Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K134004
    Date Cleared
    2014-02-14

    (49 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K060504, K080283, K070371, K080538, K110571, K121224

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CHM Device is a remote monitoring software solution intended to collect and store biometric data from physiological measurement devices intended for use in the home. The CHM Device also allows for the automated transmission of the biometric data to a remote secure server via an existing mobile telecommunications and/or internet infrastructure.

    The stored biometric data is accessible by clinicians for analysis and intervention. Patients can also review the stored biometric data and receive educational and motivational content from clinicians.

    The CHM Device can be used as a standalone device or in conjunction with supported patient monitoring devices, such as a glucometer, weight scale, pulse oximeter, and blood pressure monitor.

    The CHM Device is not intended for use in surgical rooms, intensive care units, intermediate or step-down units or emergency vehicles. It is not interpretive, nor is it intended for diagnosis or as a replacement for the oversight of healthcare professionals. It does not provide real-time or emergency monitoring.

    Device Description

    The CHM is a software platform for the collection and display of biometric data, primarily from externally supported patient monitoring devices, both to the patient and to the clinician. The CHM Device may also be used as a standalone device. The CHM Device uses existing Internet and telecommunications architecture (cell phones and computers) for the automated transmission of medical data to a remote secure server from where it can be viewed remotely by clinicians and patients for the purposes of storage and basic analysis. The CHM Device also provides educational and motivationalities allowing the clinician to send tasks, recommendations, surveys, and educational and motivational messages to patients.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Verizon Wireless Converged Health Management Device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Ensure changes to the software have not introduced new faults.Regression testing was performed and demonstrated that changes to the software did not introduce new faults.
    Ensure that a change to one part of the software does not affect other parts of the software.Usability testing was performed and demonstrated that changes to one part of the software did not affect other parts.
    Ensure that biometric data was transferred accurately from the Telcare and Genesis Health glucometers to the server infrastructure and into the CHM platform.Additional verification and validation activities were performed and demonstrated accurate transfer of biometric data from Telcare and Genesis Health glucometers to the server infrastructure and into the CHM platform.
    All verification and validation activities, as required by the risk analysis, were performed.All verification and validation activities, as required by the risk analysis, were performed.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    The document states that "regression and usability testing was performed" and "additional verification and verification activities were performed." However, specific sample sizes for a 'test set' (e.g., number of data points, number of users, number of transfers) are not explicitly mentioned.

    The data provenance (country of origin of the data, retrospective or prospective) is not mentioned in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    The document does not mention using experts to establish ground truth in the context of the performance data. The testing appears to be focused on software functionality and data transfer accuracy against design specifications rather than against expert interpretations of medical conditions.

    4. Adjudication Method for the Test Set:

    An adjudication method (e.g., 2+1, 3+1, none) is not applicable or mentioned given the nature of the testing described, which focuses on software functionality and data transfer accuracy rather than diagnostic performance requiring expert consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a "remote monitoring software solution" and not an interpretive or diagnostic AI. It collects and stores biometric data for clinicians to analyze. The document explicitly states: "It is not interpretive, nor is it intended for diagnosis or as a replacement for the oversight of healthcare professionals."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The performance evaluation described is a standalone algorithm-only performance assessment in the sense that the testing verified the software's ability to accurately transfer and handle data. There is no mention of human-in-the-loop performance measurement for the device's core functionality. The clinicians still analyze the data.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

    The ground truth used for the testing appears to be the expected behavior and functional requirements of the software, specifically related to accurate data transfer and functionality after changes. This is not a "medical ground truth" established by experts for diagnostic purposes but rather an engineering and software validation ground truth. For instance, for data transfer accuracy, the ground truth would be the original data from the glucometers.

    8. The Sample Size for the Training Set:

    The document does not mention a training set because the device is a data collection and management platform, not a machine learning or AI algorithm that requires training on a dataset to learn patterns or make predictions.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122458
    Date Cleared
    2013-07-30

    (351 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K060504, K080283, K070371, K080538

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CHM Device is a remote monitoring software solution intended to collect and store biometric data from physiological measurement devices intended for use in the home. The CHM Device also allows for the automated transmission of the biometric data to a remote secure server via existing mobile telecommunications and/or Internet infrastructure.

    The stored biometric data is accessible by clinicians for analysis and intervention. Patients can also review the stored biometric data and receive educational content from clinicians.

