Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K062297, K083097, K102394, K111810, K160433

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems using VAPROX HC Sterilant are intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.

    Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle.

    The V-PRO 1 Plus and V-PRO maX Sterilizers' Non Lumen Cycle can sterilize: t

    Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid, semi-rigid and flexible endoscopes.

    • The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).

    The V-PRO maX Sterilizer's Flexible Cycle can sterilize:

    Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:

    1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either:
    • A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length
      • Or two lumens with:

    • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length

    • And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length

    • The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
    1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. ++ The flexible endoscope may contain either:

    · A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length

    • Or two lumens with:
      • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
      • And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length.

    ** The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Sterilizers' Lumen Cycle can sterilize: *

    Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

    • Single channeled devices with a stainless lumen that is > 0.77 mm ID and ≤ 500 mm in length
    • Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
    • · Triple channeled devices with stainless lumens that are either:

    ≥ 1.2 mm ID and ≤ 275 mm in length ≥ 1.8 mm ID and ≤ 310 mm in length or

    2.8 mm ID and 1 mm (~3/64") ID and 1.8 mm (~5/64") ID and

    Device Description

    The V-PRO 1, V-PRO 1 Plus, V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems, using VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1, V-PRO 1 Plus, and the V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems are programmed to enact one (the Lumen Cycle), two (the Lumen Cycle and the Non Lumen Cycle) or three (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycles, respectively.

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting studies for the STERIS Amsco V-PRO Low Temperature Sterilization Systems.

    Here's the breakdown:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaResults
    Determination of Worst Case (Most Resistant) Material for VHPTest: The most resistant material (MRM) is the material that is most difficult to sterilize (the material with the largest number of non-sterile results)
    Controls: Controls shall perform as intended.Pass: As identified within previous V-PRO Sterilizer submissions, the most resistant material (MRM) for surface sterilization is Kraton.
    Medical Device Material CompatibilityThe devices shall be processed for 50 V-PRO 60 Sterilizer Lumen Cycles (worst case). The devices shall show no structural defects (crazing, cracks or chipping). The devices shall function following processing.Pass: This testing verified compatibility of the tested 20 materials with the V-PRO Sterilizers.
    Cytotoxicity Evaluation of Medical DevicesAll Worst Case Cycles shall proceed to completion. The cytotoxic endpoint for each medical device shall be determined. Cytotoxicity Assay Controls shall perfrom as intended.Pass: Testing demonstrates that various medical devices containing [unspecified, but likely referring to the materials tested in compatibility study] were determined to be non-cytotoxic following processing in three (3) V-PRO 60 Sterilizer Lumen Cycles.
    Hydrogen Peroxide Residual Analysis of Processed Medical DevicesThe residual hydrogen peroxide level for each medical device shall be determined and compared to the allowable hydrogen peroxide limits for mucosal and internal tissue contact. Each device shall demonstrate exhaustive extraction by showing less than a 10% increase for two (2) consecutive sample time points. Hydrogen Peroxide Residual Analysis Controls shall perform as intended.Pass: Testing demonstrates that the hydrogen peroxide residuals from the medical devices composed of the 19 proposed compatible materials are from 2660-fold lower to greater than 13,000-fold lower than the tolerable exposure limits for mucosal and internal tissue contact when processed in three (3), worst case, back-to-back Lumen Cycles in the V-PRO 60 Sterilizer.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document primarily describes validation studies for the V-PRO sterilization systems, which are typically performed in a prospective manner.

    • Determination of Worst Case Material for VHP: This refers to previous submissions, but the current document specifically highlights Kraton as the "most resistant material." No specific sample size is given for how many materials were evaluated to determine Kraton as the worst case in previous studies.
    • Medical Device Material Compatibility: "20 materials" were tested.
    • Cytotoxicity Evaluation of Medical Devices: "various medical devices containing [unspecified material components]" were tested.
    • Hydrogen Peroxide Residual Analysis of Processed Medical Devices: "19 proposed compatible materials" were tested across "three (3), worst case, back-to-back Lumen Cycles" in the V-PRO 60 Sterilizer.

