Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K984119
    Date Cleared
    1999-01-15

    (58 days)

    Product Code
    Regulation Number
    862.1410
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    CFM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Iron measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis, and chronic renal disease.

    Device Description

    The Wako L-type Fe test is an in vitro diagnostic assay for the quantitative determination of iron in serum. When a sample is mixed with the Buffer, serum protein is denatured by the action of surfactant contained in the Buffer and transferrin-bound iron is liberated. All the Fe" ions released are reduced to Fe2+ by L-ascorbate and form a chelate with bathophenanthroline disulfonic acid disodium salt. The serum iron can be determined by measuring the absorbance of the red chelate solution.

    AI/ML Overview

    The provided text describes the Wako L-type Fe test, an in vitro diagnostic assay for the quantitative determination of iron in serum. The document states that the safety and effectiveness of this new assay are demonstrated by its substantial equivalency to the Wako Fe B test.

    Here's the breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Substantial EquivalencyDemonstrated by substantial equivalency to the Wako Fe B test. (No specific thresholds or metrics for equivalency are given beyond this statement.)
    Precision"Precision studies indicate acceptable values can be obtained on a day to day basis." (No specific coefficients of variation or standard deviations are provided.)
    Minimum Detectable Level"estimated to be 1.2 mg/dL."
    Linearity"determined to be linear to 1000 mg/dL."

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the study relies on substantial equivalency to a predicate device and not expert-established ground truth for a test set.

    4. Adjudication method for the test set:

    • Not applicable as the study relies on substantial equivalency to a predicate device and not independent adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is an in vitro diagnostic device, not an AI-assisted imaging or diagnostic tool relying on human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device described is an in vitro diagnostic assay, meaning it's a lab test. Its performance is inherent to the chemical reaction and measurement process, not an algorithm that operates independently in a standalone fashion like a software device. The "algorithm" in this context is the chemical method itself. The reported performance metrics (precision, linearity, minimum detectable level) are intrinsic to this chemical "algorithm."

    7. The type of ground truth used:

    • For the Wako L-type Fe test, the "ground truth" for demonstrating its effectiveness is primarily based on comparison to a legally marketed predicate device (Wako Fe B test), and its ability to accurately measure iron concentrations using its described chemical method (bathophenanthroline as a chromogen). The document does not mention external pathology or outcomes data as the ground truth for the device's validation.

    8. The sample size for the training set:

    • Not applicable. This is an in vitro diagnostic device and does not involve machine learning models requiring training sets.

    9. How the ground truth for the training set was established:

    • Not applicable. This is an in vitro diagnostic device and does not involve machine learning models requiring training sets.
    Ask a Question

    Ask a specific question about this device

    K Number
    K981581
    Device Name
    IRON
    Date Cleared
    1998-06-08

    (35 days)

    Product Code
    Regulation Number
    862.1410
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    CFM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Iron assay is used for the quantitation of iron in human serum. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease.

    Device Description

    Iron is an in vitro diagnostic assay for the quantitative determination of iron in human serum. The Iron assay is a clinical chemistry assay which utilizes an acidic media to release ferric iron from the transferrin. The ferric iron is converted to the ferrous form by the action of hydroxylamine hydrochloride. The released ferrous iron reacts with FERENE® to produce a colored Iron-FERENE complex. The absorbance of the Iron-FERENE complex is measured at 600 nm and is proportional to the concentration of iron present in the sample. Thiourea and detergent are added to reduce copper interference and turbidity, respectively.

    AI/ML Overview

    The provided text describes a 510(k) submission for an in vitro diagnostic assay called "Iron." The purpose of the submission is to demonstrate substantial equivalence to a predicate device, not to establish new acceptance criteria or conduct a multi-reader multi-case (MRMC) study. The information provided is characteristic of a clinical chemistry assay, which focuses on analytical performance rather than diagnostic accuracy involving human readers.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a table format. Instead, it presents performance characteristics and asserts substantial equivalence to a predicate device. The primary performance metrics are correlation with the predicate, precision, linearity, and limit of quantitation.

    Performance CharacteristicAcceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance
    Correlation CoefficientExpected to be high (close to 1)0.9938
    SlopeExpected to be close to 10.973
    Y-interceptExpected to be close to 0-2.592 ug/dL
    Within-run PrecisionExpected to be low %CVLevel 1: 0.7% (total %CV)
    Level 2: 1.7% (total %CV)
    LinearitySufficiently wide range to cover clinical needsUp to 1778.5 ug/dL
    Limit of Quantitation (Sensitivity)Sufficiently low to detect clinically relevant concentrations3.8 ug/dL

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "comparative performance studies" and "precision studies" but does not specify the sample size for these studies.

    The data provenance is not explicitly stated regarding country of origin. The studies appear to be retrospective as they are comparing the new assay to an existing, established predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This information is not applicable and therefore not provided in the document. For a clinical chemistry assay like this, "ground truth" is typically established by the reference method (the predicate device) or by established analytical standards, not by human expert consensus or interpretation of test results.

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided in the document. Adjudication methods (like 2+1, 3+1) are common in studies involving human interpretation (e.g., radiology for image-based diagnostics) to resolve discrepancies among multiple expert readers. This is not relevant for a quantitative chemical assay.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study focuses on the impact of a device on human reader performance, typically for diagnostic imaging or similar interpretation tasks. The Iron assay is an in vitro diagnostic for quantitative determination, not an interpretive tool for human readers. Therefore, the effect size of how much human readers improve with AI vs. without AI assistance is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, in essence, the entire performance evaluation of the Iron assay is a standalone performance assessment. Clinical chemistry assays are designed to operate independently and provide a quantitative result. The performance characteristics (correlation, precision, linearity, sensitivity) described are the standalone performance of the assay itself, without human interpretation as part of the primary output.

    7. The Type of Ground Truth Used:

    For the comparative performance studies, the ground truth was the results obtained from the Boehringer Mannheim® Iron assay on the Hitachi® 717 Analyzer (K854298), which served as the reference or predicate method. For precision, linearity, and sensitivity, the ground truth would be based on known concentrations in control materials or spiked samples.

    8. The Sample Size for the Training Set:

    This information is not provided and is generally not applicable in the context of this device. Clinical chemistry assays like the Iron assay are typically developed and validated using well-established chemical principles and analytical methodologies. While there's an optimization process during development that might involve testing numerous samples, the concept of a "training set" in the machine learning sense (where an algorithm learns from labeled data) is not relevant to this type of device.

    9. How the Ground Truth for the Training Set Was Established:

    As noted above, the concept of a "training set" with established ground truth, as understood in machine learning/AI, is not applicable to this clinical chemistry assay. The performance is assessed against established analytical methods and standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1