Search Results
Found 2 results
510(k) Data Aggregation
(149 days)
Pitkar External Fixation System- Rail Mechanism consist of Rail Fixation System and ALFA Fixator System. Their components are intended to be used on adult or pediatric patients as required.
A. Rail fixation System
1.Fracture fixation
2.For correcting diaphyseal deformities. metaphyseal deformities with or without shortenings
3.Correction of bony or soft tissue deformities
4.Limb Lengthening
5.For treating non-union and mal-union
6.Bone transport application
B. ALFA Fixator System
1.Open and closed fracture fixation
2. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction:
3.Pseudoarthrosis of long bones;
4.Limb lengthening;
5.Infected fractures or non-unions; and
6.Correction of long bone deformities.
Note: For Rail Fixation System and ALFA Fixator System
- For Adult Assembly: Population to be adult males and females for given indications
- For Pediatric Assembly: Population to be between age of 5 to 10 for given indications
- It is recommended to avoid operating on children under the age of five, in case of pediatric application
Pitkar External Fixation System- Rail Mechanism. The subject application has two devices covered that fall under Rail fixation category. Both the systems have Adult & Pediatric apparatus. Both the systems have different Rail design & clamps design. They are designated as: 1. Rail Fixation System 2. Alfa Fixator System Thus, there will be two subject devices for approval under this heading as both these systems have same application, device description, Indications, material composition etc.
This document describes the 510(k) premarket notification for the "Pitkar External Fixation System - Rail Mechanism." This device is intended for orthopedic procedures such as fracture fixation, deformity correction, and limb lengthening.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Test Type) | Device Performance (Reported Results) |
---|---|
Axial Load Test (ASTM F1541-02) | Demonstrated against applicable standard ASTM F1541-02 |
Torsion Test (ASTM F1541-02) | Demonstrated against applicable standard ASTM F1541-02 |
Fatigue Test (ASTM F1541-02) | Demonstrated against applicable standard ASTM F1541-02 |
Cantilever Bending Test (ASTM F1541-02) | Demonstrated against applicable standard ASTM F1541-02 |
4-Point Bending Test (ASTM F1541-02) | Demonstrated against applicable standard ASTM F1541-02 |
Biocompatibility (ISO 10993-1) | Meets ISO 10993-1 standard requirements; no further testing required. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each non-clinical test. It mentions that the device performance was "demonstrated against applicable standards ASTM F1541-02" for mechanical tests and "meets the ISO 10993-1 standard requirements" for biocompatibility.
The data provenance is retrospective, as the justification for biocompatibility relies on the device being comprised of "the same material and have the same manufacturing, body contact, and sterilization properties as other commercially available devices and materials with a long history of clinical use in orthopedic implants." The mechanical tests are also non-clinical and performed under controlled laboratory conditions rather than on patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not applicable as the described tests are non-clinical engineering and materials tests, not studies involving expert medical interpretation or diagnosis.
4. Adjudication Method for the Test Set
This information is not applicable as the described tests are non-clinical engineering and materials tests, not studies requiring adjudication of various expert opinions.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for a physical medical device (external fixation system), and the performance evaluation is based on non-clinical mechanical and biocompatibility testing, not on human interpretation of images or data.
6. Standalone (Algorithm Only) Performance Study
A standalone performance study was not done. This submission is for a physical medical device, not an algorithm or AI software. All tests conducted were on the physical components of the device.
7. Type of Ground Truth Used
The ground truth used for the performance evaluation consists of established engineering standards and material science principles.
- For mechanical tests (Axial Load, Torsion, Fatigue, Cantilever Bending, 4-Point Bending), the ground truth is adherence to the specifications and performance requirements outlined in ASTM F1541-02.
- For biocompatibility, the ground truth is compliance with the requirements of ISO 10993-1 and the demonstrated safe use of similar materials in predicate devices.
8. Sample Size for the Training Set
This information is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a training set. The device's performance is not learned from data but inherent in its design, materials, and manufacturing.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(77 days)
The Smith & Nephew, Inc. External Fixation Instrumentation is intended for use in:
- Post-Traumatic joint contracture which has resulted in loss of range of motion ( not applicable for Smith & Nephew Rail System)
- Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- Open and closed fracture fixation
- Pseudoarthrosis of long bones
- Limb lengthening by epiphyseal or metaphyseal distraction (not applicable for COMPASS Universal Hinge or JET-X Fixator)
- Correction of bony or soft tissue deformities (not applicable for COMPASS Universal Hinge)
- Correction of segmental bony or soft tissue defects
- Joint arthrodesis (not applicable for Smith & Nephew Rail System)
- Infected fractures or nonunions
- Mini external fixator systems are indicated for the management of comminuted intraarticular fractures of the distal radius (not applicable for Smith & Nephew Rail System)
- Calandruccio devices are indicated for arthrodesis of the ankle or subtalar joints. As well as some select fractures, nonunion, or osteotomy of the distal tibia; and acute transverse fractures or nonunion of the distal tibia (not applicable for Smith & Nephew Rail System)
Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. External Fixation Instrumentation. The subject devices are accessory devices and are intended to be used to assist in the implantation of the Smith & Nephew External Fixation Systems and their cleared Indications for Use. Smith & Nephew External Fixation Instruments can be organized into instrument families which are categorized as follows: Tightening/Insertion, Alignment, Guides, and Drills.
The provided text is a 510(k) Premarket Notification for the "Smith & Nephew, Inc. External Fixation Instrumentation." This submission is for surgical instruments, not a device that generates performance metrics in the way an AI diagnostic or therapeutic device would.
Therefore, the requested information regarding acceptance criteria, device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, and ground truth establishment is not applicable to this type of regulatory submission.
This 510(k) submission establishes substantial equivalence to legally marketed predicate devices based on similarities in:
- Raw materials
- Manufacturing processes
- Sterilization procedures
- Nature of body contact
- Design and function
The FDA's review for this type of device focuses on ensuring these accessory instruments are safe and effective for their intended use alongside the cleared Smith & Nephew External Fixation Systems, and that they are substantially equivalent to existing predicate instruments. It does not involve a performance study with quantitative acceptance criteria as would be expected for a diagnostic or AI-driven device.
Ask a specific question about this device
Page 1 of 1