510(k) Data Aggregation

    K Number
    K092905
    Manufacturer
    Date Cleared
    2010-03-04

    (164 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K023595

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Administration of fluids from a container to a patient's vascular system through a cannula inserted into a vein.

    Device Description

    The AutoStart Burette is an in-line burette used for intravenous administration of fluids (either gravitational or via an infusion pump) from a container to a patient's vascular system through a cannula inserted into a vein. This is a sterile, single-use device. The device is fitted with a float system which that allows the primary infusion to restart following addition to and dispensation of medication from the burette. There is no need to manually restart the infusion as is the case with conventional in-line burettes.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    The provided document (K092965) for the Analytica AutoStart Burette does not contain a specific table of acceptance criteria and reported device performance in the format typically seen for algorithm-based devices. Instead, it demonstrates substantial equivalence based on adherence to an FDA recognized standard and risk assessment.

    Here's a breakdown of the information that can be extracted or inferred regarding performance and acceptance based on the submission:

    Acceptance Criteria (Inferred from regulatory compliance)Reported Device Performance
    Compliance with ISO 8536-5:2004The AutoStart Burette complies with ISO 8536-5:2004 Infusion equipment for medical use Part 5 – Burette infusion sets for single use, gravity feed – the FDA recognized safety and performance standard for these devices.
    Equivalence of Indications for UseIdentical indications for use as the predicate device (Tuta Healthcare In-Line Burette 150mL, K023595).
    Equivalent Materials and ComponentsManufactured from medical grade polymers, equivalent to the predicate device.
    Equivalence of OperationBoth devices are 150mL capacity, graduated, flow regulated in-line burettes.
    No Significant New Safety ConcernsRisk assessment performed in conformance with ISO 14971:2007, and no significant new safety concerns raised by the design.
    AutoStart Flotation Feature FunctionalityThe device is fitted with a float system which allows the primary infusion to automatically restart following addition to and dispensation of medication. This is a key distinguishing feature and implies performance in this aspect.

    2. Sample Size for Test Set and Data Provenance

    This document is for a medical device (burette) and not an AI/algorithm-based device that would typically involve test sets of data (e.g., medical images, patient records). Therefore, the concepts of "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) are not applicable in this context. The evaluation is based on physical device testing and compliance with established standards, not analysis of clinical data in the traditional sense for AI.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. As this is not an AI/algorithm-based device, there is no need for experts to establish ground truth from a test set of data. The "ground truth" for this device's performance is its adherence to engineering specifications and regulatory standards.

    4. Adjudication Method

    Not applicable for the reasons stated above.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI-assisted device, so there would be no human readers improving with or without AI assistance.

    6. Standalone Performance Study

    Yes, in a way, standalone testing of the device's physical and functional characteristics was performed to demonstrate compliance with ISO 8536-5:2004. This standard outlines tests for various aspects of infusion sets, including:

    • Flow rate consistency: Ensuring the device delivers fluids at regulated rates.
    • Capacity and Graduation Accuracy: Verifying the 150mL capacity and the accuracy of markings.
    • Material Biocompatibility: Assessed implicitly through the use of "medical grade polymers."
    • Leakage and Integrity: Implied by meeting a safety standard.
    • Functionality of the AutoStart feature: Although specific test results are not detailed, the description of the feature implies it underwent testing to ensure it works as intended.

    7. Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on:

    • Engineering Specifications and Design Documents: The design of the Analytica AutoStart Burette.
    • Regulatory Standards: ISO 8536-5:2004 ("Infusion equipment for medical use Part 5 – Burette infusion sets for single use, gravity feed").
    • Predicate Device Characteristics: The performance and safety profile of the Tuta Healthcare In-Line Burette (K023595).

    8. Sample Size for Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set of data.

    9. How Ground Truth for Training Set Was Established

    Not applicable.