Search Results
Found 2 results
510(k) Data Aggregation
(141 days)
The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional, and three-dimensional imaging of anterior ocular structures. The device is indicated for visualizing anterior and posterior ocular structures, including cornea, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head. The CIRRUS normative databases are quantitative tools indicated for the comparison of retinal nerve fiber layer thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of normal subjects. The CIRRUS OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. The CIRRUS HD-OCT is indicated as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.
The CIRRUSTM HD-OCT is a computerized instrument that acquires and analyzes crosssectional tomograms of anterior and posterior ocular structures (including cornea, retina, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. Using this noninvasive optical technique. CIRRUS HD-OCT produces high-resolution cross-sectional tomograms of the eye without contacting the eye. It also produces images of the retina and layers of the retina from an en face perspective (i.e., as if looking directly in the eye).
The CIRRUS HD-OCT is offered in four models, Model 4000, 400, 5000 and 500. In the CIRRUS HD-OCT Models 4000 and 5000, the fundus camera is a line scanning ophthalmoscope. The CIRRUS HD-OCT Models 400 and 500 are similar to the Models 4000 and 5000 except that they provide the fundus image using the OCT scanner only.
The acquired imaging data can be analyzed to provide thickness and area measurements of regions of interest to the clinician. The system uses acquired data to determine the fovea location or the optic disc location. Measurements can then be oriented using the fovea and/or optic disc locations. The patient's results can be compared to subjects without disease for measurements of RNFL thickness, neuro-retinal rim area, average and vertical cup-to-disc area ratio, cup volume, macular thickness and ganglion cell plus inner plexiform layer thickness.
In addition to macular and optic disc cube scans, the CIRRUS HD-OCT also offers scans for OCT angiography imaging, a non-invasive approach with depth sectioning capability to visualize microvascular structures of the eye.
Anterior segment scans enable analysis of the anterior segment including Anterior Chamber Depth. Angle-to-Angle and automated measurement of the thickness of the cornea with the Pachymetry scan.
Here's an analysis of the provided text, outlining the acceptance criteria and the study details for the Carl Zeiss Meditec Inc. Cirrus HD-OCT with Software Version 8.
Acceptance Criteria and Device Performance:
The document primarily focuses on demonstrating repeatability and reproducibility of measurements, and comparability between the new device (CIRRUS HD-OCT with Software Version 8, specifically Models 4000 and 5000) and the predicate device (Visante OCT). Specific acceptance criteria are not explicitly stated as numerical targets in quantifiable terms for all parameters. Instead, the study aims to show that the new device's measurements are consistent and comparable to the predicate device. The tables report the demonstrated performance rather than predefined acceptance criteria.
Table of Acceptance Criteria and Reported Device Performance (Inferred from the study's objective to demonstrate comparability, repeatability, and reproducibility):
Measurement Parameter | Acceptance Criteria (Inferred) | CIRRUS 4000 Reported Performance (Repeatability SD/CV% / Reproducibility SD/CV%) | CIRRUS 5000 Reported Performance (Repeatability SD/CV% / Reproducibility SD/CV%) | Comparison to Visante OCT (Mean Difference / 95% LOA) |
---|---|---|---|---|
Anterior Chamber Scans (Normal Cornea Group) | Demonstrates repeatability, reproducibility, and comparability to predicate. | See Table 1 | See Table 2 | See Table 7 |
CCT | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 8.806 / 1.619%; Reproducibility SD: 9.514 / 1.750% | Repeatability SD: 9.749 / 1.774%; Reproducibility SD: 11.897 / 2.165% | C4000: 9.4 (16.4); 95% LOA: -23.3, 42.1 |
Angle to Angle | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 0.187 / 1.517%; Reproducibility SD: 0.265 / 2.150% | Repeatability SD: 0.171 / 1.423%; Reproducibility SD: 0.300 / 2.494% | C4000: 0.665 (0.395); 95% LOA: -0.126, 1.456 |
ACD | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 0.066 / 2.291%; Reproducibility SD: 0.068 / 2.366% | Repeatability SD: 0.034 / 1.199%; Reproducibility SD: 0.046 / 1.601% | C4000: -0.067 (0.077); 95% LOA: -0.220, 0.087 |
