Search Results
Found 4 results
510(k) Data Aggregation
(58 days)
The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerosis disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS SOLID, CLASSIC and MICRO crowns support removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. *The 2.00 Max Crown has not been tested to support removal of stenovenous dialysis fistulae (AV shunt).
The Stealth 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS SOLID, CLASSIC and MICRO crowns support removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. *The 2.00 Max Crown has not been tested to support removal of stenotic materiovenous dialysis fistulae (AV shunt).
The Diamondback 360® Peripheral Orbital Atherectorny System Exchangeable Series is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS SOLID, CLASSIC and MICRO crowns support removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. *The 2.00 Max Crown has not been tested to support removal of stenovenous dialysis fistulae (AV shunt).
The OAS is designed to remove or reduce occlusive material and restore luminal patency by using an orbiting, diamond-coated, eccentrically mounted crown. The document lists the main components of the OAS, including a reusable saline pump, a single-use orbital atherectomy device, a single-use atherectomy lubricant, and a single-use atherectomy guide wire.
The provided text (K220568 510(k) Summary) describes a medical device, the Diamondback 360® Peripheral Orbital Atherectomy System, and its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study proving that the device meets specific acceptance criteria related to AI or algorithmic performance.
The document outlines:
- Device Description: A percutaneous orbital atherectomy system designed to remove occlusive material from peripheral arteries and arteriovenous dialysis fistulae.
- Comparison to Predicate Device: The current device is "identical" to the predicate device in terms of indication for use, vessel diameter range, principles of operation, sterilization methods, single-use nature, and performance specifications. The only described difference is a change in the diamond size used in the crown coating (from 30-micron to 70-micron diamonds). A minor update to the Indications for Use is mentioned, stating that the 2.00 Max Crown has not been tested for removal of stenotic material from AV shunts.
- Testing Conducted: Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility) and bench testing (stall, tight stenosis, life, coating integrity, tensile verification, orbit characterization, temperature and flow verification, particulate testing, corrosion testing, GLP animal testing). These tests are stated to "verify that the design meets all product specifications and address the potential safety hazards."
Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving algorithmic performance because the document does not discuss:
- Acceptance criteria for an AI/algorithmic medical device.
- A test set, training set, or data provenance for an AI/algorithmic study.
- Expert involvement for ground truth establishment or adjudication for an AI/algorithmic study.
- MRMC comparative effectiveness studies or standalone algorithm performance.
- The type of ground truth used in the context of AI/algorithmic performance.
The document is a 510(k) submission for a physical medical device (atherectomy system) that functions mechanically, not through an AI/algorithmic component. The testing described is typical for a mechanical medical device to demonstrate safety and effectiveness and substantial equivalence to a predicate device.
Ask a specific question about this device
(16 days)
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
The Diamondback 360 Peripheral Orbital Atherectomy System (OAS) is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.
The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Diamondback 360 uses an eccentrically rotating sanding surface (crown) to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body.
The Diamondback 360 Peripheral Orbital Atherectomy System consists of the following components:
-
- Orbital Atherectomy Device (OAD)
-
- Atherectomy Guide Wire (e.g., ViperWire Advance)
-
- Saline Infusion Pump (SIP)
-
- Atherectomy Lubricant (e.g., ViperSlide)
Here's an analysis of the provided text regarding the acceptance criteria and study for the Diamondback 360® Peripheral Orbital Atherectomy System:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly list quantitative acceptance criteria with specific numerical targets. Instead, it states that the device "met the established specifications necessary for consistent performance during its intended use" and "met all predetermined acceptance criteria of design verification and validation testing as specified by applicable standards, test protocols, and/or customer inputs."
However, based on the performance bench testing conducted, we can infer the categories of acceptance criteria:
Acceptance Criteria Category | Reported Device Performance |
---|---|
Biocompatibility | All tests (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility) showed materials are non-toxic and non-sensitizing, consistent with intended use. |
Durability/Life Cycle | System Life Testing met established specifications. |
Functionality (High Load) | Stall Testing met established specifications. |
Compatibility | Introducer Compatibility Testing met established specifications. |
Fluid Management | Contrast Injection Testing met established specifications. |
Thermal Performance | Temperature Testing met established specifications. |
Mechanical Strength | Tensile Testing met established specifications. |
Mechanical Flexibility | Flexibility Testing met established specifications. |
Power/Movement Delivery | Delivered Torque Testing met established specifications. |
Operational Mechanism | Orbit Testing met established specifications. |
Safety Mechanism | Guide Wire Brake Testing met established specifications. |
Fluid Dynamics | Flow Testing met established specifications. |
Navigation/Advancement | Track Testing met established specifications. |
Packaging Integrity | Packaging/Simulated Distribution Testing met established specifications. |
2. Sample Size Used for the Test Set and Data Provenance:
The document describes bench testing and biocompatibility testing, not clinical trials involving human subjects or data sets in the traditional sense of AI/diagnostic device evaluation. Therefore, there is no "test set" of patient data, nor is there information on data provenance (country of origin, retrospective/prospective) for such a test set. The "test set" here refers to the physical devices and materials undergoing various engineering and biological assessments.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
This is not applicable as the study involved bench testing and biocompatibility testing of a physical medical device, not the evaluation of an algorithm against a ground truth established by experts.
