Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K032330
    Device Name
    CT OIA
    Date Cleared
    2004-03-02

    (217 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Thermo Electron CT OIA assay is an Optical ImmunoAssay (OIA) test for the rapid, qualitative detection of chlamydial antigen from female endocervical swab specimens. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Chlamydia trachomatis antigen. The assay is intended for in vitro diagnostic use with symptomatic females in populations at risk for sexually transmitted diseases.

    CT OIA test results are presumptive evidence for either the presence or absence of C. trachomatis. Definitive laboratory evidence for the presence/ absence of C. trachomatis would need additional testing. CT OIA test results should not preclude empiric treatment of women with overt symptoms of PID. Performance for use in asymptomatic male and female populations has not been established.

    Device Description

    The CT OIA test involves the qualitative extraction of antigen specific to the Chlamydia genus. The Optical ImmunoAssay technology enables the direct visual detection of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When an extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antigen. After washing, the substrate is added, increasing the thickness (mass enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the surface retains the original gold color indicating a negative result.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Thermo Flectron CT OIA device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for clinical performance (e.g., "Sensitivity must be >= X%, Specificity must be >= Y%"). Instead, it reports the observed performance characteristics. Given this, the table will present the key performance metrics as reported in the study:

    MetricReported Device Performance (%)
    Clinical Sensitivity73.8
    Clinical Specificity98.4
    Overall PPV87.3
    Overall NPV96.2
    Reproducibility87.7

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Sensitivity and Specificity: 767 symptomatic female patients were included in the final analysis. (Initially, 885 female patients were enrolled, but 118 were excluded).
    • Sample Size for Reproducibility: Reproducibility testing was performed on three blinded panels, though the exact number of individual samples within these panels isn't specified (71/81 successful tests are mentioned, suggesting 81 blinded samples).
    • Data Provenance: The study was a "multicenter study at four geographically diverse clinical sites" located in the Northwest, Midwest, Mid-Atlantic, and Southeast regions of the United States. It was a prospective study as patients were enrolled and tested.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not specify the number or qualifications of experts involved in establishing the ground truth.

    4. Adjudication Method for the Test Set

    The document does not explicitly mention an adjudication method. For the clinical sensitivity and specificity study, the CT OIA test was compared to a "commercially available LCx nucleic acid amplification test for C. trachomatis" as the primary reference method. "Secondary confirmation testing of positive OIA was done by a commercially available PCR test." This suggests a two-step approach for ground truth, but not an adjudication process among human readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The study focuses on the standalone performance of the device compared to a reference method, not on how human readers' performance improves with or without AI assistance.

    6. Standalone (Algorithm only without human-in-the-loop performance) Study

    Yes, a standalone study was conducted. The performance metrics (sensitivity, specificity, PPV, NPV) reported are for the CT OIA assay itself, in comparison to a reference method (LCx and PCR), without human interpretation as part of the primary device output. The device itself is an "Optical ImmunoAssay technology [that] enables the direct visual detection of a physical change in the optical thickness... visually perceived as a color change. A positive result appears as a purple spot on the predominant gold background." While a human observes this color change, the reported performance is for the diagnostic test's ability to accurately reflect the presence of the antigen, not for a human's ability to interpret an image provided by an AI.

    7. Type of Ground Truth Used

    The primary ground truth for the clinical sensitivity and specificity study was a commercially available LCx Nucleic Acid Amplification Test (NAAT) for C. trachomatis. Positive OIA results were also subject to secondary confirmation testing by a commercially available PCR test.

