Search Results
Found 2 results
510(k) Data Aggregation
(115 days)
The iASSIST Knee System is a computer assisted stereotaxic surgical instrument system intended to assist the surgeon in preparing the bone surfaces for the positioning of orthopedic implant system components intra-operatively. It involves surgical instruments and inertial sensors to determine alignment axes in relation to anatomical landmarks and to precisely position alignment instruments and cut guides relative to these axes.
The present iASSIST Knee System is designed for Total Knee Arthroplasty.
As in the predicates, the iASSIST Knee System consists of Pods (tracking sensors), a computer system, software, and surgical instruments designed to assist the surgeon in the placement of Total Knee Replacement components. The Pods combined with the surgical instruments provide positional information to help orient and locate the main femoral and tibial cutting planes as required in knee replacement surgery. This includes means for the surgeon to determine and thereafter track each of the bones' alignment axes relative to which the cutting planes are set. The computer system and software components control and sequence the functions of the Pods per the applicable knee surgery steps via wireless communication.
The provided text does not contain information about specific acceptance criteria or an associated study for the iASSIST Knee System that would allow me to populate all the requested fields. The document primarily describes the regulatory submission (510(k)) and its general performance testing, but lacks the detailed quantitative data, sample sizes, and expert information typically found in a study proving acceptance criteria.
However, based on the information available, I can extract the following:
1. Table of Acceptance Criteria and Reported Device Performance
Not available in the provided text. The text generally states that the device "is safe and effective and that the performances of the iASSIST Knee System are acceptable," but does not list specific quantitative criteria or performance metrics (e.g., accuracy, precision) with corresponding reported values.
2. Sample Size Used for the Test Set and Data Provenance
The text mentions "full simulated use on cadaveric specimens" for the Validation Lab. It does not specify the number of cadaveric specimens used (the sample size) or their country of origin. The study appears to be prospective in nature, as it involves actual testing on specimens.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not available in the provided text.
4. Adjudication Method for the Test Set
Not available in the provided text.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The iASSIST Knee System is described as a computer-assisted surgical instrument system, not an AI diagnostic or interpretive tool that would involve "human readers" or AI assistance in the sense of image interpretation for diagnosis. Its purpose is to assist the surgeon in bone surface preparation and positioning of orthopedic implants.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device is inherently designed for "human-in-the-loop" performance, assisting a surgeon intra-operatively. Therefore, a standalone performance study in the absence of a surgeon would not be relevant or meaningful for this device's intended use. The validation lab involves "full simulated use on cadaveric specimens," implying the system is used as intended with human involvement.
7. The Type of Ground Truth Used
For the "Validation Lab" testing on cadaveric specimens, the ground truth would likely be established through:
- Direct measurement: Using highly accurate external instruments or established anatomical reference points to measure the actual alignment axes and cutting planes achieved by the system and surgical instruments on the cadaver bones.
- Expert surgical assessment: Evaluation by experienced orthopedic surgeons to determine if the achieved bone preparation and alignment match the surgical plan and accepted orthopedic standards.
The text does not explicitly state which method was used, but given the nature of the device, it would involve objective measurements against defined surgical parameters.
8. The Sample Size for the Training Set
Not applicable. The iASSIST Knee System is a stereotaxic surgical instrument system, not a machine learning model that requires a distinct "training set" in the conventional sense of AI/ML development. The software capabilities are likely developed and refined through engineering, modeling, and iterative testing, rather than being "trained" on a large dataset of patient images or outcomes.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as described in point 8. The "ground truth" for the development of such a system would involve engineering specifications, biomechanical principles, and established surgical techniques for total knee arthroplasty.
Ask a specific question about this device
(351 days)
PERSEUS is a computer-controlled system, intended to assist in distal femoral resection and tibial resection during Total Knee Arthroplasty, determining reference alignment axes in relation to anatomical and instrumentation structures during stereotaxic orthopaedic surgical procedures.
PERSEUS is a computer assisted system, that helps surgeon in the positioning of implant components, during total knee implant surgical procedure, according to the conventional reference axes in relation to anatomical landmarks. Perseus System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format. The device assists the surgeon in: Establishing the mechanical axis of the femur, determining the varus/valgus angle and the flexion/extension angle of the cutting block relative to the femur. Establishing the mechanical axis of the tibia, determining the varus/valgus angle and the posterior slope angle of the cutting block relative to the tibia. Perseus system utilizes triaxial accelerometer and triaxial gyroscope to determine, through limb movement, reference axes of femur and tibia and relative orientation of distal femoral and tibial cutting guide with respect to frontal and sagittal planes of the limb.
