Search Results
Found 26 results
510(k) Data Aggregation
(105 days)
MQH
Denudation Pipettes are used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART).
Denudation Pipettes are not intended for biopsy of cells from oocytes or embryos.
The Denudation Pipettes are extruded polycarbonate capillary tubes that are pulled at one end to form a tapered tip. They have an outer diameter (OD) of 0.9mm at the proximal end and fit into an actuating device. All tips are approximately 90mm in length and depending on the size of the pipette have a volumetric capacity of 18.6-21.9µl.
The Denudation Pipettes are supplied in a range of inner diameter (ID) sizes at the distal end as shown below:
- 80μm, 130μm, 140μm, 150μm, 160μm, 170μm, 180μm, 190μm, 210μm, 230μm, 250μm, 270μm, 290μm.
The Denudation Pipettes are radiation sterilized and provided in a sterile pouch, supplied in two packaging options. Each pouch contains 5 or 10 pipettes. All pipettes are intended for single use and disposable.
This 510(k) clearance letter is for Denudation Pipettes, a Class II medical device used in assisted reproductive technologies. This type of device is an instrument and does not fit the typical format for AI/ML-enabled devices so I have focused on the relevant mechanical and biological performance characteristics.
Acceptance Criteria and Device Performance
There is no single study described that proves the device meets all acceptance criteria. Instead, a series of non-clinical performance tests were conducted.
Here's a table summarizing the acceptance criteria and the reported device performance for the Denudation Pipettes:
Acceptance Criteria Category | Specific Test/Characteristic | Acceptance Criteria | Reported Device Performance | Comments |
---|---|---|---|---|
Biocompatibility/Embryo Safety | One-cell Mouse Embryo Assay (MEA) | ≥ 80% embryos developed to expanded blastocyst at 96 hours | ≥ 80% embryos developed to expanded blastocyst at 96 hours | Meets criteria. The subject device's MEA assessment time (96h) is shorter than the predicate (120h), but it's noted as the standard assessment time for this assay. |
Material Toxicity | Endotoxin (LAL, USP) |
Ask a specific question about this device
(173 days)
MQH
V-DENUPET is used for the removal of cumulus oocyte complexes (COC) from an oocyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) as well as for the handling of oocytes and embryos during assisted reproductive techniques (ART). V-DENUPET is not intended for biopsy of cells from oocytes or embryos.
The V-DENUPET is a polycarbonate micropipette. All tips are 90 mm in length, and depending on the size of the tip have a volumetric capacity of 20-25 µl. All pipettes have an outer diameter of 900 µm at the proximal end that is connected to an aspiration device.
The V-DENUPET is supplied in a range of inner diameter sizes at the distal end as shown below:
- Sizes 125 µm, 135 µm, 140 µm, 150 µm, and 175 µm are suitable for oocyte . denudation.
- Sizes 175 µm, 200 µm, 275 µm, 300 µm, and 600 µm are suitable for oocyte . and embryo handling.
V-DENUPET micropipettes are radiation sterilized and provided in a sterile pouch containing a polypropylene vial containing 10 micropipettes. Each pipette is for single-use only.
The provided text describes the V-DENUPET, a micropipette used in Assisted Reproductive Technologies (ART). However, it does not include acceptance criteria or a study proving the device meets acceptance criteria in the context of AI/algorithm performance. The document is an FDA 510(k) summary for a medical device (a micropipette), focusing on non-clinical performance testing for physical characteristics, sterility, and biocompatibility, not for an AI device.
Therefore, I cannot provide the requested information about acceptance criteria for an AI device, study details, sample sizes, expert qualifications, or MRMC studies, as this information is not present in the provided text.
The closest relevant information from the document is related to the device's performance criteria, not an AI algorithm:
Acceptance Criteria and Reported Device Performance (Non-AI device)
Acceptance Criteria (V-DENUPET) | Reported Device Performance (V-DENUPET) |
---|---|
Sterilization Validation | Per ISO 11137-1:2006 and ISO 11137-2:2013 |
Package Integrity Testing | Seal strength (ASRM F88/F88M-15 equivalent), Burst (ASTM 2054-07), Dye Penetration (ASTM F1929-98) |
Shelf Life Testing (Device Function) | |
- MEA testing (1-Cell) | ≥80% hatched blastocysts at 96 hours (at Time 0 and 36 months) |
- Endotoxin (LAL, USP) |
Ask a specific question about this device
(262 days)
MQH
EZ-Tip pipettes are for denudation, i.e. removing the cumulus from an occyte prior to Intracytoplasmic Sperm Injection (ICSI) and In Vitro Fertilization (IVF) and for handling gametes, embryos and biopsied cells (polar bodies, blastomeres and trophectoderm) during assisted reproductive techniques (ART). EZ-Tips are not intended for biopsy of cells from oocytes or embryos.