    The provenance of the data (country of origin) is not explicitly stated, but as this is an FDA submission for a U.S. company (STERIS Corporation, Mentor, OH), it's highly probable the studies were conducted in the U.S. or at facilities compliant with U.S. regulatory standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This type of information is not applicable to the studies described. The studies are engineering and chemical validation tests for a sterilization system, not diagnostic device studies that would require expert human review or interpretation of medical images or patient data to establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable as the studies are not diagnostic assessments requiring adjudication of human reader interpretations. The acceptance criteria are based on measurable physical and chemical properties and biological effects (sterilization efficacy, material integrity, cytotoxicity, residuals).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a sterilization system, not an AI-powered diagnostic tool used by human readers. Therefore, no MRMC study or AI assistance improvement metrics are relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a sterilization system and there is no algorithm or AI component involved in its operation or performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in these validation studies is established by objective scientific and engineering measurements and standards:

    • Sterilization effectiveness: Implicit in the "Worst Case Material" determination, where the goal is sterilization. Sterilization is typically confirmed by biological indicators demonstrating the inactivation of a specific resistant microorganism (e.g., Geobacillus stearothermophilus) to a defined sterility assurance level (SAL). While not explicitly detailed here, this is a standard method for sterilization validation.
    • Material compatibility: Visual inspection for defects (crazing, cracks, chipping) and functional testing of devices after processing.
    • Cytotoxicity: In vitro assays to assess cellular toxicity, with controls to ensure assay validity.
    • Hydrogen peroxide residual analysis: Chemical analytical methods to quantify residual hydrogen peroxide, compared against established biological tolerance limits for mucosal and internal tissue contact.

    8. The sample size for the training set

    This is not applicable. The device is a sterilization system, not a machine learning model. There is no concept of a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K062297, K083097, K102394, K111810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle can sterilize:(a)

    • Lumened and non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: (a)
      • o single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • o dual channeled devices with stainless steel lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
      • o triple channeled devices with stainless steel lumens that are
        • ≥ 1.2 mm ID and ≤ 275 mm in length
        • ≥ 1.8 mm ID and ≤ 310 mm in length
        • or
        • 2.8 mm ID and 1 mm ID and 1 mm ID and 1 mm ID and 1 mm (~3/64") ID and

    Device Description

    The V-PRO 1, V-PRO 1 Plus, V-PRO maX and V-PRO 60 Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems are programmed to enact one (the Lumen Cycle), two (the Lumen Cycle and the Non Lumen Cycle) or three (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle) cycles, respectively.

    The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    AI/ML Overview

    The STERIS V-PRO Low Temperature Sterilization Systems were studied to verify two modifications: the removal of eye-contacting limitations and the addition of polyurethane as a compatible material for the Lumen Cycle.

    Here's a breakdown of the acceptance criteria and study details:

