Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230121
    Date Cleared
    2023-05-15

    (118 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Non-Sterile Nitrile Patient Examination Glove is made from synthetic rubber latex. It is single use and powderfree variation of the class I Nitrile Patient Examination Gloves which is coated by on-line polymer with mild on-line chlorination process. These processes modify the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder. The available sizes are X-Small, Medium, Large and X-Large. The Nitrile Examination Gloves Powder Free (Blue & Black) have been tested for Chemotherapy Drugs and Fentanyl Citrate.

    AI/ML Overview

    This document describes the testing for Nitrile Examination Gloves, focusing on their resistance to chemotherapy drugs and Fentanyl Citrate. The study assesses the gloves' physical properties, biocompatibility, and chemical permeation alongside the reported device performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Chemotherapy and Fentanyl Citrate Permeation (Minimum Breakthrough Detection Time)

    Chemotherapy DrugsConcentrationAcceptance Criteria (from ASTM D6978-05 implied by claim of use)Reported Device Performance (Blue Glove)Reported Device Performance (Black Glove)
    Carmustine (BCNU)3.3 mg/mlNo specific public acceptance criterion provided but tested22.6 minutes21.8 minutes
    Cisplatin1.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Cyclophosphamide (Cytoxan)20.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Dacarbazine10.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Doxorubicin HCl2.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Etoposide20.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Fluorouracil50.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Ifosfamide50.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Mitoxantrone2 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Paclitaxel6.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Thiotepa10.0 mg/mlNo specific public acceptance criterion provided but tested43.9 minutes17.7 minutes
    Vincristine Sulfate1.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Fentanyl Citrate Injection100mcg/2mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    WarningAdvised against use if permeation time is too lowCarmustine, Thiotepa warningCarmustine, Thiotepa warning

    Physical Properties, Dimensional, and Biocompatibility Testing

    TestMethodAcceptance CriteriaReported Device Performance
    Freedom From HolesASTM D6319-19, D5151-19Requirement inspection level G-1, AQL 2.5Pass
    Dimension (Length, Thickness, Width)ASTM D6319-19Length: min. 220mm, 230mm. Thickness: min. 0.05mm. Width: 70-120mm (based on size)Pass (Meets requirements for various sizes)
    Physical Properties (Before Aging)ASTM D6319-19Tensile Strength: 14 MPa min. Ultimate Elongation: 500% min.Pass (Meets 14MPa min, 500% min)
    Physical Properties (After Aging)ASTM D6319-19Tensile Strength: 14 MPa min. Ultimate Elongation: 400% min.Pass (Meets 14MPa min, 400% min)
    Residual Powder ContentASTM D6319-19, D6124-06Not more than 2 mg per glovePass
    Primary Skin Irritation TestISO 10993-10Not a primary skin irritantPass (Did not cause an irritant response)
    Skin Sensitization TestISO 10993-10Not a contact sensitizerPass (Did not produce a skin sensitization effect)
    In Vitro Cytotoxicity TestISO 10993-5:2009No adverse biological reactionModerate cytotoxicity at 6.0 cm²/mL; No reactivity at 3.0 cm²/mL
    Acute Systemic ToxicityISO 10993-11No adverse biological reactionPass (No adverse biological reaction)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of gloves) used for the permeation, physical property, or biocompatibility tests.
    The data provenance is not explicitly mentioned (e.g., country of origin, retrospective or prospective). However, the tests were conducted according to international standards (e.g., ASTM, ISO), which implies a standardized, prospective testing approach in a laboratory setting. The applicant is based in Malaysia.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study involving expert readers or ground truth established by medical professionals. The "ground truth" for this device is determined by objective, standardized laboratory measurements (e.g., chemical permeation time, physical properties, biological reactions) as defined by the referenced ASTM and ISO standards.


    4. Adjudication Method for the Test Set

    Not applicable. There was no human adjudication process involved as the "ground truth" is determined by objective laboratory measurements and adherence to specified test standards.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data. The device in question is examination gloves, and its performance is evaluated through physical, chemical, and biological testing, not by human readers interpreting cases.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is a medical device (examination gloves), not an AI algorithm. Therefore, the concept of "standalone" performance for an algorithm does not apply. All listed tests are "standalone" in the sense that they assess the intrinsic properties of the glove without human intervention in the measurement process itself, though human technicians conduct and analyze the tests.


    7. The Type of Ground Truth Used

    The ground truth used for this study is based on:

    • Performance Standards: Defined by international standards such as ASTM D6978-05 (for chemotherapy drug permeation), ASTM D6319-19, ASTM D5151-19, ASTM D6124-06 (for physical properties and hole freedom), and ISO 10993-10, ISO 10993-5:2009, ISO 10993-11 (for biocompatibility).
    • Objective Laboratory Measurements: Direct measurements of breakthrough time, dimensional properties, tensile strength, elongation, residual powder, and biological response to the glove materials.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set mentioned or implied in this submission, as the device is not an AI/ML algorithm that requires training data.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1