The EZ-Grip provides aspiration and expulsion capabilities to plastic pipettes when fitted during assisted reproduction procedures.
The EZ-Tip pipette is an extruded medical grade polycarbonate capillary that is pulled at one end to form a tapered tip. It has an outer diameter (OD) of 0.9 mm at the proximal end and fits to an actuating device such as the EZ-Grip. All tips are approximately 90 mm in length and depending on the size of the tip they have a volumetric capacity of 15.9 - 25.4 µl.
Pipette Tips are supplied in a range of inner diameter (ID) sizes at the distal end as shown below:
- Size 75 µm, 200 µm, 250 µm, 290 µm, 600 µm are suitable for specimen handling
- Sizes 125 µm, 135 µm, 145 µm, 155 µm, 170 µm are suitable for denudation
The EZ-Tip is supplied sterile in one of two packaging options; individually blister packed or pouch packed in a vial containing 20 tips. All pipettes are intended single use and disposable.
The EZ-Grip is a hand held, reusable actuator or pipeter for plastic pipettes. It consists of a machined aluminium barrel containing stainless steel and PTFE internal working mechanisms, a titanium plunger wire and medical grade silicone and nylon seals. The plunger mechanism is designed to be compatible with 0.9 mm OD plastic pipettes and it has an aspiration volume range of 0.2 µl to 3 µl with a blow-out volume of 1.4 µl above the adjusted aspiration volume setting.
The EZ-Grip is supplied non-sterile with validated protocols for cleaning and sterilization included in the Instructions for Use.
The provided document describes the non-clinical testing and assessment of the EZ-Tip and EZ-Grip devices to demonstrate their substantial equivalence to predicate devices (The Stripper micropipet and micropipeter tips and The Stripper device, respectively). The study does not involve human subjects or AI, but rather focuses on physical, chemical, and biological performance characteristics relevant to assisted reproductive techniques (ART).
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
EZ-Tip | ||
Mouse Embryo Assay (MEA) | ≥80% hatched blastocysts at 120 h | Passed (three lots demonstrated ≥80% hatched blastocysts at 120 h) |
Endotoxin (LAL) | ≤20 EU/device | Passed (three lots demonstrated ≤20 EU/device) |
Sterilization | Compliance with ISO11137-1:2006 and ISO11137-2:2012 | Sterilization validated in compliance with these standards |
Packaging Integrity | Compliance with ISO11607-2:2006, ASTM F1886/F1886M - 09, ASTM F1929-12, ASTM F1140-07, and ASTMF1980:07(2011) | Validation studies conducted per these standards; accelerated aged samples conditioned per ASTMF1980:07(2011) |
Shelf-Life (EZ-Tip Vial of 20) | Supported by packaging and performance tests | Three years (supported by real-time and accelerated aging) |
Shelf-Life (EZ-Tip Singles) | Supported by packaging and performance tests | Five years (supported by real-time and accelerated aging) |
EZ-Grip | ||
Mouse Embryo Assay (MEA) (Plunger Wires) | ≥80% hatched blastocysts at 120 h | Passed (three lots of plunger wires demonstrated ≥80% hatched blastocysts at 120 h) |
Reprocessing Validation | Compliance with FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" | Reprocessing validated in accordance with this guidance |
Device Life (Simulated Use) | Performs according to product specifications after routine maintenance and reprocessing | Confirmed to perform after 16,000 actuations (measuring aspirated volume, delivered volume, force) |
2. Sample Size Used for the Test Set and Data Provenance
- EZ-Tip MEA & Endotoxin Testing: Three lots of EZ-Tip were tested.
- EZ-Grip MEA Testing (Plunger Wires): Three lots of EZ-Grip plunger wires were tested.
- EZ-Grip Device Life Bench Testing: Not explicitly stated as a number of devices, but it involved simulating one year of use (16,000 actuations) on at least one device.
- Provenance: All tests described are non-clinical, laboratory-based assessments of the physical device components and their interaction with biological samples (mouse embryos). The data provenance is from laboratory testing and not from human clinical data, nor is country of origin of the data specified beyond the company being in the UK.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable as the described studies are non-clinical and do not involve human diagnostic assessment or expert-determined ground truth in the typical clinical sense. The "ground truth" here is objective laboratory measurements (e.g., percentage of hatched blastocysts, endotoxin levels, physical performance metrics).
