Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213867
    Manufacturer
    Date Cleared
    2023-03-28

    (473 days)

    Product Code
    Regulation Number
    868.5160
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K000706, K122445, K170872, K131945, K151570

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carestation750/750c anesthesia systems are intended to provide monitored anesthesia care, general inhalation anesthesia and/ or ventilatory support to a wide range of patients (neonatal, pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics. The systems are intended to be operated by a clinician qualified in the administration of general anesthesia.

    Device Description

    The GE Carestation 750/750c anesthesia machines (Carestation 750 series) are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonate, pediatric, and adult). The anesthesia systems are suitable for use in a patient environment such as hospitals, surgical centers, or clinics. They represent one of the systems in a long line of products based on the Datex-Ohmeda Aestiva (K000706), Aespire View (K122445), Aisys CS2 (K170872), Avance CS2 (K131945), Carestation 620/650/650c (Carestation 600 series) (K151570) Anesthesia Systems. The Carestation 750 series anesthesia systems are intended to be operated by a clinician qualified in the administration of general anesthesia.

    The Carestation™ 750/750c anesthesia systems combine advanced anesthesia delivery, patient monitoring. and care information management. The contemporary, compact design allows for easy mobility and addresses many ergonomic considerations including an effective cable management solution, aesthetic covers, and an expandable work surface area. Optional integrated features include auxiliary common gas outlet, auxiliary 02 outlet, auxiliary 02+Air outlet, suction control, and respiratory gas monitoring. The system provides integration of ventilation and gas delivery on a 15-inch color graphical touchscreen interface. This system also features electronic gas mixing of oxygen and a balance gas of either N2O or Air. The Carestation 750 series represents one of the systems in a long line of products based on the Datex-Ohmeda Aestiva (K000706), AespireView (K122445), Aisys CS2 (K170872), Avance CS2 (K131945), and Carestation 600 Series (K151570) Anesthesia Systems.

    This anesthesia system is designed for mixing and delivering inhalation anesthetics (Isoflurane, Sevoflurane, or Desflurane), Air, O2, and N2O. This anesthesia system has Recruitment maneuvers, a feature to perform automated lung recruitment maneuver in a single step or in multi steps.

    This anesthesia system uses electronic flow valve ventilation technology offering Volume Control Ventilation with tidal volume compensation and electronic PEEP. This technology also features Pressure Control Ventilation, optional Pressure Support Ventilation with an Apnea Backup (PSVPro™) that is used for spontaneously breathing patients, Synchronized Intermittent Mandatory Ventilation (SIMV) modes, Pressure Control Ventilation-Volume Guarantee (PCVG), Continuous Positive Airway Pressure + Pressure Support Ventilation (CPAP + PSV), and VCV Cardiac Bypass. In Volume Control Ventilation, a patient can be ventilated using a minimal tidal volume of 20 ml. In Pressure Control Ventilation, volumes as low as 5 ml can be measured. These advanced features allow for the ventilation of a broad patient range. The device includes the following basic components:

    The Carestation 750 series anesthesia systems supply set flows of medical gases to the breathing system using an electronic gas mixer (O2 with Air or O2 with N2O). Gas flows are adjusted by the user on the touchscreen, the flows are displayed on the system graphical user interface assembly as numerical digits and as electronic representations of flow meters. The system provides an option for auxiliary mixed Oxygen + Air flow delivery where 02 with Air are blended and delivered to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. An optional auxiliary O2 supply includes a separate O2 flow tube and needle valve flow control that delivers O2 flow to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. The gas flow from the optional auxiliary O2 subsystem does not flow through the electronic gas mixer.

    The Carestation 750 series models include up to 3 breathing gases with O2 and Air as standard, and N20 as an optional breathing gas. The systems include two vaporizer positions compatible with, Isoflurane, Sevotlurane, and Desflurane vaporizers. The Carestation 750 is available with up to three back-up gas cylinder connections. The Carestation 750 series systems are also available in pendant (Carestation 750c) models.

