Search Results
Found 3 results
510(k) Data Aggregation
(89 days)
Scaling:
- Interdental junction treatment
- Tooth neck and subgingival treatment
- Treatment of large deposits
- Treatment of coating and tobacco stains
- Interproximal treatment
- Prosthesis conservative/restorative:
- Inlay/onlay condensation
- Amalgam plugging
- Loosening prostheses (bridge, crown, post, pivot ... )
Endodontics:
- Canal preparation
- Canal cleaning
- Canal filling
- Gutta percha condensation
- Treatment resumption
- Retro surgery
- Micro retro surgery
- Surface smoothing after burring
Periodontics:
- Root planning
- Initial therapy
- Treatment of periodontal pockets
- Treatment of furcations
- Maintenance therapy
The NEWTRON CAN A Module is an Ultrasonic Control Module intended to be integrated in a Dental Delivery System. The NEWTRON CAN A Module contains an Embedded Software. The NEWTRON CAN A Module is an Ultrasonic Control Module intended to be integrated in a Dental Delivery System manufactured by the American Company named A-DEC. The A-DEC Dental Delivery System provides different function to the NEWTRON CAN A Module such as Ultrasonic Power Setting, Light Command through a communication frame. The communication frame is designed and imposed by A-DEC, the Manufacturer of the Dental Delivery System. The values of Ultrasonic Power Setting and the Light Command are coded and communication frame. Ultrasonic Power Setting and the Light Command are driven by the user with a touch pad embedded in the A-DEC Dental Delivery System. The only parameters of the NEWTRON CAN A Module that can be modified by the A-DEC Dental Delivery System are the values of Ultrasonic Power Setting and the Light Command. The NEWTRON CAN A Module converts a low voltage Power Supply provided by the A-DEC Dental Delivery System into an Ultrasonic Electric Signal. The NEWTRON CAN A Module is intended to be used in combination with a SATELEC Dental Ultrasonic Handpiece and a SATELEC range of Dental Tips. The electrical signal emitted by the NEWTRON CAN A Module supplies the Dental Ultrasonic Handpiece equipped with SATELEC Dental Tip. The amplitude of the ultrasonic micro-vibrations depends on the electrical Signal provided by the NEWTRON CAN A Module (amplitude). The SATELEC Dental Ultrasonic Handpiece is connected to the Dental Delivery System via a Handpiece cord. Mechanical micro- vibrations are transmitted to a Dental tip attached to the end of the SATELEC Dental Ultrasonic Handpiece.
This document is a 510(k) summary for the SATELEC NEWTRON CAN-A ultrasonic scaler. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed independent study on specific acceptance criteria for device performance metrics like sensitivity, specificity, or accuracy, which are typical for diagnostic AI devices.
Instead, the "acceptance criteria" here refer to conformity with established standards for medical devices and demonstration of similar performance characteristics to the predicate device. The "study" refers to the various tests and comparisons performed to prove this conformity and similarity.
Here's an analysis based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present acceptance criteria and device performance in a typical clinical metrics table (e.g., sensitivity, specificity). Instead, it demonstrates "substantial equivalence" to a predicate device (SP NEWTRON Module, K033764) by showing that the new device (NEWTRON CAN A) has equivalent technological characteristics and meets relevant safety and performance standards.
The acceptance criteria are implied by the "Result / Impact of the differences on Safety / Effectiveness" column in Table 01, which consistently states "Identical" or "No Impact." The "reported device performance" are the characteristics of the NEWTRON CAN A.
