Search Results
Found 2 results
510(k) Data Aggregation
(28 days)
Celerity 5 HP Biological Indicator:
The Celerity 5 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:
- Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems
- Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear®
- Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®.
When used in conjunction with the Celerity® Incubator, the Incubator provides a fluorescent result within 5 minutes.
Celerity 5 HP Challenge Pack:
The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.
The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
Celerity 20 HP Biological Indicator
The Celerity 20 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:
- Lumen, Non-Lumen, Fast Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems.
- Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear
- Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®
When used in conjunction with the Celerity HP Incubator, the Celerity 20 HP BI provides a fluorescent result within 20 minutes.
Celerity 20 HP Challenge Pack
The Celerity 20 HP Challenge Pack is intended for qualification testing of the V-PRO® Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.
The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The challenge pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
Celerity 5 HP BI
The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 5 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.
Celerity 5 HP Challenge Pack
The Celerity 5 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator, a Celerity HP Chemical Indicator, and a layer of absorptive foam within a Tyvek pouch.
The intended change to the devices is to update the secondary container and product labeling. Specifically, the clear cellophane wrapper around the product carton will be replaced by a clear, laminated pouch. Labeling on the carton and the Instructions for Use (IFU) will be updated to specify that the product must remain stored in its original laminated pouch to maintain the claimed shelf-life. Additionally, a storage disclaimer label will be added to the exterior of the pouch to notify users of this change. Finally, this submission is intended to establish a 3-month in-use shelf-life following pouch opening, not to exceed the original expiration date.
Celerity 20 HP BI
The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 20 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.
Celerity 20 HP Challenge Pack
The Celerity 20 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator, a Celerity HP Chemical Indicator, and a layer of absorptive foam within a Tyvek pouch.
The intended change to these devices is to update the secondary container and product labeling. Specifically, the clear cellophane wrapper around the product carton will be replaced by a clear, laminated pouch. Labeling on the carton and the Instructions for Use (IFU) will be updated to specify that the product must remain stored in its original laminated pouch to maintain the claimed shelf-life. Additionally, a storage disclaimer label will be added to the exterior of the pouch to notify users of this change. Finally, this submission is intended to establish a 3-month in-use shelf-life following pouch opening, not to exceed the original expiration date.
The provided text describes premarket notification for biological indicators and doesn't involve an AI/ML powered medical device. Therefore, many of the requested criteria, such as "multi-reader multi-case (MRMC) comparative effectiveness study" or "number of experts used to establish ground truth", which are specific to AI/ML powered devices, are not applicable here.
However, I can extract information related to the acceptance criteria and the study proving the device meets those criteria, as much as applicable for a non-AI medical device.
Device: Celerity 5 HP Biological Indicator (LCB052) & Challenge Pack (LCB059), and Celerity 20 HP Biological Indicator (LCB044) & Challenge Pack (LCB045)
Type of Device: Sterilization Process Indicator (Biological Indicator - BI). These devices contain a specific type of spore (Geobacillus stearothermophilus) and media to confirm the effectiveness of a sterilization process. A positive result (fluorescent signal) indicates sterilization failure, while a negative result indicates successful sterilization.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria and results are provided for two distinct product lines: Celerity 5 HP and Celerity 20 HP. Note that the "Result" for all tests is simply "PASS", indicating that the criteria were met.
For Celerity 5 HP Biological Indicator and Challenge Pack:
Test | Acceptance Criteria | Result |
---|---|---|
Reduced Incubation Time Testing | All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5-minutes | PASS |
Simulated Use Testing | All processed BIs shall be sterile following full cycle exposure sterilizer cycles. | PASS |
Cap Media Testing | All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators. | PASS |
Stability Testing: Population | The mean initial population at the start of the stability study must be 1.0-4.0 x 10⁶ CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population. | PASS |
Stability Testing: D-value | D-value between 6 and 20 seconds. | PASS |
Stability Testing: Survival/Kill | Demonstration of one all survive time and one all kill time point. | PASS |
Stability Testing: RIT | All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5 minutes. | PASS |
Stability Testing: Media Testing | All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators. | PASS |
For Celerity 20 HP Biological Indicator and Challenge Pack:
Test | Acceptance Criteria | Result |
---|---|---|
Reduced Incubation Time Testing | All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutes | PASS |
Simulated Use Testing | All processed BIs shall be sterile following full cycle exposure sterilizer cycles. | PASS |
Cap Media Testing | All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators. | PASS |
Stability Testing: Population | The mean initial population at the start of the stability study must be 1.0-4.0 x 10⁶ CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population. | PASS |
Stability Testing: D-value | D-value between 6 and 20 seconds. | PASS |
Stability Testing: Survival/Kill | Demonstration of one all survive time and one all kill time point. | PASS |
Stability Testing: RIT | All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutes. | PASS |
Stability Testing: Media Testing | All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators. | PASS |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size (e.g., number of BIs or challenge packs) used for each specific test (Reduced Incubation Time, Simulated Use, Cap Media, Stability testing). It refers to "All BI lots" for Reduced Incubation Time and "All processed BIs" for Simulated Use, and implicitly "BIs" for Stability Testing. This suggests a statistically relevant number of units from manufactured lots were tested.