    1. Acceptance Criteria and Reported Device Performance

    Device ModificationVerification / Validation ActivityAcceptance Criteria (Implied)Reported Device Performance
    Removal of Eye-Contacting Limitations
    Compare the residual hydrogen peroxide level for materials after processing under worst-case conditions in the V-PRO Sterilizers and the STERRAD 100NX Sterilizer.Residual hydrogen peroxide levels in V-PRO processed samples should be statistically less than or equal to materials similarly processed in the predicate device (STERRAD 100NX Sterilizer).Pass: "V-PRO 60 Sterilizer (worst-case) processed samples hydrogen peroxide residues were statistically less than or equal to materials similarly processed in the STERRAD 100NX Sterilizer."
    Compare the dilution required to achieve non-cytotoxic results for materials after processing under worst-case conditions in the V-PRO Sterilizers and the STERRAD 100NX Sterilizer.The cytotoxic response for materials processed in the V-PRO Sterilizer cycles (worst-case) should be equivalent to or less than the cytotoxic response for the same materials processed in the predicate device (STERRAD 100NX Sterilizer Cycles).Pass: "The cytotoxic response for all materials processed in the V-PRO 60 Sterilizer cycles (worst-case) was equivalent to or less than the cytotoxic response for the same materials processed in STERRAD 100NX Sterilizer Cycles."
    Confirm that the cytotoxic component present in each extract (if the neat sample is cytotoxic) is hydrogen peroxide (H2O2) by the addition of catalase to the extracts.The cytotoxic component should be identified as hydrogen peroxide (H2O2)."The cytotoxic component for all V-PRO 60 Sterilizer (worst-case) processed materials was identified as hydrogen peroxide (H2O2)." (This is an observation supporting the mechanism, not a pass/fail criteria itself, but confirms understanding of the cytotoxic agent).
    Determine the cytotoxicity for medical devices after processing in three (3) back-to-back V-PRO 60 Sterilizer (worst-case) Lumen Cycles.Following exposure to three (3) back-to-back cycles, devices should be non-cytotoxic at the neat dilution.Pass: "Following exposure to three (3) back-to-back cycles, devices are non-cytotoxic at the neat dilution."
    Compare the residual hydrogen peroxide level for medical devices after processing in three (3) back-to-back V-PRO 60 Sterilizer (worst-case) Lumen Cycles to hydrogen peroxide levels known to cause eye biocompatibility concerns.Following exposure to three (3) back-to-back cycles, device hydrogen peroxide residues should be below levels known to cause biocompatibility concerns (specifically eye biocompatibility concerns).Pass: "Following exposure to three (3) back-to-back cycles, device hydrogen peroxide residues are below levels known to cause biocompatibility concerns."
    Evaluate V-PRO Sterilizer materials processed for 3 consecutive cycles for ocular irritation per ISO 10993-10.Materials coupons processed for 3 consecutive cycles in the V-PRO maX Lumen (worst-case) Cycle should be non-irritants per ISO 10993-10 evaluation.Pass: "Materials coupons processed for 3 consecutive cycles in the V-PRO maX Lumen (worst-case) Cycle are non-irritants per ISO 10993-10 evaluation."
    Addition of Polyurethane as a compatible material for V-PRO Sterilizers' Lumen Cycle
    Evaluate polyurethane material compatibility after processing the worst-case V-PRO Sterilizer Cycle.No material degradation or cosmetic changes should be observed for the polyurethane-containing device.Pass: "No material degradation or cosmetic changes were observed for the polyurethane-containing device."
    Sterilization efficacy for polyurethane in Lumen Cycle is demonstrated by previously completed testing in the V-PRO maX and V-PRO 60 Sterilizers’ (worst-case) Non Lumen Cycle.Sterilization efficacy for polyurethane should be demonstrated to meet the required sterility assurance level (SAL). (Implied from the "Pass" result, as specific SAL is not stated here but is a standard requirement for sterilization validation.)Pass: (Specific sterility efficacy data is not provided in this table, but the "Pass" indicates that prior testing supports this. This would typically involve bacterial spore inactivation.)

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state the exact "sample size for the test set" in terms of number of unique devices or materials for each specific test. However, it indicates:

    • For residual hydrogen peroxide and cytotoxicity comparison: "materials" were processed.
    • For in-device cytotoxicity and tissue residue: "medical devices" were processed.
    • For ocular irritation: "Materials coupons" were processed.
    • For polyurethane compatibility: "polyurethane-containing device" and "polyurethane material" were evaluated.

    Data Provenance: The studies were conducted by STERIS Corporation for a 510(k) premarket notification. This implies the data is prospective for the purpose of demonstrating the safety and effectiveness of the proposed changes. The country of origin of the data is not specified but is presumably the USA, where STERIS Corporation is based.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The tests described are primarily laboratory-based evaluations of chemical residues, cytotoxicity, material compatibility, and biological irritation, rather than assessments requiring interpretation by clinical experts. For biocompatibility tests like ocular irritation per ISO 10993-10, trained toxicologists or evaluators would interpret the results.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are laboratory-based experimental validations, not clinical studies requiring independent interpretation or adjudication of outcomes by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. This type of study is typically performed to evaluate the diagnostic accuracy of imaging or other interpretive devices, often with and without AI assistance to human readers. The STERIS V-PRO System is a sterilizer, and its performance is evaluated through physical, chemical, and biological efficacy studies, not reader interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone study was not done because the device itself is a sterilizer, not an algorithm. Its performance is inherent in its operation (sterilization efficacy, material compatibility, residue levels), not in an algorithmic output.

    7. Type of Ground Truth Used

    The ground truth used in these studies is based on:

    • Quantitative measurements: Residual hydrogen peroxide levels (chemical analysis).
    • Biological assays: Cytotoxicity tests (e.g., using live cells to detect toxic effects), ocular irritation tests (per ISO 10993-10 guidelines, likely involving animal models or in vitro alternatives).
    • Visual assessment: Material degradation or cosmetic changes.
    • Sterilization efficacy: Implied to be based on inactivation of biological indicators with a known sterility assurance level (SAL).

