510(k) Data Aggregation

    K Number
    K210859
    Date Cleared
    2021-11-05

    (227 days)

    Product Code
    Regulation Number
    882.4560
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open or percutaneous spine procedures. It is indical for any medical condition in which the use of stereotaxtic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures, such as:

    • · Pedicle Screw Placement (Thoracic and Lumbosacral spine)
    • Iliosacral Screw Placement
      The NextAR Spine platform is intended to be used in combination with NextAR™M Stereotaxic instruments and / or Medacta preoperative planning. In the case of pre-operative planning software is used preoperatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.
    Device Description

    The NextARTM Spine Platform is a CT based computer-assisted surgical navigation platform used in either open/mini open or percutaneous spine surgery procedure and includes the following components:

    • Navigation software which displays information to the surgeon; ●
    • Augmented Reality glasses; ●
    • Optical tracking system; ●
    • PC based hardware platform; ●
    • Fiducial Block: ●
    • Adaptor for sensor:
    • Spine attachment instruments ●
    • Reusable surgical instruments for spine surgery procedures. ●
      The system operates on the common principle of stereotaxic technology in which markers are mounted on the bones and an infrared camera is used to monitor the spatial location of the instruments. Tracking sensors attached to the bones enable the surgeon to view the position and orientation of the instrumentation relative to the intra-operative data in real-time while performing the surgical procedure. The tracking sensors, the fiducial block, and a group of pins and drills are provided sterile.
      The NextAR™ Spine Platform aids the surgeon in executing the surgical plan by visualizing all the information in real time on a screen monitor.
      The NextAR Spine system is a surgical navigation platform which uses the information of either an intra-operative scan or pre-operative CT in combination with an intra-operative 3D-CArm scan in order to register the spine to navigation elements.
      The registration can be performed with one of the following approaches:
    • Direct 3D: based on the use of an intra-operative 3D-CArm scan ●
    • . 3D-3D: based on the use of a pre-operative CT scan and an intra-operative 3D-CArm scan
      The system's navigation technology is based on an active infrared camera coupled with an active tracker (Target). These elements allow, by means of the different registration approaches and use of compatible instruments, to accurately prepare trajectories in the vertebrae and/or to implant screws while visualizing information in real time on a screen monitor.
    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text:

    1. Table of Acceptance Criteria & Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating numerical acceptance criteria for clinical performance. However, based on the performance data section, the device's accuracy was a key performance metric evaluated. The acceptance criteria for accuracy are not explicitly stated in numerical terms in this document, but the study described aimed to confirm its performance comparable to the predicate.

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    AccuracyComparable to predicate navigation systems in spine procedures."Accuracy test" was conducted and the "Cadaver study" supports the device's ability to precisely locate anatomical structures and aid in screw placement. While specific numerical accuracy metrics (e.g., error in mm) are not stated in this summary, the conclusion states substantial equivalence has been demonstrated through performance evaluations.
    Software ValidationSoftware functions as intended and meets specifications.Performed.
    BiocompatibilityMeets ISO 10993-1:2009 standards.Performed.
    Sterilization ValidationDevice can be effectively sterilized.Performed.
    Shelf-life TestingDevice maintains integrity over its shelf life.Performed.
    Shipping TestDevice withstands shipping conditions.Performed.
    Electrical SafetyConforms to IEC 60601-1:2005 and amendments.Performed.
    Electromagnetic Compatibility (EMC)Conforms to IEC 60601-1-2:2014.Performed.
    Mechanical & Optical PropertiesMeets performance expectations.Performed.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions a "Cadaver study" as part of the performance testing.

    • Sample Size: The exact number of cadavers used is not specified in the provided text.
    • Data Provenance: The study was a "Cadaver study," indicating it was conducted on human cadavers. This is a form of retrospective data in terms of patient-specific outcomes, as the cadavers represent once-living individuals, but the experimental setup is controlled like a prospective study. The country of origin is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not provide details on the number of experts or their specific qualifications (e.g., radiologists with X years of experience) used to establish ground truth within the cadaver study. It states the system is intended to aid surgical placement, implying the ground truth would typically be related to the accuracy of instrument placement or trajectory relative to anatomical landmarks, ideally verified by expert assessment or post-procedure imaging analysis.

    4. Adjudication Method for the Test Set:

    The document does not specify any adjudication method (e.g., 2+1, 3+1) for the test set (cadaver study).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size regarding human readers improving with AI vs. without AI assistance. The device is a navigation platform for surgeons, implying human-in-the-loop assistance, but a formal MRMC study as typically performed for diagnostic AI is not described.

    6. Standalone (Algorithm Only) Performance:

    The document does not explicitly describe a standalone (algorithm only without human-in-the-loop performance) study. The device is designed as a "surgical navigation platform" to "aid the surgeon" and display information "in real-time on a screen monitor," indicating an inherent human-in-the-loop design.

    7. Type of Ground Truth Used:

    For the "Accuracy test" and "Cadaver study" related to placement, the ground truth would likely involve:

    • Precise measurements of instrument tip position or trajectory relative to anatomical landmarks.
    • Verification using high-resolution imaging (e.g., CT scan) of the cadaver after "screw placement" to assess deviation from planned trajectories or ideal anatomical placement.
      The document does not explicitly state the method for establishing ground truth, but for a navigation system, it would generally be based on anatomic truth verified by imaging or detailed measurements.

    8. Sample Size for the Training Set:

    The document does not provide any information regarding the sample size for a training set. As a substantial equivalence determination, the focus is on performance testing and comparison to predicates, not on the details of algorithmic training. The "Dedicated algorithm" is mentioned as a technological characteristic, implying machine learning components, but its training details are not disclosed here.

    9. How the Ground Truth for the Training Set Was Established:

    Since there is no information on a training set, the document does not provide details on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210427
    Date Cleared
    2021-04-20

    (68 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    Reference Devices :

    K12115, K132878, K141988, K153664, K162061, K171170, K171758, K193365

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (78-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Midline Cortical (MC) Screw System is intended to be used as part of the M.U.S.T. Pedicle Screw system (K12115, K132878, K141988, K153664, K162061, K171170, K171758, K193365) for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. pedicle screw system includes cannulated or non-cannulated poly-axial pedicle screws (K121115, K132878, K153664), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115, K171758), straight and pre-bent rods (K12115, K141988, K162061), lateral connectors (K162061) and cross connectors (K132878, K193365). The M.U.S.T. pedicle screw system also includes the enhanced screws and rods designed for percutaneous surgery (K141988).

