Search Results
Found 1 results
510(k) Data Aggregation
K Number
K221693Device Name
Elecsys Anti-HCV II (08837031190)
Manufacturer
Date Cleared
2023-01-04
(208 days)
Product Code
Regulation Number
N/AWhy did this record match?
Product Code :
MZO
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Immunoassay for the in vitro qualitative detection of antibodies to hepatitis C virus (HCV) in human adult and pediatric (ages 18 months through 21 years) serum and plasma (potassium EDTA, lithium heparin, sodium heparin, and sodium citrate). Assay results, in conjunction with other laboratory results and clinical information, may be used to aid in the presumptive diagnosis of HCV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis C infection. The test does not determine the state of infection or associated disease.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
Device Description
Immunoassay for the in vitro qualitative detection of antibodies to hepatitis C virus (HCV) in human adult and pediatric (ages 18 months through 21 years) serum and plasma (potassium EDTA, lithium heparin, sodium heparin, and sodium citrate). Assay results, in conjunction with other laboratory results and clinical information, may be used to aid in the presumptive diagnosis of HCV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis C infection. The test does not determine the state of infection or associated disease.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
The Elecsys Anti-HCV II Immunoassay employs "ECLIA" technology and is a qualitative serologic, two step sandwich assay. The assay detects total antibodies to HCV in serum and plasma samples. The total duration of the assay is 18 minutes. The basic device methodology is as follows:
1. 1st incubation: 50 µL of sample, 55 µL of a reagent containing biotinylated HCV antigens, and 55 µL of a reagent containing HCV antigens labeled with a ruthenium complex react to form a sandwich complex.
2. 2nd incubation: After addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
3. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission, which is measured by a photomultiplier.
4. Results are determined automatically by the Elecsys software by comparing the electrochemiluminescence signal obtained from the sample with the cut-off value obtained by the anti-HCV calibration.
Ask a Question
Ask a specific question about this device
Page 1 of 1