4. Adjudication Method for the Test Set
This information is not applicable as the described studies are non-clinical and do not involve human diagnostic assessment requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No. A MRMC comparative effectiveness study was not done. This study is focused on the non-clinical performance of medical devices for ART, not comparative effectiveness of human readers or AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
No. This is not an AI/algorithm-based device. It is a physical medical device (pipettes and a pipettor).
7. The Type of Ground Truth Used
- For Mouse Embryo Assays (MEA): The "ground truth" is the observed biological outcome of mouse embryo development (percentage of hatched blastocysts), which is a direct biological measure.
- For Endotoxin Testing: The "ground truth" is the measured endotoxin level, a chemical quantification.
- For Sterilization, Packaging, and Bench Testing: The "ground truth" is compliance with established international standards and internal product specifications for physical and functional performance.
8. The Sample Size for the Training Set
Not Applicable. There is no "training set" as this is not an AI/machine learning study.
9. How the Ground Truth for the Training Set was Established
Not Applicable. There is no "training set" as this is not an AI/machine learning study.
Ask a specific question about this device
(214 days)
MQH
InteGlobal MicroTips TPLG-130 are used for Denuding, handling blastomeres, oocytes and oocyte stripping. LifeGlobal MicroTips TPLG-180, TPLG-225 and TPLG-275 are used for handling embryos and blastocysts.
Not Found
This document does not contain information about acceptance criteria or a study proving that a device meets such criteria.
The provided text is an FDA 510(k) clearance letter for the "LifeGlobal Micro Tips" device. This letter confirms that the device has been found substantially equivalent to a legally marketed predicate device, allowing it to be marketed. It discusses regulatory compliance, labeling requirements, and contact information for further inquiries.
However, it does not include any details on:
- Specific performance acceptance criteria (e.g., accuracy, precision, sensitivity, specificity).
- Any studies (clinical or non-clinical) that were conducted to evaluate the device's performance against such criteria.
- The methodologies for any such studies (sample size, data provenance, ground truth establishment, expert qualifications, adjudication, MRMC studies, standalone performance).
- Any reported device performance metrics.
Therefore, I cannot provide the requested table or answer any of the questions regarding acceptance criteria and studies.
Ask a specific question about this device
(267 days)
MQH
The Embryo Biopsy Pipette is intended to aspirate a blastomere or trophectoderm to diagnose genetic disorders prior to embryo selection.
The Polar Body Biopsy Pipette is intended to aspirate polar bodies to diagnose genetic disorders prior to embryo selection.
The Testicular Sperm Extraction Pipette is intended to extract sperm cells from biopsied tessue.
The Embryo Biopsy, Polar Body Biopsy, and Testicular Sperm Extraction Pipettes are 970 microns (um) outer diameter borosilicate glass tubing that is bent and tapered. The Embryo Biopsy Pipette is tapered to a flat, smooth tip inner diameter of 30 um and 35 um. It is intended to be used for blastomere or trophectoderm aspiration. The Polar Body Biopsy Pipette is tapered to an inner diameter of 20 um and beveled to a spike to assist in piercing through the zona pellucida to biopsy the polar body. The Testicular Sperm Extraction Pipette is tapered to a flat angle with an inner diameter of 6.5 um. These pipettes are intended to extract sperm cells from biopsied testicular tissue. The device is supplied gamma sterilized and is intended for one-time use.
The document describes three medical devices: Embryo Biopsy Pipette, Polar Body Biopsy Pipette, and Testicular Sperm Extraction Pipette. The acceptance criteria and the studies performed to demonstrate equivalence are detailed as follows:
1. Table of Acceptance Criteria and Reported Device Performance
Test Name | Acceptance Criterion | Reported Device Performance |
---|---|---|
MEA (Mouse Embryo Assay) | ≥80% 1-cell stage embryos must develop to blastocyst within 96 hours. | The predetermined acceptance criterion was met. |
LAL (Limulus Amebocyte Lysate) |
Ask a specific question about this device
(270 days)
MQH
The Flexipet® Denuding Pipettes are intended to be used for blastomere or polar body manipulation, occyte and embryo manipulation, or denuding.
The Flexipet® Manipulation Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, blastocyst handling, and manipulation of the oocyte-cumulus complex.
The Flexipet® Denuding Pipettes are polycarbonate tubes with inner diameters measuring 600 um (microns) at the proximal end which taper in diameter in a range from 120 µm to 170 µm on the distal end. The micropipettes are all 3.5 inches in length and constructed from polycarbonate. The devices are supplied gamma sterilized and are intended for one-time use.