    8. Sample Size for the Training Set

    This information is not applicable as the V-PRO Sterilization System is a physical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160433
    Manufacturer
    Date Cleared
    2016-07-06

    (141 days)

    Product Code
    Regulation Number
    880.6860
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle can sterilize:(a)

    • Lumened and non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices, including single, dual and triple channeled rigid endoscopes, with the following configurations: (a)
      • o single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • o dual channeled devices with stainless steel lumens that are > 0.77 mm ID and 2.8 mm ID and 1 mm ID and 1 mm ID and
    Device Description

    The V-PRO 1 Plus and V-PRO maX Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Devices that meet the criteria described within the indications for use and composed of compatible materials are placed within the sterilization chamber and the appropriate cycle selected through the graphic user interface. The preprogrammed cycles all utilize a conditioning phase which prepares the chamber, a sterilize phase in which loaded devices are exposed to the sterilizing agent, VHP, at low pressure and an aeration phase to remove residual hydrogen peroxide. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1 Plus Low Temperature Sterilization System provides two cycle options, Lumen and Non Lumen. The sterilizer contains a 136 L welded aluminum chamber with two shelves for device loading, an onboard printer to provide cycle printouts, a cup interface where the VAPROX HC cups are loaded, a display for the graphic user interface and the components necessary to achieve the desired cvcle parameters (injection cylinder, vaporizer, vacuum pump and catalytic converter). The side panels of the sterilizer are composed of stainless steel.

    The V-PRO maX Low Temperature Sterilization System provides three cycle options, Lumen, Non Lumen and Flexible. The sterilizer contains a 136 L welded aluminum chamber with two shelves for device loading, an onboard printer to provide cycle printouts, a cup interface where the VAPROX HC cups are loaded, a display for the graphic user interface and the components necessary to achieve the desired cycle parameters (injection cylinder, vacuum pump and catalytic converter). The side panels of the sterilizer are composed of plastic.

    The purpose of this pre-market notification is to modify the maximum load weight for the Non Lumen Cycle from 19.65 to 50 lbs, to add titanium diffusion-restricted areas to the indications for use of the Non Lumen Cycle, and to remove the term "reusable" from indications for use statement.

    AI/ML Overview

    The FDA 510(k) summary for the V-PRO® 1 Plus and V-PRO® maX Low Temperature Sterilization Systems describes performance testing to support the substantial equivalence claim. Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the "PASS" result in the context of sterilization, meaning the device successfully sterilized the test articles under specified conditions. The document acts as a summary, so detailed quantitative acceptance criteria (e.g., specific log reduction of microorganisms) are not explicitly stated in this high-level summary but are inherent to sterilization validation.

    TestAcceptance Criteria (Implied)Reported Device Performance
    ½ Cycle Modified Total Kill Endpoint VerificationAll sterile results within validation load at standard and at least one lower sterilant injection weight. Partial positives/all survive at lower injection weights.PASS
    ½ Cycle Sterilization Verification of Cycle ClaimsReproducibly sterilizes non-lumened devices, mated surfaces, and device tray accessories contact sites under worst-case conditions in ½ Cycle.PASS
    Simulated Use TestVerified ability of Non Lumen Cycle to sterilize medical devices under worst-case processing conditions.PASS
    Final Process QualificationAll cycle specifications met through manual inspection of process parameter data.PASS
    Sterilizer Accessories PerformanceReproducibly sterilizes SCBIs under worst-case ½ cycle conditions; PIs exhibit passing color change; SCBIs negative for growth under full cycle conditions; Chemical indicators show complete color change; Test articles effectively sterilized when packaged as specified.PASS

    2. Sample Size for the Test Set and Data Provenance

    The document does not explicitly state a numerical sample size for the test set in terms of the number of items sterilized in total. However, it provides details on the load configurations used in the validation studies:

    • Lumen Cycle: Maximum of twenty (20) lumens per load, using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
    • Non Lumen Cycle: Validation load consisting of two instrument trays for a total weight of 50 lbs. (This is a modification from the previous 19.65 lbs of two instrument trays and two pouches).
    • Flexible Cycle (Config 1): Two flexible endoscopes (with light cord/mat) with no additional load.
    • Flexible Cycle (Config 2): One flexible endoscope (with light cord/mat) and additional non-lumened instruments (instrument tray and one pouch) for a total load weight of 24.0 lbs.

