Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170047
    Date Cleared
    2017-09-09

    (247 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K063641, K111463, K113077, K073152, K982140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The All-in-One Health Monitor PC-303 is a device designed for spot-checking measuring of the patient's physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP);

    Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement.

    This device is applicable for Adult and Pediatic (age>3 years old) use in clinical institutions and has no conditions or factors of contraindication.

    Device Description

    The All-in-One Health Monitor, PC-303 is mobile modular designed patient monitor. It monitors the patient's blood oxygen saturation (SpO2) and pulse rate (PR) non-invasively by the photoelectric method. It can also measure non-invasive blood pressure (NIBP, the pressures of systolic, diastolic and mean) by the oscillating method and body temperature (TEMP) by the infrared radiation energy technology.

    Additionally, the device is also capable to link and communicate with the compatible Blood Glucose Meter and ECG monitor to make the measurement.

    This device system measure the patient's physical parameters with variety modules. It takes digital signals from NIBP, SpO2, Temperature module through COM ports, and receives ECG data, blood glucose data from compatible legally marketed ECG Monitor or BLOOD GLUCOSE MONITORING Device. The software to process the data, then display parameters on the screen and storage some useful information.

    This device is applicable for use in clinical institutions and has no conditions or factors of contraindication.

    AI/ML Overview

    The provided document describes the All-in-One Health Monitor, PC-303, and its substantial equivalence to a predicate device (UP-7000 Patient Monitor, K123711). The acceptance criteria are primarily based on meeting relevant international standards for medical electrical equipment and specific performance characteristics in comparison to the predicate device.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are not explicitly stated as distinct numerical targets in a single table within the document but are inferred from the comparison with the predicate device and the adherence to recognized international standards. The "reported device performance" is the claim of compliance with these standards and the equivalent performance to the predicate.

    Parameter/CharacteristicAcceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (PC-303)
    General
    SafetyCompliance with IEC 60601-1"All the requirements of IEC 60601-1 were evaluated in this test report" (with specific exclusions)
    EMCCompliance with IEC 60601-1-2"Testing was performed to ascertain Electromagnetic Compatibility to other electrical devices... Emissions tests... Immunity tests... Harmonics current... Voltage fluctuation" - all passed.
    BiocompatibilityCompliance with ISO 10993-5, ISO 10993-10"Test article Contact Lens Case extract did not show potential toxicity to L-929 cells." "Test article extracts showed no significant evidence of causing ocular irritation." "Test article extract showed no significant evidence of causing skin sensitization."
    PerformanceEquivalent to predicate device"Each function operated within the design parameters, within the performance criteria established by the pertinent IEC standards." "established the equivalency to the predicate device."
    SpO2 Measurement
    Patient TypeAdult and PediatricAdult and Pediatric patients
    Measurement Accuracy±3% (70%-100%), Undefined (0-70%)±3% (during 70%-100%), Undefined (during 0-70%)
    Pulse Rate Range30 bpm-240 bpm30 bpm-240 bpm
    Pulse Rate Accuracy±2bpm or ±2% (whichever is greater)±2bpm or ±2% (whichever is greater)
    AlarmsFunctional alarms (Predicate has High/Low Limit)No Alarm (for SpO2 and Pulse Rate)
    NIBP Measurement
    MethodOscillometric methodOscillometric method
    Patient TypeAdult and Pediatric PatientsAdult and Pediatric patients
    Mean Deviation±5 mmHg±5 mmHg
    Standard Deviation$\le$8 mmHg$\le$8 mmHg
    Cuff Pressure Range0 to 300mmHg0 to 300mmHg
    Over Pressure ProtectionInternal operating software with limits, circuitryCuff pressure exceeds 300mmHg (Adult & Pediatric modes) at any time. (Similar to predicate's circuitry)
    AlarmsFunctional alarms (Predicate has High/Low Limit)No Alarm (for NIBP Systolic/Diastolic)
    Temperature Measurement
    Patient TypeAdult, PediatricAdult, Pediatric
    Unit of Measure°C or °F°C or °F
    Measurement SiteEar (for PC-303)Ear
    Measurement Range32.0°C to 43.0°C (90°F to 109.5°F)32.0°C to 43.0°C (90°F to 109.5°F)
    Measurement Accuracy±0.2°C (36.0-39.0°C), ±0.3°C (rest)±0.2°C (36.0°C to 39.0°C), ±0.3°C (the rest); ±0.4°F (96.8°F to 102.2°F), ±0.5°F (the rest)
    Blood Glucose (External Linked)
    IntegrationCommunication with compatible system"receives data from Blood Glucose Meter via the USB data interfaces for data communication"
    PerformanceAs per G-777G BLOOD GLUCOSE MONITORING SYSTEM (K113077)The linked device is a legally marketed system (K113077)
    ECG (External Linked)
    IntegrationCommunication with compatible system"receives data from Easy ECG Monitor via the USB data interfaces for data communication"
    PerformanceAs per PC-80 Easy ECG Monitor (K073152)The linked device is a legally marketed system (K073152)

    2. Sample size used for the test set and the data provenance

    The document states that "Clinical testing was performed for the NIBP functions and are submitted as part of this submission." However, it does not specify the sample size used for this clinical testing or the data provenance (e.g., country of origin, retrospective/prospective).

    For biocompatibility testing, the "In Vitro Cytotoxicity Test" used L929 mouse fibroblast cells, and the "Tests for irritation and skin sensitization" used guinea pigs. These are standard biological test methods, not human clinical sample sizes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number or qualifications of experts used to establish ground truth for any of the performance or clinical studies.

    4. Adjudication method for the test set

    The document does not describe any adjudication method used for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This device is a patient monitor with built-in and external measurement capabilities for physiological parameters (SpO2, NIBP, TEMP, Blood Glucose, ECG). It is not an AI-assisted diagnostic device, and therefore, an MRMC comparative effectiveness study involving human readers improving with AI assistance would not be applicable and was not reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a medical instrument that provides physiological measurements. Its performance is inherent in its measurement capabilities and adherence to performance standards. The "standalone" performance is essentially its ability to accurately measure the parameters as specified, which is verified through testing against standards and comparison to a predicate device. The document mentions "Performance testing was performed to ensure that all of the device operated within specified parameters independently and with all other functions of the device operating." This indicates standalone performance testing of its direct measurement capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the measurements would typically be established by highly accurate reference instruments or established clinical methods, as dictated by the specific IEC/ISO/ASTM standards (e.g., for NIBP, a reference sphygmomanometer; for SpO2, a co-oximeter with induced hypoxia studies; for temperature, a reference thermometer). The document implies adherence to these standards for validation rather than explicitly stating ground truth derivation methods for each measurement. For NIBP, the reference to "clinical testing" would imply comparison to a clinically accepted gold standard.

    8. The sample size for the training set

    This device is a physiological monitor, not an AI/machine learning device that typically requires a "training set" in the context of algorithm development. Therefore, the concept of a training set sample size does not apply here.

    9. How the ground truth for the training set was established

    As the device does not employ a machine learning algorithm requiring a "training set," this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1