Pachymetry Scans (Normal Cornea Group) | Demonstrates repeatability, reproducibility, and comparability to predicate. | See Table 1 | See Table 2 | See Table 7 |
Center Pachymetry | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 3.359 / 0.635%; Reproducibility SD: 3.719 / 0.703% | Repeatability SD: 1.197 / 0.226%; Reproducibility SD: 1.628 / 0.308% | C4000: 1.4 (4.1); 95% LOA: -6.8, 9.7 |
... (Other Pachymetry Zones) | Similar criteria as Center Pachymetry. | See Table 1 (various values) | See Table 2 (various values) | See Table 7 (various values) |
Anterior Chamber Scans (Corneal Pathology Group) | Demonstrates repeatability, reproducibility, and comparability to predicate in pathological cases. | See Table 3 | See Table 4 | See Table 8 |
CCT | Similar criteria as Normal Cornea. | Repeatability SD: 10.023 / 1.923%; Reproducibility SD: 14.069 / 2.699% | Repeatability SD: 12.061 / 2.267%; Reproducibility SD: 18.951 / 3.561% | C4000: 8.2 (20.0); 95% LOA: -31.7, 48.2 |
... (ATA, ACD, Pachymetry Zones) | Similar criteria. | See Table 3 | See Table 4 | See Table 8 |
Pachymetry Scans (Post-LASIK Group) | Demonstrates repeatability, reproducibility, and comparability to predicate in post-LASIK cases. | See Table 5 | See Table 6 | See Table 9 |
Center Pachymetry | Similar criteria. | Repeatability SD: 1.793 / 0.385%; Reproducibility SD: 2.000 / 0.430% | Repeatability SD: 1.784 / 0.383%; Reproducibility SD: 2.068 / 0.445% | C4000: 2.2 (5.9); 95% LOA: -9.6, 14.0 |
Angle Study (Glaucoma Suspects/Patients) | Demonstrates repeatability, reproducibility, and comparability to predicate for angle measurements. | See Table 10 | See Table 11 | See Table 12 |
TISA 500 Nasal (Wide Angle to Angle Scan) | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 0.020 / 13.590%; Reproducibility SD: 0.032 / 21.484% | Repeatability SD: 0.025 / 16.801%; Reproducibility SD: 0.030 / 19.614% | C4000: Not directly provided for TISA (Only for AC Angle) |
AC Angle Nasal (Wide Angle to Angle Scan) | Low SD, CV% for repeatability/reproducibility; Small mean difference and narrow LOA compared to Visante. | Repeatability SD: 4.128 / 11.479%; Reproducibility SD: 4.626 / 12.862% | Repeatability SD: 3.427 / 9.475%; Reproducibility SD: 4.861 / 13.442% | C4000: -1.887 (7.155); 95% LOA: -16.196, 12.422 |
... (Other angle measurements) | Similar criteria. | See Table 10 | See Table 11 | See Table 12 |
Study Details:
-
Sample sizes used for the test set and data provenance:
- Anterior Chamber and Pachymetry Scans:
- Normal Cornea group: 46 subjects (Group 1)
- Post-LASIK group: 40 subjects (Group 2)
- Corneal Pathology group: 45 subjects (Group 3)
- Age Range (all groups): 25 to 69 years.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "non-significant risk clinical study," implying prospective data collection in a clinical setting.
- Angle Study:
- Angle Study: 27 subjects, ranging from 43 to 77 years (mean 62 years).
- Specific eye counts per measurement type:
- 26 eyes for Wide Angle-to-Angle scan
- 27 eyes for HD Angle scan
- Data Provenance: Not explicitly stated (e.g., country of origin). Described as a "non-significant risk clinical study," implying prospective data collection.
- OCT Angiography: Series of case studies. No specific sample size is provided beyond "case studies."
- Anterior Chamber and Pachymetry Scans:
-
Number of experts used to establish the ground truth for the test set and their qualifications:
- Anterior Chamber and Pachymetry Scans: No external experts were used for "ground truth" in the sense of an adjudicated diagnosis. The study focused on comparing measurements between devices and assessing repeatability/reproducibility. The predicate device (Visante OCT) served as the reference for comparability. One operator acquired data on the Visante OCT. Three operators acquired data on the CIRRUS HD-OCT devices.
- Angle Study: The study focused on device measurement comparison, repeatability, and reproducibility. One operator acquired data on the Visante OCT. Three operators acquired data on the CIRRUS HD-OCT devices. The study population had a variety of angle configurations (Grade II to Grade IV) as assessed by gonioscopy using the Shaffer method, but the gonioscopy results themselves were not used as a direct "ground truth" for individual measurement values from the OCT devices.
- OCT Angiography: The "findings demonstrate that the CIRRUS OCT Angiography... can give non-invasive three-dimensional information regarding retinal microvasculature." This was compared with "fluorescein angiography images." The number and qualifications of experts interpreting either the OCTA or fluorescein angiography images for these case studies are not specified in the provided text.