4. Adjudication Method for the Test Set:
This is not applicable for the same reasons as #3. The results of the mechanical and biological tests are objective measurements, not subject to expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is an orbital atherectomy system, a physical interventional medical device, not an AI or diagnostic tool that assists human readers. No MRMC study was conducted or would be relevant for this type of device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable. The device is a physical orbital atherectomy system, not a software algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the performance evaluation of this device is based on established engineering specifications, material science standards, and biological safety standards. For example:
- Biocompatibility: Conformance to ISO standards for biocompatibility, assessed through laboratory tests (cytotoxicity, sensitization, etc.).
- Mechanical Performance: Measurements against predefined engineering specifications for torque, tensile strength, flexibility, etc., derived from industry standards and internal design requirements.
- System Life: Durability testing against a specified number of cycles or operational hours.
8. The Sample Size for the Training Set:
This is not applicable. The device is a physical medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set was Established:
This is not applicable for the same reason as #8.
Ask a specific question about this device
(67 days)
The Stealth 360° Orbital PAD System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
The Stealth 360° Orbital PAD System is an orbital atherectomy system (OAS) that is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.
The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Stealth 360° uses an electrically driven shaft to apply a diamond coated, eccentrically rotating surface to ablate stenotic material. The stenotic particles that are removed are small enough to be absorbed by the body.
The Stealth 360° Orbital PAD System consists of the following components:
-
- Orbital Atherectomy Device (OAD)
-
- Atherectomy Guide Wire
-
- Saline Infusion Pump (SIP)
-
- Atherectomy Lubricant (e.g. ViperSlide)
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Stealth 360° Orbital PAD System:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Reported Device Performance (as per 510(k) Summary) |
---|---|
Mechanical/Physical Performance | "All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Stealth 360° Orbital PAD System met the established specifications necessary for consistent performance during its intended use." Specific tests include: |
- Life/Stall Testing
- Orbit Testing
- Switch Logic/End of Life Testing
- Distribution Testing
- Shelf Life Testing |
| Biocompatibility | "ISO MEM Elution Assay"
"ISO Intracutaneous Reactivity Test" |
| Electrical Safety | "Electrical Safety Testing" |
| Electromagnetic Compatibility (EMC) | "Electromagnetic Compatibility (EMC) Testing" |
| Software/Firmware Performance | "Firmware testing: Modular, Integration, Functional" |
| Overall Equivalence | "The Stealth 360° Orbital PAD System with the CSI developed PCBA met all predetermined acceptance criteria of design verification and validation testing as specified by applicable standards, test protocols, and/or customer inputs. Testing results demonstrate that the Stealth 360° Orbital PAD System with the CSI developed PCBA is substantially equivalent to the legally marketed predicate device and does not raise any new safety or effectiveness questions." |
2. Sample Size Used for the Test Set and Data Provenance
The provided 510(k) summary focuses on performance bench testing rather than clinical study data involving human subjects or extensive retrospective/prospective data. Therefore, the concept of "test set sample size" in the context of patients/data provenance as typically understood for AI/diagnostic devices is not directly applicable here.
- Sample Size: Not explicitly stated in terms of individual devices tested for each bench test, but implied to be sufficient for each specific test mentioned (e.g., "Life/Stall Testing," "Orbit Testing"). These would likely involve a statistically relevant number of devices or components to demonstrate reliability and performance metrics, but the exact number isn't quantified in the summary.
- Data Provenance: This relates to bench testing performed by the manufacturer, Cardiovascular Systems, Inc. The data is thus internal, from a controlled environment, rather than clinical patient data from specific countries. It is prospective in the sense that the tests were designed and executed to verify the modified device's performance against pre-defined criteria.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. This device is a mechanical atherectomy system, and the "testing" described is bench testing of physical, electrical, and firmware components, along with biocompatibility tests. The "ground truth" for these tests is established by engineering specifications, international standards (e.g., ISO), and functional requirements, not by expert interpretation of medical images or patient outcomes.