    The data provenance is not explicitly stated as retrospective or prospective, nor is a country of origin identified for the data. Given it's a pre-market notification for a medical device by the STERIS Corporation (Mentor, OH, USA), it is highly likely the studies were prospective, conducted by the manufacturer, and likely performed in the USA or under relevant international standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention the use of human experts to establish ground truth for the test set. Sterilization validation typically relies on objective measures, primarily biological indicators (BIs) containing highly resistant bacterial spores (e.g., Geobacillus stearothermophilus) and chemical indicators (CIs). The "ground truth" for sterilization is the complete inactivation of these spores, indicated by no growth.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the ground truth for sterilization is objective (growth/no growth of biological indicators), not based on human interpretation or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a sterilization system, not an AI-assisted diagnostic or interpretive device that would involve human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is a standalone system. The "algorithm" here is the pre-programmed sterilization cycle, and its performance is assessed directly against the ability to sterilize. There is no human intervention in the sterilization process itself; the human selects the cycle and loads the sterilizer. The performance data presented (e.g., ½ Cycle Modified Total Kill Endpoint Verification, Simulated Use Test) are indeed standalone performance of the device without human interpretation of the sterilization outcome itself.

    7. Type of Ground Truth Used

    The primary ground truth for sterilization efficacy is the inactivation of biological indicators (BIs). The document explicitly mentions:

    • "reproducibly sterilized under worst case ½ cycle conditions" for STERIS VERIFY® V24 Self-Contained Biological Indicator (SCBI).
    • "all SCBIs were negative for growth under worst case full cycle conditions."

    Additionally, chemical indicators (CIs) are used to confirm exposure to the sterilant, showing "complete color change."

    8. Sample Size for the Training Set

    Not applicable. This device is a physical sterilizer and its associated pre-programmed cycles, not a machine learning model that requires a "training set" in the conventional AI sense. The "training" of the cycles would refer to the developmental and engineering process of determining optimal parameters, which involves a different type of experimental design and validation than data-driven machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reasons as point 8. The "ground truth" for developing the sterilizer's parameters (e.g., sterilant injection, exposure time, pressure, temperature) would have been established through extensive empirical experimentation and adherence to sterilization standards (e.g., AAMI, ISO) to achieve a validated Sterility Assurance Level (SAL) of 10^-6 or better. This involves progressively testing parameters with biological indicators to determine the minimum conditions required for sterility.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140498
    Manufacturer
    Date Cleared
    2014-07-24

    (147 days)

    Product Code
    Regulation Number
    880.6860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K102330, K083097, K102394, K111810, K112760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 60 Sterilizer's Lumen Cycle can sterilize:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Non-lumened devices including non-lumened rigid and semi-rigid endoscopes
    • Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
      • single or dual lumen devices with stainless steel lumens that are ≥ 0.77 mm (~1/32") internal diameter (ID) and ≤ 410 mm (16-9/64") in length
      • triple lumen devices with stainless steel lumens that are
        • ≥ 1.2 mm (~3/64") ID and ≤ 275 mm (~10-55/64") in length
        • ≥ 1.8 mm (~5/64") ID and ≤ 310 mm (~12-13/64") in length
        • or ≥ 2.8 mm (~7/64") ID and ≤ 317 mm (12-31/64") in length

    The V-PRO 60 Sterilizer's Non Lumen Cycle can sterilize

    • Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    The V-PRO 60 Sterilizer's Flexible Cycle can sterilize:

    • One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:
      • single or dual lumen device with lumens that are ≥ 1 mm (~3/64") ID and ≤ 990 mm (38-63/64") in length
    Device Description

    The V-PRO 60 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1, Amsco V-PRO 1 Plus and Amsco V-PRO maX Sterilizers. As with the predicate device (K102330), the V-PRO 60 Sterilizer has three preprogrammed cycles: the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle. The V-PRO 60 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities. The V-PRO 60 Sterilizer uses VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its three preprogrammed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    AI/ML Overview

