Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K964953
    Manufacturer
    Date Cleared
    1997-02-13

    (70 days)

    Product Code
    Regulation Number
    888.4540
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    HXJ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Like the UroTac Bone Anchor System, the In-Tac Bone Anchor System is intended for soft tissue fixation to the pubic bone by means of bone anchors threaded with suture. The In-Tac Bone Anchor System is indicated for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

    Device Description

    Like the UroTac Bone Anchor System, the In-Tac Bone Anchor System consists of two components: a bone anchor, and a bone anchor inserter The In-Tac Bone Anchor System has the same overall design and technological characteristics as the configuration cleared under K962372, but implements the following changes: (1) the sutures will be provided pre-threaded for user convenience, and the retractable anchor shield will be incorporated into the disposable anchor assembly; (2) the inserter will be made from a stronger grade of stainless steel; and (3) the inserter incorporates a force measuring spring to ensure the anchors are deployed only if sufficient insertion force is applied.

    AI/ML Overview

    This document describes a 510(k) notification for the In-Tac Bone Anchor System, which is a modification of the previously cleared UroTac Bone Anchor System. This submission focuses on establishing substantial equivalence to the predicate device, K962372.

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance

      Acceptance CriteriaReported Device Performance
      Biocompatibility (materials commonly used in medical devices)Bench testing demonstrated materials are biocompatible.
      Device safety and effectiveness (modifications have no effect on safety and effectiveness)Bench testing demonstrated no effect on device safety and effectiveness.
      Substantial equivalence to predicate device (UroTac Bone Anchor System cleared under K962372)Based on materials, design, technological characteristics, and performance, the In-Tac Bone Anchor System is substantially equivalent.
    2. Sample size used for the test set and the data provenance
      The document does not specify a separate "test set" in the context of clinical or diagnostic performance evaluation. The evaluation appears to be based on bench testing for biocompatibility, safety, and effectiveness. Therefore, there's no information on sample size or data provenance in the way one might consider for a diagnostic device (e.g., patient data, country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
      Not applicable. This device is not a diagnostic tool requiring expert ground truth for a test set. The evaluation relies on engineering and material science principles via bench testing.

    4. Adjudication method for the test set
      Not applicable. As described in point 3, this device is not a diagnostic tool.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
      Not applicable. This device is a bone anchor system for surgical fixation, not an AI-assisted diagnostic or therapeutic tool for which an MRMC study would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
      Not applicable. This is a mechanical surgical implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
      The "ground truth" for this device's evaluation is primarily established through:

      • Biocompatibility standards: Common usage of materials in medical devices and specific biocompatibility testing.
      • Engineering principles and functional testing: Bench testing to demonstrate that modifications do not negatively impact safety and effectiveness, and that the device performs as intended (e.g., sufficient insertion force).
      • Comparison to predicate device: Establishing that the modified device's characteristics and performance are "substantially equivalent" to a previously cleared device.
    8. The sample size for the training set
      Not applicable. This document describes a medical device, not a machine learning model, so there is no concept of a "training set" for an algorithm.

    9. How the ground truth for the training set was established
      Not applicable. As noted in point 8, there is no training set for this type of device.

    Overall Context:

    This 510(k) submission is for a modified version of an already cleared device. The focus is on demonstrating substantial equivalence to the predicate device (UroTac Bone Anchor System, K962372). This is typically achieved through engineering analyses, material science evaluations (biocompatibility), and bench testing, rather than extensive clinical trials or diagnostic performance studies involving patient data or human readers. The modifications are described as improving convenience and material strength, with the claim that these changes had no effect on device safety and effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962372
    Manufacturer
    Date Cleared
    1996-08-19

    (61 days)

    Product Code
    Regulation Number
    888.4540
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    HXJ

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroTac Bone Anchor System is intended for soft tissue fixation to the pubic bone by means of bone anchors threaded with suture. The UroTac Bone Anchor System is indicated for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility.

    Device Description

    Consisting of: UroTac Bone Anchor Device for soft tissue to bone attachment. UroTac Bone Anchor Inserter Instrument used to insert bone anchor in desired situs.

    AI/ML Overview

    This looks like a 510(k) submission for a medical device called the UroTac Bone Anchor System. While it describes the device's function, materials, intended use, and comparison to predicate devices, it does not contain any information about acceptance criteria, device performance metrics, or study details such as sample sizes, expert qualifications, or ground truth establishment.

    The document states: "The mean failure load of the UroTac Bone Anchor, when pulled from pubic bone, has been tested in a manner consistent with the Guidance Document for Testing Bone Anchor Devices (Draft November 1, 1993). The mean fixation strength of the device was demonstrated to be substantially equivalent to that of the Mitek predicate device."

    This indicates that a non-clinical test was performed to assess fixation strength and compare it to a predicate device, but no specific numerical acceptance criteria or detailed study results are provided in this excerpt.

    Therefore, I cannot fill in the requested table or answer most of the questions based on the provided text.

    Here's an attempt to address what can be inferred:


    Acceptance Criteria and Device Performance (Based on Inferred Information):

    Acceptance CriteriaReported Device Performance
    Substantial equivalence in mean fixation strength to Mitek GII Bone Anchor.Mean fixation strength was demonstrated to be substantially equivalent to that of the Mitek predicate device.

    Detailed Study Information:

    1. Sample size used for the test set and the data provenance: Not specified in the provided text. The testing was "non-clinical," implying laboratory or bench testing rather than human clinical trials.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/Not specified. "Ground truth" for a non-clinical, mechanical strength test would typically be objective physical measurements, not expert consensus.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/Not specified. This is relevant for clinical studies with subjective interpretations, not mechanical testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a bone anchor, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Objective, measured mechanical force (failure load/fixation strength).
    7. The sample size for the training set: Not applicable. This is a physical device, not a machine learning model requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of what is known from the text regarding the study:

    • Study Type: Non-clinical (bench/laboratory) testing.
    • Purpose: To demonstrate the mean fixation strength of the UroTac Bone Anchor when pulled from pubic bone.
    • Methodology: Conducted "in a manner consistent with the Guidance Document for Testing Bone Anchor Devices (Draft November 1, 1993)."
    • Comparison: Compared to the Mitek GII Bone Anchor (predicate device).
    • Conclusion: The device's mean fixation strength was "substantially equivalent" to that of the Mitek predicate device, and the tests "did not yield any data suggesting safety or effectiveness issues which are other than those presented by the Mitek predicate device."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1