Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202618
    Date Cleared
    2021-09-24

    (379 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K170595, K153293

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IV Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. Extension Sets may be used for direct injection, intermittent infusion, continuous infusion or aspiration of fluids and medications. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

    Device Description

    IV Administration Sets are gravity, single use, disposable, IV sets and extension sets. IV Administration Sets are used to deliver fluids from a container into a patient's vascular system. Extension sets are connected to primary IV administration sets to add length and provide clamping capabilities or added to an intravascular catheter hub as a conduit for flow to and from the catheter and are used for direct injection, intermittent infusion, continuous infusion or aspiration of fluids and medications. These sets may be comprised of various components which are broadly used throughout industry including insertion spike, drip chamber, clamp, luer access device, check valve, tubing, luer connections (connector/adaptor), y-connector, and flow regulator.

    AI/ML Overview

    This is a 510(k) summary for IV Administration Sets by B. Braun Medical Inc. The document discusses the substantial equivalence of their proposed device (K202618) to two predicate devices (K170595 and K153293). It details the non-clinical performance testing conducted.

    Here's an analysis of the provided text in response to your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" alongside specific quantitative performance metrics as typically seen in device efficacy studies (e.g., Sensitivity, Specificity, Accuracy for diagnostic devices). Instead, it lists various non-clinical tests performed to demonstrate substantial equivalence to predicate devices and conformity to established standards. The "performance" is reported as a conclusion that the tests were "successfully completed" or that "verification results confirm that the differences in the set components do not raise new or different questions of safety and effectiveness."

    Here's a compilation of the tests and the implicit performance:

    Acceptance Criterion (Implied Standard/Test)Reported Device Performance
    IV Administration Sets (Proposed K202618 vs. Predicate K170595)
    Priming Test / Air VisualizationVerification results confirm no new safety/effectiveness concerns
    Flow Rate and OcclusionVerification results confirm no new safety/effectiveness concerns
    Positive PressureVerification results confirm no new safety/effectiveness concerns
    Negative Pressure (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Dynamic TensileVerification results confirm no new safety/effectiveness concerns
    Static TensileVerification results confirm no new safety/effectiveness concerns
    Leakage (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Clamp Pressure Resistance (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Slide Clamp Pressure (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Stability and Accuracy (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Particulate Contamination (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Microbial Ingress (Additional test performed for proposed device)Verification results confirm no new safety/effectiveness concerns
    Y-Connector Injection Site Assembly Pressure and Reseal (Additional test)Verification results confirm no new safety/effectiveness concerns
    Biocompatibility (Conformity to ISO 10993-1)Conformed to ISO 10993-1 (Same as predicate)
    Sterilization (Ethylene Oxide, SAL 10-6)Ethylene Oxide, SAL 10-6 (Same as predicate)
    For Components with Flow Regulators (Applicable to proposed device)
    Flow Rate Accuracy Test (ISO 8536-13:2016)Successfully completed
    Flow Rate Stability Test (ISO 8536-13:2016)Successfully completed
    Positive Pressure Test of Rate Control Device at Various Dial Settings (ISO 8536-13:2016)Successfully completed
    Negative Pressure Test of Rate Control Device at Various Dial Settings (ISO 8536-13:2016)Successfully completed
    For Clamps and Flow Regulators (Applicable to proposed device)
    Clamp Pressure Resistance (ISO 8536-14:2016)Successfully completed
    Slide Clamp Pressure (ISO 8536-14:2016)Successfully completed
    General Tests
    Particulate Matter in Injections (USP)Successfully completed

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the sample sizes used for the non-clinical tests. Non-clinical studies for devices like IV administration sets typically involve testing a statistically significant number of units to ensure reliability, but the exact numbers are not disclosed in this summary.

    The data provenance is from non-clinical performance testing conducted by B. Braun Medical Inc. The studies are by nature prospective as they are specifically conducted to support the 510(k) submission. There is no information regarding the country of origin of the data beyond the manufacturer being B. Braun Medical Inc. (located in Allentown, Pennsylvania, USA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable in the context of this 510(k) summary. The ground truth for the non-clinical performance tests is based on objective measurements against established engineering standards (e.g., ISO, USP) and internal B. Braun procedures, not on expert consensus or interpretation.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in diagnoses or assessments. The tests described here are physical and mechanical performance tests with objective pass/fail criteria based on standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done. This device is an IV administration set, a physical medical device, not an AI-powered diagnostic or assistive tool for human readers. Therefore, the concept of human readers improving with or without AI assistance is irrelevant here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    There was no standalone (algorithm only) performance study done. As mentioned, this is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance tests is derived from established industry standards (ISO, USP) and internal B. Braun procedures, which outline specific test methodologies, parameters, and pass/fail criteria. For instance, for flow rate, the ground truth would be the expected flow rate range defined by the standard under specific conditions. For tensile strength, it would be the minimum force the material must withstand.

    8. The Sample Size for the Training Set

    This is not applicable. This is a physical medical device, and the evaluation involves non-clinical performance testing, not machine learning or AI that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1