The Flexipet® Manipulation Pipettes are polycarbonate tubes with inner diameters measuring 600 um (microns) at the proximal end which taper in diameter in a range from 80 um to 600 um on the distal end. The micropipettes are all 3.5 inches in length and constructed from polycarbonate. The devices are supplied gamma sterilized and are intended for one-time use.
1. Table of Acceptance Criteria and Reported Device Performance:
Test Name | Acceptance Criteria | Reported Device Performance |
---|---|---|
MEA (Mouse Embryo Assay) | ≥80% 1-cell embryos developed to blastocyst within 96 hours. | The predetermined acceptance criterion was met. |
LAL (Limulus Amebocyte Lysate) |
Ask a specific question about this device
(205 days)
MQH
Biopsy pipettes are used in assisted reproduction techniques (ART) for the removal of blastomere(s) from embryos or polar bodies from oocytes, which may be done in order to perform preimplantation genetic diagnosis (PGD) on the genetic materials in the biopsied cell(s).
The Biopsy pipette is part of the RI Pipettes range. Other pipettes available in this range include ICSI Injection, ICSI Holding, and Assisted Hatching pipettes.
Biopsy pipettes are offered in a range
- ID sizes - 14μm, 20μm, 25μm, 30μm, 40μm .
- bend angles 0°, 15°, 20°, 25°, 30°, 35°, 40°, 45° .
- . tip profiles - flat , non-spiked (bevelled), spiked
All pipettes in this range are manufactured from very fine borosilicate glass, packed individually into a Twista-Pak (primary packaging) and sealed in a medical pouch. They are then batch sterilised by gamma irradiation.
Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:
Device: RI Pipettes - Biopsy
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Test) | Reported Device Performance |
---|---|
Non-pyrogenic (LAL) |
Ask a specific question about this device
(232 days)
MQH
The manufactured microtools are to be used for Assisted Reproductive Technology (ART) and Reproductive Medicine (RM), and the specific pipette indication is as follows:
Injection Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories for intracytoplasmic sperm injection (ICSI)
Holding Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories for holding oocyte or embryo during ICSI or biopsy
Biopsy Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories for removal of blastomere(s) from embryos, which may be done in order to perform preimplantation genetic diagnosis on the genetic material in the biopsied cell(s)
Polar Body Biopsy Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories for removal of polar bodies from occytes, which may be done in order to perform pre implantation genetic diagnosis on the genetic material in the biopsied cell(s)
Assisted Hatching Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories to create a defect in the zona pellucida chemically using Tyrode's acid solution in order to perform assisted hatching of embryos
Denuding Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories to denuding cumulus cells from oocytes by delivering enzymes to aid separation
Partial Zona Pipette: tool used in IVF/Assisted Reproduction Technology (ART) laboratories to cut zona pellucida to create a defect in the zona pellucida mechanically in order to perform assisted hatching of embryos mechanically
The following Jieying Microtools are made of glass capillary tubing. They range in sizes and have various degrees of beveled angles as necessary for each individual application. They are gamma radiated and tested for endotoxin and mouse embryo. These devices are assisted reproduction microtools/pipettes that are used in the laboratory to denude, micromanipulate, or hold human gametes or embryos for assisted hatching, intracytoplasmic sperm injection (ICSI), or other assisted reproduction methods.
- a. Injection pipette;
- b. Holding pipette;
- c. Biopsy pipette;
- d. Polar body biopsy pipette;
- e. Denuding pipette;
- f. Partial zona dissection (PZD) pipette;
- g. Assisted hatching (AH) pipette.
The provided document describes Jieying Laboratory Inc.'s micro-manipulation pipettes and their equivalence to predicate devices, focusing on the safety and performance characteristics for regulatory approval.
Acceptance Criteria and Device Performance:
Acceptance Criteria | Reported Device Performance |
---|---|
LAL (Limulus Amboycte Lysate) Testing: | Each batch performs LAL testing. Level of endotoxin units must be |
Ask a specific question about this device
(253 days)
MQH
Blastomere Biopsy Pipettes: The intended use of the Blastomere Biopsy Pipettes is for removal of blastomere(s) from embryos, which may be done in order to perform preimplantation genetic diagnosis on the genetic material in the biopsied cell(s).
Polar Body Biopsy Pipettes: The intended use of the Polar Body Biopsy Pipettes is for removal of polar bodies from oocytes, which may be done in order to perform preimplantation genetic diagnosis on the genetic material in the biopsied cell(s).
