Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133540
    Device Name
    FLOCAP
    Manufacturer
    Date Cleared
    2014-03-28

    (130 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K021576

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FLOCAP is to provide a semi-quantitative visualization of the CO2 in the patient airway. It is an adjunct in patient assessment, to be used in conjunction with other methods to determine clinical signs and symptoms by or on the order of a physician.

    The FLOCAP has a visual indicator to visually detect the end of exhalation.

    For use up to 24 hours.

    For patients greater than 15 kg (33 lbs.)

    Environment of use - hospital, sub-acute, pre-hospital, transport

    Device Description

    The proposed FLOCAP is comprised of several components:

    • Housing with standard 15 mm / 22 mm fittings to connect to ventilatory assist devices and a face . mask or endotracheal tube
    • . Colorimetric litmus media which has been treated with chemical to detect the presence of CO2 by a change in pH
    • . A spinner / vane which spins when there is expiratory flow. It is to detect presence of expiratory flow
    AI/ML Overview

    The FLOCAP device, a colorimetric CO2 detector, was deemed substantially equivalent to the predicate device, Mercury Medical STAT CO2 (K021576), based on non-clinical performance testing and comparative specifications.

    Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Predicate: Mercury Medical STAT CO₂)Reported Device Performance (FLOCAP)
    Indications for UseSemi-quantitative visualization of CO₂ in patient airway; adjunct in patient assessment.Identical: Semi-quantitative visualization of CO₂ in patient airway; adjunct in patient assessment, with a visual indicator for end of exhalation.
    Patient Use / Duration if useSingle patient use, disposable, < 24 hoursIdentical: Single patient use, disposable, < 24 hours
    Environment of UseHospital, TransportHospital, Sub-acute, Pre-hospital, Transport (Clarification, considered similar)
    Patient PopulationGreater than 15 kg (33 lbs.)Identical: Greater than 15 kg (33 lbs.)
    ContraindicationsSimilar contraindications providedSimilar contraindications, with addition of "hypercarbia" which was deemed not to alter the contraindications materially.
    Principle of OperationColorimetric, pH sensitive dyeIdentical: Colorimetric, pH sensitive dye
    PlacementBetween manual resuscitator/pressure source and patient face mask/endotracheal tubeIdentical: Between manual resuscitator/pressure source and patient face mask/endotracheal tube
    Standard 15 / 22 mm connections per ISO 5356-1YesYes
    Internal Volume (dead space)25 ml25 ml (Identical)
    Weight22 gr23 gr (Similar)
    Compliance per ISO 9360-1Not provided0.44 ml/kPa
    Leakage per ISO 9360-1Not provided0.0 ml/min
    Pressure Drop per ISO 9360-1@ 60 lpm - 3.0 cm H₂O@ 30 lpm - 0.7 cm H₂O;@ 60 lpm - 2.7 cm H₂O;@ 90 lpm - 5.7 cm H₂O (Similar)
    Shelf-life2 years2 years (Identical)
    Packaging and method to "activate"Polybag, remove plastic tabFoil pouch, nothing required to be done to activate (Considered not to raise new concerns)
    Detected % CO₂ ranges and Colors0% CO₂ - Blue; 1.0-2.0% CO₂ - Green; > 5.0% CO₂ - Permanent Yellow0% CO₂ - Purple; 1.0-2.0% CO₂ - Beige; > 5.0% CO₂ - Yellow (Different colors, but similar functionality of color change for CO₂ detection)
    Method of Communicating Meaning of Color ChangesMatching Colored Label on the Outside of the deviceIdentical: Matching Colored Label on the Outside of the device
    Meanings of detecting patient exhalationColor changeColor change; Spinning vane - visual indicator (detects exhalation flow > 1 lpm) (Added feature, enhances detection)
    StandardsISO 5356-1ISO 5356-1, ISO 9360-1 HME in part (Adherence to relevant standards)
    Filter elementPresentNot present (Considered not to raise new concerns as patient secretions mitigation is similar to predicate's instructions)
    Pull tab to activatePresentNot present (Packaging demonstrated safe storage without deterioration)

    Study Information

    1. Sample size used for the test set and the data provenance: Not explicitly stated for specific tests. The summary indicates "comparative specifications" and "specific tests related to the device" were performed. The testing seems to be bench testing against the predicate's specifications rather than a clinical study with a patient test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as detailed clinical ground truth establishment by experts is not described for the non-clinical performance testing.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: An MRMC comparative effectiveness study was not performed. This is a non-AI medical device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Standalone performance was assessed through various non-clinical bench tests (e.g., color change response time, leakage, minimum static flow indication, anti-fog, shelf-life, drop test, duration of use, burst pressure, compliance, pressure drop, internal volume, usability).
    6. The type of ground truth used: For the bench testing, the "ground truth" was established based on engineering specifications, physical measurements, and comparison to the predicate device's characteristics and performance parameters.
    7. The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1