Search Results
Found 1 results
510(k) Data Aggregation
(229 days)
The TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-12® disinfectant, which is designed to be used only with the TD 200® disinfector. The disinfectant bottles cannot be reused in the system.
TD-12® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 200® disinfector for high-level disinfection of TEE ultrasound probes.
TD-12® high level disinfectant and TD 200® disinfector is intended for use by qualified individuals trained in its use.
TD-12® disinfectant should be used with the following contact conditions in TD 200® disinfector:
High-level disinfectant: TD-12®
Time: 3 minutes
Temperature: 38°C
Minimum Recommended Concentration: 1.75% peracetic acid
The TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant provides high-level disinfection of transesophageal (TEE) ultrasound probes when used according to the operating instructions and when used with TD-12® disinfectant. The TD 200® disinfector is for use only with TD-12® disinfectant. The TD-12® is for use only in the TD 200® disinfector. Thus, the TD 200® disinfector and TD-12® disinfectant represent a dedicated system. Each soiled TEE probe is bedside cleaned and manually cleaned according to the TEE probe manufacturer's instructions before insertion into the TD 200® disinfector. A fresh, unopened bottle of TD-12® disinfectant is loaded into the TD 200® disinfector. The TD 200® disinfector heats the TD-12® disinfectant to at least 38 C, soaks the TEE probe at least three (3) minutes, and then thoroughly rinses the disinfectant off the TEE probe before the cycle is complete. The TD 200® disinfector prints a verification report indicating a successful disinfection cycle, the time and the average temperature during the disinfection. The TEE probe is then removed from the TD 200 disinfector and dried according to the TEE probe manufacturer's instructions. The TD 200 disinfector is ready for a new cycle immediately after the preceding cycle is completed. A chemical indicator is used to ensure that the solution is above the MRC.
The provided document describes the FDA 510(k) clearance for the TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant. This device is a high-level disinfection system for transesophageal (TEE) ultrasound probes. The acceptance criteria and the study that proves the device meets them are detailed primarily through non-clinical performance testing rather than human-in-the-loop studies involving expert readers.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are demonstrated through its ability to achieve high-level disinfection, stability of the disinfectant, and safety aspects. The performance is reported as a "complete kill of microorganisms" or "measurable complete kill."
Acceptance Criteria Category | Specific Criteria/Test | Reported Device Performance (TD 200® with TD-12®) |
---|---|---|
High-Level Disinfection Efficacy | Bactericidal Activity: | |
Staphylococcus aureus | 3.0 min at 37°C (Complete Kill/Total Kill) | |
Salmonella enterica | 3.0 min at 37°C (Complete Kill/Total Kill) | |
Pseudomonas aeruginosa | 3.0 min at 37°C (Complete Kill/Total Kill) | |
Tuberculocidal Activity: | ||
Mycobacterium terrae | 3.0 min at 37°C (Complete Kill/Total Kill) | |
Fungicidal Activity: | ||
Trichophyton mentagrophytes | 3.0 min at 37°C (Complete Kill/Total Kill) | |
Sporicidal Activity: | ||
Bacillus subtilis | 5.0 hrs at 37°C (Complete Kill/Total Kill) | |
Clostridium sporogenes | 5.0 hrs at 37°C (Complete Kill/Total Kill) | |
Virucidal Activity: | ||
Poliovirus Type 1 | 3.0 min at 37°C (Complete Inactivation) | |
Herpes Simplex Virus Type 1 | 3.0 min at 37°C (Complete Inactivation) | |
Human Influenza Virus A | 3.0 min at 37°C (Reported) | |
Adenovirus Type 1 | 3.0 min at 37°C (Reported) | |
Simulated Use Performance | Efficacy in automated system under simulated conditions | In all cases, a measurably complete kill of microorganisms after TEE probe processing. |
Clinical In-Use Testing | Efficacy with soiled TEE probes after routine clinical use | In all cases, there was a complete kill of microorganisms after TEE probe processing. |
Biocompatibility | Non-sensitizer, non-irritant, non-cytotoxic, and safe residue levels | TD-12® is not a sensitizer, nor an irritant and is non-cytotoxic. Residue levels on TEE probes are not likely to have toxic effects on humans. |
Material Compatibility | No observable effects on materials from disinfectant exposure | No observable effects from exposure to TD-12® disinfectant on materials. Material compatibility performed with GE, Philips, and Siemens TEE probes showed no cosmetic or functional deterioration. |
Disinfectant Stability | Shelf life of disinfectant | Stable for a shelf life of at least twelve (12) months. |
Electrical Safety | Compliance with electrical safety standards | Passed all testing in accordance with UL 61010-1. 3rd Edition. |
Software Validation | Consistent performance as intended | All qualification activities successfully completed, consistently performs as intended. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify a precise numerical sample size for "Clinical In-Use Testing" or "Automated Simulated use studies." It states "In all cases" there was a complete kill, implying all tested units/probes/cycles met the criteria. For the specific test organisms, the results are presented for each organism type (e.g., Bacillus subtilis, Mycobacterium terrae).
- Data Provenance: The document does not explicitly state the country of origin for the data or whether the studies were retrospective or prospective. Given it's an FDA submission, the implication is that the studies were conducted under controlled conditions to support regulatory clearance, likely prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This device focuses on disinfection efficacy rather than diagnostic accuracy involving human image interpretation. Therefore, the "ground truth" is established through microbiological and chemical analysis, not expert human readers.
- Experts: Not applicable in the context of human interpretation. The "experts" would be qualified microbiologists, chemists, and engineers conducting the tests according to established standards.
- Qualifications: Not specified for individuals, but implied by the nature of regulatory testing (e.g., adherence to Good Laboratory Practices or similar standards).
4. Adjudication Method for the Test Set
Not applicable. The "ground truth" for disinfection efficacy is determined by objective laboratory measurements (e.g., microbial counts, chemical residue analysis), not subjective human judgment requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a high-level disinfector, not an AI-powered diagnostic imaging tool for human readers. There are no "human readers" involved in the direct purpose of the device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The entire performance evaluation of the TD 200® Disinfector is "standalone" in the sense that its primary function (disinfection) is automated and assessed directly for its output (microbial kill, residue levels). There is no "human-in-the-loop" performance component for the disinfection process itself. The "Software Validation" section details the functioning of the algorithm/software controlling the device.
7. The Type of Ground Truth Used
The ground truth used is primarily microbiological efficacy (complete kill of specified microorganisms for bactericidal, tuberculocidal, fungicidal, sporicidal, and virucidal activity), supported by chemical analysis (disinfectant residue levels, stability) and mechanical/electrical performance (software validation, electrical safety, material compatibility). No pathology reports or outcomes data on patients were used; the focus is on the device's ability to disinfect a medical instrument.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/Machine Learning device that undergoes a "training phase" with a dataset in the typical sense. Its design and performance are based on chemical, microbiological, and engineering principles rather than learning from data.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for an AI model. The "ground truth" for the device's development would be established through scientific literature, industry standards (e.g., for disinfection efficacy), and internal research and development.
Ask a specific question about this device
Page 1 of 1