Characteristic / Acceptance Criterion (Predicate Device Value) | NEWTRON CAN A Performance | Result / Impact on Safety/Effectiveness |
---|---|---|
Intended Use (Identical to Predicate) | Identical | Identical |
Knumber (K033764) | K233922 | Identical |
Product Code (ELC) | ELC | Identical |
Common or Usual Name (Ultrasonic Scaler) | Ultrasonic Scaler | Identical |
Classification Name (21 CFR 872.4850) | 21 CFR 872.4850 | Identical |
Regulation Identification (Ultrasonic Scaler definition) | Identical | Identical |
Regulatory Class (II) | II | Identical |
Height (mm) (33) | 36.6 | No Impact (#01) |
Depth (mm) (50) | 59.5 | No Impact (#02) |
Width (mm) (60) | 85.95 | No Impact (#03) |
Weight (g) (130) | 160 | No Impact (#04) |
Mounting Unit (Embedded in Dental Delivery System) | Embedded | Identical |
Typical Voltage (VAC) (24) | 24 | Identical |
Minimum Voltage (VAC) (21.5) | 18.8 | No Impact (#05) |
Maximum Voltage (VAC) (27.6) | 28.5 | No Impact (#06) |
Input Frequency (Hz) (50 or 60) | 50 or 60 | Identical |
Maximum Current Consumption (AC) (1) | 1.2 | No Impact (#07) |
Maximum Power Consumption (VA AC) (26) | 29 | No Impact (#08) |
Maximum Idle Current Consumption (mA AC) (100) | 100 | Identical |
Maximum Idle Power Consumption (VA AC) (2) | 2 | Identical |
Electrical Safety Class (B type) | B type | Identical |
Fire aspects (for Casing) (UL94-V0) | UL94-V0 | Identical |
Electrical Technology (NEWTRON Technology) | NEWTRON Technology | Identical |
Minimum Output Frequency (KHz) (28) | 28 | Identical |
Maximum Output Frequency (KHz) (36) | 36 | Identical |
Minimum Output Power in Handpiece (mW) (10) | 10 | Identical |
Maximum Output Power in Handpiece (W) (12) | 12 | Identical |
Minimum Output Current in Handpiece (mA) ($8 \pm 10%$) | $8 \pm 10%$ | Identical |
Maximum Output Current in Handpiece (mA) ($100 +20% -10%$) | $100 +20% -10%$ | Identical |
Power Factor (More than 0.65) | More than 0.65 | Identical |
Standard LED Ring Output Current (mA DC) (Not applicable for predicate) | 45 | No Impact (#10) |
Low Voltage LED Ring Output Current (mA DC) (800) | $135 \pm 15$ | No Impact (#11) |
Maximum Open Voltage (V DC) (5) | 14 | No Impact (#12) |
Maximum Short-circuit Current (mA DC) (1000) | 180 | No Impact (#13) |
Quantity of Connectors (2) | 3 | No Impact (#14) |
Power Supply Connector (1) | 1 | No Impact (#15) |
Command Connector (1 (one the same connector)) | 1 | No Impact (#16) |
Handpiece Connector (1) | 1 | Identical |
Type of Command / Input Setting (Digital Signal stepped in Volts 0 to 5 Volts) | Digital Values 0 to 100 Contained in the communication frame | No Impact (#17) |
Where used (Dental Office) | Dental Office | Identical |
Thermal Cooling (Natural Convection) | Natural Convection | Identical |
Operating Temperature (+10°C to +30°C) | +10°C to +30°C | Identical |
Storage Temperature (-20°C to +70°C) | 0°C to +50°C | No Impact |
Operating Humidity (30% to 75%) | 30% to 75% | Identical |
Storage Humidity (10% to 100%) | 10% to 95% | Identical |
Operating Pressure (800 hPa to 1060 hPa) | 800 hPa to 1060 hPa | Identical |
Storage Pressure (500 hPa to 1060 hPa) | 500 hPa to 1060 hPa | Identical |
Maximum Operating Altitude (Less than 2000 m) | Less than 2000 m | Identical |
Safety Standard (IEC 60601-1) | IEC 60601-1 | Identical |
EMC Standard (IEC 60601-1-2) | IEC 60601-1-2 | Identical |
Location of Fixation (Fixed in Dental Delivery System) | Fixed | Identical |
Principles of fixation (Mounting bracket fixed with 2 screws) | Mounting bracket fixed with 2 screws | Identical |
2. Sample Size Used for the Test Set and Data Provenance
The document describes performance testing based on engineering bench tests and compliance with recognized standards rather than a clinical study with a "test set" of patients or data.
- Sample Size for Test Set: Not applicable in the context of a clinical test set. The "test set" would be the device itself and its components undergoing engineering validation.
