Found 2 results

510(k) Data Aggregation

    K Number
    K220955
    Date Cleared
    2023-06-23

    (448 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K141214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The non-prefilled nebulizer is indicated for use when humidity needs to the air flow going to a patient in the form of an aerosol. The large volume nebulizer also allows for the provision of humidified oxygen at select oxygen concentrations from 28% to 98%. This product is for single patient use and is designed to be used in hospital, nursing homes, extended care facilities, outpatient clinics as prescribed by a healthcare professional.

    Device Description

    The large volume nebulizer is a non-prefilled reservoir nebulizer for supplying humidity for inhalation therapy. The device features a wing nut style connector that fits standard flow metered medical gas sources and includes a 500mL capacity jar with minimum and maximum fill lines. Large volume nebulize an internal venturi nozzle to draw the solution up from the jar through a small plastic pickup tube and into the gas stream to be aerosolized. A rotating collar sets the delivered oxygen concentration by controlling the size of the room air opening around the venturi. A bull-nose style output connector is used to connect 22mm aerosol tubing for delivery to the patient.

    AI/ML Overview

    This document is a 510(k) Summary for the Hudson RCI Variable Concentration Large Volume Nebulizer, a medical device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving device performance against specific acceptance criteria for a new and innovative AI-powered product.

    Therefore, the requested information categories related to AI/algorithm performance, such as sample size for test sets and training sets, number of experts for ground truth, adjudication methods, MRMC studies, and standalone algorithm performance, are not applicable to this document. This document describes a traditional medical device, not an AI/ML-driven one.

    Here's a response tailored to the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding reported device performance values in the manner typically seen for novel AI systems with quantitative metrics for accuracy, sensitivity, specificity, etc. Instead, it states that tests were performed to demonstrate safety based on current industry standards and concludes that the device is substantially equivalent to predicate devices. The "performance" here refers to meeting established safety and functional requirements rather than predictive accuracy.

    Here's an interpretation based on the provided text, outlining the tests performed and the general "acceptance" that they indicate substantial equivalence:

    Test (Acceptance Criteria Implicitly Met)Reported Device Performance (Implied)
    Biocompatibility (ISO 10993-1, -5, -10, -11, -18)Compliant with relevant ISO standards for external communicating, prolonged/permanent contact devices that indirectly contact tissue/bone/dentin, and indirect gas pathway.
    PackagingSatisfactory for maintaining device integrity and sterility/cleanliness.
    Environmental Conditioning (high and low humidity)Withstood environmental conditions without adverse impact on performance.
    AgingDemonstrated satisfactory performance over its useful life.
    oxygen entrainmentAchieved specified oxygen concentrations (FIO2 28-98%)
    Lift testingPerformed as intended during operation, likely referring to mechanical stability or water lift.
    Humidity outputProduced adequate and consistent humidity output.
    Useful life testingMaintained performance throughout its intended useful life.
    Cleaning processDemonstrated appropriate cleanability or non-cleanability (given it's single-use).
    Comparison to predicate devices (K041418, K141214)Demonstrated similar design, dimensions, materials, intended use, and technological characteristics, leading to a conclusion of substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is typically found in detailed study reports or clinical trial summaries, which are not part of this 510(k) Summary. The document describes pre-clinical (bench) testing, not human subject data. Therefore, concepts like "test set," "data provenance," "country of origin," and "retrospective/prospective" studies as they apply to clinical data are not present. The "tests were performed" implies an internal engineering and quality assurance process.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. "Ground truth" in the context of expert consensus is relevant for diagnostic algorithms or subjective assessments. For a physical device like a nebulizer, the "ground truth" is typically established by objective measurements against engineering specifications and validated test methods overseen by qualified engineers and technicians. The document does not specify the number or qualifications of individuals involved in setting or verifying these engineering standards or conducting the tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., in medical image analysis). For the physical and chemical testing of a nebulizer, the "adjudication" is through meeting predefined analytical standards and specifications, not through expert consensus on subjective findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are specific to evaluating diagnostic technologies, often AI-driven, where human readers interpret cases. This document is for a physical medical device (nebulizer) and does not involve human "readers" or an AI component.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm and does not have a "standalone" or "human-in-the-loop" mode for performance evaluation in the context of AI. It is a mechanical device that delivers aerosolized humidity.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this type of device, the "ground truth" is based on:

    • Engineering specifications and standards: Performance metrics like oxygen entrainment percentages, humidity output, and mechanical integrity are objectively measured against predefined engineering targets.
    • Material compatibility standards: Biocompatibility is assessed against ISO 10993 series standards.
    • Safety standards: General device safety is evaluated against recognized industry standards and regulatory requirements.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this device is not an AI/ML model that undergoes a training process.

