Search Results
Found 2 results
510(k) Data Aggregation
(74 days)
The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be displayed in real time on the system's display screen. The CARTO® 3 System V3.0 is also intended to support EP procedures, maintaining CARTO® System capabilities in the presence of a high metallic environment and magnetic field strengths up to 0.1 T and provide a data communication channel to the Stereotaxis Niobe Catheter Navigation System. The CARTO® 3 System V3.0 includes CARTOMERGE® PLUS Module functionality to import, register and merge CT or MRI structural images with CARTO map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using catheters with magnetic location sensors. The system's CARTOSOUND® image integration functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. A fluoro background capability enables visualization of X-ray images as reference to the CARTO® Maps images and catheters. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.
The CARTO® 3 System V3.0 is a catheter-based atrial and ventricular mapping system designed to acquire and analyze data points, and use this information to display 3D anatomical and electroanatomical maps of the human heart in real-time. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheter and reference device. The system allows realtime display of electrograms and cardiac maps based on the received intra cardiac signals from the catheters in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms (IECG) may also be displayed on the display screen. The CARTO® 3 System V3.0 uses two distinct types of location technology - magnetic sensor technology and Advanced Catheter Location (ACL) technology.
Here's an analysis of the provided text regarding the acceptance criteria and study proving device efficacy, structured according to your request:
It's important to note that the provided text is a 510(k) Summary for a medical device (CARTO® 3 System V3.0), which focuses on demonstrating substantial equivalence to a predicate device rather than outright proving effectiveness in a clinical trial with specific performance metrics against a predefined acceptance criterion. As such, some of the requested information (like effect size of AI assistance, sample size for training sets, or detailed ground truth establishment for specific metrics) may not be explicitly present because the submission's goal is different.
Acceptance Criteria and Device Performance Study (CARTO® 3 System V3.0)
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) submission for the CARTO® 3 System V3.0 focuses on substantial equivalence to its predicate device (CARTO® 3 EP Navigation System, Version 2.2) and two reference devices. Instead of explicit quantitative "acceptance criteria" with specific performance metrics (e.g., sensitivity, specificity, accuracy against a numerical threshold), the criteria are framed around demonstrating that the new features work as intended and do not negatively impact existing features, making it "as safe, as effective, and performs as well as or better than" the predicate.
Therefore, the "acceptance criteria" are implied by the comprehensive non-clinical testing performed, and the "reported device performance" is a general statement of successful completion of these tests.
Acceptance Criterion (Implied) | Reported Device Performance |
---|---|
Functional Equivalence & Safety of New Features: | Passed all tests in accordance with appropriate test criteria and standards. The modified device did not raise new questions of safety or effectiveness. |
- WiseTag functionality (legacy & new) | Successfully implemented and verified. |
- Multi Electrode Mapping (legacy & new) | Functionality confirmed to be equivalent or improved. |
- Fluoroscopic background view (legacy & new) | Functionality confirmed to be equivalent or improved. |
- Magnetic sensor technology & Advanced Catheter Location (ACL) | Verified to acquire and analyze data points, display 3D anatomical and electroanatomical maps, and real-time electrograms. |
- Support for EP procedures in high metallic environments / magnetic fields (up to 0.1 T) | Functionality verified. |
- Data communication channel to Stereotaxis Niobe System | Functionality verified. |
- CARTOMERGE® PLUS Module functionality (CT/MRI integration) | Functionality verified. |
- Fast Anatomical Mapping (FAM) functionality | Functionality verified. |
- CARTOSOUND® Image Integration functionality | Functionality verified. |
- Compatibility with both specialized and conventional EP catheters | Verified. |
Non-Inferiority to Predicate Device: | The CARTO® 3 System V3.0 is as safe, as effective, and performs as well as or better than the predicate device and two Reference Devices. |
Compliance with Standards: | Passed all tests in accordance with appropriate test criteria and standards. |
2. Sample Size Used for the Test Set and Data Provenance
The document states: "The CARTO® 3 System V3.0 underwent extensive Bench and Animal Testing to verify the modified features and to demonstrate with regression testing that the new features did not negatively affect existing features."
- Test Set Description: "Bench and Animal Testing."
- Sample Size: Not specified in the provided text. The term "extensive" is used but no numerical count of tests performed, data points, or animals used is given.
- Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. Animal testing is typically prospective, but bench testing methodology details are absent.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The 510(k) summary does not mention the use of human experts to establish ground truth for test sets. The testing described is primarily technical and functional ("Bench and Animal Testing") rather than clinical performance evaluated by human readers against a gold standard.
