Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190560
    Date Cleared
    2019-05-02

    (58 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K102715, K151305

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model 8006CA reusable, non-sterile, regional oximetry sensor is intended for use as an absolute realtime adjunct monitor of regional hemoglobin oxygen saturation (rSO2) of blood underneath the sensor of adult and pediatric patients weighing ≥ 88 pounds (≥ 40 kilograms). The sensor and/or the patient interface may be repositioned or replaced with another 8006CA sensor without baseline re-establishment.

    Nonin's Model 8006PI single patient use, non-sterile disposable patient interface is designed for use with Nonin's Model 8006CA sensor.

    The sensor and patient interface is in hospitals, long-term care, medical facilities, sleep laboratories, and subacute environments.

    Device Description

    The Model 8006CA Reusable Regional Oximetry Sensor and Model 8006PI Disposable Patient Interface work together for use with the Nonin Medical SenSmart Model X-100 Universal Oximetry System (Model X-100).

    AI/ML Overview

    The Nonin Medical, Inc. Model 8006CA Reusable Regional Oximetry Sensor and Model 8006PI Disposable Patient Interface have undergone testing to establish substantial equivalence to predicate devices (Model 8004CA and Model 8204CA Regional Oximetry Sensors).

    Here is a summary of the acceptance criteria and performance based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Predicate Device)Reported Device Performance (Model 8006CA/8006PI)
    Functional and Safety Testing
    Electrical SafetyComplies with IEC 60601-1Pass
    Temperature and HumidityComplies with IEC 60601-1, EN 1789, IEC 60601-1-12Pass
    Atmospheric Pressure (Altitude)Complies with IEC 60601-1, IEC 60601-1-12Pass
    Electromagnetic Immunity and EmissionsComplies with IEC 60601-1-2Pass
    PerformanceComplies with ISO 80601-2-61, IEC 60601-1, IEC 60601-1-6, IEC 60601-1-12, IEC 62304, ANSI/AAMI EC13, ISO 14155Pass
    Ingress ProtectionComplies with IEC 60601-1, ISO 80601-2-61Pass
    Mechanical DurabilityComplies with IEC 60601-1, ISO 80601-2-61Pass
    BiocompatibilityComplies with ISO 10993-1Biocompatible
    Regional (rSO2) Accuracy
    rSO2 Accuracy (Arms)Absolute: Right 4.1, Left 3.8, Both 3.9, Hypercapnia 5.1, Hypocapnia 3.3 (from K102715)Not explicitly re-tested, but "critical sensor optics technology... remains unchanged" and "prior clinical testing... is applicable." It "performs equivalently to the same specifications."
    Declared range50 - 100% (from K102715)"Performs equivalently to the same specifications."
    Inter-/Intra-sensor repeatability± 2 digits (from K102715)"Performs equivalently to the same specifications."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "prior clinical testing provided in K102715 is applicable for the proposed Model 8006CA sensor and Model 8006PI patient interface" because "The critical sensor optics technology of the proposed devices remains unchanged from the predicate devices." Therefore, the clinical test set data for Regional (rSO2) Accuracy testing is from the K102715 submission for the predicate Model 8004CA Sensor. The specific details of the sample size, data provenance (country of origin, retrospective/prospective), etc., for that original clinical study are not provided in this document (K190560).

    The current submission (K190560) relies on "detailed device comparison, analysis and testing" and the equivalency of the critical sensor optics.

    For the Functional and Safety Testing, the tests were performed on the proposed devices (Model 8006CA and Model 8006PI) and are likely prospective tests conducted specifically for this submission. The sample sizes for these engineering tests are not specified but are typically small (e.g., a few units) sufficient to demonstrate compliance with standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The regional oximetry accuracy is derived from clinical testing, but the specifics of expert involvement in establishing ground truth (e.g., from an arterial blood gas reference for SpO2) are not detailed within this 510(k) summary. It refers back to a previous submission (K102715).

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method for the test set.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The device is an oximeter, which provides a direct measurement, not an interpretative tool that requires human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the functional, safety, and performance tests assess the standalone performance of the device (Model 8006CA/8006PI) in conjunction with the Nonin Medical SenSmart Model X-100 Universal Oximetry System. The regional oximetry accuracy cited from the predicate device's clinical testing also represents standalone algorithm performance in measuring rSO2.

    7. The Type of Ground Truth Used

    For the regional oximetry accuracy (rSO2), the ground truth typically involves a reference method for measuring regional oxygen saturation, often arterial blood gas analysis or other established physiological measurements, in a controlled hypoxemic study. However, the specific type of ground truth used for the predicate device's clinical testing (K102715) is not explicitly stated in this document. Given the nature of oximetry, it would likely be a direct physiological measurement.

    8. The Sample Size for the Training Set

    The document does not mention a training set sample size. Since the validation relies on the "critical sensor optics technology... remains unchanged" from predicate devices and the device is a measurement sensor rather than an AI/machine learning algorithm that requires extensive training data, the concept of a "training set" in the context of machine learning is not directly applicable here.

    9. How the Ground Truth for the Training Set was Established

    As no training set is discussed or implied for an AI/ML model, this information is not applicable here. The device's performance is based on its physical and optical design, not on learned patterns from a dataset.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1