Search Results
Found 5 results
510(k) Data Aggregation
(403 days)
When used as a vertebral body replacement, the TASMIN® R devices are indicated for use to replace a vertebral body that has been resected or excised due to tumor or traumaffracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental internal fixation.
When used as an intervertebral fusion device, the TASMIN® R devices are intended for use at one level in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used in patients who have had six months of non-operative treatment. The devices are intended for use with a supplemental internal fixation system and with autograft to facilitate fusion.
The basic shape of the TASMIN® R devices is a hollow structural frame having a rounded, tapered leading face. The upper and lower aspects of the implant are open with peaked teeth that assist in anchoring and seating the implant between the vertebral bodies. There are lateral fenestrations for bony in-growth. The device is available in a variety of sizes and angulations thereby enabling the surgeon to choose the size best suited to the individual pathology and anatomical condition.
The TASMIN® R device is a spinal intervertebral body fixation orthosis and intervertebral body fusion device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from recognized standards) | Reported Device Performance (TASMIN® R) |
---|---|
Mechanical performance, specifically static and dynamic compression, torsion, and compression shear properties, as per ASTM F2077. | Device performance demonstrated substantial equivalence to predicate devices. |
Subsidence properties, as per ASTM F2267. | Device performance demonstrated substantial equivalence to predicate devices. |
Material properties (PEEK-OPTIMA® LT1, Invibio® an ASTM F2026; Tantalum an ASTM F560) | TASMIN® R is manufactured from PEEK-OPTIMA® LT1 as described by ASTM F2026. Integral marker pins are manufactured from tantalum as described by ASTM F560. |
Basic design (hollow structural frame) | TASMIN® R has a hollow structural frame with a rounded, tapered leading face. Upper and lower aspects are open with peaked teeth. Lateral fenestrations are present. |
Sizes (widths, lengths, and heights) | Sizes are within the range(s) offered by predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The documentation refers to "worst case TETRIS™ and TASMIN® R devices" for finite element analysis and "worst case TASMIN® R devices" for physical testing. It does not specify the exact number of devices or scenarios considered. It is implied that representative "worst-case" configurations were selected for the evaluation.
- Data Provenance: The studies are described as "Finite element analysis simulations" and "Static and dynamic compression testing." This indicates the data was generated from engineering analyses and laboratory bench testing performed by the sponsor, SIGNUS Medizintechnik GmbH. The country of origin of the data is not explicitly stated, but the sponsor is based in Germany. The data is prospective in the sense that it was generated specifically for the premarket submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as the studies performed were bench testing and finite element analysis, not clinical studies requiring expert consensus on patient data. The "ground truth" for these types of studies is established by adherence to recognized ASTM standards and engineering principles.
4. Adjudication Method for the Test Set
This information is not applicable for bench testing and finite element analysis. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies for interpretation of imaging or clinical outcomes.
5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not conducted. The device is a medical implant, not an AI-assisted diagnostic or therapeutic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. The device is a physical implant, not an algorithm.
7. The Type of Ground Truth Used
The ground truth used for these studies was derived from:
- Recognized Industry Standards: ASTM F2077 (for compression, torsion, compression-shear) and ASTM F2267 (for subsidence properties). These standards define the methodology and expected performance for intervertebral body fusion devices.
- Engineering Principles: Finite element analysis relies on established biomechanical and engineering principles to simulate mechanical behavior.
- Material Specifications: ASTM F2026 for PEEK-OPTIMA® LT1 and ASTM F560 for Tantalum, defining the accepted properties of the materials.
8. The Sample Size for the Training Set
This information is not applicable. The studies performed were bench testing and simulations, not machine learning studies that require a training set.
9. How the Ground Truth for the Training Set was Established
This information is not applicable as there was no training set for a machine learning algorithm. The "ground truth" for the device's performance was established by adherence to ASTM standards and engineering principles as described in point 7.
Ask a specific question about this device
(28 days)
When used as a vertebral body replacement, the TETRIS™ II devices are indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental internal fixation.
When used as an intervertebral fusion device, the TETRIS™ II devices are intended for use at one level in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used in patients who have had six months of non-operative treatment. The devices are intended for use with a supplemental internal fixation system and with autograft to facilitate fusion.
The basic shape of the TETRIS™ II devices is a hollow structural frame having a rounded, tapered leading face. The upper and lower aspects of the implant are open. Surface spikes assist in the positive anchorage and seating of the implant between the vertebral bodies. The device is available in a variety of sizes and angulations thereby enabling the surgeon to choose the size best suited to the individual pathology and anatomical condition.
The information provided in the document describes the TETRIS™ II devices, which are spinal vertebral body replacement and intervertebral fusion devices. The FDA 510(k) summary focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific acceptance criteria in a clinical setting.