    8. Sample Size for the Training Set

    The document does not mention a separate "training set" or "training data" in the context of developing the CT OIA assay. This assay is a diagnostic test based on Optical ImmunoAssay technology, likely developed and validated through laboratory experimentation and calibration rather than machine learning on a large dataset. Therefore, the concept of a training set for an algorithm is not applicable here.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a training set as understood in AI/ML is not applicable. The device's development would involve establishing its analytical performance characteristics (e.g., detection limits, cross-reactivity) through laboratory studies using characterized samples, rather than establishing "ground truth" for a training set in a clinical context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033865
    Manufacturer
    Date Cleared
    2004-01-21

    (40 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DakoCytomation Ltd IDEIA™ PCE Chlamydia is an immunoassay utilising dual amplification technology for the qualitative detection of Chlamydia lipopolysaccharide antigen in human urethral and endocervical swabs and male urines.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the IDEIA™ PCE Chlamydia immunoassay. It does not contain the detailed study information regarding acceptance criteria or device performance that you are requesting.

    The letter states that the device is "substantially equivalent" to legally marketed predicate devices, which means it has been determined to be as safe and effective as a previously cleared device. However, it does not provide the specific performance data, acceptance criteria, or study details that would typically be found in a clinical study report or a more comprehensive submission document.

    Therefore, I cannot provide the information requested in your prompt based on the provided text. To answer your questions, I would need a different type of document, such as a detailed clinical study report, a premarket notification summary, or the full 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033079
    Date Cleared
    2003-11-26

    (58 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinity Biotech Captia™ Chlamydia IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative detection of IgG antibodies in human serum to Chlamydia for the determination of immunological experience.

    Device Description

    The Chlamydia IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative detection of IgG antibodies inhuman serum to Chlamydia for the determination of immunological experience.

    The Chlamydia IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Chlamydia. Purified Chlamydia antigen (strain LGV II) is attached to a solid phase microtiter well. Diluted test sera are added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is preset it will bind to the antibody attached to the antigen on the well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period, the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study proving the device meets them:

    Acceptance Criteria and Device Performance for Trinity Biotech Chlamydia IgG ELISA Test Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the reported performance relative to a predicate device and established precision standards.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Comparative Agreement- % Agreement Positive: High agreement with predicate IFA kit.
    • % Agreement Negative: High agreement with predicate IFA kit.
    • % Agreement: High overall agreement with predicate IFA kit. | % Agreement Positive: 92.1% (95% CI: 86.3% - 97.8%)
      % Agreement Negative: 98.0% (95% CI: 96.2% - 99.8%)
      % Agreement: 96.5% (95% CI: 94.5% - 98.5%) |
      | Precision | Coefficient of Variation (CV) 0.989 between sites.
      97.1% (133/137) agreement between three sites (excluding equivocals). |

    2. Sample Size Used for the Test Set and Data Provenance

    • Comparative Agreement Study (Table 1):
      • Test Set Sample Size: 355 sera (86 positive by IFA, 16 equivocal by IFA, 253 negative by IFA).
      • Data Provenance: Retrospective. Sera were from "normal individuals of various ages, gender, and geographical areas." The studies were conducted at R&D laboratories of commercial companies located in Maryland and New York, affiliated with the manufacturer.
    • Precision Studies (Tables 2, 3, 4):
      • Test Set Sample Size: 7 sera, each assayed 10 times in 3 different assays at 2 different sites (total 60 tests per sera for inter-site precision, 30 tests per sera for inter-assay precision within each site).
      • Data Provenance: Retrospective (sera likely selected for different activity levels), performed at 2 sites affiliated with the manufacturer.
    • Paired Serum Analysis:
      • Test Set Sample Size: 9 serum pairs.
      • Data Provenance: Retrospective, serum pairs showing changes in Complement Fixation (CF) titer.
    • Reproducibility Study:
      • Test Set Sample Size: 50 different sera.
      • Data Provenance: Retrospective, conducted at three different sites: two R&D laboratories at commercial companies (Maryland and New York, affiliated with the manufacturer) and one large clinical laboratory (Pennsylvania).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of "experts" in the context of establishing ground truth in the traditional sense (e.g., medical professionals reviewing cases).