-
Adjudication method for the test set:
- Anterior Chamber and Pachymetry Scans: Not applicable for establishing ground truth, as the study focused on device measurement comparison and repeatability/reproducibility across devices and operators. Measurements were generated by "manual placement of software tools."
- Angle Study: Similar to the above, not applicable for establishing "ground truth" through adjudication. Measurements were generated by "manual placement of software tools (Angle tool; TISA tool)" by the operators who acquired the data.
- OCT Angiography: Not specified.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study with human readers and AI assistance was not explicitly reported or performed in the provided text. The study primarily focused on the device's technical performance (repeatability, reproducibility, and agreement with a predicate device) rather than its impact on human reader diagnostic accuracy with or without AI. The device itself (CIRRUS HD-OCT with Software Version 8) is an imaging and measurement device, not explicitly an "AI" diagnostic tool in the context of human-in-the-loop performance measurement.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, in a sense, the primary studies for Anterior Chamber, Pachymetry, and Angle measurements assessed the standalone performance of the device's measurement algorithms by evaluating their repeatability, reproducibility, and agreement with predicate device measurements. These measurements are generated by the device's software (e.g., "manual placement of software tools" by an operator, but the calculation is still algorithmic). The OCT Angiography section also mentions the device's ability to provide information on microvasculature without human "in-the-loop" interpretation for the basic image generation. However, it's not described as an "AI algorithm" in the common sense of diagnosing from images.
-
The type of ground truth used:
- Anterior Chamber, Pachymetry, and Angle Studies: The "ground truth" was effectively the measurements obtained from the predicate device (Visante OCT). The purpose was to show substantial equivalence and comparability, not to determine diagnostic accuracy against an independent, gold-standard clinical pathology or outcome.
- OCT Angiography: Comparisons were made with fluorescein angiography images. Fluorescein angiography is a clinical gold standard for visualizing retinal and choroidal vasculature.
-
The sample size for the training set:
- The document does not provide information on the training set size. The studies described are validation studies for the device's technical performance and comparability to a predicate. The device incorporates "proprietary algorithms" and "normative databases," which would have been developed using some form of training data, but details about this training data are not included in the provided 510(k) summary.
-
How the ground truth for the training set was established:
- Not described in the provided text. As mentioned above, the 510(k) summary focuses on the validation studies, not the development or training of any underlying algorithms or normative databases. The "normative databases" would inherently rely on data from "normal subjects," but the specifics of their ground truth establishment are not given.
Ask a specific question about this device
(238 days)
The IOLMaster 700 is intended for biometric measurements and visualization of ocular structures. The measurements and visualization assist in the determination of the appropriate power and type of intraocular lens. The IOLMaster 700 measures:
- Lens thickness
- Corneal curvature and thickness
- Axial length
- Anterior chamber depth
- Pupil diameter
- White-to-white distance (WTW)
For visualization, the IOLMaster 700 employs optical coherence tomography (OCT) to obtain two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
The IOLMaster 700 device is a computerized biometry device consisting of an OCT system, a Keratometer system, and a camera for the purposes of:
- measuring distances within the human eye along the visual axis (e.g. axial length, lens thickness, anterior chamber depth),
- measuring the corneal surface with a keratometer,
- measuring distances at the front of the eye with a camera (e.g. white-to-white distance).
The IOLMaster 700 is used for visualization and measurement of ocular structures mainly required for the preparation of cataract surgeries to calculate the refractive power of the intraocular lens (IOL) to be implanted. The IOLMaster 700 device includes a swept source frequency domain optical coherence tomography (OCT) module capable of acquiring tomograms of the eye. The axial measurements are based on those tomograms.
The IOLMaster 700 device is operated via multi touch monitor and alternatively with computer mouse and keyboard. A joystick on the measuring head is used for manual alignment of the device to the patient's eye.
Here's a breakdown of the acceptance criteria and study details for the IOLMaster 700 device, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state quantitative acceptance criteria (e.g., "AL difference must be less than 0.05 mm"). Instead, it focuses on demonstrating comparability to predicate devices and showing good repeatability and reproducibility. The "Limits of Agreement" from the Bland-Altman analysis (which is implied by the report of mean difference and limits of agreement) can be considered as the implicit acceptance ranges for comparability.