    The system uses touchscreen technology, hard keys, and a Comwheel to access system functions, menus, and settings on a 15'' color graphical user interface assembly (aka display). The graphical user interface assembly is mounted on an arm on the left side of the machine. It can be rotated via the arm toward, or away from, the system to adjust the horizontal position. The arm is available allowing the display to be tilted up or down to adjust the vertical viewing angle or be tilted left or right to adjust the horizontal position of the display. An optional arm can be raised/lowered and rotated 360 degrees. A split screen field can be set to show gas trends, Spirometry loops, a Paw gauge, airway compliance, and optional ecoFlow information. If none is selected, the waveforms expand to fill the split screen area.

    The Carestation 750 series systems accept Tec 7, and Tec 820/850 series vaporizers on a 2position Selectatec manifold. Safety features and devices within the systems are designed to decrease the risk of hypoxic mixtures, multiple anesthetic agent mixtures, complete power failure, or sudden gas supply The Carestation 750 series systems are available with optional integrated respiratory gas failures. monitoring which can be physically integrated into the system, receive electronic power from the Carestation 750/750c, and communicate measured values to the Carestation 750/750c for display on the system graphical user interface assembly. When supplied as an option, integrated respiratory gas monitoring is provided via the GE CARESCAPE series (EsCAiO or E-sCAiOV) respiratory airway modules (GE Healthcare Finland Oy, CE 0537) which is identical to the module used on Avance CS2.

    The Anesthesia Ventilator used in the Carestation 750 series is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. This version of the GE 7900 ventilator is equipped with a built-in system for monitoring inspired oxygen (using an optional 02 cell or optional integrated gas module), patient airway pressure, and exhaled volume. Flow sensors in the breathing circuit are used to monitor and control patient ventilation.

    This allows for the compensation of gas and tubing compression losses, fresh gas contribution, and small gas leakage from the breathing absorber, bellows, and pneumatic system connections. User settings and microprocessor calculations control breathing patterns. The user interface keeps ventilation settings in memory. The user may change settings with a simple ventilation parameter setting sequence. A bellows contains breathing gasses to be delivered to the patient and provides a barrier keeping patient gas separate from the ventilatory drive gas. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that occurs is outward from the patent breathing circuit.

    This ventilator comes with a standard ventilation mode as well as optional ventilation modes.

    Standard ventilation modes:

    • VCV (Time Cycled, Volume Controlled Ventilation) .
    • . PCV (Time Cycled, Pressure Control Ventilation)

    Optional ventilation modes:

    • VCV-SIMV (Synchronized Intermittent Mandatory Ventilation Volume Control)
    • . PCV-SIMV (Synchronized Intermittent Mandatory Ventilation Pressure Control)
    • PSVPro (Pressure supported ventilation with apnea backup)
    • . PCV-VG (Pressure Controlled ventilation - Volume Guaranteed)
    • . PCV-VG-SIMV (Synchronized Intermittent Mandatory Ventilation, Pressure Controlled ventilation - Volume Guaranteed)
    • . CPAP+PSV (Continuous Positive Airway Pressure/Pressure Support)

    The system can include an internal, factory installed, suction regulator and control visible from the front of the machine. It can mount different monitors using an arm or shelf mounts. The mounting is achieved through a combination of GE Healthcare adapters and other third-party mounts, including one that allows for the physical integration of the GE Monitor Series B650 (K102239). The Carestation 750 system also includes an optional cable management solution, which can help user to manage the various cables attached to the system.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Carestation 750/750c anesthesia system. It primarily focuses on demonstrating substantial equivalence to a predicate device (Carestation 620/650/650c) through technological characteristic comparison and bench testing against recognized standards.

    Therefore, the document does not contain the kind of detailed information typically found in a study proving a device meets acceptance criteria related to a specific performance metric or clinical outcome, especially for AI or algorithmic performance. There is no information about:

    • Specific acceptance criteria for device performance in terms of diagnostic accuracy or clinical effectiveness. The acceptance criteria mentioned are related to compliance with quality assurance measures and recognized safety standards.
    • A study that proves the device meets acceptance criteria in a clinical setting with patient data, experts, or specific performance metrics like sensitivity, specificity, or effect sizes.
    • Sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, or standalone algorithm performance.
    • Ground truth types or sample sizes for training sets in the context of AI/algorithms.