    The M.U.S.T. Midline Cortical (MC) Screw System introduce new polyaxial screws and reduction screws (solid and cannulated), developed with a focus on Midline Cortical Trajectory approach. The devices subject of this submission are:

    • M.U.S.T. MC polyaxial screws Solid
    • M.U.S.T. MC polyaxial screws Cannulated
    • M.U.S.T. MC polyaxial reduction screws Solid
    • M.U.S.T. MC polyaxial reduction screws Cannulated

    Intended purpose and the performance specification of the devices are equivalent to the ones of the current US cleared portfolio: polyaxial screws are already used in the MUST implant construct.

    The M.U.S.T. Midline Cortical (MC) Screws and the Inlay are manufactured from Ti-6Al-4V ELI (ISO 5832-3 Implants for surgery -Metallic materials - Part 3: Wrought titanium 6-vanadium alloy + ASTM F136 Standard Specification for Wrought Titanium-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401); the Tulip and the Setscrew are made of Co-Cr-Mo alloy according to ISO 5832-12:2019, Implants for Surgery - Metallic materials -Part 3: Wrought cobalt-chromium-molybdenum alloy, the same material of the previous cleared M.U.S.T. Pedicle Screw (K153664, K121115, K171170).

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the M.U.S.T. Midline Cortical (MC) Screw System:

    1. Table of acceptance criteria and the reported device performance:

    The provided document does not explicitly state specific quantitative acceptance criteria for many of its tests. Instead, it describes various tests performed and implies that the results of these tests met the necessary standards for substantial equivalence. For instance, mechanical evaluations were conducted "according to ASTM F1717-18" without specifying the pass/fail thresholds. The "geometrical analysis" was performed "to ensure that the implant has sufficient fixation" and "in comparison to approved and marketed implants," implying an acceptance of similarity or non-inferiority.

    To create a table, I will infer the performance reported based on the conclusion of substantial equivalence.

    Evaluation/TestAcceptance Criteria (Inferred)Reported Device Performance/Outcome
    Non-Clinical Studies
    DESIGN VALIDATION
    Wetlab (Surgical Technique)Satisfactory performance and usability by experienced surgeons according to the MUST MC surgical technique.Performed by 3 experienced surgeons according to MUST MC surgical technique; implies satisfactory validation.
    Geometrical Analysis (Implant Fixation)Sufficient fixation in bone, comparable to approved and marketed implants, concerning important thread dimensions.Performed in comparison to approved and marketed implants to ensure sufficient fixation; implies comparable and sufficient fixation.
    PERFORMANCE TESTING
    Worst Case Definition for ASTM F1717Identification of appropriate implant types and sizes representing the "worst case" for mechanical testing.Worst case defined for mechanical testing of the spinal construct in a vertebrectomy model.
    Mechanical Evaluation (Static Compression Bending)Compliance with ASTM F1717-18 standard test methods for spinal implant constructs in a vertebrectomy model. (Implicitly meeting or exceeding predicate's performance or standard thresholds).Performed according to ASTM F1717-18; implies compliance and satisfactory mechanical performance.
    Mechanical Evaluation (Dynamic Compression Bending)Compliance with ASTM F1717-18 standard test methods for spinal implant constructs in a vertebrectomy model. (Implicitly meeting or exceeding predicate's performance or standard thresholds).Performed according to ASTM F1717-18; implies compliance and satisfactory mechanical performance.
    Mechanical Evaluation (Static Torsion)Compliance with ASTM F1717-18 standard test methods for spinal implant constructs in a vertebrectomy model. (Implicitly meeting or exceeding predicate's performance or standard thresholds).Performed according to ASTM F1717-18; implies compliance and satisfactory mechanical performance.
    PYROGENICITY
    Endotoxins-Mediated Pyrogenicity Assessment (Implants)Assessment indicating acceptable pyrogenicity levels as per documented reports. (Implicitly non-pyrogenic)."ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT REPORT FOR: 'MUST PEDICLE SCREWS IMPLANTS SURGICAL KITS'" (TF VI-PS-01, RAS-01.008.141 Rev. 5, Dated: October 17 2020) was provided.
    Endotoxins-Mediated Pyrogenicity Assessment (MC ModularAssessment indicating acceptable pyrogenicity levels as per documented reports. (Implicitly non-pyrogenic)."ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT FOR: 'MUST MC MODULAR TAP'" (TF VS-PS-01, RAS-01.008.260 Rev. 0, Dated: February 18 2020) was provided.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size:
      • Wetlab (Surgical Technique): The "cadaver labs were performed by 3 experienced surgeons." This refers to the number of users of the devices, not a number of distinct device units or anatomical samples tested. The number of cadavers used is not specified.
      • Geometrical Analysis: Not specified, but it would involve measurements of the (new) M.U.S.T. MC implants and approved/marketed implants.
      • Mechanical Performance Testing (ASTM F1717): Not specified. ASTM F1717 standards generally define specific sample sizes (e.g., minimum of 5 samples per test condition for static tests, 3-6 for dynamic fatigue). The document states "Worst Case definition... for mechanical testing," implying specific samples were chosen.
      • Pyrogenicity Assessment: Not specified, but would align with standard biological testing protocols.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not explicitly stated. These were non-clinical studies conducted for regulatory submission. Given the manufacturer is Medacta International SA (Switzerland) and its US counterpart Medacta USA, the testing could have been conducted in any accredited lab globally. They are prospective tests conducted on the device components and system.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Number of Experts: 3 experienced surgeons were used for the wetlab (surgical technique validation).
    • Qualifications of Experts: They are described as "experienced surgeons." No further detailed qualifications (e.g., years of experience, specific sub-specialty) are provided in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The document mentions "3 experienced surgeons" performed the wetlab. However, it does not describe an adjudication method for their observations or findings. It simply states the lab was performed "according to MUST MC surgical technique," implying their use was for validation of the technique and device performance, not necessarily for a consensus-driven "ground truth" establishment in the typical sense of diagnostic accuracy.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was NOT done.
    • This device is a medical implant (spinal screw system), not an AI/software device that assists human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No, a standalone (algorithm only) performance study was NOT done.
    • As explained above, this is a physical medical implant, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the non-clinical studies:

    • Wetlab: "Performance by 3 experienced surgeons according to MUST MC surgical technique" implies the ground truth for surgical technique validation was established through expert application/observation and adherence to predefined protocols.
    • Geometrical Analysis: Ground truth was the geometric properties of approved and marketed implants used for comparison, and engineering specifications.
    • Mechanical Testing (ASTM F1717): Ground truth was defined by the requirements and methodologies of the ASTM F1717-18 standard and comparisons to the mechanical characteristics of the predicate device.
    • Pyrogenicity: Ground truth was established by standard biological assessment methods for endotoxins, validated against established limits.

