Search Results
Found 1 results
510(k) Data Aggregation
(487 days)
Prolong is used as part of a climax control program for males who suffer from premature ejaculation. It is designed to increase the time between arousal and ejaculation through the use of the training sessions using the stimulating vibratory effects of the device on the penis as part of the "start-stop" technique. Prolong is intended to be available as an over-the-counter device.
The PROLONG™ device consists of the PROLONG™ vibrating device and a 4-ounce bottle of water-based lubricant. The lubricant was FDA cleared under K013086. The PROLONG™ vibrating device includes a white silicone body and a vibrator unit as shown in Figure 1. The vibrator unit comprises a 3-Volt coin (pancake) motor connected to a battery pack of two battery cells (Type PX625A) shrink wrapped together and an Acrylonitrile Butadiene Styrene (ABS) slide switch unit to activate the motor. The device should be disposed of at the end of the six week treatment.
Acceptance Criteria and Device Performance for PROLONG™
1. Table of Acceptance Criteria and Reported Device Performance
This device is not a diagnostic AI device, so the acceptance criteria are not in terms of traditional metrics like sensitivity, specificity, or accuracy. Instead, the acceptance criteria relate to safety, biocompatibility, and demonstrated effectiveness in improving time to ejaculation and patient satisfaction.
Acceptance Criterion | Reported Device Performance | Study that Proves the Device Meets Acceptance Criteria |
---|---|---|
Biocompatibility: Patient-contacting materials are non-cytotoxic, non-sensitized, and produce a negligible irritation response. | Extracts were determined to be non-cytotoxic, non-sensitized, and produced a negligible irritation response. | Biocompatibility testing according to ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and skin sensitization). |
Shelf Life (Battery): Battery capacity sustains specified use for 3 years. | Battery testing demonstrated the shelf life of the battery is 3 years, and devices performed satisfactorily over a 6-week period near end of shelf life. | Battery testing over a 6-week period with devices being switched on and off, conducted on units nearing or at the end of their three-year shelf life. |
Electromagnetic Compatibility (EMC) & Electrical Safety: Device complies with applicable EMC and electrical safety standards. | Device complies with all applicable parts of IEC 60601-1, EN 60601-1-2, 55014-1, 61000-4-2, 61000-4-3, 61000-6-1, 61000-6-3, 61000-4-8, and ANSI/AAMI ES60601-1. | Testing of the battery and motor against the requirements of IEC 60601-1, EN 60601-1-2, 55014-1, 61000-4-2, 61000-4-3, 61000-6-1, 61000-6-3, 61000-4-8, and ANSI/AAMI ES60601-1. |
Mechanical Safety: Device withstands forces encountered during use. | Pull testing and mechanical durability testing were conducted. (Results stated as "demonstrated that the device will withstand forces encountered during use"). | Mechanical Safety - Pull testing and mechanical durability testing. |
Effectiveness (Increase in time to ejaculation): Demonstrated improvement in Intra-Vaginal Ejaculatory Latency Time (IELT) or masturbation time. | For 23 subjects completing the 6-week study, there was a median increase in time to ejaculation of 3 minutes. | Clinical audit/evaluation. |
Effectiveness (Reduction in distress): Reduction in distress related to premature ejaculation. | Overall improvement in both treatment groups for distress parameters (anxiety, depression) as measured by BAI and BDI-II. | Clinical audit/evaluation (comparison of pre- and post-treatment scores). |
Effectiveness (Improvement in sexual satisfaction): Improvement in parameters of sexual satisfaction. | Overall improvement in both treatment groups for parameters of sexual satisfaction (sexual function, sexual relationship) as measured by GRISS. High levels of satisfaction on a 10-point Likert scale. | Clinical audit/evaluation (comparison of pre- and post-treatment scores, Likert satisfaction survey). |
Safety (Adverse Events): No adverse events during use. | No reported (0%) adverse events for the Prolong hand-held device in either lubricant group. | Clinical audit/evaluation. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 33 subjects were enrolled in the clinical audit/evaluation. 23 subjects completed the full six-week treatment period and were included in the primary outcome analysis for time to ejaculation.
- Data Provenance: The data was collected from a single-site, prospective, open-label clinical audit/evaluation. The country of origin is not explicitly stated, but the contact information for Ergon Medical Ltd. is in the UK, and a European CE Mark is mentioned, suggesting the study likely took place in the UK or Europe.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This device does not involve a "ground truth" established by experts in the typical sense of a diagnostic AI device. The outcomes are self-reported (e.g., IELT/masturbation time recorded by subjects with a stopwatch, questionnaire scores for distress and satisfaction) or objective measurements (e.g., adverse events). Therefore, no experts were used to establish a ground truth in this context.
4. Adjudication Method for the Test Set
Not applicable. There was no expert adjudication process as the outcome measures were either objective measurements (e.g., stopwatch time, adverse events) or standardized questionnaire scores. Subjective satisfaction was rated by the subjects themselves on a Likert scale.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The study was a clinical audit/evaluation of the device's performance, not a comparison of human readers with and without AI assistance.
However, a comparison with Cognitive Behavioral Therapy (CBT) was presented, referencing a study by Zamar (2012) using Prolong in 58 subjects against men having CBT.
- Effect Size of human readers improvement with AI vs without AI assistance: Not applicable as this is not an MRMC study related to "human readers" and AI. The comparison was device vs. non-device therapy. The Zamar (2012) study showed a difference of 6 minutes and 12 seconds (6.2 minutes) in delayed time to ejaculation between the Prolong group and the CBT group. In the Prolong group, 61% of men delayed time to ejaculation from 0.8 minutes to 8.8 minutes on average. In the CBT group, 40% of men delayed time to ejaculation to 2.6 minutes on average.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the clinical audit/evaluation assesses the standalone performance of the PROLONG™ device used by subjects as instructed. It measures the device's effectiveness and safety without human-in-the-loop (e.g., a clinician interpreting an AI output). The device itself is the intervention.
7. The Type of Ground Truth Used
The "ground truth" for the effectiveness of the PROLONG™ device was established through:
- Subject-reported outcomes:
- Direct measurement: Intra-Vaginal Ejaculatory Latency Time (IELT) or masturbation time recorded by subjects using a stopwatch.
- Standardized questionnaires: Scores from validated psychological inventories (Golombock Rust Inventory of Sexual Satisfaction (GRISS), Beck Anxiety Inventory (BAI), and Beck Depression Inventory (BDI-II)).
- Subjective satisfaction: 10-point Likert satisfaction scale.
- Safety data: Reporting of adverse events by subjects.
- Bench testing: For biocompatibility, electrical safety, mechanical safety, and shelf life, the "ground truth" was established by adherence to recognized international and national standards (e.g., ISO, IEC, ANSI/AAMI).
8. The Sample Size for the Training Set
This device is not an AI/ML algorithm that requires a "training set" in the traditional sense. The device is a physical product, and its design and function are based on engineering principles and clinical research, not machine learning model training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this physical device.
Ask a specific question about this device
Page 1 of 1