Here's a detailed breakdown of the acceptance criteria and the studies proving the device's performance, based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
| Measure | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| HKA (Hip-Knee-Ankle) angle deviation | within 3° (for final leg alignment) | Average post-operative HKA angle deviation: 0.3 ± 1.5° |
| Average post-operative HKA alignment deviation: 0.7° (range 0-2°) | ||
| Femoral distal resection deviation | within 2° (for distal femoral resection, frontal and lateral planes) | Femoral distal resection deviation: < 2° (frontal and lateral planes) |
| Femoral implant flexion | Not explicitly stated | Average femoral implant flexion: 0.8 ± 0.8° |
| Tibial slope | Not explicitly stated | Average tibial slope: 0.1 ± 1.7° |
| Blood loss | Reduced blood loss compared to conventional technique | Confirmed reduced blood loss compared to conventional technique |
2. Sample Sizes Used for Test Sets and Data Provenance
The document mentions several studies, primarily clinical evaluations. The provenance is Europe (aftermarket clinical evaluation). All studies appear to be prospective clinical evaluations or retrospective analysis of post-operative data.
- Study 1 (Aftermarket Clinical Evaluation - Europe):
- Test Set Sample Size: 3 cohorts of 10 patients each (total 30 patients)
- EM Perseus: 10 patients
- EM Nav: 10 patients (surgical navigation system)
- IM Conv: 10 patients (conventional mechanical instrumentations)
- Data Provenance: Europe (aftermarket clinical evaluation), the specific country is not mentioned. Data is from patients operated by the same surgical team.
- Test Set Sample Size: 3 cohorts of 10 patients each (total 30 patients)
- Study 2 (Aftermarket Clinical Evaluation - Europe):
- Test Set Sample Size: 2 cohorts of 18 patients each (total 36 patients)
- Perseus: 18 patients
- Conventional instrumentation: 18 patients
- Data Provenance: Europe, multicentric study involving two centers.
- Test Set Sample Size: 2 cohorts of 18 patients each (total 36 patients)
- Study 3 (Aftermarket Clinical Evaluation - Europe):
- Test Set Sample Size: Single cohort of 20 patients.
- Data Provenance: Europe.
- Study 4 (Retrospective analysis of post-operative X-rays):
- Test Set Sample Size: 20 cases.
- Data Provenance: Not explicitly stated, but likely from European clinical practice given the context of aftermarket clinical evaluations.
- Study 5 (Retrospective analysis of post-operative X-rays):
- Test Set Sample Size: 53 patients operated with Perseus vs. 52 patients operated with conventional technique (total 105 patients).
- Data Provenance: Not explicitly stated, but likely from European clinical practice.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not specify the number of experts, their qualifications, or their role in establishing ground truth for the test sets (patient data). It mentions that "Clinical evaluation has been done with the same performance requirement when possible or, in alternative, looking at final leg alignment with post-operative x-rays." This suggests that the ground truth for clinical outcomes like HKA angle, femoral implant flexion, and tibial slope was derived from post-operative X-rays, which are generally interpreted by qualified medical professionals (e.g., orthopedic surgeons or radiologists), but the specific details are not provided.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method (like 2+1 or 3+1) for the clinical outcomes or the interpretation of post-operative X-rays. The analysis of X-rays for angular deviations is likely based on standard radiographic measurement techniques.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study is explicitly described. The clinical studies compare patient outcomes (e.g., HKA alignment, blood loss) between those treated with Perseus and those with conventional techniques or other navigation systems, not specifically focusing on how human readers improve with vs. without AI assistance in interpretation.
6. Standalone (Algorithm Only) Performance Study
Yes, standalone performance was evaluated through bench tests and pre-clinical evaluation testing:
- Bench tests: Conducted to quantify device accuracy and repeatability. These involved controlled setups to verify:
- Measurement repeatability with different cursor positions.
- Error introduced by mispositioning of ankle instrumentation.
- Repeatability with different sensors.
- Repeatability with different instrument positioning.
- Pre-clinical evaluation testing: Verified accuracy, repeatability, and reproducibility in a simulated OR setup and consistency of the procedure with different users/experience levels.
These tests evaluate the algorithm and device's performance in a controlled environment, separate from human-in-the-loop clinical scenarios.
7. Type of Ground Truth Used
- Bench Tests & Pre-clinical Evaluation: Ground truth was established using controlled variables within appropriate test setups, likely using precise measurement tools, and simulated surgical scenarios with known desired outcomes.
- Clinical Studies: The primary ground truth for measuring clinical outcomes (e.g., HKA angle, femoral distal resection deviation, tibial slope) was derived from post-operative X-rays. Blood loss was also measured as an outcome.
8. Sample Size for the Training Set
The document does not provide details on a specific "training set" for the PERSEUS device. PERSEUS is described as a "computer-controlled system" using inertial sensors and physical positions for positional information. While such systems are developed based on algorithms, the document focuses on the validation studies rather than the development phase's data sets. Therefore, the sample size for an explicit training set is not reported.
9. How the Ground Truth for the Training Set Was Established
Since no specific training set is mentioned in the document, the establishment of ground truth for a training set is not described. The device's operation, based on accelerometers and gyroscopes, suggests it relies on engineering principles and sensor data processing rather than learning from a labeled training dataset in the way a typical AI/ML system for image classification might. The "performance tests" and "bench tests" served to validate the system's accuracy against established physical measurement standards and simulated scenarios.
Ask a specific question about this device
Page 1 of 1