    8. The sample size for the training set:

    • Not applicable. This device is a physical medical implant, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. (See point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K203482
    Date Cleared
    2021-01-26

    (60 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Extension and M.U.S.T. Long Tab is intended to be used as part of the M.U.S.T. Pedicle Screw system (K121115, K132878, K141988, K153664, K162061, K171170, K171758, K193365) for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. pedicle screw system includes cannulated or non-cannulated poly-axial pedicle screws (K121115, K132878, K153664), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115, K171758), straight and pre-bent rods (K121115, K141988, K162061), lateral connectors (K162061) and cross connectors (K132878, K193365). The M.U.S.T. pedicle screw system also includes the enhanced screws and rods designed for percutaneous surgery (K141988).

    The M.U.S.T. Extension and M.U.S.T. Long Tab introduce new sizes of sterile implantable devices intended to be used for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The devices subject of this submission are:

    • New size of Straight Rods (both Titanium and CoCr version): length 600mm Hex interface at the end
    • New size of Straight Rods (Anodized Titanium version): length 600mm
    • Pre-contoured Rods (both Titanium and CoCr version)
    • New sizes of Lateral Connectors: length 80; 100; 125 & 150mm
    • New sizes of Enhanced Cannulated Pedicle Screws: Ø8; Ø9; Ø10 length from 30 to 100mm
    • New cannulated pedicle screws with thread 4 leads (15 and 25 mm)
    • Setscrews in Titanium
    AI/ML Overview

    This document is a 510(k) Premarket Notification from Medacta International SA for their M.U.S.T. Pedicle Screw Extension and Long Tab Implants. It focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than providing detailed clinical study data on device performance against specific acceptance criteria for AI/ML-enabled devices.

    Therefore, the requested information regarding acceptance criteria, study details, human reader improvement with AI assistance, standalone algorithm performance, and ground truth establishment cannot be found in the provided text. The document describes a traditional medical device submission (pedicle screws) which relies on bench testing and comparison to predicates, not AI/ML performance studies.

    However, based on the structure of the provided document, I can infer the "acceptance criteria" and "study" are related to the engineering and design validation of the medical device, rather than the performance of an AI/ML algorithm.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document describes several non-clinical studies and tests with implied acceptance criteria, but it does not present them in a clear table format with specific quantitative performance results against those criteria. Instead, it lists the types of tests performed.

    Acceptance Criteria (Implied)Reported Device Performance (Implied)
    Design Validation:Design Validation:
    Sufficient fixation in bone (Geometrical Analysis MUST LT Thread Geometry)A geometrical analysis was performed in comparison to approved and marketed implants, related to the most important dimensions of the threads (assumed to meet acceptance criteria through comparison).
    Operability with MIS percutaneous technique (Cadaver surgery)Cadaver surgery performed according to the well-known MIS percutaneous technique by experienced surgeons (assumed successful operation; no specific outcome metrics provided).
    Sufficient flexural strength of the tulip (Tab welding MUST LT Flexion strength test)Flexion strength test performed to characterize the specific design of the tulip in the dedicated breaking area (assumed to meet acceptance criteria for breaking area design; no explicit quantitative results provided).
    Performance Testing:Performance Testing:
    Adequate mechanical performance under static and dynamic axial compression (ASTM F1717-18)Mechanical evaluation performed according to ASTM F1717-18 (Endolab 970 200220 30 1413 part 1 rev.0 and 970 200220 30 1414 part 1 rev.0) for worst-case definition (assumed to demonstrate acceptable mechanical performance comparable to predicate; no specific quantitative results provided).
    Adequate mechanical performance under static torsion (ASTM F1717-18)Mechanical evaluation performed according to ASTM F1717-18 (Endolab 970 200220 30 1413 part 2 rev.0 and 970 200220 30 1414 part 2 rev.0) (assumed to demonstrate acceptable mechanical performance comparable to predicate; no specific quantitative results provided).
    Adequate axial tensile force resistance (ASTM F543-17, ASTM F2193-18a)Mechanical evaluation performed according to ASTM F543-17 and ASTM F2193-18a (Endolab 970.200227.120.868-part1 rev. 0) (assumed to demonstrate acceptable resistance; no specific quantitative results provided).
    Adequate torsional force resistance (ASTM F543-17, ASTM F2193-18a)Mechanical evaluation performed according to ASTM F543-17 and ASTM F2193-18a (Endolab 970.200227.120.867-part2 rev. 0 and 970.200227.120.868part2 rev. 0) (assumed to demonstrate acceptable resistance; no specific quantitative results provided).
    Acceptable torsional yield strength, maximum torque, and braking angle (ASTM F543-17 A1)Mechanical evaluation performed according to ASTM F543-17 - A1 (Endolab 970.200227.120.867-part3 rev. 0 and 970.200227.120.868-part3 rev. 0) (assumed to demonstrate acceptable properties; no specific quantitative results provided).
    Acceptable risk of breakage during insertion (ASTM F543-17)Characterization and evaluation performed according to ASTM F543-17 (Medacta test report ASTM F543 Summary Report) (claimed acceptable risk; no specific methodology or results provided besides the report reference).
    Biocompatibility:Biocompatibility:
    Non-pyrogenic (Pyrogenicity Assessment)Pyrogenicity Assessment Reports for MUST PEDICLE SCREWS IMPLANTS SURGICAL KITS (TF VI-PS-01, RAS-01.008.141 Rev. 5) and MUST LT IMPLANTS (VI-PS-01) (RAS-01.008.277, Rev. 0) were conducted (assumed to demonstrate non-pyrogenicity).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not explicitly stated for each test, but standard ASTM methods often define minimum sample sizes for mechanical testing. For the cadaver surgery, it states "experienced surgeons," implying at least one or more instances.
    • Data Provenance: Not specified, but given the company is Medacta International SA (Switzerland) and the testing labs are mentioned (Endolab), the data is presumed to be from their internal testing facilities or contracted labs, likely within Europe. No mention of patient data or clinical data is made, so "retrospective or prospective" is not applicable in the context of this submission. This is a bench test and design validation submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not an AI/ML study involving human expert ground truth. The "ground truth" for these tests is defined by engineering standards (e.g., ASTM) and physical measurements. For the cadaver surgery, it states "experienced surgeons" performed the procedure, but their role was to evaluate the instrument's use in a simulated environment, not to establish a "ground truth" for an AI model.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is for resolving disagreements in expert labeling for AI/ML ground truth, which is not relevant here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a physical pedicle screw system, not an AI/ML software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is based on:

