Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K022363
    Date Cleared
    2002-09-19

    (59 days)

    Product Code
    Regulation Number
    870.3800
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K014161

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SJM® Tailor™ annuloplasty band is indicated for use in the repair of a mitral or tricuspid valve that is diseased or damaged due to acquired or congenital processes.

    Device Description

    The Tailor band is a flexible band fabricated with a medical grade silicone rubber core, surrounded by polyester fabric.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device, the SJM® Tailor™ annuloplasty band. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting new clinical studies to establish acceptance criteria and performance. Therefore, most of the requested information regarding acceptance criteria, study details, and human-AI comparative effectiveness is not applicable to this document.

    Here's a breakdown of the parts that can be addressed based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify "acceptance criteria" in a quantitative sense for performance metrics like sensitivity, specificity, or accuracy. Instead, the "acceptance criteria" are implied by demonstrating substantial equivalence to the predicate device through various non-clinical tests.

    CharacteristicAcceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Product LabelingSubstantially Equivalent to predicateSubstantially Equivalent
    Intended UseIdentical to predicateIdentical
    Physical CharacteristicsSubstantially Equivalent to predicateSubstantially Equivalent
    Anatomical SitesIdentical to predicateIdentical
    Target PopulationIdentical to predicateIdentical
    Performance TestingIdentical to predicateIdentical
    Safety CharacteristicsIdentical to predicateIdentical
    Band end evaluationDemonstrated satisfactory performance in comparison to predicatePerformed
    GappingDemonstrated satisfactory performance in comparison to predicatePerformed
    Band releaseDemonstrated satisfactory performance in comparison to predicatePerformed
    Suture RetentionDemonstrated satisfactory performance in comparison to predicatePerformed
    Suture pullDemonstrated satisfactory performance in comparison to predicatePerformed
    Biocompatibility evaluationDemonstrated satisfactory performancePerformed
    Sterility Assurance evaluationDemonstrated satisfactory performancePerformed
    Manufacturing Process ValidationDemonstrated satisfactory performancePerformed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for the test set: Not applicable. This submission relies on non-clinical testing and comparison to a predicate device, not a patient-based test set in the context of AI or diagnostic performance. The tests mentioned are for material and mechanical properties, not clinical assessment on patients.
    • Data provenance: Not applicable. The tests performed are laboratory-based and do not involve patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. There is no "ground truth" establishment by experts in the context of image interpretation or diagnosis for this device's non-clinical testing. The tests are physical and microbiological validations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no "adjudication method" in the context of clinical interpretation or expert review for this type of non-clinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for heart valve repair (an annuloplasty band), not an AI diagnostic or assistance tool that would undergo MRMC studies involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles for material integrity, mechanical strength, and biocompatibility, as well as validated testing methods. It's not a clinical "ground truth" derived from patient-specific data.

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1