Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994001
    Date Cleared
    2000-02-09

    (77 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    POLYGLYCOLIC ACID is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures.

    Device Description

    POLYGLYCOLIC ACID is a synthetic absorbable sterile surgical suture composed of a synthetic polyglycolic acid polymer and is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures. POLYGLYCOLIC ACID meets all requirements as described in the United States Pharmacopeia (U.S.P.) monograph for Absorbable Surgical Suture. Offered as monofilament or multifilament in form, and it is offered uncoated or coated with polycarpolactone and calcium stearate, it could also be undyed or dyed with an FDA listed color additive, D&C Violet No. 2 (21CFR74.3602). POLYGLYCOLIC ACID will be offered with or without standard needles attached.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the POLYGLYCOLIC ACID surgical suture:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (USP Standards)Reported Device Performance
    Suture DiameterMeet U.S.P. RequirementsMeets U.S.P. Requirements
    Suture LengthMeet U.S.P. RequirementsMeets U.S.P. Requirements
    Knot Pull Tensile StrengthMeet U.S.P. RequirementsMeets U.S.P. Requirements
    Needle Attachment StrengthMeet U.S.P. RequirementsMeets U.S.P. Requirements
    Original Tensile Strength (at 2 wks)Retains approximately 65% (implied USP standard for absorbable sutures)Retains approximately 65%
    Original Tensile Strength (at 3 wks)Retains approximately 40% (implied USP standard for absorbable sutures)Retains approximately 40%
    Suture AbsorptionAbsorption essentially complete between 60 and 90 days (implied USP standard for absorbable sutures)Absorption essentially complete between 60 and 90 days
    SterilityEthylene Oxide Gas sterilization (implied by predicate, conformity to USP)Ethylene Oxide Gas
    Material CompositionComposed of glycolide polymer (implied by product name and predicate comparison)Composed of glycolide polymer
    Biocompatibility(Implicitly covered by USP standards and substantial equivalence to predicate)(Not explicitly stated, but implied by USP and substantial equivalence)
    PackagingDry packaged in Aluminum Foil and Polyester tear open packaging.Dry packaged in Aluminum Foil and Polyester tear open packaging.
    LabelingIn conformance with CDRH instructions and FDA Guidance documentsIn conformance with CDRH instructions and FDA Guidance documents

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "Non-Clinical Testing was conducted on the subject device to prove conformance to the requirements of U.S.P. standards." This implies a sample size adequate for the tests required by USP, but the specific number of units tested is not provided.
    • Data Provenance: Not explicitly stated regarding country of origin. The study is described as "Non-Clinical Testing," which is typically laboratory-based and not tied to specific patient populations or geographical locations. It is a retrospective study in the sense that the tests were performed and then reported, but it's not a retrospective analysis of patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not applicable. The ground truth for the non-clinical testing is based on the United States Pharmacopeia (U.S.P.) standards. These are established, widely accepted objective performance metrics for medical devices, particularly sutures. No human experts are described as establishing "ground truth" for these physical and chemical tests.

    4. Adjudication Method for the Test Set

    • Not applicable. The evaluation relies on objective measurements against pre-defined U.S.P. standards. There is no human adjudication process described for interpreting the results of these standardized tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and assessing the impact of AI assistance on their performance. The device in question is a surgical suture, which is a physical product, and its evaluation relies on non-clinical performance characteristics.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not applicable. This device is a surgical suture, not an algorithm or AI system. The concept of "standalone performance" for an algorithm does not apply here. The evaluation is of the physical product's performance against established standards.

    7. The Type of Ground Truth Used

    • U.S.P. Standards: The ground truth for the device's performance is objective compliance with the United States Pharmacopeia (U.S.P.) monograph for Absorbable Surgical Suture. These standards define the acceptable range for physical properties like diameter, tensile strength, length, and absorption characteristics.

    8. The Sample Size for the Training Set

    • Not applicable. The device is a surgical suture, not an AI or machine learning model. Therefore, there is no "training set" of data in the context of AI development.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI model, the question of establishing its ground truth is not relevant here.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1