    • Engineering Standards: Adherence to established ASTM (American Society for Testing and Materials) standards for spinal implants and medical bone screws to define acceptable mechanical properties.
    • Physical Measurements/Analysis: Geometrical analysis of threads and characterization of mechanical properties (tensile strength, torsion, etc.).
    • Risk Analysis: Identification of new risks and verification against predefined acceptance criteria based on this analysis, likely conforming to ISO standards for medical device risk management.
    • Biocompatibility Testing: Demonstrated through pyrogenicity assessment reports.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As above, no AI/ML device with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200792
    Date Cleared
    2020-10-05

    (193 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K12115, K132878, K141988, K153664, K162061, K171170

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MySpine S2-Alar/Alar-Iliac is intended for use with M.U.S.T. Pedicle Screw System and its cleared indications for use. MySpine S2-Alar/Alar-Iliac guides (referred to from this point on as, MySpine guides) are intended to be used as anatomical perforating guides, specific to a patient's anatomy, to assist intra-operatively in the preparation of the screw trajectory in S1, S2 and in the Ilium. The guides are created using a surgical planning software which pre-operatively plans the positions of the components based upon radiological images of the patients' anatomical landmarks and the surgical equipment selected. MySpine guides are intended for single use only.

    Device Description

    The MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides are a line extension to Medacta's MySpine Pedicle Screw Placement Guides. Identical to the other Medacta MySpine products, the MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides are a patient matched, pedicle targeted, technology involving the production of patient specific guides for placement of the M.U.S.T. Pedicle Screw System (K12115, K132878, K141988, K153664, K162061, and K171170). Specifically, the subject MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides are intended to be used as anatomical perforating guides to assist intra-operatively in the preparation of the screw trajectory in S1, S2 and in the Ilium. The MySpine software platform allows the surgeon to complete 3D pre-operative planning based on the patient's spinal CT scans. CT images are used to create a 3D model of the vertebrae that will represent the template used to generate the corresponding MySpine Screw Placement Guides fitting the patient's vertebral anatomy. The MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides as well as their bone models are single-use and they can be provided in sterile or non-sterile version.

    AI/ML Overview

    The provided text describes a medical device, the "MySpine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides," and its substantial equivalence determination by the FDA. However, the document does NOT contain information about specific acceptance criteria for a device's performance (like sensitivity, specificity, accuracy, etc.) nor does it usually detail specific studies that prove the device meets such criteria in terms of quantitative metrics suitable for the requested table.

    Instead, this document focuses on demonstrating substantial equivalence to a predicate device based on similar technological characteristics (manufacturing process, material, biocompatibility, device usage, sterility, shelf life, packaging) and performance data from non-clinical studies (software validation, cadaver testing, guide accuracy, stability assessment). The "guide accuracy" mentioned is a general category and not a specific set of acceptance criteria with reported performance.

    Therefore, I cannot fill in the table of acceptance criteria and reported device performance, nor can I answer questions related to sample size, expert qualifications, or ground truth establishment for a standalone algorithm performance study, because this information is not present in the provided text. The document explicitly states: "No clinical studies were conducted." and it does not describe an AI algorithm with human-in-the-loop performance.

    Here's an overview of what can be extracted or inferred based on the provided text, and where information is missing:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated with quantifiable metrics (e.g., "accuracy must be > 95%"). The document talks about "software validation," "cadaver testing," "guide accuracy," and "stability assessment" as general categories of testing performed, aiming to show safety and effectiveness comparable to the predicate.
    • Reported Device Performance: No specific quantitative performance metrics (e.g., accuracy percentages, error margins) are reported in the document.

    Therefore, this table cannot be filled based on the provided input.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified. The document mentions "Cadaver testing" and "Guide accuracy" testing, but no details on the number of cadavers or cases used for these tests are provided.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as no detailed test set and ground truth establishment methodology is described in the provided text. The document refers to "surgical planning software" that "pre-operatively plans the positions of the components based upon radiological images," but this is part of the device's function, not a ground truth establishment for an AI algorithm's performance evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable as no detailed test set and ground truth establishment methodology is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The document explicitly states: "No clinical studies were conducted." The device (surgical guides) assists surgeons, but the documentation does not describe a study involving human readers and AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The device is a physical surgical guide produced using software, not a standalone AI algorithm that provides diagnostic or prognostic outputs. The "MySpine software platform" is integral to creating the guides. "Software validation" was performed, but no standalone algorithm performance as typically understood in AI/ML context (e.g., measuring diagnostic accuracy against ground truth without human intervention) is described.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not explicitly stated for any performance evaluation in terms of AI. For the device itself and its intended use, the anatomical landmarks from patient CT scans and selected surgical equipment (per planning software) would form the basis for the design of the patient-specific guides. The "guide accuracy" testing would presumably involve comparing the guide's output (screw trajectory) against the planned trajectory or anatomical reality, but the specific ground truth methodology for this testing is not detailed.

    8. The sample size for the training set

    • Not applicable. The document describes a medical device (surgical guides) and the software that creates them for individual patients. It does not refer to a machine learning model that requires a "training set" in the conventional sense for AI performance. The software uses patient-specific CT data for planning.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no mention of a "training set" for an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193365
    Date Cleared
    2020-01-27

    (54 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132878, K171170, K121115, K153664, K141988

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/ilum) and anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion in skeletally mature patients.

    Device Description

    The purpose of this submission is to add the M.U.S.T. MC Cross Connectors to the M.U.S.T. Pedicle Screw System (K132878, K171170, K121115, K153664, K141988) for the stabilization and fusion of the non-cervical spine. The M.U.S.T. Pedicle System includes a wide range of implants: Rods, Polyaxial and Mono-axial Pedicle Screws, Hooks, Connectors of various design and size, that are used in combination to compose a spinal construct.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the M.U.S.T. Pedicle Screw System, which is a medical device for spinal fixation. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than performing novel clinical studies to prove performance against specific acceptance criteria.

