510(k) Data Aggregation
(219 days)
The software performs digital enhancement of a radiographic image generated by an x-ray device. The software can be used to process adult and pediatric x-ray images. This excludes mammography applications.
Eclipse software runs inside the ImageView product application software (also namely console software). The Eclipse image processing software II with Smart Noise Cancellation is similar to the predicate Eclipse image processing software (K180809). Eclipse with Smart Noise Cancellation is an optional feature that enhances projection radiography acquisitions captured from digital radiography imaging receptors (Computed Radiography (CR) and Direct Radiography (DR). The modified software is considered an extension of the software (it is not stand alone and is to be used only with the predicate device supports the Carestream DRX family of detectors, this includes all CR and DR detectors. The primary difference between the predicate and the subject device is the addition of a Smart Noise Cancellation module. The Smart Noise Cancellation module consists of a Convolutional Network (CNN) trained using clinical images with added simulated noise to represent reduced signal-to-noise acquisitions. Eclipse with Smart Noise Cancellation (modified device) incorporates enhanced noise reduction prior to executing Eclipse II image processing software.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Based on the provided text, the device Eclipse II with Smart Noise Cancellation is considered substantially equivalent to its predicate Eclipse II (K180809) due to modifications primarily centered around an enhanced noise reduction feature. The acceptance criteria and the study that proves the device meets these criteria are inferred from the demonstrated equivalence to the predicate device and the evaluation of the new Smart Noise Cancellation module.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly tied to the performance of the predicate device and the new feature's ability to maintain or improve upon key image quality attributes without introducing new safety or effectiveness concerns.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Diagnostic Quality Preservation/Improvement: The investigational software (Eclipse II with Smart Noise Cancellation) must deliver diagnostic quality images equivalent to or exceeding the predicate software (Eclipse II). | Clinical Evaluation: "The statistical test results and graphical summaries demonstrate that the investigational software delivers diagnostic quality images that exceed the quality of the predicate software over a range of exams, detector types and exposure levels." |
No Substantial Residual Image Artifacts: The noise reduction should not introduce significant new artifacts. | Analysis of Difference Images: "The report focused on the analysis of the residual image artifacts. In conclusion, the images showed no substantial residual edge information within regions of interest." |
Preservation/Improvement of Detectability: The detectability of lesions should not be negatively impacted and ideally improved. | Ideal Observer Evaluation: "The evaluation demonstrated that detectability is preserved or improved with the investigational software for all supported detector types and exposure levels tested." |
No New Questions of Safety & Effectiveness: The modifications should not raise new safety or effectiveness concerns. | Risk Assessment: "Risks were assessed in accordance to ISO 14971 and evaluated and reduced as far as possible with risk mitigations and mitigation evidence." |
Overall Conclusion: "The differences within the software do not raise new or different questions of safety and effectiveness." | |
Same Intended Use: The device must maintain the same intended use as the predicate. | Indications for Use: "The software performs digital enhancement of a radiographic image generated by an x-ray device. The software can be used to process adult and pediatic x-ray images. This excludes mammography applications." (Stated as "same" for both predicate and modified device in comparison chart) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The text mentions "a range of exams, detector types and exposure levels" for the clinical evaluation, and "clinical images with added simulated noise" for the CNN training.
- Data Provenance: Not explicitly stated. The text mentions "clinical images," implying real-world patient data, but does not specify the country of origin or whether it was retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Not explicitly stated. The text mentions a "clinical evaluation was performed by board certified radiologists." It does not specify the number involved.
- Qualifications of Experts: "Board certified radiologists." No specific years of experience are provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. The text mentions images were evaluated using a "5-point visual difference scale (-2 to +2) tied to diagnostic confidence" and a "4-point RadLex scale" for overall diagnostic capability. It does not describe a method for resolving discrepancies among multiple readers, such as 2+1 or 3+1.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- MRMC Comparative Effectiveness Study: Yes, a clinical evaluation was performed by board-certified radiologists comparing the investigational software to the predicate software. While it doesn't explicitly use the term "MRMC," the description of a clinical evaluation by multiple radiologists comparing two versions of software suggests this type of study was conducted.