- Data Provenance: The tests are likely performed by the manufacturer, SATELEC-ACTEON GROUP, based in France (Z.I DU PHARE, MERIGNAC, Gironde, FR). The data comes from the results of the specified engineering and biocompatibility tests. This is not retrospective or prospective clinical data but rather controlled laboratory test results.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This type of information is not relevant for the provided document. The device is an ultrasonic scaler. Its performance is evaluated through compliance with electrical, mechanical, and biocompatibility standards, and comparison to a predicate device. "Ground truth" in the sense of expert consensus on medical images or patient outcomes, as would be relevant for an AI diagnostic device, is not established here. The "ground truth" for these tests are the objective measurement results compared against the requirements of the standards (e.g., maximum current, voltage, biocompatibility indices).
4. Adjudication Method for the Test Set
Not applicable. There is no "adjudication method" in the context of human expert review for establishing ground truth for a test set, as this is a physical device testing submission.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. These studies are typically conducted for AI-powered diagnostic tools to assess the impact of AI assistance on human reader performance in interpreting medical images or data. The device in question is an ultrasonic scaler, a therapeutic dental tool, not an AI diagnostic system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, in a way. The "performance data" described (Electromagnetic Compatibility, Electrical Safety, Software Verification and Validation, Biocompatibility Validation, Performance Testing bench) are all standalone tests of the device's technical specifications and compliance with standards, without human interaction as part of a clinical diagnosis or treatment decision process. The "algorithm" here refers to the embedded software and control systems of the device.
7. The Type of Ground Truth Used
The "ground truth" for this submission is:
- Engineering Specifications: Objective measurements of electrical output (frequency, power, current), physical dimensions, voltage, current consumption, etc., that meet specified design parameters.
- Regulatory Standards: Compliance with recognized international standards for medical electrical equipment (IEC 60601-1, IEC 60601-1-2), software (FDA guidance), and biocompatibility (ISO 10993 series).
- Predicate Device Equivalence: The characteristics and performance being substantially equivalent to the legally marketed predicate device (SP NEWTRON Module, K033764).
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML device that requires a training set of data for a machine learning model. The "NEWTRON Technology" and "feedback system" mentioned are likely traditional control algorithms, not AI that learns from data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML model for this device.
Ask a specific question about this device
(231 days)
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
The Satelec New Device is a Dental Ultrasonic Generator. The Satelec New Device uses Piezoelectric Technology. The Satelec New Device uses a Satelec Dental Piezoelectric Handpiece (NEWTRON SLIM or NEWTRON SLIM B.LED). The Satelec New Device is designed to be used with Satelec Dental Tips previously cleared with Predicate Devices (K050895, cleared April 20, 2005 and K113430, cleared February 23, 2012).
The provided text describes the regulatory submission for the Satelec NEWTRON P5 and NEWTRON P5 B.LED ultrasonic scalers, focusing on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific acceptance criteria through novel performance data.
Here's an analysis of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative "acceptance criteria" for the new device, nor does it present device performance against a pre-defined set of criteria in a table format. Instead, it relies on demonstrating similarity to legally marketed predicate devices.
The performance data that is reported is in the context of similarity:
Acceptance Criteria (Implied: Similarity to Predicate) | Reported Device Performance (NEWTRON P5 vs. Satelec SUPRASSON P5 NEWTRON K050895) |
---|---|
Irrigation Flow | Similar measured values |
Current delivered in the Piezoelectric Handpiece | Similar measured values |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document states, "Clinical Data is not needed for this 510(k) process." Performance data was primarily based on bench testing. Therefore, there is no sample size for a clinical test set, nor is there information on data provenance in terms of country of origin or retrospective/prospective nature for clinical data. The bench tests involved comparing measured values, so the "sample size" would relate to the number of units tested during the bench comparison. This specific number is not provided, but it would typically be a small N for engineering verification.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. Since no clinical data or human-in-the-loop performance study was conducted, there are no experts establishing ground truth for a test set in the traditional sense of medical image analysis or clinical outcomes. The "ground truth" for the performance evaluation appears to be the measured physical parameters from the predicate device.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. As no clinical test set requiring expert interpretation or adjudication was used, there is no mention of an adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an ultrasonic scaler. The submission focuses on demonstrating substantial equivalence based on technological characteristics and bench performance, not on AI assistance or human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device (ultrasonic scaler), not an AI algorithm. The performance evaluation was done through bench testing of its physical characteristics (irrigation flow, current) by comparing it to a predicate device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance evaluation in this 510(k) submission is the measured physical properties and performance characteristics of legally marketed predicate devices. The new device aims to be "similar" in these aspects. specifically:
- Bench Test Measurements: Irrigation Flow and Current delivered in the Piezoelectric Handpiece (for K050895).