    Therefore, the requested information regarding acceptance criteria, device performance, study details (sample size, data provenance, expert validation, ground truth, MRMC studies, standalone performance, training set) typically associated with a de novo or PMA submission for a new device's efficacy and safety is not applicable to this 510(k) summary.

    Here's how the provided information aligns with the request, highlighting what is and isn't available from this document:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable in this Context. For a 510(k), the "acceptance criteria" for the device's performance are primarily established by the performance of the predicate devices. The study performed is to demonstrate that the new device (M.U.S.T. MC Cross Connectors) performs equivalently to the predicate devices and meets relevant industry standards rather than achieving specific performance metrics against a disease outcome.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable for clinical data.
    • Mechanical Testing:
      • Sample Size: The document states that "worst-case implants" were tested, but it does not specify the exact number of samples used for each test (Static compression/bending, Dynamic compression bending).
      • Data Provenance: The provenance of the data is from internal testing conducted by Medacta International SA. The country of origin for the testing facility is not explicitly stated, but Medacta International SA is based in Switzerland. The studies are prospective in the sense of being planned tests on newly manufactured components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a mechanical device, not a diagnostic or AI-driven system that requires expert interpretation for ground truth establishment. The "ground truth" for mechanical testing is derived from engineering principles and validated test standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials or studies involving expert review of images or clinical outcomes. For mechanical performance testing, the results are typically quantitative and objective, based on physical measurements and material properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. An MRMC study is relevant for diagnostic devices, particularly those involving image interpretation by human readers, often comparing human performance with and without AI assistance. This device is a surgical implant, not a diagnostic tool, and involves no human "readers" in the context of interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This question is also relevant for AI/software as a medical device (SaMD). The M.U.S.T. Pedicle Screw System is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Mechanical Testing Standards. For the non-clinical performance tests, the "ground truth" is implied by the adherence to and successful completion of established industry standards for spinal implant testing, specifically ASTM F1717 (Static compression/bending and Dynamic compression bending) and European Pharmacopoeia §2.6.14 / USP chapter and for pyrogenicity. The acceptance criteria are embedded within these standards.

    8. The sample size for the training set

    • Not Applicable. The concept of a "training set" doesn't apply to the mechanical testing of a physical medical device like this, which doesn't involve machine learning algorithms.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an algorithm, this question is not relevant.

    Summary of Relevant Information from the Document for a 510(k):

    The document explicitly states:

    • "No clinical studies were conducted." This reinforces that the focus is on non-clinical, mechanical performance and comparison to predicate devices.
    • Performance Tests: The device underwent "Non-Clinical Studies" including:
      • Static compression / bending per ASTM F1717
      • Dynamic compression bending per ASTM F1717
      • Cadaver studies (though no specifics are given beyond "Cadaver studies O")
      • Pyrogenicity testing (Bacterial Endotoxin Test per European Pharmacopoeia §2.6.14/USP chapter and pyrogen test per USP chapter ).
    • Acceptance Criteria for Non-Clinical Tests: "Testing was conducted according to written protocols with acceptance criteria that were based on standards." The implicit acceptance criteria are that the device meets or exceeds the performance of legally marketed predicate devices, as defined by the standards (e.g., ASTM F1717) and the specific testing protocols.

    In essence, for a 510(k) for a physical implant, the "study that proves the device meets the acceptance criteria" is primarily the non-clinical performance testing demonstrating that the device is substantially equivalent to existing, legally marketed predicate devices, by meeting relevant recognized standards. This is a comparison of technological characteristics and performance, not a clinical efficacy study with human subjects.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173472
    Date Cleared
    2018-04-19

    (161 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    Reference Devices :

    K121115, K132878, K141988, K153664, K162061, K171170

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MySpine is intended for use with M.U.S.T Pedicle Screw System and its cleared indications for use. MySpine Drill Pilot is intended as a thoracic and lumbar posterior pedicle targeting guide for patients requiring spinal fusion between the levels of T1 to L5. The device is intended for perforating a guiding hole to assist in the positioning of pedicle screws in the vertebral body.

    MySpine MC is intended as a lumbar and sacral posterior pedicle targeting guide for patients requiring spinal fusion between the levels of L1 to S1.

    The device is provided with two options:

    • Drill based
    • K-wire based

    MySpine MC drill based are intended for perforating a guiding hole to assist in the positioning of pedicle screws in the vertebral body.

    MySpine MC k-wire based are intended for the placement of K-wires to assist in the positioning of pedicle screws in the vertebral body.

    Use of the guides involves a surgical planning software used pre-operatively to plan the surgical placement of the components on the basis of patient radiological images with identifiable placement anatomical landmarks and surgical equipment components. These components include patient-specific guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan.

    MySpine MC and Drill Pilot guides are intended for single use only.

    Device Description

    The MySpine Pedicle Screw Placement Guides – MC (Midline Cortical) and Drill Pilot Instruments are a line extension to Medacta's MySpine Pedicle Screw Placement Guides. The MySpine Pedicle Screw Placement Guides – MC and S1 Guide Instruments are for use in spinal levels L5 to S1. The MySpine Pedicle Screw Placement Guides - Drill Pilot Guide Instruments are for use in spinal levels T1 to S1.

    The MySpine product line is a patient matched, pedicle targeted, technology involving the production of patient specific guides for placement of the M.U.S.T. Pedicle Screw System based on the patient's anatomy. The MySpine Pedicle Screw Placement Guides - MC and Drill Pilot Instruments are intended to be used with Medacta's M.U.S.T. Pedicle Screw System (K121115, K132878, K141988, K153664, K162061, and K171170) for the following indications: degenerative disc disease (DDD), spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudo-arthrosis, and failed previous fusion.

    The MySpine platform allows the surgeon to complete pre-operative planning in 3D, based on the patient's spinal CT scans. Medacta International SA uses these images in combination with SW Mimics® (Materialize®) and W Solidworks® (Dassault Systèmes®) to import DICOM images from a patient's CT scans and then process the scans into accurate 3D models. CT imaging is used to create a 3D model of the vertebrae according to the patient's anatomy. The subsequent vertebral model represents the template used to generate the corresponding MySpine Screw Placement Guides to fit the patient's vertebral anatomy.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the MySpine Pedicle Screw Placement Guides - MC and Drill Pilot Instruments. It describes the device, its intended use, and argues for its substantial equivalence to predicate devices.

    However, the document does not contain specific acceptance criteria or performance data for an AI-powered device. The MySpine system described is a patient-matched surgical guide system that uses pre-operative CT scans to create 3D models and fabricate patient-specific guides. It utilizes software for planning, but it is not an AI/ML-driven device as understood in the context of diagnostic or prognostic AI.