    • For the Comparative Agreement Study: The "ground truth" for comparison was the results of a commercial IFA kit. The document states, "Please be advised that "% agreement positive" and "% agreement negative" refer to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease." This clarifies that the IFA kit was the reference standard, not a true clinical ground truth established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by a single predicate device (commercial IFA kit) or by the device itself for precision and reproducibility. There was no human adjudication process described. Equivocal results from the predicate IFA were excluded from agreement calculations, but this is not an adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This device is an in-vitro diagnostic ELISA kit, not an AI-assisted diagnostic tool that requires human readers for interpretation in the manner typically associated with MRMC studies in medical imaging. The comparison is between two laboratory assays.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This is an inherently standalone device as it is an ELISA kit. Its performance is measured directly through its reaction with serum samples, producing a measurable color intensity that is read photometrically. The results are quantitative and then interpreted qualitatively (positive/negative/equivocal) based on pre-defined cut-offs. There is no "human-in-the-loop" once the assay is performed and read by the instrument, beyond potentially interpreting the final qualitative result.

    7. The Type of Ground Truth Used

    • For Comparative Agreement, Paired Serum Analysis: The ground truth was the results of a predicate commercial IFA kit (for agreement) or a Complement Fixation (CF) titer (for seroconversion) for Chlamydia. It is explicitly stated that there was no attempt to correlate the assay's results with disease presence or absence, and therefore, no true clinical or pathological ground truth was established in these comparative studies.
    • For Precision and Reproducibility: "Expected results" for reproducibility were derived from "previous Trinity Biotech ELISA testing of the samples." For precision, the study aimed to evaluate the consistency of the device's own measurements.

    8. The Sample Size for the Training Set

    Not applicable. This is a traditional in-vitro diagnostic kit, not a machine learning model. There is no concept of a "training set" in the context of developing this device. The development process would involve iterative optimization of assay components and protocols, rather than training an algorithm on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K982210
    Manufacturer
    Date Cleared
    1998-07-09

    (16 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Syva MicroTrak II Chlamydia EIA is a n enzyme immunoassay intended for use in the qualitative detection of Chlamydia in female endocervical, male urethral, male urine and ocular specimens. The assay may be used to detect the presence of chlamydia where infection is suspected or likely to exist

    Device Description

    The Syva MicroTrak II Chlamydia EIA has been modified in that the preservative system in the Enzyme/Antibody reagent has been changed to meet USP Challenge test requirements and to be effective against pseudomonads and staphylococci. The modified preservative system meets those design requirements without affecting the performance characteristics, claims, and labeling of the assay as compared to the product with the unmodified preservative system.

    AI/ML Overview

    The provided text describes a 510(k) submission for a device modification, specifically a change in the preservative system of the Syva MicroTrak II Chlamydia EIA. The core statement in the summary is: "The modified preservative system meets those design requirements without affecting the performance characteristics, claims, and labeling of the assay as compared to the product with the unmodified preservative system."

    This indicates that the acceptance criteria for this modification are that the new device's performance is non-inferior or substantially equivalent to the previous version of the device. The study would have demonstrated that the change in preservative did not adversely affect the analytical performance (sensitivity, specificity, accuracy) of the EIA.

    However, the provided text does not contain the detailed study information required to fill out the requested table and answer all questions comprehensively. It only states that the performance characteristics were not affected.

    Based on the available information, here's what can be inferred and what is missing:


    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Modified Device vs. Unmodified Device)
    Primary Goal: Maintain equivalent performance characteristics (sensitivity, specificity, accuracy) to the unmodified device."meets those design requirements without affecting the performance characteristics, claims, and labeling of the assay as compared to the product with the unmodified preservative system."
    Secondary Goal: Meet USP Challenge test requirements for the preservative system.The modified preservative system "has been changed to meet USP Challenge test requirements and to be effective against pseudomonads and staphylococci."