- Effect Size of Human Readers Improvement with AI vs. without AI Assistance: The text states, "The statistical test results and graphical summaries demonstrate that the investigational software delivers diagnostic quality images that exceed the quality of the predicate software over a range of exams, detector types and exposure levels." This indicates an improvement in diagnostic image quality with the new software (which incorporates AI - the CNN noise reduction), suggesting that human readers benefit from this enhancement. However, a specific effect size (e.g., AUC improvement, percentage increase in accuracy) is not provided in the summary.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Partially. The "Ideal Observer Evaluation" seems to be a more objective, algorithm-centric assessment of detectability, stating that "detectability is preserved or improved with the investigational software." Also, the "Analysis of the Difference Images" checked for artifacts without human interpretation as the primary outcome. However, the overall "diagnostic quality" assessment was clinical, involving human readers.
7. The Type of Ground Truth Used
- Type of Ground Truth: The text implies a human expert consensus/evaluation as the primary ground truth for diagnostic quality. The "5-point visual difference scale" and "4-point RadLex scale" evaluated by "board certified radiologists" serve as the basis for assessing diagnostic image quality. For the "Ideal Observer Evaluation," the ground truth likely involved simulated lesions.
8. The Sample Size for the Training Set
- Training Set Sample Size: Not explicitly stated. The text mentions "clinical images with added simulated noise" were used to train the Convolutional Network (CNN).
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: The ground truth for training the Smart Noise Cancellation module (a Convolutional Network) was established using "clinical images with added simulated noise to represent reduced signal-to-noise acquisitions." This suggests that the model was trained to learn the relationship between noisy images (simulated low SNR) and presumably clean or less noisy versions of those clinical images to perform noise reduction. The text doesn't specify how the "clean" versions were obtained or verified, but it implies a supervised learning approach where the desired noise-free output served as the ground truth.
Ask a specific question about this device
(40 days)
The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projections wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications
The modified DRX Plus 3543C is a scintillator-photodetector device (Solid State X-ray Imager) utilizing an amorphous silicon flat panel image sensor. The modified detector is redesigned with the intent to reduce weight and increase durability, while utilizing a non-glass substrate material and cesium iodide scintillator. The modified detector, like the predicate is designed to interact with Carestream's DRX-1 System (K090318).
The modified DRX Plus 3543C Detector, like the predicate, creates a digital image from the x-rays incident on the input surface during an x-ray exposure. The flat panel imager absorbs incident x-rays and converts the energy into visible light photons. These light photons are converted into electrical charge and stored in structures called "pixels." The digital value in each pixel of the image is directly related to the intensity of the incident x-ray flux at that particular location on the surface of the detector. Image acquisition software is used to correct the digital image for defective pixels and lines on the detector, perform gain and offset correction and generate sub-sampled preview images
The provided text describes a 510(k) submission for a medical device, the Lux 35 Detector, which is a digital X-ray flat panel detector. The submission aims to demonstrate substantial equivalence to a predicate device (DRX Plus 3543 Detector). The information focuses on design modifications and non-clinical testing.
Here's an analysis of the acceptance criteria and study details based on the provided text, highlighting where information is present and where it is not:
Device: Lux 35 Detector (Carestream Health, Inc.)
Study Type: Non-clinical (bench) testing, specifically a Phantom Image Study, to demonstrate substantial equivalence of image quality to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" for image quality in a tabular format with pass/fail thresholds. Instead, it provides a qualitative comparison of image attributes. The closest interpretation of "acceptance criteria" is that the modified device's image quality needed to be "equivalent to just noticeably better than" the predicate.
Acceptance Criterion (Inferred) | Reported Device Performance (Lux 35 Detector vs. Predicate) |
---|---|
Image Detail Performance | Ratings for detail were "significantly greater than 0," indicating images were equivalent to or better than predicate. |
Image Sharpness Performance | Ratings for sharpness were "significantly greater than 0," indicating images were equivalent to or better than predicate. |
Image Noise Performance | Ratings for noise were "significantly greater than 0," indicating images were equivalent to or better than predicate. |
Appearance of Artifacts | Qualitative assessment, results not numerically quantified but implied to be equivalent or better given overall conclusion. |
DQE (Detective Quantum Efficiency) | 55% (RQA-5, 1 cycle/mm, 2.5 µGy) for Lux 35 vs. 26% (RQA-5, 1 cycle/mm, 3.1 µGy) for Predicate. This represents "improved image quality." |
MTF (Modulation Transfer Function) | 62% (RQA-5, 1 cycle/mm) for Lux 35 vs. 54% (RQA-5, 1 cycle/mm) for Predicate. This represents "improved image quality." |
Overall Image Quality Comparison | "Greater than 84% of all responses were rated 0 or higher in favor of the modified DRX Plus 3543C panel." "All ratings for the attributes (detail contrast, sharpness and noise) were significantly greater than 0 indicating that the modified DRX Plus 3543C images were equivalent to just noticeably better than the predicate images." "The image quality of the modified device is at least as good as or better than that of the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated. The text mentions "a Phantom Image Study" but does not quantify the number of images or runs.