    Therefore, I cannot extract the requested information about acceptance criteria and a study proving device performance for an AI/ML device from this document. The document primarily focuses on the physical design, manufacturing process, materials, and intended use of the surgical guides, along with non-clinical performance tests relevant to mechanical and design validation, and a statement that no clinical studies were conducted for this particular submission.

    To answer your request, if this were an AI or ML-enabled device, the information would typically include:

    1. A table of acceptance criteria and the reported device performance: This would detail metrics like sensitivity, specificity, accuracy, precision, recall, F1-score, AUC, etc., along with the target thresholds (acceptance criteria) and the actual performance achieved.
    2. Sample size used for the test set and the data provenance: Number of cases/patients in the test set, and details like country of origin, whether it was retrospective or prospective data, etc.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: For example, 3 radiologists with 5+ years of experience each.
    4. Adjudication method for the test set: How disagreements among experts establishing ground truth were resolved (e.g., 2+1 means 2 experts agree, if not, a 3rd expert adjudicates; 3+1 means 3 experts agree, if not, a 4th adjudicates).
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Details on the study design and the effect size (e.g., human readers improved AUC by X with AI assistance).
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The results of the algorithm's performance without human intervention.
    7. The type of ground truth used: For instance, expert consensus, pathology reports, clinical outcomes data, follow-up imaging, etc.
    8. The sample size for the training set: Number of cases/patients used to train the AI model.
    9. How the ground truth for the training set was established: Similar to the test set, but often less stringent or by different methods depending on the training methodology.

    Since the provided document does not pertain to an AI/ML device, I am unable to provide the specific details requested. The "Performance Data" section (page 7) lists only non-clinical studies related to design validation, stability, post-operative evaluation (accuracy of the physical guide), and wear tests for the physical instruments. It explicitly states "no clinical studies were conducted."

    Ask a Question

    Ask a specific question about this device

    K Number
    K171640
    Date Cleared
    2017-10-23

    (143 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132878, K141988, K153664, K162061

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    M-ARS ACL: Anatomic Ribbon Surgery System - M-ARS ACL Tibial Pull Suture Plate (PSP): Reconstructive treatment of ruptured anterior and posterior cruciate ligaments by means of autologous grafts.

    M-ARS ACL: Anatomic Ribbon Surgery System: M-ARS ACL Extracortical Femoral Button: Reconstructive therapy of ruptures to the anterior cruciate ligament by means of autologous grafts.

    Device Description

    The M-ARS ACL: Anatomic Ribbon Surgery System components are implantable devices used for the reconstructive treatment of ligament ruptures as well as the fixation of an implanted Anterior Cruciate Ligament graft by means of suspensory extracortical fixation.

    The M-ARS ACL: Anatomic Ribbon Surgery System Extracortical Fixation Button component is an implantable extracortical suspension device used to secure the graft fixation on the femoral side during ACL reconstruction surgery. The M-ARS ACL: Anatomic Ribbon Surgery System Tibial Pull Suture Plate (PSP) component is an implantable extracortical suspension device used to secure the graft fixation on the tibial side during ACL reconstruction surgery.

    The M-ARS ACL: Anatomic Ribbon Surgery System Extracortical Fixation Button and Tibial Pull Suture Plate are manufactured with Titanium-6 Vanadium Extra Low Interstitial (Ti-6Al-4V ELI).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "M-ARS ACL: Anatomic Ribbon Surgery System," which is a medical device for reconstructive treatment of ruptured anterior and posterior cruciate ligaments. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific performance criteria through a study with acceptance criteria.

    Therefore, many of the requested categories (e.g., acceptance criteria table, sample sizes, expert ground truth, MRMC study, standalone performance, training set data) are not applicable as this submission is for a medical device that relies on the established safety and effectiveness of predicate devices, and the studies conducted are primarily non-clinical and comparative in nature, not performance studies with specific statistical acceptance criteria for a new AI or diagnostic device.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document as it is a 510(k) submission for a surgical implant device, not a diagnostic or AI device that typically presents performance metrics against acceptance criteria. The focus is on demonstrating substantial equivalence to predicate devices through material, design, and functional comparisons, along with non-clinical testing for safety.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size (Test Set): Not applicable for the type of studies performed. The document mentions "design comparison" and "cadaver workshop." These are not statistical studies with defined test set sample sizes in the context of device performance metrics.
    • Data Provenance: Not applicable. The studies are non-clinical (design comparison, cadaver workshop).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Experts and Qualifications: For the "cadaver workshop," the text mentions "surgeon evaluations to verify the design features and surgical technique." The specific number or qualifications of these surgeons are not detailed. This workshop serves to evaluate usability and design, not to establish a "ground truth" in the way it would be for a diagnostic AI system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. There is no mention of a formal adjudication process for the non-clinical studies described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices or AI tools, which is not the nature of the M-ARS ACL system.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance Study: No, a standalone performance study was not done. This device is a surgical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Type of Ground Truth: Not applicable in the traditional sense of a diagnostic or AI device. For the non-clinical studies, "ground truth" would be related to engineering specifications, material properties, and surgical technique verification by surgeons. There's no pathological or outcomes data used to establish "ground truth" for the device's technical performance in this context.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This device is not an AI or machine learning algorithm requiring a training set.

    9. How the ground truth for the training set was established:

    • Training Set Ground Truth Establishment: Not applicable.

    Summary of Device Performance and Acceptance Criteria (as described in the context of a 510(k) submission for this device):

    The document does not explicitly state "acceptance criteria" for performance metrics in the way one would for a diagnostic test (e.g., sensitivity, specificity thresholds). Instead, the "performance" is demonstrated through:

    • Design Comparison: Geometrical and comparative analysis between the M-ARS ACL System and predicate devices to show mechanical equivalence. The acceptance criterion here is implied as being "substantially equivalent" in design and mechanical properties to legally marketed predicate devices.
    • Cadaver Workshop: Surgeon evaluations to verify design features and surgical technique. The implied acceptance criterion would be that surgeons can successfully use the device and technique, and that the design is appropriate for its intended use.
    • Substantial Equivalence Evaluation: A comparative analysis, especially for the Tibial Pull Suture Plate, against the predicate device. This ensures that any differences in design do not raise new questions of safety or effectiveness.
    • Biocompatibility and Sterilization: The device materials (Titanium Alloy ELI) have an extensive history of use in medical devices and have been previously reviewed by the FDA, implying they meet established biocompatibility standards. The device is sterile and single-use, meeting standard regulatory requirements for such implants.
    • Pyrogenicity: Medacta uses Bacterial Endotoxin Test (LAL test) and Pyrogen Test (USP chapter ), indicating compliance with standards for sterility and pyrogenicity, though they do not label the devices as non-pyrogenic or pyrogen-free.