    The CHM Device can be used as a standalone device or in conjunction with supported patient monitoring devices, such as a glucometer, weight scale, pulse oximeter, and blood pressure monitor.

    The CHM Device is not intended for use in surgical rooms, intensive care units, intermediate or stepdown units or emergency vehicles. It is not interpretive, nor is it intended for diagnosis or as a replacement for the oversight of healthcare professionals. It does not provide real-time or emergency monitoring.

    Device Description

    The CHM Device is a software platform for the collection and display of biometric data, primarily from externally supported patient monitoring devices, both to the clinician. The CHM Device may also be used as a standalone device. The CHM Device uses existing Internet and telecommunications architecture (cellphones and computers) for the automated transmission of medical data to a remote server from where it can be viewed remotely by clinicians and patients for the purposes of storage and basic analysis. The CHM Device also provides educational and motivational functionalities allowing the clinician to send tasks, recommendations, surveys, and educational and motivational messages to patients.

    AI/ML Overview

    The Verizon Wireless Converged Health Management Device (K122458) is a software solution for remote monitoring, not an AI/ML diagnostic tool. Therefore, many of the typical acceptance criteria and study components requested for AI/ML devices, such as performance metrics (e.g., sensitivity, specificity), ground truth establishment by experts, and MRMC studies, are not applicable.

    The submission focuses on software verification and validation, demonstrating that the device functions as intended and safely transmits data.

    Here's a breakdown of the relevant information from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Software Functionality: The CHM Device performs appropriately per defined specifications.Software verification and validation testing, including usability validation, was performed successfully, demonstrating that the CHM Device performs appropriately per defined specifications.
    Input Requirements: The CHM Device meets all input requirements.Software verification and validation testing... demonstrating that the CHM Device... meets all input requirements.
    Intended Use: The CHM Device fulfills the device's intended use.Software verification and validation testing... demonstrating that the CHM Device... fulfills the device's intended use.
    Safety Mitigations: The CHM Device correctly incorporates all required safety mitigations.Software verification and validation testing... demonstrating that the CHM Device... correctly incorporates all required safety mitigations.

    2. Sample size used for the test set and the data provenance

    Not applicable for this type of software device. The "test set" would primarily involve internal software testing and validation rather than a clinical dataset of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for this device is its adherence to software specifications and functional requirements, which are typically established by software engineers and quality assurance personnel, not medical experts.

    4. Adjudication method for the test set

    Not applicable. Software testing for functionality does not typically involve an adjudication method by medical experts in the way an AI diagnostic tool would.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This device is a data collection and transmission platform, not an interpretive or diagnostic AI. Therefore, an MRMC study related to human reader improvement with AI assistance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the software verification and validation described is a standalone assessment of the algorithmic/software performance. It focuses on the device's ability to collect, store, and transmit data accurately and according to specifications, independent of human interpretation of that data.

    7. The type of ground truth used

    The ground truth used for this device's performance validation is its defined software specifications and functional requirements. The testing confirmed that the software behaved as designed and intended.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable. As this device is not an AI/ML algorithm requiring training data, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100524
    Date Cleared
    2010-05-24

    (89 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K080538

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polytel® APT receives data wirelessly from compatible devices to transmit over the Internet or common telephone lines. The APT is an optional accessory to other Polytel® devices, including the various GMA (glucose meter accessory) models, blood pressure meter, weight scale and spot-check SpO2. The APT is intended to aid people at home and health care professionals to review and evaluate historical blood glucose, weight and blood pressure test results, to support effective health care management.

    The APT does not measure, interpret, or make any decisions with respect to any of the data it transports. All patient medical diagnosis and treatment are to be performed under the supervision and oversight of appropriate healthcare professional. This device is not intended as a substitute for medical care.

    Any device certified by Polymap Wireless as compatible can use the Polytel® APT to forward its data.

    The device is not intended for emergency calls, and may not be used to send any real-time alarms or time-critical data. The device is not intended for use in systems set up for patients who need direct medical supervision or who might need emergency intervention.

    The APT and its compatible devices must be used in conjunction with a subscription to a compatible monitoring service. The monitoring service may distribute the APT to the patient; alternatively, the patient may acquire the APT and other compatible devices and subscribe to a compatible service.

    A list of compatible devices and services will be available on the Polymap Wireless website.

    Device Description

    The Polytel® APT device performs transmission of physiological patient information to and from wireless patient monitors and a remote data server healthcare facility over an encrypted dialup Internet connection. The APT, with its built-in telephone modem, transmits data over the public switched telephone network and the dial-up Internet.