- Technological Characteristics: Piezoelectricity Technology, materials (self-extinguishing material UL94V-0), user interface principles, energy sources, output energy, and light function principles (for K071424).
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
Not applicable. As this is not an AI/machine learning device, there is no training set or corresponding ground truth establishment process.
Ask a specific question about this device
(151 days)
The medical device is used in association with a dental ultrasound handpiece to which an ultrasound instrument is attached. It is designed for the treatment of scaling (prophylaxis), periodontics, endodontics and prosthesis (preservation and restoration dentistry).
The Satelec New Device is a Dental Ultrasonic Generator. The Sateler New Device uses Piezoelectric Technology.The Satelec New Device uses a Satelec Dental Piezoelectric Plezonecht - Technology. The Satelec New Device can use the same Handpiece Harroplece (NEWTRON OElin). In Option the School (K050895, cleared April 20, 2005). The Satelec New Device is designed to be used with Satelec Dental Tips previously cleared with The Satelec New Device is Gesigned to be association of the St. 430, cleared February 23, 2012) or other new Satelec Dental Tips.
The provided text is a 510(k) Summary for the NEWTRON BOOSTER, an ultrasonic scaler. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than providing new clinical study data to prove safety and effectiveness for novel devices.
Therefore, the document explicitly states that no new clinical studies were conducted to establish acceptance criteria or demonstrate device performance. Instead, the device's characteristics and performance are compared to those of already cleared predicate devices.
Here's a breakdown of why the requested information about acceptance criteria and studies cannot be fully provided from the given text:
1. A table of acceptance criteria and the reported device performance:
- This information is not present. The submission focuses on demonstrating similarity in technological characteristics, materials, design, and energy sources to predicate devices. There are no specific performance metrics or acceptance criteria presented in a table format.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not applicable as no new test set or study was conducted to generate performance data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable as no new test set was created or evaluated by experts for ground truth establishment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable as no new test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an ultrasonic scaler, not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is hardware (ultrasonic scaler), not an algorithm. Its performance is demonstrated through substantial equivalence to existing devices.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable as no new ground truth was established for a new study. The "ground truth" in this context is the established safety and effectiveness of the predicate devices.
8. The sample size for the training set:
- Not applicable as no new training set was used.
9. How the ground truth for the training set was established:
- Not applicable.
Summary of the document's approach:
The submission for the NEWTRON BOOSTER relies entirely on demonstrating substantial equivalence to two predicate devices:
- SUPRASSON P5 NEWTRON (K020892, cleared April 20, 2005)
- ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012)
The manufacturer asserts that the New Device is substantially equivalent by comparing the following:
- Intended Use: "The Satelec New Device Indications for Use is similar to the Satelec Predicate Devices."
- Technological Characteristics:
- Uses the same Piezoelectricity Technology.
- Casings are made of the same self-extinguishing material (UL94V-0).
- Same Principle of User interface.
- Uses the same or a derivated Dental Ultrasonic Handpiece (Satelec NEWTRON and NEWTRON SLIM Handpiece, which are based on the predicate device handpieces).
- Energy Source:
- Uses the same input energy source (Electric Power Supply).
- Delivers the same output energy source (ultrasonic micro-vibration).
- Delivers the same Handpiece Current Values.
- Delivers similar irrigation Flow values.
- External Structure: Very similar, made from the same raw material (UL94-V0) and uses a single function footswitch.
- Patient Contact Materials: The materials in contact with the patient are the same.
Conclusion stated in the document:
"Because of the used technologies, characteristics and performances are similar to the Satelec Predicate Devices, the characteristics of the Satelec New Device do not affect the Safety or the Effectiveness of the Predicate Devices, the Safety and Effectiveness is the same as of the Predicate Devices."