    This document describes the V-PRO® 60 Low Temperature Sterilization System, a vaporized hydrogen peroxide sterilizer. The acceptance criteria are based on achieving sterility and maintaining material compatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied)Reported Device Performance
    Sterilization Efficacy
    AOAC Sporicidal TestAll carriers processed should be sterile.All 720 carriers processed using 3 lots of EOSL sterilant were sterile.
    Determination of D-value and Total Kill EndpointGreater than a 12 log reduction of the most resistant organism.Greater than a 12 log reduction of the most resistant organism is achieved within all cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) of the V-PRO 60 Sterilizer.
    ½ Cycle Modified Total Kill Endpoint VerificationAll sterile results within the validation load.Modified total kill end point analysis was demonstrated for all three V-PRO 60 Sterilizer cycles. The standard injection weight of 1.1 g and at least one lower injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle. Partial positives or all survive results were seen at lower injection weights.
    ½ Cycle Sterilization Verification of Cycle ClaimsReproducible sterilization of specified devices/lumens.• The Lumen Cycle reproducibly sterilizes single, dual and triple lumen devices under worst case conditions in ½ Cycle.
    • The Flexible Cycle reproducibly sterilizes 1 x 990 mm flexible endoscope lumens under worst case conditions in ½ Cycle.
    • The Non Lumen Cycle reproducibly sterilizes non-lumened devices under worst case conditions in ½ Cycle.
    ½ Cycle Verification of Mated SurfacesSterile efficacy for mated surfaces under worst case conditions.Sterile efficacy was demonstrated for mated surfaces under worst case conditions.
    Simulated Use TestSterilization of medical devices under worst case conditions.Simulated use testing verified the ability of the V-PRO 60 Sterilizer cycles to sterilize medical devices under worst case processing conditions.
    In Use TestSterilization of patient-soiled, clinically-cleaned instruments.The in use investigation demonstrated the ability of the V-PRO 60 Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.
    Device Compatibility & Integrity
    BiocompatibilityBiocompatibility after processing.Cytotoxicity and residue analysis of 23 materials have demonstrated biocompatibility after processing in the V-PRO 60 Sterilizer.
    Medical Device Material CompatibilityCompatibility with materials of construction after multiple cycles.Evaluation of medical devices after multiple cycles in the V-PRO 60 Sterilizer has demonstrated compatibility with 23 materials of construction.
    Process Control & Verification
    Final Process QualificationSuccessful qualification for all cycles.The V-PRO 60 Sterilizer final process qualification was successful for all three (3) sterilizer cycles. All three lots of CI exhibited complete color change. All three SCBI PIs exhibited a passing color change and all SCBIs were negative for growth. Manual inspection of the process parameter data confirmed that all cycle specifications were met.
    Conformity to Standards (Safety and EMC)Compliance with EN 61010-1:2001 and EN61326-1:2006.The V-PRO 60 Low Temperature Sterilization System has been tested for conformity and is certified to EN 61010-1:2001 and EN61326-1:2006.

    Study Proving Device Meets Acceptance Criteria:

    The document summarizes the "Summary of Nonclinical Tests" performed on the V-PRO 60 Low Temperature Sterilization System to demonstrate its substantial equivalence to the predicate device, the V-PRO maX Low Temperature Sterilization System. The studies aim to prove the device's ability to sterilize various medical devices effectively and safely.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides specific sample sizes for particular tests:

    • AOAC Sporicidal Test: "All 720 carriers processed using 3 lots of EOSL sterilant."
    • Validation Loads:
      • Lumen Cycle: "a maximum of twelve (12) lumens per load... consisting of one instrument tray with instrument organizers and mat and two pouches for a total weight of 11 lbs (5.0 kg)."
      • Non Lumen Cycle: "a validation load consisting of one instrument tray with instrument organizers and mat and one pouch for a total weight of 12 lbs (5.4 kg)."
      • Flexible Cycle: "one flexible endoscope, packaged into a tray with silicone mat, instrument organizers and light cord (if not integral to scope) and no additional load."
    • Biocompatibility and Material Compatibility: "23 materials" were analyzed.

    Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. However, based on the nature of medical device clearance submissions to the FDA (United States Food and Drug Administration), these are typically prospective laboratory and simulated-use studies conducted by the manufacturer or authorized testing facilities, aiming to meet regulatory requirements prior to market entry.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the "number of experts" or their "qualifications" used to establish ground truth. The studies involve standardized microbiological and engineering tests rather than expert interpretation of results in a clinical setting. For sterilization studies, the ground truth is typically established by laboratory methods such as growth/no growth of biological indicators (e.g., Geobacillus stearothermophilus) following exposure to the sterilization process.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests described are objective performance evaluations (e.g., sterility, log reduction, material compatibility) rather than interpretation-based assessments that would require an adjudication method like 2+1 or 3+1.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The device is a sterilizer, and its performance is evaluated through objective physical and biological tests, not through human reader interpretation of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies described are standalone performance evaluations of the sterilization device. There is no "human-in-the-loop" interaction in the output or interpretation of the sterilization cycle results beyond operating the device and reading standard indicators. The device itself is the "algorithm only" in the context of its function.