Instead, the performance data provided is related to bench testing and finite element analysis (FEA), which compares the new device to its predicates.
Here's a breakdown based on the provided text:
Acceptance Criteria and Device Performance for TETRIS™ II Devices
The primary acceptance criteria for the TETRIS™ II devices in this submission hinge on demonstrating that their technological characteristics and mechanical performance are equivalent to predicate devices, ensuring they are "as safe and as effective."
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Technological Characteristics | The TETRIS™ II devices must possess the same technological characteristics as the predicate devices. This includes intended use, basic design (hollow structural frame), material (PEEK polymer and/or titanium alloy), and sizes (widths, lengths, and heights). | The TETRIS™ II devices possess the "same technological characteristics" as the predicate devices. Specifically, they match in intended use, basic design (hollow structural frame), material (PEEK and/or titanium alloy), and available sizes (widths, lengths, and heights within the predicate range). |
Mechanical Performance (Worst Case) | Finite element analysis (FEA) simulations of the worst-case TETRIS™ II devices should demonstrate equivalence to, or no degradation compared to, the worst-case predicate devices under ASTM F2077 prescribed tests (compression, torsion, and compression shear). | FEA simulations of the worst-case TETRIS™ II and TETRIS™ devices were compared for ASTM F2077 prescribed tests (compression, torsion, and compression shear). The results "demonstrate that the additional TETRIS™ II devices do not create a new worst case device." |
Interpretation of Performance | The performance data should support the conclusion that the new devices are "as safe and as effective" as the predicates. | The FEA results lead to the conclusion that Tetris II devices are "as safe and as effective as the predicates." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: This notification does not detail a clinical test set with human or animal subjects. The "test set" in this context refers to the simulated models used in the finite element analysis. The specific number of models or simulations is not provided, but it states "Finite element analysis simulations of the worst case TETRIS™ and TETRIS™ II devices were compared."
- Data Provenance: The data provenance is from finite element analysis (FEA) simulations, which are computational models. It is not clinical data, and thus concepts like "country of origin" or "retrospective/prospective" don't apply in the traditional sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. The "ground truth" for the FEA simulations is based on engineering principles, material properties (ASTM F136 for Ti-6Al-4V, ASTM F2026 for PEEK-OPTIMA® LT1, ASTM F560 for Tantalum), and the defined ASTM F2077 standards for mechanical testing of spinal implants. There's no indication of human experts establishing a ground truth in a diagnostic or clinical sense for these engineering simulations.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. As this involves engineering simulations and comparison to established ASTM standards, a human adjudication method (like 2+1, 3+1 consensus) is not relevant. The outcome is determined by adherence to engineering analysis principles and the resulting numerical data.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This submission relies on engineering bench testing and FEA, not human reader studies of clinical cases.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: Not applicable in the context of an algorithm or AI. The "performance" assessment refers to the mechanical integrity and equivalence of the physical device design via FEA.
7. The Type of Ground Truth Used
- Type of Ground Truth: The ground truth is based on established engineering standards and material properties. Specifically, the mechanical performance is evaluated against the requirements set forth in ASTM F2077 (Standard Test Methods for Static, Dynamic, and Torsional Axial Compression, Distraction, and Shear Testing of Spinal Arthrodesis Implants) and the material specifications of ASTM F136, ASTM F2026, and ASTM F560. The "ground truth" is that the device, when subjected to these simulated load conditions, performs comparably to the predicate devices and does not create a new "worst case."
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This submission does not involve machine learning or AI, so there is no "training set." The FEA models are built based on geometries and material properties, not trained on data.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no training set.
Ask a specific question about this device
(55 days)
The Spinal USA VBR System is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture).
The Spinal USA VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column
The Spinal USA VBR System is intended to be used with supplemental spinal fixation systems that have been cleared for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). In addition, the Spinal USA VBR System is intended for use with bone graft.
The Spinal USA VBR System is a vertebral body replacement device that is implanted into the vertebral body space to improve stability of the spine. The system consists of a straight, curved, round, and trapezoidal implants. The various implants via shallsin, curved, curved, curved, the individual patient anatomy. All components are made from medical grade titanium or titanium or titanium or titanium or titanium or titanium or titanium or titanium or tit alloy described by such standards as ASTM F136 or ISO5832-3. The products are supplied clean and "NON-STERILE".
This document is a 510(k) summary for the Spinal USA VBR System, a vertebral body replacement device. It focuses on demonstrating substantial equivalence to previously cleared devices rather than a detailed performance study with acceptance criteria in the typical sense of AI/software devices. Therefore, many of the requested categories are not applicable or cannot be extracted directly from this type of regulatory submission.
