Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113362
    Manufacturer
    Date Cleared
    2012-02-10

    (87 days)

    Product Code
    Regulation Number
    888.1100
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K081051, K110992, K061345, K994363, K001918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Surgical accessories are indicated for facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.

    Device Description

    The SpineVu Endoscopic Spine System (SESS) is a collection of arthroscopic surgical accessories provided sterile (irradiated) and intended for single-use only. As a group, the accessories are provided to facilitate delivery of an endoscope and other instruments to the targeted treatment site. The SpineVu Endoscopic Spine System (SESS) consists of the following devices: enVue Cannula, enVue Sheath, 16G Introducer Cannula with Stylet, Guidewire, Dilator, Beveled Cannula, Ball-Tipped Probe, Infusion Cannula, Suction Cannula.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "SpineVu Endoscopic Spine System (SESS)". It outlines the device's description, intended use, technological comparison to predicate devices, and non-clinical performance data.

    However, the document does not contain the specific information requested regarding acceptance criteria and the study that proves the device meets those criteria, such as a table of accuracy metrics, sample sizes for test/training sets, expert qualifications, or details on ground truth establishment.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through a comparison of technological characteristics and non-clinical performance testing.

    Here's what can be extracted from the provided text, and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Implicit): The implicit acceptance criteria are that the modified SpineVu Endoscopic Spine System (SESS) performs according to its stated intended use and functions as intended, meeting design specifications and external standard requirements. This is a general statement rather than specific quantitative metrics.

    • Reported Device Performance: The document states: "All data fell well within product specifications and external standard requirements. Results of non-clinical testing demonstrated that the SpineVu Endoscopic Spine System (SESS) is substantially equivalent to the predicate devices for its intended use."

      There is no table explicitly detailing acceptance criteria alongside reported performance for each criterion with quantitative results. The non-clinical tests conducted are listed as:

      • Visual and Dimensional Verification
      • Device to Device Compatibility Testing
      • Tensile Testing
      • Flow Rate Testing
      • Luer Attachment Testing
      • Leakage Testing
      • Jaw & Trigger Force Testing
      • Jaw Cycle Integrity Testing
      • Jaw Activation Testing
      • Biocompatibility Testing
      • Design Validation Testing
      • Packaging Testing
      • Sterility Testing

    2. Sample size used for the test set and the data provenance:

    • Not specified. The document only mentions "non-clinical testing" without detailing the methodologies or sample sizes for each test. Data provenance (e.g., country of origin, retrospective/prospective) is also not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is a 510(k) for a surgical accessory based on substantial equivalence and non-clinical (engineering/lab) testing, not a clinical study involving diagnosis or interpretation by experts to establish ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable, for the same reasons as point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a manual surgical accessory, not an AI-powered diagnostic tool. Therefore, no MRMC comparative effectiveness study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical surgical accessory, not an algorithm.

    7. The type of ground truth used:

    • For the non-clinical tests listed, the "ground truth" would be established by engineering specifications, validated test methods, and industry standards for mechanical performance, material properties, sterility, and biocompatibility. It is not based on expert consensus, pathology, or outcomes data in the context of diagnostic accuracy.

    8. The sample size for the training set:

    • Not applicable. This device is a manual surgical accessory. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set.

    Summary of Missing Information:

    The provided 510(k) summary focuses on demonstrating substantial equivalence through comparison to predicate devices and non-clinical (benchtop) testing results. It does not contain the detailed clinical study information (like test/training set sample sizes, expert qualifications, ground truth establishment for diagnostic accuracy, or AI performance metrics) that would typically be found for a device requiring performance evaluation in a clinical or diagnostic context. The acceptance criteria are implicitly defined by compliance with product specifications and external standards, rather than specific quantitative performance metrics from a user study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1