    7. The Type of Ground Truth Used:

    The primary type of ground truth used would be:

    • Microbiological Negativity/Positivity: For sterilization efficacy tests (AOAC Sporicidal, D-value/Total Kill, ½ Cycle Modified Total Kill Endpoint, ½ Cycle Sterilization Verification, Simulated Use, In Use), the ground truth is established by culturing biological indicators or test organisms and observing the presence or absence of growth. "Sterile" or "no growth" is the ground truth for effective sterilization.
    • Physical/Chemical Analysis: For biocompatibility and material compatibility, the ground truth is established through laboratory analyses (e.g., cytotoxicity assays, residue analysis, visual inspection for degradation) and quantitative measurements.
    • Process Parameter Verification: For Final Process Qualification, the ground truth is established by objective measurements of process parameters (time, temperature, pressure, sterilant injection weight) and the observed color change of chemical indicators.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical device (sterilizer) and not an AI/machine learning model that would require a 'training set' of data in the conventional sense. The development of the sterilizer's cycles and parameters is based on engineering principles and iterative testing, not on data training in the way an algorithm learns from a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no "training set" in the context of an AI/machine learning algorithm for this device. The physical and biological principles governing sterilization, along with extensive engineering development and testing, define the performance of the sterilizer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131120
    Manufacturer
    Date Cleared
    2013-07-16

    (85 days)

    Product Code
    Regulation Number
    880.6860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K083097, K102394, K111810, K102330, K112760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco V-PRO I, V-PRO I Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.
    The Amsco V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle, the subject of this submission, can sterilize:

    • Lumened and non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
      • single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • dual channeled devices with stainless steel lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
      • triple channeled devices with stainless steel lumens that are either
        • ≥ 1.2 mm ID and ≤ 275 mm in length
        • ≥ 1.8 mm ID and ≤ 310 mm in length
        • or
        • ≥ 2.8 mm ID and ≤ 317 mm in length
    • The validation studies for all channel/lumen configurations were conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
      The Amsco V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Non Lumen Cycle, cleared under K083097, K102394 and K111810, can sterilize:
    • Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
      • The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 Ibs.
        The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, cleared under K102330 and K112760, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:
      1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.
        The flexible endoscopes may contain either:
    • a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
    • or two lumens with:
      • one lumen that is ≥ 1 mm ID and ≤ 998 mm in length
      • and the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
    • The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
      1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.
        The flexible endoscope can contain either:
    • a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
    • or two lumens with:
      • one lumen that is ≥ 1 mm ID and ≤ 998 mm in length
      • and the other lumen that is > 1 mm and ≤ 850 mm in length
    • The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 Ibs.
    Device Description

    The Amsco V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems are vaporized hydrogen peroxide sterilizers. The V-PRO Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The pre-programmed V-PRO cycles all utilize a conditioning phase, a sterilization phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Amsco V-PRO 1, V-PRO 1 Plus, and V-PRO maX Low Temperature Sterilization Systems. It focuses on updating the indications for use statement for the Lumen Cycle.

    Here's the breakdown of the acceptance criteria and the study information based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria are implicitly defined by the successful demonstration of sterilization capability under various test conditions for the device's Lumen Cycle, especially for the modified stainless steel lumen claims. The "Conclusion" column in the table below explicitly states "PASS" for all tests, indicating that these criteria were met.