4. Adjudication Method for the Test Set
- Not applicable. As the testing involves objective, quantifiable bench tests (e.g., electrical measurements, mechanical endurance, orbital rotation, elution assays), there is no need for expert adjudication in the way it would be applied to subjective clinical evaluations or image interpretations. Test results are compared directly against pre-defined acceptance limits.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- No. An MRMC study is relevant for diagnostic devices, particularly those involving human readers interpreting outputs (e.g., radiologists reading images with or without AI assistance). This 510(k) is for a mechanical surgical device and its modification. Clinical effectiveness is typically established via larger clinical trials, not solely MRMC studies, although this 510(k) relies on substantial equivalence primarily through bench testing. The summary does not provide any information about human readers or their improvement with AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done
- Not applicable. This device is a mechanical and electrical system, not an AI algorithm. Its performance is inherent to its design and manufacturing, and while it has "firmware," it doesn't function as a standalone diagnostic algorithm in the way this question implies for AI or software as a medical device. Its operation is always "human-in-the-loop" as it's a surgical tool operated by a physician.
7. The Type of Ground Truth Used
- For the bench tests: Engineering specifications, design requirements, and compliance with recognized industry standards (e.g., ISO for biocompatibility, electrical safety standards) serve as the "ground truth." For example, a "Life/Stall Test" would have an established number of cycles or operating hours the device must withstand before failure, and passing this threshold is the "ground truth."
8. The Sample Size for the Training Set
- Not applicable. This device does not use a "training set" in the context of machine learning. The design and validation are based on traditional engineering principles, physical testing, and prior knowledge of the predicate device.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set, this question is not relevant.
Ask a specific question about this device
(29 days)
The Stealth 360° Orbital PAD System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
The Stealth 360° Orbital PAD System is an orbital atherectomy system (OAS) that is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.
The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Stealth 360° uses an electrically driven shaft to apply a diamond coated, eccentrically rotating surface to ablate stenotic material. The stenotic particles that are removed are small enough to be absorbed by the body.
The Stealth 360° Orbital PAD System consists of the following components:
-
- Orbital atherectomy device (OAD)
-
- Atherectomy guide wire
-
- Externally operated saline pump
-
- Atherectomy Lubricant (e.g. ViperSlide)
The provided text describes the regulatory clearance of the Stealth 360° Orbital PAD System and mentions performance data used to support its substantial equivalence to a predicate device. However, it does not contain the requested information regarding acceptance criteria, a detailed study proving the device meets those criteria, or specific performance metrics in the format requested.
The submission is for a modification to an existing device where the "crown blanks" will be molded instead of machined. The primary goal of the submission is to demonstrate that this modified device is substantially equivalent to the predicate device (K110389, also the Stealth 360° Orbital PAD System).
Here's an analysis based on the provided text, highlighting what is missing for each requested point:
1. A table of acceptance criteria and the reported device performance
- Missing. The document states that "performance bench testing" was conducted, specifically "Life/Stall Testing" and "Orbit Testing," to confirm performance characteristics compared to the predicate device. However, it does not provide specific acceptance criteria (e.g., "device must operate for X hours without failure") nor report any quantitative device performance metrics from these tests. It only states that the testing "demonstrate[s] that the modified Stealth 360° Orbital PAD System is as safe and effective as the predicate device."
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Missing. No information on sample sizes for any test sets is provided. The tests mentioned ("Life/Stall Testing" and "Orbit Testing") appear to be bench tests, not clinical studies on human subjects. Therefore, data provenance related to human subjects (country, retrospective/prospective) is not applicable or provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Missing. This information is relevant for studies involving expert review or interpretation, typically in diagnostic or imaging devices. This document describes an atherectomy device and physical bench testing, not a study requiring expert ground truth establishment in this manner.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Missing. As above, this is not applicable for the type of bench testing described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Missing. This type of study is completely unrelated to the device described (an atherectomy system) and the type of testing performed (bench testing for substantial equivalence of a manufacturing change). This question is typically relevant for AI-powered diagnostic tools.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Missing. This is also irrelevant to the device and testing described. The Stealth 360° Orbital PAD System is a mechanical atherectomy device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Missing. For the "Life/Stall Testing" and "Orbit Testing," the "ground truth" would likely be engineering specifications or performance baselines established by the predicate device's performance. However, these specific "ground truths" or benchmarks are not detailed.
8. The sample size for the training set
- Missing. There is no mention of a "training set" as this is not an AI/machine learning device.
9. How the ground truth for the training set was established
- Missing. This is also irrelevant as there is no "training set."
In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence through bench testing to a predicate device, primarily to address a manufacturing change (molded vs. machined crown blanks). It does not provide the detailed study results, acceptance criteria, or expert review/AI-related information requested.
Ask a specific question about this device
Page 1 of 1