Search Results
Found 1 results
510(k) Data Aggregation
(118 days)
The DePuy Mitek INTRAFIX™ Advance Tibial Sheath and Screw System is indicated for fixation of tissue including ligament, or tendon to bone during cruciate ligament reconstruction.
The INTRAFIX™ Advance Tibial Fastener System is an implant system used for fixation of tissue including ligament, or tendon to bone during cruciate ligament reconstruction. The INTRAFIX™ Advance Sheaths are available in absorbable BR and non-absorbable Polypropylene materials. The INTRAFIX™ Advance Screws are available in absorbable BR and non-absorbable PEEK materials. The INTRAFIXTM Advance Tibial Sheaths and Screws are supplied sterile ready to use.
The provided document describes the INTRAFIX™ Advance Tibial Fastener System, a medical device used for fixation of tissue during cruciate ligament reconstruction. The information indicates that this is a 510(k) premarket notification for a Class II medical device, meaning it is substantially equivalent to legally marketed predicate devices.
The document does not describe a study in the context of an AI/ML powered device, or a diagnostic product that would typically involve acceptance criteria related to accuracy metrics (e.g., sensitivity, specificity, AUC) and ground truth establishment by experts. Instead, the "study" referred to here is non-clinical testing to ensure the physical and mechanical performance of the implant.
Therefore, many of the requested points, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, and standalone performance for an algorithm, are not applicable to this type of device submission as described in the provided text.
Here's an attempt to answer the relevant questions based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document mentions "Verification activities were performed on the implant and / or its predicate. Testing assessments include pull out testing, and insertion / failure torque." However, it does not explicitly state specific acceptance criteria values (e.g., "minimum pull-out strength of X Newtons") nor specific performance values obtained from the tests. It only states that "Results of performance testing have demonstrated that the proposed devices are suitable for their intended use."
Without explicit numbers from the document, a table with specific criteria and performance cannot be created.
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the provided document. The document refers to "the implant and / or its predicate" for testing, implying physical samples of the device were tested.
- Data Provenance: Not applicable in the context of data for an AI/ML model. The "data" here would be the physical testing results from the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in the context of a physical implant's mechanical performance is usually established through standardized engineering test methods (e.g., ASTM standards) rather than expert consensus on medical images or pathology.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in expert interpretations of data, which is not relevant to mechanical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This is a submission for a physical medical implant, not an AI/ML-powered diagnostic or assistive device. Therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No. This is not an algorithm.
7. The type of ground truth used
For this device, the "ground truth" for non-clinical testing would be definitions or measurements established by recognized engineering standards and biocompatibility requirements. The document states "Verification activities were performed on the implant and / or its predicate. Testing assessments include pull out testing, and insertion / failure torque." and "The proposed device also met requirement of bacterial endotoxin testing." This implies adherence to predefined physical and biological safety standards.
8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of machine learning.
9. How the ground truth for the training set was established
Not applicable. This device does not involve a "training set."
Ask a specific question about this device
Page 1 of 1