510(k) Data Aggregation

    K Number
    K103167
    Device Name
    STELLAR 150
    Date Cleared
    2011-03-01

    (125 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    BREAS VIVO 40 SYSTEM (K090113), ResMed VPAP III ST-A/KIDSTA MASK SYSTEM (K060105), Draeger Carina Home (K060705

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stellar 150 is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use with an uncuffed or deflated tracheostomy. Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.

    Device Description

    The Stellar 150 is a pressure controlled ventilator using a single limb vented circuit, product code MNT. A microprocessor controlled blower generates the required airway pressure between 3-40 cm H2O. Comparing to previous ResMed devices, CPAP and Bi-level modes are implemented as in VPAP III ST-A / Kidsta mask system, which is indicated for patients above 40 lb (18 kg). With ongoing technological progress, the new device is further suitable to include a population above 30 lb (13 kg) for CPAP and Bilevel modes, which is comparable to Draeger Carina Home. The new device also includes a volume assured pressure support mode, indicated for patients above 66 lb (30 kg), which is substantially equivalent to Breas Vivo 40.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Stellar 150 device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly present a table of specific numerical acceptance criteria with corresponding performance data for the Stellar 150. Instead, it states that "All tests confirmed the predetermined acceptance criteria." The performance is described in terms of compliance with established standards and substantial equivalence to predicate devices.

    Acceptance Criteria CategoryDescription of Performance / Compliance
    General Safety and Performance- Tested to appropriate ISO and IEC standards and other applicable requirements, passing all test protocols.
    • Complies with IEC 60601-1:1988 (and its amendments) and IEC 60601-1-2:2007.
    • Complies with ISO 10651-6:2004.
    • Complies with FDA guidance documents: "Draft Reviewer Guidance for Ventilators (July 1995)" and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)". |
      | Risk Analysis Outcomes | - Design and Verification activities performed as a result of risk analysis.
    • All tests confirmed predetermined acceptance criteria (implying risks were mitigated and performance met safety thresholds). |
      | Pressure Performance | - Performance testing comprises pressure performance.
    • Microprocessor controlled blower generates required airway pressure between 3-40 cm H2O. |
      | Trigger and Cycling | - Performance testing comprises trigger and cycling. |
      | Volume Assured Pressure Support Ventilation | - Performance testing comprises volume assured pressure support ventilation. |
      | Substantial Equivalence | - Side-by-side testing demonstrated substantial equivalence to predicate devices (BREAS VIVO 40 SYSTEM, ResMed VPAP III ST-A/KIDSTA MASK SYSTEM, Draeger Carina Home) in terms of:
      • Same intended use
      • Same operating principle
      • Similar technologies
      • Same manufacturing process |
        | Indication Expansion (CPAP/Bi-level) | - New device further suitable for patients above 30 lb (13 kg) for CPAP and Bi-level modes, comparable to Draeger Carina Home. |
        | Indication (Volume Assured Pressure Support) | - New device includes a volume assured pressure support mode for patients above 66 lb (30 kg), substantially equivalent to Breas Vivo 40. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data or a specific cohort of patients for testing. The performance data described refers to engineering verification and validation activities conducted on the device itself.

    • Sample Size: Not applicable/not specified for a clinical test set. The testing refers to device units.
    • Data Provenance: Not applicable/not specified in terms of patient data. The provenance is from internal design, verification, and validation activities.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The study described is not a clinical study involving expert ground truth for patient outcomes.

    4. Adjudication Method for the Test Set

    This information is not provided in the document, as it does not describe a clinical study requiring adjudication of patient cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

    No MRMC comparative effectiveness study is mentioned in the document. The study described focuses on technical performance and substantial equivalence to predicate devices, not on human reader performance with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The device itself is a continuous ventilator, not explicitly an "algorithm" in the sense of an AI diagnostic tool. The performance described is inherent to the device's operation. Therefore, a standalone algorithm performance study is not directly applicable or described. The performance data reflects the device operating as intended without human intervention in its core functions (like pressure generation, trigger, and cycling).

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Engineering Specifications and Design Requirements: The device was designed and tested against predetermined acceptance criteria derived from risk analysis and product design requirements.
    • International Standards (ISO, IEC): Compliance with these standards serves as a benchmark for safety and performance.
    • FDA Guidance Documents: Adherence to FDA guidelines is a form of regulatory "ground truth" for manufacturing and submission.
    • Predicate Device Performance: Substantial equivalence claims are based on comparing the Stellar 150's performance and characteristics to those of legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This information is not applicable. The Stellar 150 is a hardware-based medical device with microprocessor control, not an AI/ML algorithm that requires a "training set" of data in the typical sense. Its control logic and parameters are developed through engineering design and testing, not statistical learning from a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the reasons stated in point 8.