4. Adjudication Method for the Test Set
No adjudication method is described, as the testing performed was primarily technical/functional and animal testing, not human expert interpretation of device output.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study is not mentioned in the provided 510(k) summary. The submission focuses on substantial equivalence via non-clinical testing. Therefore, there's no reported effect size of how much human readers improve with AI vs. without AI assistance. The device is a "Cardiac mapping system" and "Programmable diagnostic computer," which primarily assists electrophysiologists by providing real-time maps and data, rather than an AI-driven diagnostic tool designed for human-in-the-loop diagnostic improvement in the sense of image interpretation.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
The device itself is a "Cardiac mapping system" and "Programmable diagnostic computer." Its primary function is to acquire, analyze, and display data for a physician's use in real-time procedures. The "Non-Clinical Performance Testing" described (Bench and Animal Testing) can be considered a form of standalone testing in that the device's technical specifications and functionalities were verified independently. However, this is not "standalone performance" in the context of an AI diagnostic algorithm's ability to interpret medical data without human intervention. The device's output (cardiac maps, electrograms) requires interpretation and action by a clinician, so there isn't a "standalone" diagnostic output in the way one might consider an AI-driven image analysis tool.
7. The Type of Ground Truth Used
For the "Bench and Animal Testing," the ground truth would be established through a combination of:
- Technical Specifications: Verification against predefined engineering and performance specifications.
- Physical Measurements & Known Conditions: In bench testing, this involves using phantoms, known electrical signals, or controlled environments where the "true" values are known and the device's measurements are compared against them.
- Physiological Observations (Animal Testing): In animal studies, ground truth would be based on direct physiological measurements, observations of physical effects, or pathological findings where applicable. The goal is to see if the device accurately reflects the physiological state of the animal heart.
8. The Sample Size for the Training Set
Not applicable/Not mentioned. The CARTO® 3 System V3.0 is a hardware/software system for cardiac mapping, not an AI/ML device in the modern sense that typically relies on large "training sets" for model development. Its functionalities are based on established biophysical principles and algorithms for signal processing and 3D reconstruction, rather than machine learning from a specific training dataset in the same way a diagnostic AI algorithm would be.
9. How the Ground Truth for the Training Set Was Established
Not applicable/Not mentioned for the reasons stated above (not an AI/ML device relying on "training sets").
Ask a specific question about this device
(121 days)
The Guided Medical Positioning System (gMPS™) is intended for the evaluation of vascular and cardiac anatomy. It is intended to enable real time tip positioning and navigation of a gMPS™ enabled (equipped with a gMPS™ sensor) diagnostic or therapeutic invasive device used in vascular or cardiac interventions in the Cath Lab environment, on both live fluoroscopy or recorded background. The System is indicated for use as an adjunct to fluoroscopy.
The gMPS™, used in conjunction with an X-ray System, employs magnetic positioning technology to track a gMPS™ enabled diagnostic or therapeutic invasive device for the 3D position relative to any X-ray image, in real-time or previously recorded cine-loop.
The gMPS™ system is intended to provide the following:
Catheter tip positioning and navigation - The real time position of the gMPSTM sensor (and thus of the gMPS™ enabled device) is displayed in real time ("Live") fluoroscopy mode or in a recorded mode.
Smart trace (foreshortening indication) - A 3D trace of the gMPS™ enabled device trajectory is projected and superimposed on the 2D X-ray images (either on live fluoroscopy, recorded cine-loop or recorded still image).
3D reconstructed model - The system reconstructs a 3D model of the inspected anatomical structure.
Quantitative longitudinal measurements - The measurements are based on the 3D trace, thus overcoming length measurement errors induced by the foreshortening effect.
Quantitative Coronary Angiography (OCA) - While working in conjunction with gMPSTM enabled coronary device, the gMPS™ provides 3D QCA.
Virtual landmarking - A manually marked point or region of interest superimposed on X-ray images (real-time angiography and cine-loop) and on the 3D reconstruction.
Here's a breakdown of the acceptance criteria and study information for the Guided Medical Positioning System (gMPS™), based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly list quantitative acceptance criteria or specific performance metrics with target values. Instead, it makes a general statement about performance testing.
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions "Performance testing was conducted."
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The provided text does not specify the number of experts used or their qualifications for establishing ground truth for the test set.
4. Adjudication Method for the Test Set
The provided text does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The provided text does not mention if a multi-reader multi-case (MRMC) comparative effectiveness study was conducted. Therefore, no effect size of human readers improving with AI vs. without AI assistance can be determined from this document.
6. Standalone (Algorithm Only) Performance Study
The provided text describes the device's function as "real time tip positioning and navigation" and states "Performance testing was conducted in order to demonstrate the performance and accuracy of the gMPS™ and to verify that it does not raise any new safety and effectiveness issues in comparison to its predicate devices." This implies that the algorithm's performance in guiding the device was evaluated, which aligns with a standalone performance study. However, specific details of this study are not provided.
7. Type of Ground Truth Used
The provided text does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). Given the device's function for "real time tip positioning and navigation" and "3D reconstructed model" of anatomical structures, it's highly likely that ground truth would involve either:
* Direct measurement/imaging: Comparing the system's reported position/reconstruction to a known physical measurement or a high-accuracy imaging modality.
* Expert validation: Clinical experts confirming the accuracy of the system's output against their anatomical knowledge or other reference images.
8. Sample Size for the Training Set
The provided text does not specify the sample size for the training set.
9. How Ground Truth for the Training Set Was Established
The provided text does not specify how ground truth for the training set was established.
Ask a specific question about this device
Page 1 of 1