Search Results
Found 3 results
510(k) Data Aggregation
(469 days)
The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older.
The Mini Screws system has been designed to provide a firm anchorage point in the oral cavity for orthodontic treatment. Mini screws are manufactured with CP Ti Gr 4 and are SLA treated. The Mini Screw system is designed to be applied with different orthodontic solutions and is versatile enough to satisfy different clinical cases. Titanium orthodontic mini screws are intended to provide fixed anchorage for orthodontic treatment used by implanting in the maxillary and mandibular bone. These mini screws used for anchoring purposes can be divided into 3 parts as head, neck and body.
The provided text is an FDA 510(k) clearance letter and summary for a medical device called "Proimtech Mini Screw". This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets strict performance acceptance criteria through the kind of study typically associated with AI/ML software.
Therefore, the information required to answer the prompt regarding acceptance criteria and a study proving a device meets these criteria (especially sections related to AI/ML software performance like sample size, experts, MRMC, standalone performance, and ground truth establishment) is not present in the provided document.
The document discusses the physical characteristics, materials, and intended use of a dental mini-screw, and compares it to similar existing devices to show it is substantially equivalent, thus not requiring a new PMA (Premarket Approval Application). It does not involve an AI/ML component or a study with performance metrics in the way the prompt implies.
Here's how to interpret the available information in the context of the prompt:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria (Implicit for Substantial Equivalence): The implicit acceptance criterion for a 510(k) clearance is that the new device is substantially equivalent to a legally marketed predicate device. This means it has the same intended use, and the same technological characteristics, or if it has different technological characteristics, these differences do not raise new questions of safety and effectiveness.
- Reported Device Performance: The document states: "However, testing data such as mechanical testing provided in the submission show that these differences do not raise issues in performance." This very generally indicates that the device met certain mechanical performance benchmarks, but no specific quantitative acceptance criteria or results are provided in the public summary.
- The document also mentions: "The biocompatibility of our finished products has been carried out in accordance with EN 10993-1 Standard." and "Tests for the mechanical strength of our products have been carried out in accordance with TS EN ISO 19023 and ASTM F543 Standard." This indicates the types of tests performed for this physical device, but not the specific criteria or results.
2. Sample size used for the test set and the data provenance:
- Not applicable (N/A) / Not provided. The summary mentions "testing data" but does not detail sample sizes, data provenance (e.g., country of origin, retrospective/prospective), as it's a physical device, not an AI/ML software tested on a dataset of images/cases.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A / Not provided. This is relevant for AI/ML software that requires expert annotation for ground truth. For a physical medical device like a mini-screw, ground truth is established through material standards, mechanical testing protocols, and biocompatibility assessments, not human expert interpretation of data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- N/A / Not provided. This concept applies to human reader studies or data annotation for AI/ML, not a physical device's mechanical or material testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- N/A / Not provided. This is specific to AI/ML software. The device is a surgical implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- N/A / Not provided. This is specific to AI/ML software.
7. The type of ground truth used:
- For a physical device, "ground truth" would be established through engineering standards, material specifications (e.g., ISO 5832-2, ASTM F67-13 for material composition), mechanical testing protocols (e.g., TS EN ISO 19023, ASTM F543 for mechanical strength), and biocompatibility assessments (e.g., EN 10993-1). The document indicates these standards were followed.
8. The sample size for the training set:
- N/A / Not provided. There is no "training set" in the context of a physical device. This refers to AI/ML model development.
9. How the ground truth for the training set was established:
- N/A / Not provided. As above, no training set for a physical device.
In summary, the provided document details the regulatory clearance of a physical medical device (dental mini-screw) based on substantial equivalence, and therefore does not contain the information typically associated with acceptance criteria and studies for AI/ML software performance.
Ask a specific question about this device
(391 days)
The ACR Screw System is indicated for use as a fixed anchorage point for the attachment of orthodontic appliances to facilitate the orthodontic movement of teeth for use in patients 12 years of age and older. It is used temporarily and is removed after orthodontic treatment is complete. The screws are intended for single use only.
The ACR Screw System is indicated for use as a fixed anchorage point for the attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment is complete. The average temporary implantation period for the anchorage screw is six months. Screws are intended for single use only.