    Study Details (Based on Inference and Missing Information)

    Since this is a 510(k) for a device modification, the study would primarily be a comparative study demonstrating that the modified device performs equivalently to the predicate (unmodified) device.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: Not specified in the provided text. A typical comparative study for an EIA would involve a sufficient number of positive and negative clinical samples to demonstrate equivalent analytical performance.
      • Data Provenance: Not specified in the provided text (e.g., country of origin of the data, retrospective or prospective). Historically, such studies often involve multi-center prospective or retrospective clinical sample testing.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

      • Not specified. For Chlamydia EIA, ground truth is typically established by:
        • Confirmatory laboratory methods (e.g., PCR, culture, or another validated reference method) on the same samples.
        • Clinical diagnosis, sometimes supplemented by expert adjudication if discordant results occur.
      • The term "experts" in the context of establishing ground truth for an in vitro diagnostic (IVD) device like an EIA usually refers to the reference method itself and the laboratory personnel performing and interpreting the reference tests, rather than clinical experts adjudicating images.
    3. Adjudication Method for the Test Set:

      • Not specified. If there were discrepancies between the EIA results and the ground truth method(s), an adjudication process would typically involve retesting or further confirmatory testing to resolve conflicts. The specific method (e.g., 2+1) is not mentioned.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is an Enzyme Immunoassay (EIA) for the qualitative detection of Chlamydia. It is an in vitro diagnostic (IVD) device, not an imaging device or an AI-based diagnostic tool that would involve human readers interpreting AI outputs. Therefore, an MRMC study is not applicable.
    5. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, indirectly. An EIA device, by its nature, is a standalone analytical test. Its performance (sensitivity, specificity, accuracy) is determined by the reaction occurring in the MTP-wells, the spectrophotometric reading, and the interpretation criteria. There is no "human-in-the-loop" once the assay is run and read, other than the initial sample preparation and final interpretation of the numerical results against established cutoffs. The "algorithm" here is the assay's chemistry and readout interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not explicitly stated, but for Chlamydia EIA, the ground truth would most likely be established by confirmatory laboratory methods such as:
        • Nucleic Acid Amplification Tests (NAATs) like PCR.
        • Cell culture (historically a gold standard for Chlamydia).
        • Potentially, a combination of clinical diagnosis and a reference laboratory method for discordant results.
    7. The Sample Size for the Training Set:

      • Not applicable as this is an EIA, not a machine learning or AI-based device that requires a training set in the typical sense. The "training" for an EIA occurs during its initial design, optimization, and validation by the manufacturer using internal characterization studies, not through external data sets in an AI context.
    8. How the Ground Truth for the Training Set was Established:

      • Not applicable for the reasons mentioned above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962558
    Manufacturer
    Date Cleared
    1997-02-24

    (238 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chlamydia IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative detection of IgG antibodies in human serum to Chlamydia for the determination of immunological experience.

    Device Description

    The Chlamydia IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Chlamydia. Purified Chlamydia antigen (strain LGV II) is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided text, formatted as requested:

    Acceptance Criteria and Device Performance for Chlamydia IgG ELISA Test Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Chlamydia IgG ELISA)
    Relative Sensitivity92.1% (95% CI: 86.3% - 97.8%)
    Relative Specificity98.0% (95% CI: 96.2% - 99.8%)
    Relative Agreement96.5% (95% CI: 94.5% - 98.5%)
    Precision (CV)Intra-assay CV: typically
    Ask a Question

    Ask a specific question about this device

    K Number
    K960850
    Date Cleared
    1997-02-04

    (340 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    qualitative, paramagneticparticle, chemiluminescent enzyme immunoassays for the direction of chlamydia antigen in adult male urethral, female endocervical, and male urine specimens, using the ACCESS® Immunoassay System.