    Conclusion stated in the document: "Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics as well as performance evaluations. The M-ARS ACL: Anatomic Ribbon Surgery System is as safe and effective as the predicate devices, Karl Storz's Flipptack (K982571) and Karl Storz's Endotack (K022853)."

    In essence, the "acceptance criteria" are the demonstration of substantial equivalence to predicate devices, supported by non-clinical studies verifying design, materials, and surgical technique.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171758
    Date Cleared
    2017-08-01

    (48 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132878, K141044, K141988, K153664, K162061, K171170

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Combined Set Screws are intended to be used as part of the M.U.S.T. Pedicle Screw System (cleared under K12115, K132878, K141044, K141988, K153664, K162061, and K171170) for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. Pedicle Screw System includes: cannulated or non-cannulated poly-axial pedicle screws (K12115 and K132878), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K12115), straight and pre-bent rods (K121115 and K162061), cross connectors (K132878), hooks (K141044), enhanced screws and rods designed for percutaneous surgery (K141988 and K162061), cannulated or non-cannulated reduction screws (K153664), connectors (K162061 and K171170), and standard, enhanced, and reduction pedicle screws (K171170). The M.U.S.T. Combined Set Screws are manufactured from CoCrMo alloy (ISO 5832-12 and ASTM F1537-11) and are provided sterile and non-sterile screw is packaged individually as well as in packages of 2, 4, 6, and 8 screws. The non-sterile screw is packaged individually.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the M.U.S.T. Combined Set Screws. It focuses on demonstrating substantial equivalence to a predicate device, not on validating an AI algorithm. Therefore, many of the requested items related to AI device performance, such as human reader improvement with AI, ground truth establishment for training and test sets, and multi-reader multi-case studies, are not applicable or cannot be extracted from this document.

    However, I can extract information regarding the device's acceptance criteria, the study conducted, and other relevant details provided in the submission for the M.U.S.T. Combined Set Screws.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical "acceptance criteria" in the format typically seen for algorithm performance (e.g., sensitivity, specificity thresholds). Instead, it refers to performance studies conducted in accordance with specific ASTM standards. The acceptance criteria are implicitly that the device performs equivalently to the predicate device and meets the requirements of these standards. The reported device performance is that it met these standards and thus is "as safe and effective as the predicate devices."

    Acceptance Criteria CategoryReported Device Performance and Relevant Standard
    Mechanical Strength (Dynamic Bending)Met requirements of ASTM F1717-15 Standard Test Methods For Spinal Implant Constructs In A Vertebrectomy Model.
    Mechanical Strength (Interconnection Mechanism)Met requirements of ASTM F1798-13 Standard Guide For Evaluating The Static And Fatigue Properties Of Interconnection Mechanisms And Subassemblies Used In Spinal Arthrodesis Implants.
    Biocompatibility/Safety (Pyrogenicity)Pyrogenicity Testing (LAL Endotoxin Test) was conducted, implying it met relevant safety standards for non-pyrogenic materials.
    Structural Integrity (Sawbones Testing)Sawbones Testing was performed, implying it demonstrated structural integrity and appropriate interaction within a simulated bone environment.
    Overall EquivalenceBased on performance data, "the M.U.S.T. Combined Set Screws are as safe and effective as the predicate devices." The minor design difference (interface) "does not raise any new questions of safety and effectiveness" due to the performance testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set" in the context of an AI algorithm, nor does it provide details on sample sizes for the mechanical and pyrogenicity tests beyond indicating that the tests were performed. The "data provenance" information (country of origin, retrospective/prospective) is not applicable to the device testing described. The materials for the device itself are CoCrMo. The manufacturing company, Medacta International SA, is located in Switzerland.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The device is a physical medical implant (surgical screws), not an AI diagnostic or prognostic tool requiring expert-established ground truth from images or clinical data. The "ground truth" for this device would be its physical and mechanical properties, assessed through standardized laboratory tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable, as there is no "test set" in the context of expert review for an AI algorithm. The device's performance was evaluated through laboratory testing against established engineering and material standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The M.U.S.T. Combined Set Screws are a physical orthopedic implant, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The "device" in question is a physical hardware component, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is based on established engineering standards and material properties, specifically through performance testing in accordance with:

    • ASTM F1717-15 (for dynamic compression bending of spinal implant constructs)
    • ASTM F1798-13 (for static and fatigue properties of interconnection mechanisms)
    • Sawbones Testing (a simulated biomechanical environment)
    • LAL Endotoxin Test (for pyrogenicity/safety of materials)

    8. The sample size for the training set

    This information is not applicable. There is no "training set" as this is not an AI algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K171170
    Date Cleared
    2017-05-25

    (34 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K121118, K132878, K141988, K153664

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/lium) and nonpedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Extension - Straight Connectors and Additional Screws are intended to be used as part of the M.U.S.T. Pedicle Screws System (K121118, K132878, K141988, K153664) for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. Pedicle Screws System includes: cannulated or non-cannulated poly-axial pedicle screws (K121115and K132878), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115), straight and pre-bent rods (K121115), cross connectors (K132878), enhanced screws and rods designed for percutaneous surgery (K141988), and cannulated or non-cannulated reduction screws (K153664). The M.U.S.T. Additional Pedicle Screws (standard and enhanced pedicle) are made of Ti6Al4V ELI (ISO 5832-3, ASTM F136-13) and CoCrMo (ISO 5832-12, ASTM F1537-11), while the M.U.S.T. Additional Reduction Screws are made of Ti6A14V ELI (ISO 5832-3, ASTM F136-13). The screw shaft is color anodized to simplify the identification of the screw diameter. The pedicle screw has a dual lead thread to simplify the screw insertion and reduce the number of turns. The threads are designed with a cylindrical diameter. The M.U.S.T. Straight Connectors and Additional Screws introduces the following additional components: Pedicle screw, Enhanced Solid Pedicle screw, Enhanced Cannulated Pedicle screw, Reduction Pedicle screw, Reduction Cannulated Pedicle screw, Straight Cross Connector.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the M.U.S.T. Pedicle Screw System Extension - Straight Connectors and Additional Screws. It focuses on demonstrating substantial equivalence to previously cleared predicate devices through comparisons of technological characteristics and performance testing.