    Based on the provided document, here's what can be extracted regarding acceptance criteria and performance, albeit in a different context than what might be expected for an AI-driven device:

    1. Table of Acceptance Criteria and Reported Device Performance (based on compliance criteria):

    Acceptance Criteria (Compliance with Standards/Testing)Reported Device Performance (as stated in the document)
    Risk AnalysisPerformed
    Requirements ReviewsPerformed
    Design ReviewsPerformed
    Testing on unit level (Module verification)Performed
    Integration testing (System verification)Performed
    Performance testing (including accuracy, environmental, tip, threshold testing)Performed
    Biocompatibility Testing (PM, VOC, leachables)Performed (Classified as Limit exposure based on ISO18562-1:2017)
    Safety testing (electrical safety, EMC)Performed
    Simulated use testing (Validation)Performed
    Compliance with ANSI AAMI ES60601-1:2005/(R)2012Compliant
    Compliance with IEC 60601-1-2:2014Compliant
    Compliance with IEC 60601-1-6 Edition 3.1 2013-10Compliant
    Compliance with IEC 60601-1-8 Edition 2.1 2012-11Compliant
    Compliance with ISO 80601-2-13:2011Compliant
    Compliance with IEC 62366-1 Edition 1.0 2015-02Compliant
    Compliance with IEC 62304 Edition 1.1 2015-06Compliant
    Compliance with ISO 18562 series (parts 1, 2, 3, 4) 2017Compliant (e.g., "Classified as Limit exposure based on ISO18562-1:2017")
    Compliance with AIM 7351731 Rev. 2.00 2017-02-23Compliant
    Compliance with ISO 17664:2017Compliant

    The study that proves the device meets the acceptance criteria is described as:

    • Bench testing: "Bench testing was performed to establish substantial equivalence of the Carestation 750/750c."
    • Verification and validation testing: "Verification and validation testing was performed according to predetermined acceptance criteria, which concluded that the Carestation 750/750c is substantially equivalent to the predicate Carestation 620/650/650c."
    • Non-clinical design verification and validation tests: "The Carestation 750/750c anesthesia machines incorporate modifications to the predicate Carestation 620/650/650c. These modifications did not require clinical testing. The changes made were completely evaluated by non-clinical design verification and validation tests to verify and validate the safety and functionality of the anesthesia machines."

    Regarding the specific questions about AI/algorithmic studies, the document provides no relevant information:

    1. Sample size used for the test set and the data provenance: Not applicable/Not provided. The testing focused on engineering validation and adherence to standards, not on a clinical test set with patient data for algorithmic performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/Not provided. Ground truth in the context of clinical expert review is not mentioned.
    3. Adjudication method: Not applicable/Not provided.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an anesthesia machine, not an AI-assisted diagnostic tool for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device itself is an anesthesia machine, not a standalone algorithm.
    6. The type of ground truth used: For the engineering and standards compliance, the "ground truth" would be the specifications and requirements of the standards themselves, as well as validated engineering measurements against design specifications.
    7. The sample size for the training set: Not applicable/Not provided. This is not an AI/ML submission that would typically involve training sets of data for model development.
    8. How the ground truth for the training set was established: Not applicable/Not provided.

    In summary, this document is a regulatory submission for an anesthesia machine, which relies on demonstrating equivalence to an existing device through rigorous engineering testing and compliance with established medical device standards. It does not describe an AI or algorithm evaluation study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151570
    Manufacturer
    Date Cleared
    2016-01-21

    (225 days)

    Product Code
    Regulation Number
    868.5160
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K000706, K122445, K123195, K133576, K023366, K102239

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carestation 620/650/650c anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics. The systems are intended to be operated by a clinician qualified in the administration of general anesthesia.

    Device Description

    The GE Carestation 620/650/650c anesthesia machines (Carestation 600 series) are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric and adult). The anesthesia systems are suitable for use in a patient environment such as hospitals, surgical centers, or clinics. They represent one of the svstems in a long line of products based on the Datex-Ohmeda Aestiva (K000706). Aespire View (K122445) and Avance CS2 (K131945) Anesthesia Systems. The Carestation 600 series anesthesia systems are intended to be operated by a clinician qualified in the administration of general anesthesia.