- Data Provenance: This was a non-clinical bench testing study using phantoms. Therefore, there is no patient data or geographical provenance. The study was likely conducted at Carestream's facilities. It is a prospective study in the sense that the testing was performed specifically for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:
- Number of Experts: Not specified. The text mentions "Greater than 84% of all responses were rated 0 or higher," implying a group of evaluators, but their number is not provided.
- Qualifications of Experts: Not specified. It's unclear if these were radiologists, imaging scientists, or other relevant personnel.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not specified. The phrase "Greater than 84% of all responses were rated 0 or higher" suggests individual ratings were collected, but how conflicts or multiple ratings were aggregated or adjudicated is not detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- Answer: No. The study was a "Phantom Image Study" focused on technical image quality attributes, not human reader performance.
- Effect Size of Human Readers: Not applicable, as no MRMC study was performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Answer: Yes, in a sense. The evaluation of DQE and MTF are standalone technical performance metrics of the detector itself, independent of human interpretation. The "Phantom Image Study" also evaluates the output of the device (images) based on technical attributes, rather than a human diagnostic task.
7. The Type of Ground Truth Used:
- Type of Ground Truth: For the phantom image study, the "ground truth" for evaluating image quality attributes (detail, sharpness, noise, artifacts) is based on technical image quality metrics (DQE, MTF) and potentially expert consensus on visual assessments of phantom images against known ideal phantom characteristics. It is not based on patient outcomes, pathology, or clinical diagnoses.
8. The Sample Size for the Training Set:
- Sample Size for Training Set: Not applicable. This device is a hardware component (X-ray detector) and the study described is a non-clinical evaluation of its image quality, not an AI/algorithm that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established:
- Ground Truth Establishment for Training Set: Not applicable, as this is not an AI/algorithm that requires a training set.
Ask a specific question about this device
(29 days)
The device is designed to perform radiographic x-ray examinations on all pediatric and adult patient treatment areas.
The DRX-Revolution Mobile X-ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. The system consists of a self-contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of a standard stationary xray room. The DRX-Revolution system incorporates a flat-panel detector that can be used wirelessly for exams such as in-bed chest projections. The system can also be used to expose CR phosphor screens or films.
The Carestream DRX-Revolution Mobile X-ray System (K191025) underwent modifications compared to its predicate device (K120062). The primary changes include a different X-ray tube supplier, additional support for DRX Plus detectors, updated image acquisition software (ImageView), and a replaced high-voltage X-ray generator.
Here's an analysis of the acceptance criteria and the study proving adherence:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria | Reported Device Performance |
---|---|
Image Quality Equivalency | "Results of this data demonstrated that image quality on the modified DRX-Revolution Mobile X-ray System is equivalent to the device on the market." |
"Testing demonstrates that the modified device produces diagnostic image quality that is the same or better than the predicate." | |
"The detectors have been tested and verified to meet the requirements for integration with the DRX-Revolution Mobile X-ray System (modified) device and the DQE/MTF data demonstrates image quality is the same as or better than the predicate." | |
"The image quality of the modified device is at least as good as or better than that of the predicate device." | |
Safety and Effectiveness Equivalency | "The modified DRX-Revolution Mobile X-ray System is substantially equivalent to the predicate device currently cleared on the market (K120062)." |
"The change in X-ray tube does not significantly change the functionality of the redesigned DRX-Revolution system, nor do changes significantly affect the safety or effectiveness of the device." | |
"The generator was verified and validated and passed all testing and demonstrates there is no significant impact on clinical functionality or performance that could significantly affect safety and effectiveness." | |
"Risks were assessed in accordance to ISO 14971 and risk control options were implemented with safety by design principles and with a risk methodology that reduces risks as far as possible." | |
"Results of non-clinical testing demonstrate that the modified device is as safe and as effective as the predicate device." | |
"The subject device is expected to be safe and effective for the device indications and are substantially equivalent to the predicate." | |
Maintenance of Intended Use | "In addition, the indications for use of the modified device, as described in labeling does not change as a result of the device modification(s)." |
"The intended use remains unchanged." | |
Fundamental Scientific Technology Equivalency | "The modified DRX-Revolution employs the same fundamental scientific technology as the predicate device." |
"The fundamental scientific technology of the modified device is the same and is substantially equivalent to the predicate." | |
Hardware Components Functionality (e.g., X-ray tube) | "The change in X-ray tube does not significantly change the functionality of the redesigned DRX-Revolution system, nor do changes significantly affect the safety or effectiveness of the device." |
Detector Integration and Performance | "The detectors have been tested and verified to meet the requirements for integration with the DRX-Revolution Mobile X-ray System (modified) device and the DQE/MTF data demonstrates image quality is the same as or better than the predicate." |
Software Functionality (ImageView) | "No changes have been made between the DRX Carestream Evolution with ImageView (K163203) and the subject device with ImageView, other than some minor changes necessary for the software to function on the subject device. The image processing between the two devices is the same. This change has no clinical impact on image diagnosis, bench testing data demonstrates substantial equivalence." |
Generator Performance | "The High-voltage X-ray generator has been replaced. This generator is considered a 1:1 replacement, there was no change in performance specifications. The generator was verified and validated and passed all testing and demonstrates there is no significant impact on clinical functionality or performance that could significantly affect safety and effectiveness." |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated as a number of images or cases. The document mentions a "Phantom Imaging study."
