Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue); and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff, is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) – Extended Cuff, are disposable, single-use, sterile, blue-colored, and powder-free examination gloves made from nitrile latex.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to medical examination gloves, not an AI-powered medical device. Therefore, the document does not contain information about the acceptance criteria and study proving an AI device meets those criteria.

    The information typically found in such a submission would include:

    • Device performance: This document explicitly states the "Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)" has been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
    • Acceptance Criteria/Performance Data: The acceptance criteria for the gloves are based on relevant ASTM standards for physical properties, barrier integrity, and biocompatibility. The chemotherapy drug resistance is presented as "Minimum Breakthrough Detection Time in Minutes."
    • Study Design (Non-Clinical): The studies are non-clinical, primarily laboratory-based physical and chemical tests compliant with recognized international and national standards (e.g., ASTM, ISO).
    • Ground Truth: For these types of devices, the "ground truth" is typically defined by quantifiable physical and chemical properties measured against established standard methods (e.g., measuring milligrams of residual powder, breakthrough time of chemicals). There are no human experts, adjudication, or training/test sets in the context of AI for this type of device.

    Given this, I cannot extract the information required for an AI-powered device from this document. The sections you asked for (sample size for test/training, expert qualifications, MRMC studies, standalone performance, etc.) are specific to AI/software as a medical device (SaMD) assessments and are not present in a submission for examination gloves.

    However, I can summarize the acceptance criteria and reported performance for the examination gloves as provided in the document:

    Acceptance Criteria and Reported Device Performance (for Examination Gloves)

    This table summarizes the non-clinical acceptance criteria and the "result" or reported performance for the Sterile Nitrile Powder Free Examination Gloves.

    Test / CharacteristicAcceptance Criteria / Standard ReferenceReported Device Performance
    Freedom from holesMeets ASTM D5151-19 (R2023): AQL 1.5%Pass
    Residual PowderMeets ASTM D6124-06 (R2022): Average less than 2 mg/glovePass
    Dimensions (Length)Meets ASTM D6319-19 (R2023): Overall Length: ≥ 230 mm (Subject Device 1); ≥ 260 mm (Subject Device 2)Pass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Width)Meets ASTM D6319-19 (R2023): Specific ranges for XS, S, M, L, XLPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Thickness)Meets ASTM D6319-19 (R2023): Palm Thickness: ≥ 0.05 mm; Finger Thickness: ≥ 0.05 mmPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Tensile Strength Before AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Tensile Strength After AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Ultimate Elongation Before AgingMeets ASTM D6319-19 (R2023): ≥ 500 %Pass (Conforms)
    Ultimate Elongation After AgingMeets ASTM D6319-19 (R2023): ≥ 400 %Pass (Conforms)
    Primary Skin IrritationISO 10993-23: Not an irritantUnder the conditions of the study, not an irritant.
    Dermal SensitizationISO 10993-10: Not a sensitizerUnder the conditions of the study, not a sensitizer.
    Acute Systemic ToxicityISO 10993-11: No evidence of acute systemic toxicityUnder the conditions of this study, no acute systemic toxicity.
    Chemotherapy Drug PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)Listed in table below.
    Fentanyl Citrate PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)> 240 minutes

    Chemotherapy Drug and Fentanyl Citrate Permeation Results

    Device Performance (Minimum Breakthrough Detection Time in Minutes)

    Chemotherapy Drug and ConcentrationSubject Device Result
    Carmustine (3.3 mg/ml)38.3
    Cisplatin (1.0 mg/ml)> 240
    Cyclophosphamide (20.0 mg/ml)> 240
    Dacarbazine (10.0 mg/ml)> 240
    Doxorubicin HCl (2.0 mg/ml)> 240
    Etoposide (20.0 mg/ml)> 240
    Fluorouracil (50.0 mg/ml)> 240
    Methotrexate (25.0 mg/ml)> 240
    Mitomycin C (0.5 mg/ml)> 240
    Paclitaxel (6.0 mg/ml)> 240
    Thiotepa (10.0 mg/ml)78.6
    Vincristine Sulfate (1.0 mg/ml)> 240
    Vidaza (5-Azacytidine), 25 mg/ml> 240
    Busulfan, 6 mg/ml> 240
    Carboplatin, 10 mg/ml> 240
    Docetaxel, 10 mg/ml> 240
    Epirubicin HCl, 2 mg/ml> 240
    Gemcitabine HCl, 38 mg/ml> 240
    Ifosfamide, 50 mg/ml> 240
    Irinotecan, 20 mg/ml> 240
    Mitoxantrone HCl, 2 mg/ml> 240
    Oxaliplatin, 5 mg/ml> 240
    Vinorelbine, 10 mg/ml> 240
    Fentanyl Citrate Injection (100mcg/2ml)> 240

    Since the provided document is not for an AI device, the following points cannot be addressed:

    • Sample sized used for the test set and the data provenance: Not applicable to a physical glove.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is based on physical/chemical measurements by laboratory technicians following standard protocols.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For gloves, it's objective physical/chemical measurements according to ASTM/ISO standards.
    • The sample size for the training set: Not applicable (no AI training involved).
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1