    The Carestation 600 series anesthesia systems supply set flows of medical gases to the breathing system using needle valve flow controlled gas mixing (O2 with Air, O2 with N2O, or O2 with Air and N2O). Gas flows are adjusted by the user using needle valve control knobs on the main system frame, the flows are displayed on the system display unit as numerical digits and as electronic representations of flow meters. Some models (Carestation 650, 650c) also display the flow digits directly above the flow controls. Carestation 600 series systems are also equipped with an integrated pneumatic flow tube that indicates total mixed gas flow from the 2 or 3 needle valves in the gas mixer (prior to the vaporizer manifold). The Carestation 650/650c systems provide an option for auxiliary mixed Oxygen + Air flow delivery where O2 with Air are blended and delivered to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. An optional auxiliary O2 supply includes a separate O2 flow tube and needle valve flow control that delivers O2 flow to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. The gas flow from the optional auxiliary O2 subsystem does not flow through the total flow tube. An optional auxiliary common gas outlet (ACGO) allows the clinician to direct the fresh gas flow of O2. Air, N2O, or anesthetic agent, through the ACGO port on the front of the system. The ACGO may be used to provide fresh gas to an auxiliary manual breathing circuit.

    A large selection of frame options include central brake (Carestation 650) or individual caster brakes (Carestation 620), up to 3 breathing gases, optional storage drawers, and vaporizers are available to give the user control of the system configuration. The Carestation 600 series systems are also available in pendant and wall mount (Carestation 650c) models. All Carestation 600 series models have O2 gas and come with one or two additional gases (Air, N2O, or both). Carestation 600 series systems include two vaporizer positions and are available with up to three back-up gas cylinder connections.

    The system uses touchscreen technology, hard keys, and a ComWheel to access system functions, menus, and settings on a 15" color display. The display is mounted on an arm on the left side of the machine. It can be rotated via the arm toward, or away from, the system to adjust the horizontal position. An optional arm is available allowing it to be tilted up or down to adjust the vertical viewing angle, or be moved left or right to adjust the horizontal position of the display. The split screen field can be set to show gas trends, Spirometry loops, airway pressure (Paw) gauge, airway compliance, and optional ecoFLOW information. If none is selected, the waveforms expand to fill the split screen area.

    The Carestation 600 series systems accept Tec 6 Plus and Tec 7 vaporizers on a 2position Selectatec vaporizer manifold in the same way the Aestiva, Aespire View and Avance CS2 machines use the Tec 6 Plus or Tec 7 vaporizers. Features and devices within the Carestation 600 series systems are designed to decrease the risk of hypoxic mixtures, multiple anesthetic agent mixtures, complete power failure, or sudden gas supply failures. The Carestation 600 series systems are available with optional integrated respiratory gas monitoring which can be physically integrated into the Carestation 600, receive electronic power from the Carestation 600, and communicate measured values to the Carestation 600 for display on the system display unit. When supplied as an option, integrated respiratory gas monitoring is provided via the GE CARESCAPE series (E-sCAiO and E-sCAiOV) respiratory airway modules (K123195) which is identical to the module used on Avance CS2, or the N-CAiO respiratory airway module which was cleared as part of the GE B40 monitor (K133576).

    The Carestation 600 series Anesthesia Ventilator is used in the Carestation 600 series Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. This version of the GE 7900 ventilator (cleared on K023366) is equipped with a built-in monitoring system for inspired oxygen (using an optional O2 cell or optional integrated gas module), patient airway pressure and exhaled volume. Flow sensors in the breathing circuit are used to monitor and control patient ventilation. This allows for the compensation of gas and tubing compression losses, fresh gas contribution, and small gas leakage from the breathing absorber, bellows and pneumatic system connections. User setting and microprocessor calculations control breathing patterns. The user interface keeps ventilation settings in memory. The user may change settings with a simple ventilation parameter setting sequence. A bellows contains breathing gasses to be delivered to the patient and provides a barrier keeping patient gas separate from the ventilatory drive gas. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward from the patent breathing circuit.

    This ventilator comes with a standard ventilation mode as well as optional ventilation modes.

