
The "Summary of Clinical Tests" section on page 5, point 12, clearly states:
"The subject of this premarket submission, HM70 EVO, is not required clinical studies to support substantial equivalence."
Without a clinical study, none of the requested information regarding acceptance criteria, performance metrics, sample sizes, ground truth establishment, or expert involvement can be extracted from this document. The provided text only discusses non-clinical tests related to safety standards, acoustic output, biocompatibility, software function, cleaning, disinfection, thermal, electrical, electromagnetic, and mechanical safety.
Ask a specific question about this device
(121 days)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Peripheral vessel and Lung.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals (includes emergency room), private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
The HM70 EVO is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic). Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HM70 EVO also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HM70 EVO has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
This document pertains to the 510(k) Premarket Notification of the HM70 EVO Diagnostic Ultrasound System by Samsung Medison CO., LTD. As specified in section 12, "Summary of Clinical Tests: The subject of this premarket submission, HM70 EVO, is not required clinical studies to support substantial equivalence."
Therefore, I cannot provide details regarding acceptance criteria for clinical performance or a study demonstrating device performance based on the provided text, as no clinical studies were deemed necessary for this submission. The submission relies solely on non-clinical tests to establish substantial equivalence to predicate devices.
Here's the breakdown of the information that can be extracted from the document regarding non-clinical acceptance criteria and their assessment:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Reference/Standard | Reported Device Performance |
---|---|---|
Acoustic Output | NEMA UD 2-2004 (R2009) | Acoustic power levels are below the applicable FDA limits. |
Biocompatibility | ISO 10993-1:2009/(R)2013 | System is manufactured with materials which have been evaluated and found to be safe for the intended use. |
Software Function | Not explicitly stated, implied by overall safety and effectiveness. | Software functionality contributes to the device performing as intended. |
Cleaning & Disinfection Effectiveness | Not explicitly stated, implied by overall safety and effectiveness. | The device meets requirements for cleaning and disinfection. |
Thermal Safety | IEC 60601-1 | Device conforms to applicable FDA guidance and medical device safety standards. |
Electrical Safety | IEC 60601-1 | Device conforms to applicable FDA guidance and medical device safety standards. |
Electromagnetic Safety (EMC) | IEC 60601-1-2: 2014 (4th Edition) | Device conforms to applicable FDA guidance and medical device safety standards. |
Mechanical Safety | IEC 60601-1, IEC 60601-2-37 | Device conforms to applicable FDA guidance and medical device safety standards. |
Risk Management | ISO 14971:2007 | Application of risk management to medical devices has been followed. |
Basic Safety and Essential Performance (General) | IEC 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 | Device conforms to applicable FDA guidance and medical device safety standards. |
Basic Safety and Essential Performance (Ultrasonic Specific) | IEC 60601-2-37 Edition 2.0 2007 | Device conforms to applicable FDA guidance and medical device safety standards. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable, as no clinical studies were performed or required for this 510(k) submission. The evaluation was based on non-clinical testing against established engineering and safety standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no clinical studies were performed or required. Ground truth in this context refers to compliance with established technical specifications and safety standards, evaluated by qualified engineers and testers.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical studies were performed or required.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The HM70 EVO is a diagnostic ultrasound system, and while it may incorporate advanced imaging techniques (e.g., ElastoScan), the provided document does not indicate the presence of AI assistance requiring an MRMC study. No clinical studies were required.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The HM70 EVO is a diagnostic ultrasound system and its performance is inherently tied to human operation and interpretation. The document does not describe any standalone algorithm performance separate from the device itself. No clinical studies were required.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical tests, the "ground truth" was compliance with established national and international recognized standards and FDA guidance documents for medical devices (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA UD 2-2004). This involves objective measurements and tests to verify specified performance and safety parameters.
8. The sample size for the training set
Not applicable, as no clinical studies or AI algorithm training are described or required for this submission.
9. How the ground truth for the training set was established
Not applicable, as no clinical studies or AI algorithm training are described or required for this submission.
