Search Results
Found 2 results
510(k) Data Aggregation
(241 days)
The Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy. The Image Guidance System for Radiotherapy sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device.
The product is a medical electrical device which is composed of hardware and software, and is jointly used with the radiation therapy device. Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy.
The provided text describes several bench tests conducted to demonstrate the substantial equivalence of the "Image Guidance System for Radiotherapy" to a predicate device. This submission does not include clinical studies.
Here's an analysis of the acceptance criteria and performance based on the non-clinical tests:
1. Table of Acceptance Criteria and Reported Device Performance
Device Name: Image Guidance System for Radiotherapy (Models: IGTS, IGPS)
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Compatibility Testing with Radiotherapy Systems | ||
Targeting Accuracy (integrated with MASEP Gamma Ray SRT) | No larger than 1.0 mm | Well meets the requirements for SRS (implies ≤ 1.0 mm) |
Targeting Accuracy (integrated with Varian Linear Accelerator) | No larger than 1.0 mm | Well meets the requirements for SRS (implies ≤ 1.0 mm) |
Targeting Accuracy (integrated with Elekta Linear Accelerator) | No larger than 1.0 mm | Well meets the requirements for SRS (implies ≤ 1.0 mm) |
Optical Guided Tracking Accuracy Testing | ||
3D RMS Volumetric Accuracy | 0.15 mm for extended pyramid volume | 0.11 mm |
System Positioning Accuracy Testing | ||
Overall Positioning Accuracy | No larger than 1.0 mm | No larger than 1.0 mm |
Individual Rotational Angle Accuracy | No larger than 1.0° | No larger than 1.0° |
System Tracking Accuracy Testing | ||
Target Tracking Accuracy | Better than 1.5 mm | No larger than 1.5 mm |
System Performance Testing | ||
Mechanical Accuracy of Imaging Center | Not exceed 0.5 mm | Meets the requirement (implies ≤ 0.5 mm) |
Positioning Accuracy | Not exceed 1.0 mm | Meets the requirement (implies ≤ 1.0 mm) |
Positioning Stability | Not exceed 0.3 mm | Meets the requirement (implies ≤ 0.3 mm) |
Detection Accuracy of Target Movement (regular respiratory) | Not exceed 1.5 mm | Meets the requirement (implies ≤ 1.5 mm) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes (number of cases/patients/phantoms) used for each individual bench test. It mentions using an "anthropomorphic target phantom" and "anthropomorphic head phantom and a thoracic spine phantom" for system accuracy tests, and a "phantom used for this test provides respiratory waveforms collected from clinical patient data" for tracking accuracy.
The data provenance is from bench testing conducted by the manufacturer, Jiangsu Rayer Medical Technology Co., Ltd., in Wuxi, Jiangsu, CHINA. The data is non-clinical.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. These were non-clinical bench tests using phantoms and controlled setups, not human-annotated data with expert ground truth in the traditional sense. The "ground truth" for the tests was established by physical movements and measurements from the treatment couch or optical tracking systems during the experiments against known values.
4. Adjudication Method for the Test Set
Not applicable. As these were bench tests with physical measurements against known references (e.g., treatment couch movements, optical tracking systems), there was no need for expert adjudication of results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." The studies were bench tests demonstrating the device's performance alone and its compatibility with existing radiotherapy systems, not a comparison of human reader performance with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, standalone performance was assessed through various bench tests. The tests evaluate the intrinsic accuracy and performance characteristics of the "Image Guidance System for Radiotherapy" itself, in terms of positioning and tracking accuracy, mechanical accuracy, and compatibility with other radiotherapy devices. These tests do not involve human-in-the-loop performance evaluation in the context of diagnostic interpretation or decision-making.
7. The Type of Ground Truth Used
The ground truth used for these bench tests was based on:
- Physical measurements and known displacements: For positioning and tracking accuracy, the ground truth was provided by precise movements of a treatment couch acting as a reference, or known positions/displacements of inserts within phantoms.
- Optical tracking systems: For optical guided tracking accuracy, the ground truth was likely established by a highly accurate optical tracking system (e.g., NDI unit itself or a reference system) against which the device's tracking capabilities were validated.
- Design specifications and established standards: The acceptance criteria themselves are the ground truth, derived from engineering specifications and industry standards for medical devices, particularly in radiation therapy.
8. The Sample Size for the Training Set
The document does not mention a training set. This suggests that the device, or at least the specific aspects being tested for this submission, may not rely on a machine learning model that requires a distinct training dataset, or if it does, the details of the training set are not included in this summary. The tests focus on the physical and algorithmic performance of the system rather than the learning capabilities of a model.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set information is provided or implied by the tests described.
Ask a specific question about this device
(54 days)
The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of radiation therapy treatment process, as determined by a licensed medical practitioner.
XVI, is intended to confirm patient positioning and support decision making in response to target displacement resulting from organ deformation and anatomical movement. Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cycle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.
This Traditional 510(k) describes modifications made to the XVI kilo-voltage imaging accessory of the Elekta range of digital linear accelerators. The primary reason for these enhancements is to improve the acquisition and utilization of kV images in facilitating the correction of patient position for anatomical changes or movement. The additional features are; Symmetry™ (4D VolumeView™) to evaluate respiration induced motion, 3D shaped region of interest for registration, dual registration for quantitative information regarding both the critical structures and target position, and 3D seed registration of implanted markers. Improvements have also been made to the operator interface, connectivity with other systems through DICOM and in the provision of licensable options to tailor individual features.
The XVI system consists of a kV radiation source mounted onto the linac gantry drum and a kV radiation image detector.
The provided 510(k) summary for the K100115 submission for the Elekta XVI R4.5 describes modifications to an existing device, primarily focusing on enhanced features for image acquisition and utilization in radiation therapy.
No study details, acceptance criteria, or device performance metrics are included in the provided document.
The submission asserts that "The functionality for the XVI R4.5 is equivalent to its predicate device XVI R3.5 as part of the Elekta Synergy (K051932) in safety and effectiveness. The fundamental technical characteristics are the same as those of the predicate device and differences in operation are described in the comparison chart and discussion provided elsewhere in this 510(k) submission."
This statement indicates that the device was deemed substantially equivalent to its predicate based on the similarity of its technological characteristics and the absence of significant changes that would raise new questions of safety or efficacy. Therefore, a separate study to establish new acceptance criteria or prove performance against them was likely not deemed necessary for this particular 510(k) submission, as the predicate device's performance was relied upon.
To provide the requested information, a detailed review of the original K051932 submission for the predicate device (Elekta Synergy® and its XVI R3.5 component) would be necessary, as that is where the initial performance and safety evaluations would have been established.
Therefore, based solely on the provided K100115 document, I cannot complete the requested tables and information as no such study details are present within this summary.
Ask a specific question about this device
Page 1 of 1