    Standard ventilation mode:
    VCV (Time Cycled, Volume Controlled ventilation)

    Optional ventilation modes:
    PCV (Time Cycled, Pressure Controlled ventilation) VCV-SIMV (Synchronized Intermittent Mandatory Ventilation Volume Control) PCV-SIMV (Synchronized Intermittent Mandatory Ventilation Pressure Control) PSVPro (Pressure supported ventilation with apnea backup) PCV-VG (Pressure Controlled ventilation - Volume Guaranteed) PCV-VG-SIMV (Synchronized Intermittent Mandatory Ventilation, Pressure Controlled ventilation - Volume Guaranteed) CPAP+PSV (Continuous Positive Airway Pressure/Pressure Support)

    The system can include an internal, factory installed, suction regulator and control visible from the front of the machine. It can mount different monitors using an arm or shelf mounts. The mounting is achieved through a combination of GE Healthcare adapters and other third party mounts, including one that allows for the physical integration of the GE Monitor Series B650 (K102239).

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and study specifically for an AI/ML powered device. Instead, it describes a medical device, the Carestation 620/650/650c anesthesia system, and its premarket notification to the FDA. The document focuses on establishing substantial equivalence to a predicate device, the Avance CS2, through non-clinical testing.

    Therefore, many of the requested details regarding AI/ML device evaluation criteria, such as sample sizes for test sets, data provenance, expert ground truth adjudication, MRMC studies, standalone performance, and training set information, are not available in this document.

    However, I can extract the general acceptance criteria for the Carestation 620/650/650c device based on the non-clinical testing performed, which is focused on verifying its safety and functionality and establishing substantial equivalence to a predicate device.

    Here's the information that can be extracted or deduced from the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document broadly states that "Verification and validation testing was performed according to predetermined acceptance criteria." While specific numerical performance metrics are not provided, the types of tests conducted serve as the "acceptance criteria" categories for this device. The reported performance is that the device met these criteria.

    Acceptance Criteria CategoryReported Device Performance
    Alarms testingMet acceptance criteria
    Breathing System testingMet acceptance criteria
    Display Unit InteractionMet acceptance criteria
    Display Unit Screens & ConfigurationMet acceptance criteria
    Electrical PowerMet acceptance criteria
    Environmental - AltitudeMet acceptance criteria
    Environmental - ShippingMet acceptance criteria
    Environmental - Temperature & HumidityMet acceptance criteria
    Gas DeliveryMet acceptance criteria
    Physical Specifications & FramesMet acceptance criteria
    PneumaticsMet acceptance criteria
    Ventilation AccuracyMet acceptance criteria
    Ventilation MonitoringMet acceptance criteria
    Ventilation DeliveryMet acceptance criteria
    ScavengingMet acceptance criteria
    Waveform ComparisonMet acceptance criteria
    Volatile Organic CompoundsMet acceptance criteria
    Particulate TestingMet acceptance criteria
    Cytotoxicity, Sensitization, Irritation/Intracutaneous ReactivityMet acceptance criteria
    Electrical Safety (AAMI / ANSI ES60601-1)Met acceptance criteria
    Electromagnetic Compatibility (IEC 60601-1-2)Met acceptance criteria
    Simulated Use Testing (Validation)Met acceptance criteria
    Software ValidationMet acceptance criteria
    Risk AnalysisConducted
    Requirements ReviewsConducted
    Design ReviewsConducted
    Unit Level Testing (Module verification)Conducted
    Integration Testing (System verification)Conducted

    2. Sample size used for the test set and the data provenance

    Not applicable. The testing described is primarily non-clinical verification and validation testing of a hardware/software system, not a study involving a "test set" of patient data for an AI/ML algorithm. No specific sample sizes for these tests are mentioned beyond the implication that sufficient testing was done. The document does not refer to any patient data for testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is not an AI/ML algorithm that requires expert consensus for ground truth. Its functionality is tested against engineering specifications and regulatory standards.

    4. Adjudication method for the test set

    Not applicable, as there is no "test set" in the context of AI/ML evaluation. Product verification and validation involve testing against predefined specifications and requirements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes an anesthesia system, not an AI-assisted diagnostic device. Therefore, no MRMC study or evaluation of human reader improvement with AI is mentioned or relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the Carestation 620/650/650c as a standalone medical device. The document explicitly states that "Extensive non-clinical testing was performed..." and that "The Carestation 620/650/650c has been thoroughly tested through verification of specifications and validation, including software validation." This means its standalone performance against design specifications was evaluated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is its compliance with engineering specifications, functional requirements, and relevant voluntary and regulatory standards (e.g., AAMI / ANSI ES60601-1, IEC 60601-1-2). The testing aims to prove that the device operates as designed for its intended use.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1