    TestAcceptance Criteria (Implied)Reported Device Performance
    ½ Cycle Modified Total Kill Endpoint VerificationDemonstrate a modified total kill endpoint analysis, showing all sterile results for specified Lumen Cycle claims at standard and lower injection weights, and survival at the lowest weight.Modified total kill endpoint analysis was demonstrated. The standard injection weight of 2.1 g and a lower injection weight of 1.457 g resulted in all sterile results for the Lumen Cycle modified stainless steel lumen claims. All survival results were shown at the lowest weight evaluated (0.121 g). Partial positives results were seen at the intermediate injection weights.
    ½ Cycle Sterilization Verification of Lumen ClaimsReproducibly sterilize the modified stainless steel lumen claims under worst-case conditions in a V-PRO Lumen ½ Cycle.The V-PRO Lumen Cycle reproducibly sterilizes the modified stainless steel lumen claims under worst case conditions in a V-PRO Lumen ½ Cycle.
    ½ Cycle Sterilization Verification of Double Pouched LumensReproducibly sterilize the modified stainless steel lumen claims under worst-case conditions in a V-PRO Lumen ½ Cycle in a double pouched configuration.The V-PRO Lumen Cycle reproducibly sterilizes the modified stainless steel lumen claims under worst case conditions in a V-PRO Lumen ½ Cycle in a double pouched configuration.
    Simulated Use TestVerify the ability of the V-PRO Lumen Cycle to sterilize the modified stainless steel lumen claims under worst-case processing conditions.Simulated use testing verified the ability of the V-PRO Lumen Cycle to sterilize the modified stainless steel lumen claims under worst case processing conditions.
    In Use TestDemonstrate the ability of the V-PRO Lumen Cycle to sterilize clinically cleaned, patient-soiled medical instruments representative of the modified stainless steel lumen claims.The in use investigation demonstrated the ability of the V-PRO Lumen Cycle to sterilize clinically cleaned, patient-soiled medical instruments representative of the modified stainless steel lumen claims.
    BiocompatibilityDemonstrate biocompatibility after processing in the V-PRO Lumen Cycle through residue analysis.Residue analysis evaluation has demonstrated biocompatibility after processing in the V-PRO Lumen Cycle.
    Validation of Accessories for use with the V-PRO Lumen Cycle Modified Lumen Claims (V-PRO Sterilization Trays)Successfully sterilize the modified stainless lumens claims in V-PRO Sterilization Trays under Lumen ½ cycle worst-case conditions.The modified stainless lumens claims were successfully sterilized in the V-PRO Sterilization Trays under Lumen ½ cycle worst case conditions.
    Validation of Accessories for use with the V-PRO Lumen Cycle Modified Lumen Claims (Vis-U-All Tyvek Pouches)Successfully sterilize the modified stainless lumens claims in Vis-U-All Tyvek Pouches under ½ cycle worst-case conditions.The modified stainless lumens claims were successfully sterilized in the Vis-U-All Tyvek Pouches under ½ cycle worst case conditions.

    Study Details

    Based on the provided text, the document describes nonclinical tests primarily focused on verifying the sterilization efficacy of the device.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: The exact number of individual devices or lumens tested for each specific test is not explicitly stated in numerical form. However, for the "Lumen Cycle," the validation studies for all channel/lumen configurations were conducted using a maximum of twenty (20) lumens per load. The validation load consisted of two instrument trays and two pouches for a total weight of 19.65 lbs.
      • Data Provenance: The document does not specify the country of origin of the data. It is a 510(k) submission to the US FDA, implying the studies were conducted to meet US regulatory requirements. The studies appear to be prospective validation studies conducted by the manufacturer, STERIS Corporation, for the purpose of demonstrating the device's efficacy.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not mention the use of experts or their qualifications for establishing the ground truth. Sterilization efficacy studies like these typically rely on microbiological testing (e.g., spore-kill assays) to establish an objective "ground truth" of sterility, rather than expert human interpretation.
    3. Adjudication method for the test set:

      • Adjudication methods (like 2+1, 3+1) are typically used in studies involving human readers or subjective assessments. This document describes objective performance testing (sterilization efficacy), so such adjudication methods are not applicable and therefore not mentioned.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a study on a sterilization device, not an AI-assisted diagnostic device that would involve human readers. Therefore, an MRMC study and related effect sizes are not relevant or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense. The tests described are "standalone" performance evaluations of the sterilization system's ability to achieve sterility without direct human intervention during the sterilization cycle itself. The device is an automated sterilizer, so its performance is inherently "algorithm-only" in its operation. However, no "algorithm" in the typical AI sense is discussed.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth in these sterilization studies is microbiological sterility. The "½ Cycle Modified Total Kill Endpoint Verification" and "½ Cycle Sterilization Verification" tests indicate the use of biological indicators (spores) to definitively determine if sterilization was achieved (total kill/survival results). The "In Use Test" also mentions sterilization of "patient-soiled medical instruments," implying a microbiological assessment of sterility. Biocompatibility is assessed by "residue analysis."
    7. The sample size for the training set:

      • This document describes validation and verification studies for a physical sterilization device, not a machine learning model. Therefore, the concept of a "training set" for an algorithm is not applicable and not mentioned.
    8. How the ground truth for the training set was established:

      • As there is no training set for an AI algorithm, this information is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1