    Device Description

    The Sanofi Diagnostics Pasteur Inc. ACCESS® Chlamvdia antigen are qualitative, paramagneticparticle, chemiluminescent enzyme immunoassays for the direction of chlamydia antigen in adult male urethral, female endocervical, and male urine specimens, using the ACCESS® Immunoassay System.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and study details for the ACCESS® Chlamydia Reagents:

    Acceptance Criteria and Device Performance Study

    The ACCESS® Chlamydia Reagents are qualitative, paramagnetic-particle, chemiluminescent enzyme immunoassays intended for the detection of Chlamydia antigen. The study compared the device's performance against culture or DFA (Direct Fluorescent Antibody) as a reference standard.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" as pass/fail thresholds. Instead, it presents the reported device performance, implying these are the results achieved and deemed acceptable for substantial equivalence.

    Population CategoryPerformance MetricReported Device Performance (ACCESS® Chlamydia)
    High Risk FemalesSensitivity86%
    Specificity99.7%
    PPV96%
    NPV99%
    Low Risk FemalesSensitivity100%
    Specificity100%
    PPV100%
    NPV100%
    Symptomatic MalesSensitivity93%
    Specificity99.7%
    PPV98%
    NPV99%
    Symptomatic Males (Urine)Sensitivity88%
    Specificity99%
    PPV94%
    NPV98%
    Asymptomatic MalesSensitivity88%
    Specificity100%
    PPV100%
    NPV99.5%
    Asymptomatic Males (Urine)Sensitivity56%
    Specificity100%
    PPV100%
    NPV98%
    Comparison to Predicate DeviceConcordance (urogenital)98.1%
    Concordance (male urine)95.4%
    Precision (Within Run)Negative4.2%
    Low Positive5.6%
    High Positive3.6%
    Precision (Total)Negative11.9%
    Low Positive17.5%
    High Positive21.1%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Against Culture or DFA: 2092 urogenital specimens and 572 male urine specimens.
      • Against Predicate Device (Syva MicroTrak® II Chlamydia EIA): 1518 urogenital specimens and 303 male urine specimens.
    • Data Provenance: The document does not explicitly state the country of origin. It can be inferred that the data is retrospective clinical study data, as it describes a comparison in past clinical studies. The study was conducted by Sanofi Diagnostics Pasteur, Inc., based in Chaska, Minnesota, USA, suggesting the data is likely from the United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide details on the number of experts or their qualifications for establishing the ground truth. It simply refers to "culture or DFA" as the reference method. In clinical microbiology, these methods are typically performed by trained laboratory professionals.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method for the test set results. The ground truth was established by "culture or DFA," which are objective laboratory tests that generally do not require adjudication in the same way as subjective expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is for a diagnostic assay, comparing its performance to a reference standard (culture/DFA) and a predicate device. It does not involve human readers interpreting images or data with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    Yes, the study presents standalone performance. The ACCESS® Chlamydia Reagents are an in vitro diagnostic device that provides a direct result for Chlamydia antigen. The reported performance metrics (sensitivity, specificity, etc.) are for the device alone, without human intervention in interpreting the assay's final output. Human involvement is limited to performing the assay and reading the instrument's result.

    7. Type of Ground Truth Used

    The primary ground truth used was expert reference methods: cell culture or Direct Fluorescent Antibody (DFA). These are established laboratory techniques for the detection of Chlamydia and are considered gold standards or highly reliable reference methods.

    8. Sample Size for the Training Set

    The document does not provide any information about a specific training set or its sample size. This type of submission (510(k) summary from 1996) is for an immunoassay, which does not typically involve machine learning model training in the same way that AI/ML-based devices developed more recently do. The "development" of such assays involves reagent formulation, assay optimization, and internal validation, but not usually a distinct "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set in the context of an AI/ML model, this question is not applicable based on the provided information. The assay's performance was evaluated against the reference methods (culture/DFA) on the test set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K955627
    Date Cleared
    1996-10-07

    (301 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    used as a supplemental test to verify positive and equivocal results from female endocervical and male urtheral specimens in the VIDAS Chlamydia (CHL) assay and is substantially equivalent to cell culture for the qualitative detection of Chlamydia antigen.