Screws are essentially C-type screw head design and size. Screw diameters and lengths are offered in the following range: thread diameter (1.75 mm) and overall length (11.10 – 15.10 mm). The screw tip has a self-drilling feature for insertion and removal. It is manufactured from a machined piece of Ti-6Al-4V ELI titanium alloy (ASTM F 136-13). These devices are supplied sterilized by gamma irradiation. This device is individually packaged in a polyethylene bag.
The provided document is a 510(k) premarket notification for the ACR Screw System, an orthodontic anchorage screw. It describes the device, its indications for use, and its substantial equivalence to predicate devices, primarily through non-clinical testing. However, it does not contain a study that quantitatively proves the device meets specific acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or similar measures typically associated with diagnostic or AI-driven devices.
The document focuses on demonstrating physical and material equivalence, as well as safety and basic functionality. Therefore, I cannot fully answer your request for acceptance criteria and a study proving device performance in the way you've outlined for AI/diagnostic devices.
Here's an attempt to extract relevant information given the limitations of the provided text:
1. A table of acceptance criteria and the reported device performance
The document details performance testing related to mechanical properties, biocompatibility, sterilization, bacterial endotoxin, and shelf-life. However, explicit quantitative acceptance criteria and detailed reported performance values for each are not presented in a comparative table within the document. Instead, it states that tests "met predetermined acceptance criteria" or "all test requirements were met."
Test Category | Acceptance Criteria (Stated broadly) | Reported Device Performance (Stated broadly) |
---|---|---|
Mechanical Properties | In accordance with ISO 19023:2018 and ASTM F543-17 objectives for pull-out, torsion, and torque. | Performance "determined" and implied to be acceptable for substantial equivalence. No specific values provided. |
Biocompatibility | In accordance with ISO 10993-1, ISO 10993-5, ISO 10993-12. | Deemed biocompatible due to use of same materials and manufacturing process as previously cleared predicate device (K063495). |
Sterilization | Achievement of a Sterility Assurance Level (SAL) of $10^{-6}$ and meeting all test requirements per ISO11137-1, -2, -3. | SAL of $10^{-6}$ was achieved, and all test requirements were met. |
Bacterial Endotoxin | Test limit of 0.06 EU/mL per ANSI/AAMI ST72:2011. | Testing met predetermined acceptance criteria. |
Shelf-Life | Confirm stability and effectiveness of packaging for 3 years (sterility, peel strength, dye penetration, burst) per ISO 11607-1, ISO 11607-2, ASTM F1980-21. | Device met the acceptance criteria for each test (sterility, peel strength, dye penetration, burst) for a 3-year shelf life. |
2. Sample size used for the test set and the data provenance
The document does not specify sample sizes for mechanical, sterilization, bacterial endotoxin, or shelf-life testing. The data provenance is implied to be from internal testing by BioMaterials Korea, Inc. or their contracted labs, primarily from South Korea, as indicated by the manufacturer's and consultant's addresses. All tests are non-clinical (laboratory/bench).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable to the non-clinical testing described. Ground truth for these types of tests is established by standardized testing protocols and measurement instruments, not human expert consensus.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this is for non-clinical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical orthodontic screw, not an AI or diagnostic imaging device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the non-clinical tests (mechanical properties, biocompatibility, sterilization, etc.) is established by adherence to international standards and validated laboratory test methods (e.g., ISO and ASTM standards for material properties, sterility, and packaging integrity).
8. The sample size for the training set
Not applicable. This device is a physical product, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(109 days)
The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
The screw diameters and lengths vary as shown below. It is made of a machined piece of Ti-6Al-4V ELI titanium alloy (ASTM F 136-13).
Structure (head type) | Thread diameter and Length | Remark |
---|---|---|
Double head | - Ø1.4mm, 6.0/8.0mm | Exceptionally S5T (Double head) and S5R (Slot) |
Button head | - Ø1.6mm, 6.0/8.0/10.0mm | are Ø1.8mm, 10.0 to 16.0mm per 1.0mm. |
Wing | - Ø1.8mm, 6.0/8.0/10.0mm | |
Slot | - Ø2.0mm, 6.0/8.0/10.0/12.0mm |
These size combinations are the same across all model types (Double Head, Wing, Slot) with exception case.
The provided document is an FDA 510(k) Premarket Notification for the "ORTHO MI SYSTEM," an endosseous dental implant (orthodontic anchor screw). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical effectiveness through extensive studies like those required for PMA (Premarket Approval).
