510(k) Data Aggregation

    K Number
    K141940
    Manufacturer
    Date Cleared
    2014-11-24

    (130 days)

    Product Code
    Regulation Number
    868.5450
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132017, K131957, K131895, K110019, K063758, K131912

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used with a Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients with and without an artificial airway in place.

    Device Description

    The ConchaSmart Column is an accessory to the Neptune Heated Humidifier (K063758 and K131912). It is a single use, disposable humidifier cartridge which is intended for neonatal, infant, pediatric, and adult patients requiring heated humidification. The ConchaSmart Column is used by inserting the column into the center of the Neptune Heated Humidifier.

    AI/ML Overview

    The provided document is a 510(k) summary for the ConchaSmart Column, a respiratory gas humidifier. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that the device meets specific acceptance criteria in the context of an AI/algorithm-based medical device.

    Therefore, the requested information categories related to acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies are largely not applicable to this type of regulatory submission. This document describes a medical device, but not an AI-powered one that would typically undergo the kind of performance evaluation requested.

    However, I can extract the information that is present and indicate where the requested information is not applicable (N/A) or not provided in this document.


    1. A table of acceptance criteria and the reported device performance

    The document presents a comparison to a predicate device based on technological characteristics and performance specifications rather than explicit acceptance criteria and a study proving device performance against those criteria in the way an AI device would be evaluated. The "reported device performance" in this context refers to its specifications alongside those of the predicate.

    CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (ConchaSmart Column)
    Intended UseSame as predicate (Fisher & Paykel MR290)When used with a Neptune Heated Humidifier and Hudson RCI ventilator circuits, provides heated humidification for patients with and without an artificial airway in place.
    Patient PopulationNeonatal, Infant, Pediatric, AdultNeonatal, Infant, Pediatric, Adult
    Humidification TherapiesInvasive, Non-Invasive, High Flow Nasal Cannula TherapyInvasive, Non-Invasive, High Flow Nasal Cannula Therapy
    Humidity OutputISO 8185:2007ISO 8185:2007
    - Subglottic mode≥33mg H2O /L≥33mg H2O /L
    - Supraglottic mode≥10mg H2O /L≥10mg H2O /L
    Enthalpy Limit