The Blastomere Biopsy Pipettes are very fine glass pipettes used in IVF/Assisted Reproduction Technology (ART) laboratories for the aspiration of blastomere(s) from embryos for the purpose of preimplantation genetic diagnosis. These pipettes have an inner diameters varying from 18-42 um based customer's preferences and the stage of the embryos being biopsied. The tip may be flat or beveled at 45 degrees, then fire polished, and bended to 10 - 45 degrees or straight.
The Polar Body Biopsy Pipettes are very fine glass pipettes tools used in IVF/Assisted Reproduction Technology (ART) laboratories for the aspiration of polar bodies from oocytes and embryos for the purpose of preimplantation genetic diagnosis. These pipettes have an inner diameters varying from 13-15 um. The tip may be flat or beveled at 45 degrees, then fire polished, and bended to 10-45 degrees or straight. Some of the tip may be pulled to form a sharp spike after polishing based customer's preferences.
These devices are intended for one-time use and will be marked sterile.
These devices are manufactured entirely from borosilicate glass. They are manufactured to specific sizes or the size may be modified to meet customer specifications, following procedures of the Sunlight Quality System. The final products are batch tested as part of a quality assurance program using Mouse Embryo Toxicity testing and endotoxin testing. The acceptance specifications of these tests are ≥80% of 2-cell mouse embryos to blastocysts and endotoxin ≤ 0.5 EU/device.
The Sunlight Medical, Inc. Biopsy Pipettes (Blastomere Biopsy Pipettes and Polar Body Biopsy Pipettes) underwent a 510(k) submission (K092554) to demonstrate substantial equivalence to predicate devices rather than a direct clinical performance study with acceptance criteria in the traditional sense. The equivalence was established based on similarities in indications for use, materials, physical construction, manufacturing processes, and quality controls.
Here's an analysis based on the provided document:
1. A table of acceptance criteria and the reported device performance
The submission does not specify "acceptance criteria" for clinical performance as would be seen in a diagnostic accuracy study. Instead, the "acceptance specifications" refer to quality control tests performed on the manufactured devices. The primary "performance" reported is related to the equivalence to the predicate device and the results of these quality control tests.
Acceptance Criteria (Quality Control) | Reported Device Performance |
---|---|
Mouse Embryo Toxicity Test: ≥80% of 2-cell mouse embryos to blastocysts | Meets the acceptance specification (implied, as it's part of Quality Assurance Program) |
Endotoxin Test: ≤ 0.5 EU/device | Meets the acceptance specification (implied, as it's part of Quality Assurance Program) |
Material: Borosilicate glass | Borosilicate glass (same as predicate) |
Tip features (Beveling, Angulations, Fire Polished, Bended) | Similar to predicate, with variations in inner diameter for Blastomere and potential spike for Polar Body |
Inner Diameter (Blastomere Biopsy Pipettes) | 18-42 um (Predicate: 28-42 um) - noted as a bigger variation |
Inner Diameter (Polar Body Biopsy Pipettes) | 13-15 um (Same as predicate) |
Sterility | One-time use, marked sterile |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
There is no test set of clinical images or cases described for evaluating device performance in a clinical setting. The "test set" for the quality control criteria would be a sample of manufactured pipettes, but the number of devices tested is not specified. The provenance of the data for these quality control tests (Mouse Embryo Toxicity and Endotoxin) is from the manufacturer's internal "Quality Assurance Program".
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as there was no test set requiring expert ground truth for clinical performance evaluation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable as there was no test set requiring adjudication for clinical performance evaluation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a surgical tool (pipette) and not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance evaluation was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical tool, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for the quality control tests is based on established biological and chemical assays:
- Mouse Embryo Toxicity: The ability of 2-cell mouse embryos to develop to the blastocyst stage in the presence of the device's leachables. This is an accepted biological indicator for cytotoxicity in reproductive medical devices.
- Endotoxin Test: Quantification of bacterial endotoxins, which is a standard safety measure for medical devices that may come into contact with biological systems.
8. The sample size for the training set
This is not applicable. There is no "training set" in the context of device performance in a clinical scenario or AI training, as this is a physical medical device. The "training" for the device would pertain to the manufacturing process and quality system setup, for which sample sizes are not specified in the document.
9. How the ground truth for the training set was established
This is not applicable as there is no training set of clinical data. The "ground truth" for manufacturing consistency and quality is established through standard operating procedures, material specifications, and quality control methodologies (like the mouse embryo and endotoxin tests).
Ask a specific question about this device
(248 days)
MQH
Ask a specific question about this device
Page 1 of 3