    Device Description

    The VIDAS Chlamydia Blocking assay is a fully automated enzyme-linked fluorescent immunoassay (ELFA) and requires only the addition of the blocking and reference reagents to the VIDAS Chlamydia strips.

    AI/ML Overview

    The provided text is a summary of substantial equivalence for a medical device cleared in 1996. It does not contain the detailed information necessary to fully answer all aspects of your request, particularly regarding acceptance criteria, specific study design elements, and statistical analysis metrics. This is typical for older 510(k) summaries, which often focused on substantial equivalence rather than explicit performance targets and detailed study breakdowns as seen in more recent submissions.

    However, I can extract the available information and highlight what is missing based on your template.

    Acceptance Criteria and Study Information (Based on Available Data)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (Implied from substantial equivalence)
    Qualitative detection of Chlamydia antigen."substantially equivalent to cell culture for the qualitative detection of Chlamydia antigen."
    Performance as a supplemental test to verify positive and equivocal results from female endocervical and male urethral specimens in the VIDAS Chlamydia (CHL) assay.Implied to meet this function by demonstrating substantial equivalence to cell culture.

    Missing: Specific quantitative acceptance criteria (e.g., sensitivity, specificity thresholds, PPV/NPV targets) are not provided in this summary. The performance is broadly stated as "substantially equivalent" to cell culture.

    2. Sample Size Used for the Test Set and Data Provenance

    The summary does not provide the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the study). It refers to "the data comparing the performance... to that of cell culture" in Section 8 of the submittal, which is not provided here.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The summary does not provide this information.

    4. Adjudication Method for the Test Set

    The summary does not provide this information.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, based on the description, this was a standalone device performance study comparing the VIDAS Chlamydia Blocking assay to cell culture as the reference method. MRMC studies are typically for image-based diagnostic aids and involve human interpretation, which is not the case for this automated immunoassay.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, this was a standalone performance study of the VIDAS Chlamydia Blocking assay. It is an automated enzyme-linked fluorescent immunoassay (ELFA) and does not involve human interpretation of the assay results in a "human-in-the-loop" manner.

    7. The Type of Ground Truth Used

    The ground truth used for comparison was cell culture. The text explicitly states: "The data comparing the performance of the VIDAS Chlamydia Blocking assay to that of cell culture..." and that "The VIDAS Chlamydia Blocking assay...is substantially equivalent to cell culture for the qualitative detection of Chlamydia antigen."

    8. The Sample Size for the Training Set

    The summary does not provide information on a training set size. For an immunoassay like this, the "training" might refer to assay development and optimization, rather than a distinct training set in the machine learning sense. The available text describes the performance study rather than the development process.

    9. How the Ground Truth for the Training Set Was Established

    The summary does not provide this information, as it doesn't describe a traditional "training set" or its ground truth establishment.


    In summary: The provided text is a high-level summary from a 1996 510(k) submission focused on demonstrating substantial equivalence. It lacks the quantitative performance metrics, detailed study design parameters (like sample sizes, expert qualifications, adjudication methods), and specific acceptance criteria that are commonly found in more recent regulatory documents or required by your detailed template. The core comparison was against cell culture as the gold standard.

    Ask a Question

    Ask a specific question about this device

    K Number
    K936054
    Date Cleared
    1996-05-30

    (891 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IMx SELECT Chlamydia is a Micropaticle Enzyme Immunoassay (MEIA) for the qualitative detection of chlamydial LPS antigen and is indicated for use in testing female endocervical swab specimens, and male urethral swab specimens from symptomatic individuals to identify Chlamvia trachomatis. The IMx SELECT Chamydia Blocking Reagent may be used to verify the chlamydial specificity of the antigen detected.

    Device Description

    Micropaticle Enzyme Immunoassay (MEIA) for the qualitative detection of chlamydial LPS antigen.

    AI/ML Overview

    The provided text describes the IMx SELECT Chlamydia diagnostic device and its performance compared to cell culture for detecting Chlamydia trachomatis.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve a sensitivity of at least X%"). Instead, it presents the device's performance in terms of relative sensitivity and specificity against C. trachomatis cell culture as the reference method, with a focus on demonstrating "substantial equivalence."