Here's a breakdown of the information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
This type of information is not available in the provided 510(k) summary. For medical devices seeking 510(k) clearance, the primary "acceptance criterion" is often demonstrating substantial equivalence to a predicate device. This is typically shown through:
- Materials testing: Ensuring the device materials meet established standards (e.g., biocompatibility, mechanical properties).
- Mechanical testing: Demonstrating that the device performs equivalently or better than predicate devices under specified biomechanical loads.
- Design similarities: Highlighting that the device's design and intended use are similar to predicate devices.
The document states: "Testing in accordance with ASTM F2077-03 'Test Methods for Intervertebral Body Fusion Devices' of the Spinal USA VBR System demonstrates that the device is substantially equivalent to the Novel VBR Spinal System (K042201), Lanx VBR System (K052384), Depuy Acromed Stackable Cage (K990148), Quantum VBR Quantum Orthopedics(K050449), PEEK Tetris Signus Medical (K031757), CO VBR Scient X(K050348), Ellys and Aurys VBR Scient'X (K033109)."
This statement implies that the device met the performance requirements outlined in ASTM F2077-03, which serves as the "acceptance criteria" for mechanical performance, to establish substantial equivalence. However, the specific quantitative criteria (e.g., "withstanding X N of compression for Y cycles") and the exact performance results are not reported in this summary.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Conformance to ASTM F2077-03 "Test Methods for Intervertebral Body Fusion Devices" for mechanical performance relative to predicate devices. | The device "demonstrates that the device is substantially equivalent" to listed predicate devices through testing in accordance with ASTM F2077-03. Specific numerical performance data are not provided in this summary. |
Materials meeting medical-grade standards (e.g., ASTM F136 or ISO5832-3). | All components are made from medical grade titanium or titanium alloy described by such standards as ASTM F136 or ISO5832-3. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable in the context of a mechanical device 510(k) submission like this. The "test set" here refers to the physical devices undergoing mechanical and material testing, not a dataset of patient images or clinical outcomes. Data provenance for such mechanical testing typically relates to the lab where tests were conducted, not country of origin of patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable. Ground truth in the context of mechanical device testing refers to established engineering specifications, material standards, and biomechanical principles, not expert medical opinion on images or clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for mechanical device testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This document describes a physical medical implant, not an AI or diagnostic imaging device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. This document describes a physical medical implant, not an AI or diagnostic imaging device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance relies on established engineering standards, material specifications, and biomechanical testing protocols. The device's ability to maintain spinal stability and integrity, as well as its material properties, are objectively measured against these predefined standards. For instance, the ASTM F2077-03 standard specifies methods for evaluating intervertebral body fusion devices, and conformance to this standard serves as a form of "ground truth" for mechanical performance.
8. The sample size for the training set
This is not applicable. There is no "training set" in the context of a 510(k) submission for a mechanical implant.
9. How the ground truth for the training set was established
This is not applicable. There is no "training set" in the context of a 510(k) submission for a mechanical implant.
Ask a specific question about this device
(46 days)
The Sustain Radiolucent Spacer is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The Sustain Radiolucent Spacer is intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material.
The Sustain Radiolucent Spacer is designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
The Sustain Radiolucent Spacer is a vertebral body replacement device used to provide structural stability in skeletally mature individuals following corpectomy or vertebrectomy. The system is comprised of spacers of various fixed heights and footprints to fit the anatomical needs of a wide variety of patients. Each spacer has an axial hole to allow grafting material to be packed inside of the spacer. Protrusions on the superior and inferior surfaces of each device will grip the endplates of the adjacent vertebrae to resist expulsion.
The Sustain Radiolucent Spacer devices are made from radiolucent polymer and titanium alloy as specified in ASTM F2026 and F136.
The provided 510(k) summary for the Globus Medical Inc. Sustain Radiolucent Spacer details a medical device, not a software algorithm or AI-powered system, therefore, the requested information regarding acceptance criteria and a study proving device performance in the context of AI/ML is not applicable.
The document indicates that the device's substantial equivalence is based on its similarity to a predicate device (Globus Sustain Spacer K031302) and highlights mechanical testing according to "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", September 27, 2000. This implies that the 'acceptance criteria' were met through established biomechanical benchmarks for spinal implants, rather than performance metrics related to diagnostic accuracy, sensitivity, or specificity that would be relevant for AI studies.
Therefore, I cannot populate the table or answer the specific questions about sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as these pertain to the evaluation of AI/ML models.
Ask a specific question about this device
(175 days)
Ellys and Aurys Vertebral Body Replacements are vertebral body replacements for use in the lumbar and thoracic spine (T1-L5) to replace a damaged, collapsed or unstable vertebral body due to tumor or trauma (i.c. fracture). These are not stand-alone devices, ISOBAR Ø6.2 Hemispherical Screws with Offset Clamps and Ø5.5 Rods must be utilized to enhance the stability of the reconstruction in skeletally mature patients following full or partial corpectomy.
