Search Results
Found 1 results
510(k) Data Aggregation
(216 days)
The troCarWash™ system consists of a reusable control unit and a disposable, sterile, single-use trocar intended to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintain a clear image of the surgical site.
The troCarWash™ system is a laparoscopic lens cleaning device which is intended to be used during any laparoscopic surgical procedure where there is a potential for contamination of the distal lens of the laparoscope. Consisting of a mains powered reusable control unit, and a single use disposable trocar, obturator, and tubing set, it is intended to maintain surgical vision by removing contaminants such as condensation, blood, peritoneal fluid, smoke, fat, and tissue smears that have contaminated the distal lens of the laparoscope during surgical procedures providing a clear image of the surgical site. The disposable portion of the system is sterilized via irradiation and has limited (
The provided document describes the troCarWash™ System, a device intended to remove visual obstructions from the distal lens of a laparoscope during surgery.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with reported device performance in the format requested. However, it details various validation and testing activities with implied acceptance criteria being successful completion of these tests according to specified standards.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Packaging Validation | The troCarWash™ system passed seal strength and package integrity testing for environmental conditioning, simulated T&D, and Accelerated Aging, complying with ASTM F88/F88M, ASTM F2096, ASTM F1980, ASTM D4332, ASTM D4169, ISO 11607-1, and ISO 16269-6. |
Sterilization Validation | The troCarWash™ system, sterilized via irradiation, successfully validated its sterilization cycle using VDmax25. This included bioburden testing, bioburden recovery, verification dose testing, product sterility testing, and method suitability testing, complying with ANSI AAMI ISO 11137-1, ANSI AAMI ISO 11137-2, AAMI/ISO 11737-1, and AAMI/ISO 11737-2. |
Biocompatibility Testing | For its limited ( |
Ask a specific question about this device
Page 1 of 1