    However, we can infer the reported performance metrics from the study results. The primary comparison is between the IMx SELECT Chlamydia assay and isolation of C. trachomatis in tissue culture.

    Acceptance Criteria (Implied / Comparator Performance)Reported Device Performance (IMx SELECT Chlamydia)
    Relative Sensitivity (vs. Cell Culture with Discordant Resolution by DFA)84.1% (230/261) (95% CI: 63.6-91.4) for Symptomatic Male
    89.9% (148/164) (95% CI: 83.4-94.8) for Female Total
    86.1% (230/261) (95% CI: 83.6-91.9) for Grand Total
    Relative Specificity (vs. Cell Culture with Discordant Resolution by DFA)98.5% (982/997) (95% CI: 97.7-99.3) for Female Total
    99.0% (356/360) (95% CI: 97.4-99.7) for Symptomatic Male
    98.7% (1366/1383) (95% CI: 98.0-99.2) for Grand Total
    Positive Predictive Value(Not clearly stated as an acceptance criterion, but reported)
    Negative Predictive Value(Not clearly stated as an acceptance criterion, but reported)

    Important Note: The confidence intervals for sensitivity, particularly for "Symptomatic Male" (63.6-91.4%) and "FEMALE: Other" (83.6-98.8%), are quite wide, suggesting variability or smaller subgroups. The "Grand Total" provides a more robust overall statistical picture.

    2. Sample size used for the test set and the data provenance

    • Sample Size: A total of 1647 specimens were used for the test set.
    • Data Provenance: The specimens were obtained from patients attending 6 different clinics, including sexually transmitted disease clinics, family planning clinics, and OB/GYN clinics. The study appears to be prospective as it involves collecting specimens from patients at the clinics for testing. The country of origin is not explicitly stated, but given "Abbott Park, Ill." as the submitter's address, it is likely the United States.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth.

    4. Adjudication method for the test set

    The ground truth was established by comparing the IMx SELECT Chlamydia assay to isolation of C. trachomatis in tissue culture. For discordant results (where the IMx SELECT Chlamydia result differed from the initial cell culture result), a DFA (Direct Fluorescent Antibody) test was used for resolution. This indicates a form of 2+1 adjudication where the initial two methods (IMx SELECT Chlamydia and Cell Culture) are compared, and a third method (DFA) is used to resolve discrepancies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This study evaluates a diagnostic assay, not an AI-assisted human reader interpretation. Therefore, there is no information on how human readers improve with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance evaluation was done. The IMx SELECT Chlamydia device itself is an automated immunoassay. Its performance metrics (relative sensitivity, relative specificity) are reported for the device operating independently without human interpretation influencing the test result beyond standard laboratory procedures (e.g., sample handling, loading).

    7. The type of ground truth used

    The primary ground truth used was isolation of C. trachomatis in tissue culture, with Discordant Resolution by DFA for any discrepancies between the test device and the culture. This is a common and accepted reference standard for Chlamydia trachomatis detection.

    8. The sample size for the training set

    The document does not provide information on a specific training set or its sample size. This is a diagnostic assay (MEIA), which typically relies on established biochemical reactions rather than machine learning models that require explicit training sets. The development and validation of such assays involve different processes than those for AI algorithms.

    9. How the ground truth for the training set was established

    As there is no explicit training set mentioned for an AI/machine learning model, this question is not applicable in the context of this traditional diagnostic assay. The "ground truth" establishment for the assay development would have involved optimizing reagents and protocols to accurately detect C. trachomatis antigen through rigorous laboratory testing using well-characterized samples.

    Ask a Question

    Ask a specific question about this device

    K Number
    K940564
    Date Cleared
    1996-01-11

    (707 days)

    Product Code
    Regulation Number
    866.3120
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LJC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1