Search Results
Found 2 results
510(k) Data Aggregation
(74 days)
TrueLok Elevate is intended for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities. The TrueLok Elevate is indicated for adult and pediatric (greater than 2 through 21 years of age) patients.
The subject TrueLok Elevate is an external fixation component system (including its accessories) to be used with the Orthofix TrueLok family, for which Orthofix identified as a predicate device TrueLok Hexapod System (TL-HEX) V2.0 (K170650). The subject device consists in a further series of elements for external fixation added to the Orthofix TrueLok family with the aim of supporting the Orthofix TrueLok external fixator systems family falling within the indications for use of the more extensive, cleared indications for use of the chosen predicate device, for the specific use in bone transport treatment.
The subject TrueLok Elevate is intended for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities.
The TrueLok Elevate is indicated for adult and pediatric (greater than 2 through 21 years of age) patients.
The subject device is constituted by an external fixator and related accessories (half pins, k-wires, drill positioning guide, template and template inserts and convenience kits).
The technique for the use of the subject device consists in fixing two half pins on the first cortical of the bone segment that the surgeon decided to transport, and two half pins on both cortexes of the bone.
The positioning of the half pins is driven by a template.
During the treatment, through the knob present on the device, the bone segment is gradually pulled outward by the patient/caregiver to laterally transport the bone segment.
The subject device, as the predicate, will be implanted only by Healthcare Professionals (HCP), with full awareness of the appropriate orthopedic procedures (including application and removal), in the operating theatre only. The distraction of the limb will be activated in home by the patient/caregiver or in clinic theatre by the HCP. Treatment activation for pediatric patients in the home environment may require the assistance of a caregiver.
The Orthofix TrueLok™ Elevate is an external fixation component system intended for the treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities in adult and pediatric patients (greater than 2 through 21 years of age).
Here's an analysis of the acceptance criteria and the study performed, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly state specific pass/fail acceptance criteria values for the mechanical tests conducted. Instead, it describes comparative testing against predicate or reference devices to demonstrate similarity in performance. The assessment for each technological characteristic indicates that "no different questions have been raised" or that the subject device's indications fall within the predicate's, suggesting that the goal was to demonstrate equivalence rather than meeting pre-defined numerical thresholds for acceptance.
Test Description | Reported Device Performance |
---|---|
External Frame: | |
Static axial stiffness test (according to ASTM F1541-17 Annex 7) | Performed for subject TrueLok Elevate external fixator. Performance was compared with reference device True/Lok Monolateral/Bilateral Fixator (K941048). The implicit acceptance criterion is that its stiffness is comparable or superior to the reference device, ensuring similar mechanical behavior. |
Slipping torque on connectors test (according to ASTM F1541-17 Annex 2) | Performed for subject TrueLok Elevate external fixator. Performance was compared with reference device True/Lok Monolateral/Bilateral Fixator (K941048). The implicit acceptance criterion is that its slipping torque is comparable or superior to the reference device, ensuring secure connection and stability. |
Half Pins: | |
Static 4-point bending evaluation | Performed for subject half pins of TrueLok Elevate. Performance was compared to the predicate device Orthofix TrueLok Hexapod System (K170650). The implicit acceptance criterion is that its bending strength is comparable or superior to the predicate device, ensuring similar structural integrity. |
Torsional strength evaluation | Performed for subject half pins of TrueLok Elevate. Performance was compared to the predicate device TrueLok Hexapod System (K170650). The implicit acceptance criterion is that its torsional strength is comparable or superior to the predicate device, ensuring similar resistance to twisting forces. |
Overall Assessment: | The conclusions state: "Based upon substantial equivalences in: intended use, patient population, site of application, conditions of use, operating principles, and the non-clinical performance data, the subject TrueLok™ Elevate has been shown to be substantially equivalent to the legally marketed predicate device (K170650)." This indicates the device met the implicit acceptance criterion of performing comparably to the predicate. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the mechanical tests (e.g., number of external fixators or half pins tested).
The provenance of the data is not specified regarding country of origin or whether it was retrospective or prospective, as these were bench tests performed on devices rather than patient data.
3. Number of Experts Used to Establish Ground Truth and Their Qualifications
This information is not applicable to this type of submission. The ground truth for mechanical performance is established through standardized testing protocols (e.g., ASTM standards) and comparisons to previously cleared predicate devices, not through expert consensus on medical images or patient outcomes.
4. Adjudication Method for the Test Set
This information is not applicable to this type of submission as the mechanical tests are objectively measured, not subject to subjective interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC study was not performed. This type of study is typically relevant for AI/imaging devices where human reader performance is being evaluated with and without AI assistance. The TrueLok™ Elevate is a physical external fixation system.
6. Standalone (Algorithm Only) Performance Study
A standalone study was not performed, as the device is a physical medical device, not a software algorithm.