Measured Parameter | IOLMaster 700 vs. IOLMaster 500 (Mean Difference (SD)) | Limits of Agreement (Lower, Upper) | IOLMaster 700 vs. Lenstar LS 900 (Mean Difference (SD)) | Limits of Agreement (Lower, Upper) | IOLMaster 700 Repeatability (SD) | IOLMaster 700 Repeatability (%CV) |
---|---|---|---|---|---|---|
Biometry Mode | ||||||
Axial Length (AL) [mm] | 0.004 (0.025) | -0.045, 0.053 | N/A | N/A | 0.009 | 0.037% |
Anterior Chamber Depth (ACD) [mm] | 0.017 (0.121) | -0.221, 0.254 | N/A | N/A | 0.010 | 0.314% |
Lens Thickness (LT) [mm] | N/A | N/A | 0.020 (0.120) | -0.246, 0.256 | 0.019 | 0.410% |
Central Corneal Thickness (CCT) [µm] | N/A | N/A | 0.116 (4.492) | -8.688, 8.920 | 2.271 | 0.414% |
Keratometry Mode | ||||||
R1, Radius in Flat Meridian [mm] | 0.001 (0.044) | -0.086, 0.087 | N/A | N/A | 0.026 | 0.334% |
R2, Radius in Steep Meridian [mm] | -0.002 (0.046) | -0.093, 0.088 | N/A | N/A | 0.024 | 0.316% |
Spherical Equivalent (SE) [D] | -0.001 (0.190) | -0.374, 0.371 | N/A | N/A | 0.100 | 0.231% |
Cylinder (Cyl) [D] | 0.013 (0.318) | -0.610, 0.636 | N/A | N/A | 0.191 | 20.449% |
Axis (A) [°] | 1.407 (11.388) | -20.91, 23.73 | N/A | N/A | 5.633 | 6.93% |
WTW Mode | ||||||
White-to-White (WTW) [mm] | -0.125 (0.167) | -0.452, 0.203 | N/A | N/A | 0.090 | 0.755% |
Acceptance Criteria Implied: The study's conclusion that "The results of the IOLMaster 700 measurements are comparable to those of the predicate devices" and that it "demonstrated good repeatability and reproducibility for all parameters" indicates that the observed differences and variability fall within clinically acceptable ranges, deeming the device substantially equivalent. The p-values for paired t-tests (Table 2) consistently being above 0.05 (except for WTW) further support comparable means, although differences in variability exist.
2. Sample Size and Data Provenance
- Test Set Sample Size: A total of 120 eyes were enrolled in the study. Only one eye of each subject was designated as the study eye.
- Data Provenance: The study was a prospective non-significant risk clinical study conducted at three sites. The document does not specify the countries/regions of these sites, but it is a US FDA submission, implying compliance with US regulations.
3. Number of Experts and Qualifications for Ground Truth for Test Set
The document does not mention the use of experts to establish ground truth for the test set. The study directly compares the IOLMaster 700 measurements against the measurements obtained by the predicate devices (IOLMaster 500 and Lenstar LS 900) as the reference for comparability. For repeatability and reproducibility, the device's own measurements are compared against each other.
4. Adjudication Method for the Test Set
There is no mention of an adjudication method in the context of expert review for ground truth, as ground truth was not established by experts. For data quality during analysis, "scans were reviewed using the same quality criteria as described in the User Manual, Software Description." The specific criteria included issues like incorrect caliper placement, blurred images, closed eyelids, and distorted reflections.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This study focuses on device biometry measurement accuracy, comparability to predicate devices, and repeatability/reproducibility, not on human reader performance with or without AI assistance. The device is a biometer, not an AI-assisted diagnostic tool for interpretation.
6. Standalone Performance Study
- Yes, a standalone study was performed in the sense that the IOLMaster 700's own measurements were evaluated for repeatability and reproducibility, independent of human interpretation or "human-in-the-loop" performance. The device is intended to provide objective biometric measurements. The comparison to predicate devices also serves as a standalone performance assessment against established benchmarks.
7. Type of Ground Truth Used
The "ground truth" for the comparative study was the measurements obtained from the predicate devices:
- IOLMaster 500 (for AL, ACD, R1, R2, SE, Cyl, A, WTW)
- Lenstar LS 900 (for LT, CCT)
For repeatability and reproducibility, the device's own repeated measurements served as the basis for assessing consistency.
8. Sample Size for the Training Set
The document does not describe a "training set" in the context of an algorithm or AI development. This device is a measurement instrument based on optical principles, not a machine learning model that requires a separate training data set for its core functionality of measurement.
9. How the Ground Truth for the Training Set was Established
As there is no mention of a training set for an algorithm, this question is not applicable based on the provided document. The device's measurement algorithms are likely based on physical principles (e.g., optical coherence tomography) rather than being trained on a large dataset with established ground truth in the AI/ML sense.
Ask a specific question about this device
Page 1 of 1