    Here's an analysis of the acceptance criteria and study information provided, structured as requested:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly list specific numerical acceptance criteria. Instead, it states that the acceptance criteria were "pre-defined" and "based on the standards, FDA guidance, and comparison to the predicate device system." The reported device performance is a general statement that the device "met all acceptance criteria" and "is substantially equivalent to the predicate devices."

    Acceptance Criteria CategorySpecific Acceptance Criteria (as implied)Reported Device Performance
    Mechanical Performance (Dynamic Compression Bending)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    Mechanical Performance (Static Compression Bending)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    Mechanical Performance (Static Torsion)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    BiocompatibilityMaterial and manufacturing process identical to predicate devices, supporting biological safety.Deemed unnecessary to conduct additional testing as materials and manufacturing are identical to predicate devices.
    Technological CharacteristicsNew diameters and straight connector do not raise new questions of safety or effectiveness compared to predicates.A comparison evaluation shows no new risks associated with the subject device design.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document repeatedly mentions "the worst-case device" for each of the mechanical tests (Dynamic Compression Bending, Static Compression Bending, Static Torsion). This implies a very small sample size, potentially a single representative device or a small set of devices configured to represent the most challenging conditions. The exact number is not explicitly stated.
    • Data Provenance: The study is a laboratory-based mechanical testing study, not a clinical study involving human or animal data. Therefore, concepts like country of origin, retrospective, or prospective do not apply in the typical sense. It is an engineering verification study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable to this type of device and study. The "ground truth" for mechanical testing is established by engineering standards (ASTM F1717-12) and the inherent physical properties and design specifications of the device, rather than expert judgment or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable. Adjudication methods are typically used in clinical studies to resolve discrepancies in expert interpretations of data (e.g., imaging, clinical assessments). For mechanical testing, the "adjudication" is inherent in the adherence to established test standards and the objective measurement of physical properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not applicable to this device. This type of study evaluates the performance of AI-assisted diagnostic or classification systems in a clinical context, often involving human readers interpreting medical images. The M.U.S.T. Pedicle Screw System Extension is a physical orthopedic implant, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation (in the context of AI algorithms) is not applicable to this device. This device is a physical implant and does not involve any AI algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance data in this document is based on engineering standards and specifications. Specifically, the mechanical tests were performed according to ASTM F1717-12 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model. The acceptance criteria were based on these standards, FDA guidance, and comparison to the predicate device's established performance.

    8. The sample size for the training set

    This question is not applicable as there is no "training set" in the context of a physical device mechanical performance study. Training sets are relevant for machine learning algorithms.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162061
    Date Cleared
    2016-10-21

    (87 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K12115, K132878

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/Ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Pedicle Screw System devices are fixation devices intended for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. Pedicle Screw System includes cannulated or non-cannulated poly-axial pedicle screws (K12115, K132878), cannulated or noncannulated mono-axial pedicle screws (K132878), set screws (K121115), straight and pre-bent rods (K121115), and cross connectors (K132878).

    The new lateral connectors can be used when a certain off-set screw has to be implanted with the rod. The lateral connectors offer the same interface to the monoaxial screw (K132878). The lateral connectors are offered in three sizes (20, 40 and 60 mm sizes) to accept a 5.5mm rod and manufactured of Ti-6Al-4V ELI (ISO 5832-3, ASTM F136).

    The new anodized rods and anodized enhanced rods act as a connector between the different screws to create a stable construct. The screws are used in combination with 5.5mm titanium allov rods. This allows the surqeon to choose the construct stiffness based on the indication by maintaining the screw sizes. They are offered as a 5.5mm diameter in various lengths (100, 200, 300 and 480 mm sizes) and manufactured of Ti-6AI-4V ELI (ISO 5832-3, ASTM F136).

    The new rod-to-rod connectors can be used to connect two rods to increase the construct stability. The cross connectors have a medial/lateral length adjustable element in order to address different distances. Rod distances from 35mm to 98 mm can be addressed with the different connector sizes. The rod to rod connectors are manufactured of Ti-6AI-4V ELI (ISO 5832-3. ASTM F136) and the set-screws are made of Co-Cr-Mo (ISO 5832-12, ASTM F1537).

    AI/ML Overview

    This FDA 510(k) premarket notification describes the M.U.S.T. Pedicle Screw System and its substantial equivalence to predicate devices. However, it does not contain the specific information requested in the prompt regarding acceptance criteria, device performance, study details (sample size, data provenance, expert qualifications, adjudication method, MRMC studies, standalone performance), or training set information.

    Instead, this document focuses on:

    • Indications for Use: The system is intended for posterior non-cervical pedicle fixation and non-pedicle/anterolateral fixation for conditions like degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients.
    • Device Description: It includes cannulated/non-cannulated poly-axial and mono-axial pedicle screws, set screws, straight and pre-bent rods, cross connectors, new lateral connectors, new anodized rods, and new rod-to-rod connectors.
    • Comparison to Predicate Devices: The document asserts that the indications for use, design features, and materials of the subject devices are substantially equivalent to those of the predicate devices (M.U.S.T. Pedicle Screw System K121115, K141988, K141044, K132878; Expendium Spine System K041119; Synergy D2 Spinal Implants K984578; MOSS Miami Spinal System K983583). Detailed tables compare materials, sterilization, device usage, sizes, rod compatibility, and shelf life.
    • Performance Testing: It states that a risk analysis was conducted, and design verification was performed with written protocols and pre-defined acceptance criteria based on standards, FDA guidance, and comparison to predicate devices. Mechanical tests mentioned are:
      • Static Compression Bending Strength ASTM F1717
      • Static Torsion Strength ASTM F1717
      • Dynamic Compression Bending Strength ASTM F1717
      • Static Transverse Moment ASTM F1798
      • Four-Point Dynamic Bending Strength ASTM F2193

    Since the provided document is a 510(k) summary for a medical device and not a study report for an AI/CAD system, it does not include the detailed performance study information requested about acceptance criteria, test set, ground truth, etc., typically associated with such systems. The document confirms that specific mechanical tests were performed to support substantial equivalence, but it does not provide the quantitative acceptance criteria or the specific numerical results of these tests, nor does it refer to studies involving human readers or AI.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from this document in the manner you requested for AI/CAD systems.