    9. How the ground truth for the training set was established

    Not applicable. As there is no AI/ML model, there is no "training set" or ground truth established for it. The standards and specifications are established through medical device development processes and regulatory requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153010
    Date Cleared
    2016-07-26

    (286 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K141214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hudson RCI® AquaPak® Sterile Prefilled Nebulizer adds sterile water or saline solution in aerosol form to a patient's breathing gases. It may be used with pediatric (ages 2 years and above) and adult patients.

    Device Description

    Hudson RCI® AquaPak® Sterile Prefilled Nebulizers provide sterile water or sterile saline for inhalation therapy. Nebulizers generate aerosol, a fine mist of liquid water (or sodium chloride solution) that is suspended in the gas to be inhaled by the patient during aerosol therapy.

    The Hudson RCI® AquaPak® Sterile Prefilled Nebulizers consist of two components, a sterile adaptor and a prefilled sterile reservoir. The adaptor may be type 028 or type 033. The standard model with a yellow collar (type 028) and the quiet model with a blue collar (type 033) feature an adjustable air entrainment window, which provides a specific oxygen concentration by entraining room air into the oxygen stream. Prefilled sterile reservoirs for the Hudson RCI® AquaPak® Sterile Prefilled Nebulizers come in three sizes; 440ml, 760ml and 1070ml, and may be sterile water or sterile sodium chloride solution. The sterile reservoirs are cleared under K141214, and are presented here for reference only. Each reservoir, when combined with an adaptor component, provides nebulizer functionality when connected to a gas source.

    The primary purpose of this Traditional 510(k) submission is to change the nebulizer adaptor cleared in K141214 from non-sterile, and update the packaging materials and shelf life to maintain sterility.

    AI/ML Overview

    The document describes a 510(k) summary for the Hudson RCI® AquaPak® Sterile Prefilled Nebulizers. This is a medical device submission, and the content provided outlines the device, its intended use, and the testing performed to demonstrate its substantial equivalence to a predicate device.

    Here's an analysis of the provided text in the context of your request for acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table outlining the tests performed and a general statement about performance. Specific quantitative acceptance criteria are not explicitly listed in a separate column, but the "Principle of Test" column often hints at what was being measured. The reported device performance for these tests is generally stated as "The device passed all of the acceptance criteria for the tests" or "The packaging passed all of the acceptance criteria for the tests."

    Test PerformedImplied Acceptance Criteria (from Principle of Test)Reported Device Performance
    Packaging Configuration / Shelf life TestingEstablish integrity of the package materials and sterile barrier after shelf life testing, visual inspections and nebulization function test. (ASTM F1980-07, ASTM F2096)"The packaging passed all of the acceptance criteria for the tests."
    Sterility TestingValidates the correct dose of gamma radiation to ensure a sterile product at the point of use when manufactured, stored, and handled according to the instructions for use. (ISO 11137-1:2006, ISO 11137-2:2013)"The packaging passed all of the acceptance criteria for the tests."
    Lift PressureMeasures the amount of negative pressure required to pull liquid up for nebulization."The device passed all of the acceptance criteria for the tests."
    Oxygen EntrainmentMeasures the percentage of oxygen/air mixture when adjusted to preset oxygen percentage."The device passed all of the acceptance criteria for the tests."
    Nebulization RateDetermines the basic rate at which the liquid is aerosolized and emitted in mL/min."The device passed all of the acceptance criteria for the tests."
    Particle Size DistributionDetermines the particle size (MMAD) and geometric standard deviation (GSD) of the aerosolized liquid."The device passed all of the acceptance criteria for the tests."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document states "Bench testing has been performed". However, it does not specify the sample size used for any of the tests (e.g., how many devices were tested for lift pressure, nebulization rate, etc.). It also does not mention the country of origin of the data or whether the data was retrospective or prospective. Given it's bench testing, it's typically prospective in nature, but this is not explicitly stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable to the provided document. The document describes a technical device (nebulizer) performance assessment, not an AI or diagnostic device that requires expert-established ground truth for image or data interpretation. The ground truth for these tests would be objective measurements against established engineering and industry standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This section is not applicable for the same reasons as point 3. Adjudication methods are relevant for subjective interpretations, typically in diagnostic studies with multiple human readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The document describes the performance of a medical device (a nebulizer), not an AI algorithm designed to assist human readers or interpreters.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The document is about a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the type of testing described (packaging integrity, sterility, lift pressure, oxygen entrainment, nebulization rate, particle size distribution) would be objective measurements against predefined engineering specifications and regulatory standards. For example, sterility would be verified by microbiological testing against a defined sterility assurance level (SAL), and particle size distribution would be measured using validated instruments against a specified range. It's not expert consensus, pathology, or outcomes data in this context.

    8. The sample size for the training set

    This section is not applicable. The document describes the testing of a physical medical device. There is no mention of a "training set" as this term is typically used in the context of machine learning or AI development.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reasons as point 8.