The Ellys components are kidney shaped and are used in pairs for full corpectomy or individually when only a portion of the vertebral body is resected. The interior of the space for home graft. The implant is always implanted in the vertical position. These components are available in various heights (8mm-150mm) to accommodate the variability of patient size, anatomic variation, and the size of the vertebral body defect. The superior and inferior surfaces of these components have ridges to interface with the vertebral endplates to resist rotation and migration. The body has a multitude of holes to allow additional impaction of bone graft.
The Aurys components are trapezoidal in cross sccion and are used singly. The interior of these components have three hollow compartments that run the vertically and are provided for the placement of bone grafi. I like the Aurys components are also available in various hcights (9mm-1.50mm) and have angled ends to match the angle of the vertebral endplates. The ends have ridges similar to the Ellys to interface with the vertebral endplates. There are also holes evenly spaced on the medial/lateral surfaces to complete bone graft impaction after insertion.
Both the l'llys and Aurys components are fabricated from pure poly(ether cther ketone) (PEIFA). This material closcly matches the modulus of elasticity of cortical bone, improving the biomechanical interface and reducing the stress shickling clicci. They are fully radio-translucent, which enables optimum follow-up with diagnostic imaging, as the interbody fusion progresses. Two metal wires at the opposite ends of the spacers allow radiological confirmation of the cage position post operatively.
The provided text describes a medical device, the Ellys and Aurys Vertebral Body Replacements, and its regulatory clearance (510(k)). However, it does not contain information about acceptance criteria for performance, a study proving the device meets those criteria, or any specifics regarding device performance metrics, sample sizes, expert involvement, or ground truth establishment.
The document is a 510(k) premarket notification summary and the FDA's clearance letter. It focuses on:
- Device Description: Physical characteristics, materials, and available sizes of the Ellys and Aurys components.
- Intended Use/Indications for Use: The specific medical conditions and anatomical locations for which the device is intended.
- Equivalent Device: Listing of similar previously cleared devices.
- Nonclinical Tests: A brief mention that "Testing was performed per ASTM F1717." This indicates a standard was followed for mechanical testing, but no results or acceptance criteria are provided in this document.
Therefore, I cannot populate most of the requested table and answer the questions about the study that proves the device meets the acceptance criteria, as this information is not present in the provided text.
Here's an attempt to answer based on the limited information given, noting where information is absent:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
---|---|---|
Biomechanical Testing | Adherence to ASTM F1717 | Not specified in document |
Material Properties | PEK material properties | Not specified in document |
Radiological Confirmation | Visibility of metal wires | "Two metal wires at the opposite ends of the spacers allow radiological confirmation of the cage position post operatively." (No quantitative criteria or performance data given) |
Stability (in vivo) | Not specified in document | "These are not stand-alone devices, ISOBAR (6.2 Hemispherical Screws with Offset Clamps and 05.5 Rods must be utilized to enhance the stability of the reconstruction..." (Indicates dependency on other devices for stability, but no performance data) |
Migration/Rotation | Resistance to rotation and migration | "The superior and inferior surfaces of these components have ridges to interface with the vertebral endplates to resist rotation and migration." (No quantitative criteria or performance data) |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not specified. The document only mentions "Testing was performed per ASTM F1717" without any details on sample size or specific test results.
- Data Provenance: Not specified. Given the nature of a 510(k) summary for a physical device, testing would typically be in vitro (bench testing) or potentially ex vivo (using cadaveric spines), often conducted by the manufacturer or a contracted testing lab. There is no indication of clinical data or patient data in this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. The document describes pre-market notification for a physical medical device, not an AI/software device that would require expert-established ground truth for performance evaluation in the context of image interpretation or similar tasks. Biomechanical testing per ASTM F1717 does not typically involve expert consensus for ground truth.
4. Adjudication method for the test set
- Not applicable. This concept is relevant for studies involving human interpretation (e.g., medical imaging reviews) where discrepancies need resolution. For biomechanical testing, results are typically objective measurements.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/software device. No MRMC study was performed or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/software device. No algorithm standalone performance was done or mentioned.
7. The type of ground truth used
- For the nonclinical (biomechanical) testing, the "ground truth" would be the objective measurements obtained according to the ASTM F1717 standard's protocols, compared against performance specifications derived from the standard or predicate devices. This is not the "expert consensus, pathology, or outcomes data" typically associated with AI/diagnostic device ground truth.
8. The sample size for the training set
- Not applicable. This is not an AI/software device. There is no mention of a training set.
9. How the ground truth for the training set was established
- Not applicable. This is not an AI/software device. There is no mention of a training set or its ground truth establishment.
Ask a specific question about this device
Page 1 of 1