- Data Provenance: The document does not specify the country of origin of the data. Given the "U.S. Food & Drug Administration" letterhead and the manufacturer's address in Ohio, USA, it is highly likely that the testing was conducted in the US.
- Retrospective or Prospective: The testing appears to be prospective as it's a pre-market submission to demonstrate the device meets performance criteria for clearance. Stability testing also inherently involves prospective observation over time.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This concept is not applicable to this type of device. The "ground truth" for a biological indicator is determined by the biological and chemical reactions themselves (i.e., presence or absence of viable spores, or a color change in the chemical indicator), not by human expert interpretation of images or data. The determination of "sterile" or "non-sterile" from a biological indicator is a direct, objective measurement.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the results of biological indicators are typically objective and determined by the physical/chemical properties and reactions, not by subjective human interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device nor one that involves human "readers" interpreting medical images or data.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical, biological and chemical indicator device, not a software algorithm. The "Celerity Incubator" is an accessory that reads the fluorescent signal, it is not an AI algorithm.
7. The Type of Ground Truth Used
The ground truth for these devices is based on biological viability (presence or absence of viable Geobacillus stearothermophilus spores) for the biological indicator and chemical reaction (color change) for the chemical indicator, validated against recognized sterilization parameters (e.g., D-value, Survival Time, Kill Time). This is a direct measure of the effectiveness of the sterilization process.
8. The Sample Size for the Training Set
Not applicable. These are not AI/ML devices that require training sets. The "training" of these devices (i.e., their manufacturing and design) is based on established microbiological and sterilization science.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The performance of these biological indicators relies on their inherent design and manufacturing quality, ensuring the spores exhibit the specified resistance characteristics and the media accurately detects their viability. This is established through rigorous quality control and adherence to relevant standards for biological indicators.
Ask a specific question about this device
(108 days)
The CELERITY 20 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.
The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The challenge pack is not intended for routine monitoring of the V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the sterilizers.
The CELERITY 20 HP Challenge Pack (pack), is used by healthcare providers for qualification testing of the V-PRO Low Temperature Sterilization Systems. The pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The user places the pack into the V-PRO Sterilizer and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the CELERITY 20 HP Biological Indicator* (BI) contained in the pack are retrieved. The CI is assessed for a passing color change immediately and the BI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.
The BI is activated by sealing the vial and thus puncturing the cap to release the contained media.
The activated SCBI is incubated at 55-60 ℃ in the CELERITY HP Incubator** for a final determination of viability within 20 minutes of incubation.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Simulated Use | Demonstrate the pack shows passing results in worst-case load under worst-case sterilization conditions | PASS |
Cycle-Specific Testing | The pack provides an equivalent or greater challenge to the claimed cycle than the biological model under worst-case conditions. | PASS |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions "Performance testing... has been completed" and implies that these tests were conducted by the manufacturer for regulatory submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The tests seem to be laboratory-based performance evaluations against specific criteria rather than expert-based assessments.
4. Adjudication Method for the Test Set
This information is not provided as the tests appear to be objective performance evaluations (pass/fail) rather than requiring adjudication for subjective interpretations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and its Effect Size
An MRMC comparative effectiveness study was not performed or reported in this document. The device is a sterilization process indicator, not an AI-assisted diagnostic tool that would typically involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
The device itself is a "Biological Indicator Challenge Pack" which is a physical product, not a software algorithm. The "performance" being tested is its ability to accurately indicate sterilization conditions. Therefore, a standalone algorithm-only performance study is not applicable in this context. The "incubation" part involves a device (CELERITY HP Incubator) that reads the biological indicator, which is a form of automated assessment, but not an "algorithm only" study in the typical sense of AI.
7. The Type of Ground Truth Used
The ground truth for the "Simulated Use" and "Cycle-Specific Testing" appears to be established through:
- Worst-case sterilization conditions: This refers to the scientific and engineering parameters that define a minimal challenge for sterilization.
- Biological model: This likely refers to the known and established resistance of the biological indicator organism (e.g., bacterial spores) to the sterilization process. The acceptance criteria imply a biochemical or microbiological ground truth (i.e., whether the biological indicator's spores are killed or not under defined conditions).
8. The Sample Size for the Training Set
This product is a physical device (sterilization indicator), not an AI model that requires a training set. Therefore, the concept of a "training set" and its sample size is not applicable.
9. How the Ground Truth for the Training Set was Established
As explained above, a training set is not applicable for this device.
Ask a specific question about this device
Page 1 of 1