510(k) Data Aggregation

    K Number
    K140644
    Device Name
    PROCEM
    Manufacturer
    Date Cleared
    2014-06-19

    (98 days)

    Product Code
    Regulation Number
    872.3275
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K003567, K080032

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ProCem™ is intended for non-weight bearing applications in Otologic surgery, such as:

    1. Reconstruction of ossicular bones or mechanical coupling of the ossicular chain.
    2. Mechanical stabilization of middle ear and cochlear implants.
    3. Attachment of middle ear implants to the ossicular bones.
    Device Description

    ProCem™ is glass ionomer cement that is provided as two components, a glass powder and polyacrylic acid liquid. By mixing the two components, viscous moldable ionomeric cement is obtained which hardens in situ. ProCem™ is provided in a sterile capsule. The capsule includes the two separate components, a glass powder and polyacrylic acid liquid. An activator is used to release the liquid component prior to use. The capsule is then placed in a mixer, which mixes the glass powder and polyacrylic acid liquid to form a compliant cement. The applicator is used to apply the compliant cement to the appropriate location in the middle ear. The cement then hardens in place to provide a permanent bond.

    AI/ML Overview

    Acceptance Criteria and Device Performance for ProCem™ Otologic Bone Cement

    This document summarizes the acceptance criteria and supporting study for the ProCem™ Otologic Bone Cement, as derived from the provided 510(k) summary (K140644).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for ProCem™ Otologic Bone Cement are based on its substantial equivalence to predicate devices (SerenoCem, K003567 and EnvoyCem, K080032). The study conducted was primarily to demonstrate that ProCem™ meets physical and functional requirements comparable to these predicate devices.

    Acceptance CriteriaReported Device Performance (ProCem™)Reference to Predicate Device
    Exothermic ReactionSuccessfully performed and met requirementsComparable to SerenoCem
    Working TimeSuccessfully performed and met requirementsComparable to SerenoCem
    Snap Set TimeSuccessfully performed and met requirementsComparable to SerenoCem
    Simulated Preparation PropertiesSuccessfully performed and met requirementsComparable to SerenoCem
    BiocompatibilityDemonstrated compliance to biocompatibility testing required for intended useGIC cements (including predicates) have a long history of safe and effective use
    Sterility (SAL)10-6 SAL achieved via gamma radiationPackaged in the same manner as predicate devices to assure sterility

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes "Design verification testing" but does not explicitly state specific sample sizes for each test (exothermic reaction, working time, snap set time, simulated preparation properties). The data provenance is not specified in terms of country of origin, but the studies were conducted by Ototronix LLC, the submitter, as part of their 510(k) application. This would be considered a prospective study conducted for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    There is no mention of experts being used to establish a "ground truth" in the context of clinical performance or diagnostic accuracy. The testing described focuses on the physical and functional properties of the bone cement itself, rather than interpretation of clinical outcomes by experts.

    4. Adjudication Method for the Test Set

    Not applicable. The described testing is focused on objective physical and chemical properties of the device, not on subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This device is a bone cement, not an imaging or diagnostic AI device that would typically involve human readers.

    6. Standalone Performance Study

    Yes, a standalone study was done. The "Performance Bench Testing" describes in-house design verification testing, demonstrating the physical and functional properties of the ProCem™ device on its own. This testing was performed independent of human interaction in a clinical setting for performance evaluation.

    7. Type of Ground Truth Used

    The ground truth for the performance bench testing was based on established engineering specifications and comparison to the known performance characteristics of the predicate device (SerenoCem). This involves objective measurements of physical and chemical properties.

    8. Sample Size for the Training Set

    Not applicable. The ProCem™ Otologic Bone Cement is a physical medical device (bone cement), not a software algorithm that would require a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of medical device.