    The APT is contained in a small plastic unit, containing two standard telephone jacks for connection to standard phone outlets, a 5V power input jack for connection to an AC adapter, and four indicator lights. The indicator lights are used to show the reception of a measurement from a transmitting device, to show when the telephone line is in use, and to indicate if there has been an error (delay) in transmission.

    Two server-side components are involved in the operation of the APT, the first providing a centralized configuration database, the second providing for data collection and translation to a standard protocol.

    The APT device is not used directly on a patient, nor attached electrically to any device that is used directly on a patient, and poses no significant risk to the patient or other people within the patient's home.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Polytel® APT device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state numerical acceptance criteria in terms of performance metrics (like accuracy, sensitivity, specificity, etc.) for the Polytel® APT. This is common for devices like data transmission gateways, where the focus is on functional verification rather than diagnostic performance. Instead, the "acceptance criteria" are implied by the nature of the testing performed, which aimed to demonstrate that the device functions as intended and is substantially equivalent to predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Transmission: Successfully transmit physiological patient information from wireless monitors to a remote data server/healthcare facility.Acceptable: Bench testing using Polymap procedures and specifications, and field testing under actual use conditions demonstrated acceptable results.
    Data Integrity: Ensure data is transmitted accurately via an encrypted dialup Internet connection.Acceptable: Implied by the overall acceptable results of testing, indicating no reported issues with data integrity.
    Compatibility: Receive data wirelessly from compatible devices (e.g., GMA models, blood pressure meter, weight scale, spot-check SpO2).Acceptable: Implied by the successful field testing under actual use conditions with compatible devices.
    Safety: Pose no significant risk to the patient or other people (as it's not used directly on a patient or electrically attached to patient-contacting devices).Acceptable: Stated that the device "poses no significant risk to the patient or other people within the patient's home."
    Technological Equivalence: Similar technological characteristics (Bluetooth, dial-up modems, line-powered, 2.402-2.480 GHz frequency band) to predicate devices.Met: Explicitly stated, "The Polytel APT has technological characteristics that are very similar to those of the predicate devices, as all use Bluetooth technology and dial-up modems. All of these devices are line-powered, and each device uses the same frequency band of 2.402-2.480 GHz."
    Compliance with Performance Standards: Adherence to relevant performance standards.Acceptable: Performance standards testing was conducted, and the “results were acceptable.”

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a numerical "sample size" for the test set in terms of patients or data points. However, it mentions:

    • Bench Testing: Conducted using "Polymap procedures and specifications."
    • Field Testing: Conducted "under actual use conditions."

    The data provenance is not explicitly stated in terms of country of origin, nor is it explicitly labeled as retrospective or prospective. Given the nature of "field testing under actual use conditions," it would inherently be prospective, simulating real-world usage.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., expert consensus on a medical image) is not applicable here. The Polytel® APT is a data transmission device, not a diagnostic tool. Therefore, there's no mention of experts establishing a ground truth for a test set in the traditional sense of medical diagnosis. The "truth" in this context is whether the data is transmitted correctly and reliably.

    4. Adjudication Method for the Test Set

    No adjudication method is described, as it's not relevant for a device whose primary function is data transmission. The evaluation would have been based on successful transmission, data integrity, and functional operation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for evaluating the impact of an AI algorithm on human reader performance, typically in diagnostic tasks. The Polytel® APT is a data transmission gateway, not an AI diagnostic tool, so such a study would not apply.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in essence. The "testing" described (bench testing, field testing, and performance standards testing) evaluates the device's functional performance in a standalone capacity, without direct human intervention in its data transmission function. The device's role is to automate data transfer.

    7. The Type of Ground Truth Used

    As mentioned in point 3, the concept of "ground truth" (expert consensus, pathology, outcomes data) doesn't directly apply here for a data transmission device. Instead, the "ground truth" for the testing would have been:

    • Expected Functional Behavior: The device should transmit data correctly and reliably according to its design specifications.
    • Successful Data Reception: Data sent by the APT should be received accurately by the remote server/healthcare facility.
    • Compliance with Standards: Meeting established performance standards for wireless communication and data transfer.

    8. The Sample Size for the Training Set

    The Polytel® APT is a hardware device for data transmission, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI or machine learning. The device's functionality is based on its hardware design and firmware programming, not on being trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1