Ask a specific question about this device
(54 days)
The HS40 ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The HS40 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS40 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS40 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
The provided text is a 510(k) Premarket Notification for the Samsung Medison HS40 Diagnostic Ultrasound System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets specific performance acceptance criteria for a novel AI/ML algorithm.
Therefore, the document does not contain the acceptance criteria and the detailed study results as would be presented for a device demonstrating novel AI/ML performance. It focuses on showing the HS40 ultrasound system, including new software features and transducers, is equivalent to existing cleared ultrasound systems.
Here's an analysis based on the information provided, highlighting why the requested details are not present for an AI/ML algorithm in this specific document:
Missing Information (as expected for a 510(k) that is not for a novel AI/ML algorithm):
- Acceptance Criteria Table and Reported Device Performance: This document describes a general diagnostic ultrasound system. It outlines the intended clinical applications (Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal, Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel) and modes of operation (B-mode, M-mode, PWD, CWD, Color Doppler, etc.). There are no specific performance metrics (e.g., sensitivity, specificity, AUC) or acceptance criteria for an AI/ML algorithm.
- Sample Size for Test Set and Data Provenance: Not applicable, as no specific AI/ML performance test set is described.
- Number of Experts and Qualifications for Ground Truth Establishment (Test Set): Not applicable.
- Adjudication Method for Test Set: Not applicable.
- Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. The document states, "The subject of this premarket submission, HS40, is not required clinical studies to support substantial equivalence." This means a formal comparative effectiveness study like an MRMC study was not deemed necessary for this 510(k). The "summary of non-clinical tests" focuses on acoustic output, biocompatibility, software function, and safety standards.
- Standalone Performance (Algorithm Only): Not applicable. This is a system submission, not a standalone AI/ML algorithm. While the device mentions "added some of SW features (2DNT, BiometryAssist, E-Strain, NeedleMate, AutoIMT, etc.) that migrated from predicate HS50/60," it does not present performance data for these specific features in the way an AI/ML algorithm's performance would be evaluated. These features are likely automated measurement or image processing tools that were already cleared on other predicate devices.
- Type of Ground Truth Used: Not applicable, as there's no specific AI/ML performance outcome being measured against ground truth.
- Sample Size for Training Set & How Ground Truth for Training Set was Established: Not applicable, as this document does not describe the development or training of a novel AI/ML algorithm. It describes a general ultrasound system and its equivalence to pre-existing cleared devices.
Information from the document that is tangentially related (but not what was asked for directly related to AI/ML acceptance criteria):
- Device Name: HS40 Diagnostic Ultrasound System.
- Intended Use: "Diagnostic ultrasound imaging or fluid flow analysis of the human body." Clinical applications are listed above.
- Technology: "The HS40 employs the same fundamental scientific technology as its predicate device." This highlights the equivalence approach rather than novel technology.
- Predicate Devices: HS40 Diagnostic Ultrasound System (K180409) - Primary Predicate, HS50/60 Diagnostic Ultrasound System (K181336), SONOACE R7 Diagnostic Ultrasound System (K133505), HM70A Diagnostic Ultrasound System (K182894).
- Non-Clinical Tests: Evaluated for acoustic output, biocompatibility, software function, cleaning and disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety, conforming to applicable FDA guidance and medical device safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO10993-1, ISO14971, NEMA UD 2-2004, NEMA UD 3-2004). This speaks to the safety and fundamental performance of the ultrasound system itself, not the performance of an AI/ML algorithm.
- Regulatory Pathway: 510(k) Premarket Notification - Traditional. This pathway typically relies on substantial equivalence to a predicate, and for general ultrasound systems, does not usually require detailed clinical studies demonstrating AI/ML performance.
In summary, this 510(k) submission is for a general diagnostic ultrasound system and its associated transducers, demonstrating substantial equivalence to already-cleared predicate ultrasound systems. It does not provide the kind of performance data, acceptance criteria, or study details typically associated with the evaluation and clearance of novel AI/ML algorithms.
Ask a specific question about this device
Page 1 of 1