7. Type of Ground Truth Used
The ground truth used for demonstrating performance was based on objective mechanical measurements conducted in accordance with recognized industry standards (e.g., ASTM F1541-17) and comparisons to the established performance characteristics of legally marketed predicate devices. The "truth" is that the new device's mechanical properties fall within acceptable limits or are comparable to those of the predicate, as determined by these tests.
8. Sample Size for the Training Set
This information is not applicable. The TrueLok™ Elevate is a physical medical device, not a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(138 days)
The indications for the I.T.S. Intramedullary Nailing System (INS) - Proximal Femur include intramedullary treatment of fractures and stabilization after tumor resection in the area of the proximal femur, as well as combinations that additionally affect the shaft area using the long nails.
The indications for use of the I.T.S. INS Proximal Femur include:
All Nails:
- · Stable and unstable pertrochanteric fractures
- · Intertrochanteric fractures
- · Combinations of fractures listed above
Additionally for Long Nails:
- Subtrochanteric fractures
- · Proximal fractures as listed above associated with shaft fractures
- · Pathological fractures in regions as listed above
- · Nonunions and malunions in regions as listed above
The I.T.S. INS Proximal Femur Nail consists of the following implants:
-
Nails:
The Proximal Femur Nails are designed for the treatment of femoral fractures. The Femur Nails are available in distal diameters ranging from 9 to 14mm and lengths of 180mm (short nail), 240mm (intermediate nail), and 260 to 480mm in 20mm increments (long nail). 180mm and 240mm nails have a single, oblong hole distally, while all nails with a length over 260 have a round hole, an oblong hole and another round hole for distal fixation. The oblong hole allows for both static and dynamic locking configurations. All Trochanteric Nails are locked proximally with a Lag Screw. The set screw engages with the lag screw and allows translation while preventing rotation. Endcaps close the top of the nail and are available in 0mm, 5mm, 10mm, 15mm and 20mm lengths. -
Lag Screw
Lag screws are cannulated with a major diameter of 10.5mm and lengths ranging from 70mm to 130mm in 5mm increments. They are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136. -
Set Screw
There are two types of set screws available. Inferior short and standard with a diameter of 6.6mm. The Length of the inferior short version is 45mm and the standard lengths are 70 to 130mm in 5mm increments. They are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136. -
Distal Cortical Locking Screw
Distal locking screws are fully threaded, with a diameter of 5.0mm and lengths ranging from 25mm to 120mm in 5mm increments (2.5mm increments available from 25-70mm). Thev are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136. -
Instrumentation
I.T.S. INS Proximal Femoral Nail Instrumentation consists of implant-specific targeting devices which help facilitate insertion of the nails, lag screws, and distal locking screws. Targeting devices are manufactured from stainless steel and carbon fibre reinforced PEEK. Other instrumentation includes drills, drill guides, k-wires and screwdrivers.
All nails and screws are provided Sterile for single-use.
The provided text describes a 510(k) submission for the I.T.S. INS Proximal Femur Nail. This is a medical device for intramedullary fixation, and the FDA has determined it to be substantially equivalent to previously marketed predicate devices.
However, the text does not include information about AI/ML device performance, acceptance criteria normally associated with diagnostic algorithms (like sensitivity, specificity), or details about studies involving human readers or ground truth establishment relevant to AI. The nature of this submission is for a physical medical device (an intramedullary nail), not an AI-powered diagnostic tool.
Therefore, many of the requested categories (2, 3, 4, 5, 6, 7, 8, 9) are not applicable to the provided document.
Here's a breakdown of what can be extracted from the text, focusing on the mechanical evaluation of the physical device:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical acceptance criteria in the format typically used for AI/ML performance (e.g., minimum sensitivity). Instead, the performance is evaluated against consensus standards and compared to predicate devices for substantial equivalence.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Mechanical Performance | Fatigue Testing of the nail screw construct: Demonstrated that the I.T.S. INS Proximal Femoral Nail is substantially equivalent in mechanical performance to the predicate device. |
Performance testing following consensus standards ASTM F543 & ASTM F1264: Demonstrated that the I.T.S. INS Proximal Femoral Nail is substantially equivalent in mechanical performance to the predicate device. | |
Material, Geometry, Design, Indications, and Operational Principles | The I.T.S. INS Proximal Femur Nail is substantially equivalent in material, geometry, design, indications, and operational principles to the legally marketed predicate systems. |
2. Sample size used for the test set and the data provenance
Not applicable. This is a physical medical device, and the testing involved mechanical performance, not a test set of data for an AI algorithm.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth establishment by experts is relevant to AI/ML diagnostic performance, not the mechanical testing of an intramedullary nail.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are relevant to expert consensus for AI/ML ground truth, not mechanical device testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. There is no AI algorithm being evaluated.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the mechanical testing, the "ground truth" or standard for comparison was defined by consensus standards (ASTM F543 & ASTM F1264) and the mechanical performance of predicate devices.
8. The sample size for the training set
Not applicable. There is no AI algorithm with a training set.
9. How the ground truth for the training set was established
Not applicable. There is no AI algorithm with a training set or ground truth in this context.
Ask a specific question about this device
Page 1 of 1