Therefore, the document explicitly states: "No clinical data were necessary for the demonstration of substantial equivalence." This means there was no study performed to prove the device meets acceptance criteria in the way a clinical study for an AI/CADe device would.
Instead, acceptance criteria are met through bench testing demonstrating compliance with relevant performance and safety standards, and ensuring the device's technical characteristics and indications for use are substantially equivalent to marketed predicate devices.
Here's a breakdown of the missing and implied information based on the provided text:
Acceptance Criteria and Device Performance
Since no clinical study was conducted, there are no "acceptance criteria" in the sense of clinical performance metrics (like sensitivity, specificity, or reader improvement). The acceptance criteria are tied to bench test performance against established standards and the demonstration of substantial equivalence to predicate devices.
Acceptance Criteria (Implied from Bench Testing & Equivalence) | Reported Device Performance (from "Non-Clinical Test Conclusion") |
---|---|
Mechanical Performance: Compliance with ISO 19023:2018 (Dentistry Orthodontic anchor screws) and ASTM F543:2013 (Metallic medical bone screws). | "The test results demonstrated that the subject device complies with the following standards: ISO 19023: 2018, ASTM F543: 2013." |
Sterilization, Shelf-life, and Packaging: Compliance with various ISO and ASTM standards related to radiation sterilization, microbiological methods, packaging, and accelerated aging. | "The test results demonstrated that the subject device complies with... ISO11137-1/2/3, ISO11737-1/2, ISO 11607-1/2, ASTM F1980, ASTM F88/F88M, ASTM F1140/F1140M, ASTM F2096, ASTM F1929, ASTM D882, ASTM F1886F1886M." |
Sterilization for Non-Sterile Product: Compliance with ANSI/AAMI ST79:2017 and ISO 17665-1/2. | "The test results demonstrated that the subject device complies with... ANSI/AAMI ST79:2017, ISO 17665-1/2." |
Biological Safety: Compliance with ISO 10993-1:2018 and FDA Guidance. | "Biological assessment has been performed according to ISO 10993-1:2018... and to the FDA Guidance document." |
Substantial Equivalence: Material, Indications for Use, Operating Principles, Form, and Structure are similar to predicate device(s). Product Size, Sterilization methods, Single Use/Reuse, Packaging, and Shelf-life are comparable or within acceptable variations. | "The technological characteristics of subject device is same materials and has similar dimension as the predicate and/or reference device. The Indication for use of subject device is same as the predicate device." "The result of the comparative bench testing... demonstrate that all screws are similar performance with predicate and/or reference device." |
Study Information (as requested, but largely N/A for this 510(k) submission)
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. No clinical test set was used. The "test set" for this submission would refer to the physical devices subjected to bench testing according to the specified standards. The number of devices tested is not detailed, but it would have been sufficient to meet the requirements of the standards cited.
- Data Provenance: Not applicable. No biological or clinical data was collected from patients or subjects. The data derived from physical testing of the device hardware.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts/Qualifications: Not applicable. There was no ground truth determination by experts in a clinical context. Bench testing relies on objective measurements against engineering and materials standards.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Adjudication Method: Not applicable. No human interpretation requiring adjudication was involved.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This device is an orthodontic anchor screw, a physical implant, not an AI/CADe system. Therefore, no MRMC study was performed.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Performance: Not applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Type of Ground Truth: Not applicable in a clinical sense. The "ground truth" for this device's performance is adherence to established engineering, materials, sterilization, and biocompatibility standards, and demonstrating physical properties consistent with predicates.
-
The sample size for the training set:
- Training Set Sample Size: Not applicable. This is a physical medical device, not a machine learning algorithm requiring a training set.
-
How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable.
In summary, the provided document details a 510(k) submission for a physical medical device (orthodontic anchor screw). These submissions do not typically involve clinical trials or studies like those conducted for AI/CADe devices, and therefore, most of the questions regarding clinical performance metrics, expert adjudication, or AI model training are not applicable. The "proof" of meeting acceptance criteria lies in comprehensive non-clinical (bench) testing demonstrating compliance with relevant consensus standards and establishing substantial equivalence to previously cleared predicate devices.
Ask a specific question about this device
Page 1 of 1