- Data Provenance: The study was a "Phantom Imaging study," which implies the use of test phantoms rather than real patient data. This is typically done in a controlled laboratory environment. The country of origin is not specified but given Carestream's location (Rochester, New York), it is likely the US. The study type is retrospective, as it's bench testing to compare a modified device to an already marketed predicate.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. As the study was a non-clinical "phantom imaging study" evaluating technical image quality attributes, it might not have involved human expert readers establishing diagnostic ground truth in the traditional sense. The evaluation likely relied on quantitative measurements of image quality metrics.
4. Adjudication method for the test set:
- This information is not provided as the study was a phantom imaging study focusing on technical image quality. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies with human readers interpreting medical images.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical testing was not required to establish substantial equivalence. Bench testing was sufficient to assess the device safety and effectiveness." This device is a mobile X-ray system, not an AI-powered diagnostic software.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, in a sense, a "standalone" evaluation of the device's image quality was performed through the "Phantom Imaging study" and DQE/MTF data. This tested the device's inherent capability to produce images without direct human interpretation for diagnostic purposes, focusing on technical image quality attributes rather than diagnostic accuracy.
7. The type of ground truth used:
- The ground truth used was based on technical image quality attributes such as detail, sharpness, noise, and appearance of artifacts, as evaluated through a "Phantom Imaging study" and by DQE/MTF data. This is an objective technical assessment against established metrics for image quality, rather than a clinical ground truth like pathology or expert consensus on a diagnosis.
8. The sample size for the training set:
- This information is not applicable/not provided. This device is a hardware X-ray system with standard image processing software, not an AI/Machine Learning algorithm that undergoes a "training" phase with a large dataset. The "ImageView" software is a web-based application to improve usability, and its image processing is stated to be the same as previously cleared versions.
9. How the ground truth for the training set was established:
- This information is not applicable/not provided for the same reasons as point 8.
Ask a specific question about this device
(26 days)
The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
The DRX-Evolution is a diagnostic x-ray system utilizing digital radiography (DR) technology. The DRX-Evolution is designed for horizontal and upright projections. It consists of a high frequency x-ray generator, overhead tube crane (with x-ray tube assembly), elevating 4-way float radiographic table with detector tray, tilting Bucky receptor on an upright Wall Stand, and x-ray controls containing a power distribution unit and operator PC.
Due to the nature of the submitted document, which is a 510(k) summary for a diagnostic X-ray system, the information typically found in a study demonstrating how a device meets acceptance criteria for an AI/CADe (Computer-Aided Detection/Diagnosis) system is not present.
This document describes a conventional diagnostic X-ray system (DRX-Evolution) and its substantial equivalence to predicate devices, focusing on hardware components and intended use. It is not an AI/CADe device, and therefore, the concepts of "acceptance criteria" and a "study to prove the device meets acceptance criteria" as defined in the prompt for AI/CADe are not applicable here.
The "Discussion of Testing" mentions "Predefined acceptance criteria was met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device." However, these acceptance criteria relate to the performance of an X-ray system (e.g., image quality, radiation output, mechanical safety, software functionality), not the diagnostic performance of an AI algorithm based on sensitivity, specificity, or other metrics typically used for AI/CADe.
Therefore, I cannot provide the requested information regarding AI/CADe specific acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details, as this device does not appear to incorporate such AI technology requiring these types of evaluations.