Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K123425, K203393

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEALICOIL◊ PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL◊ PK Suture Anchor with Needles, ULTRATAPE (Blue Cobraid) are intended for use only for the reattachment of soft tissue to bone for the following indications:

    Foot & Ankle

    • Medial or lateral instability repairs/reconstructions
    • Achilles tendon repairs/reconstructions
    Device Description

    The Smith & Nephew HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is a fixation device intended to provide reattachment of soft tissue to bone. The device consists of a suture anchor with a preloaded ULTRATAPE suture with needles assembled on an insertion device.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a medical device, specifically a suture anchor. The information provided heavily focuses on regulatory compliance, device description, intended use, and comparison to predicate devices, rather than detailed performance study data against acceptance criteria typically seen for AI/ML-driven devices.

    Therefore, many of the requested details about acceptance criteria, study design for proving device performance, sample sizes, expert involvement, and ground truth establishment (especially in the context of an AI/ML algorithm) are not present in this document. This is because the device described is a physical medical device (suture anchor), not an AI/ML diagnostic or predictive tool.

    I will address the elements based on the information that can be inferred for a physical device, and clearly state when information is not available or not applicable given the nature of the device.


    Device: HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue); HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid)
    Device Type: Soft Tissue Fixation Device (Suture Anchor)

    Overall Conclusion from Document: The device is deemed substantially equivalent to a legally marketed predicate device (HEALICOIL PK Suture Anchor, K152566). This determination is based on similarities in indications for use, design features, operational principles, material biocompatibility, and composition. The changes (new inserter and addition of needles to the preloaded suture) are considered minor and do not raise new questions of safety or effectiveness.


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred for a physical device like a suture anchor)Reported Device Performance (Summary from document)
    Mechanical Performance:
    - Fixation Strength/Integrity"met the required specifications"
    - Insertion & Fixation Capability"met the required specifications"
    - Cyclic Loading Endurance"met the required specifications"
    - Needle Attachment Strength"met the required specifications"
    Biocompatibility:"material biocompatibility and composition" (stated as similarity to predicate)
    - Material Safety
    Functional Equivalence:"substantially equivalent in intended use and fundamental scientific technology"
    - Similar Intended Use"similar in intended use, indications for use, materials, design, and manufacturing process"
    - Similar Indications for Use"Similar to predicate"
    - Similar Materials"Similar to predicate"
    - Similar Design"Similar to predicate"
    - Similar Manufacturing Process"Similar to predicate"

    Note: The document provides a high-level summary that "all components met the required specifications for the completed tests" and that "The results for all testing were passing." Specific quantitative acceptance criteria values (e.g., minimum load force, cycle count) and reported numerical performance metrics are not provided in this summary.


    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified in the document. The document states "Mechanical testing performed on the device and all components" but does not give sample sizes (e.g., number of anchors tested for each type of mechanical test).
    • Data Provenance: Not specified. For mechanical testing of a manufactured physical device, the data would typically originate from in-vitro lab testing, not patient data. The document does not mention the country of origin of the data or whether it was retrospective or prospective, as these concepts are generally not applicable to the type of testing described for this device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable to this type of device. The "ground truth" for a physical medical device like a suture anchor is established through engineering specifications, material science standards, and biomechanical testing standards, not expert consensus on diagnostic images or patient outcomes in the way an AI/ML device would require. Performance is measured against physical test standards (e.g., ASTM, ISO standards for implantable devices), not judgments made by experts in a conciliation process.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human readers or evaluators to resolve discrepancies in interpretation (e.g., in radiology image reading). For the mechanical testing of a suture anchor, performance is objectively measured against pre-defined engineering and material specifications. There is no subjective interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. This device is a physical medical implant (suture anchor), not an AI-assisted diagnostic or treatment planning tool. Therefore, MRMC studies and the concept of human readers improving with AI assistance do not apply here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. This device is a physical medical implant, not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device's performance is based on predefined engineering specifications, material properties, and biomechanical testing standards (e.g., tensile strength, fatigue life, pull-out strength in a synthetic bone model). This is not derived from expert consensus, pathology, or outcomes data in the usual sense. The document states "all components met the required specifications," implying these engineering-defined benchmarks served as the ground truth.

    8. The sample size for the training set

    • This question is not applicable to this device. A "training set" refers to data used to train an AI/ML algorithm. This device is a manufactured physical product. Its design and validation rely on engineering principles, material science, and mechanical testing, not a data-driven training process in the AI/ML context.

    9. How the ground truth for the training set was established

    • This question is not applicable, as there is no "training set" for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1