(560 days)
Medentika abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Medentika abutments for the Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm and TX 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika TiBase is intended for use with the Straumann® CARES® System. All digitally designed copings and/or crowns are intended to be sent to Straumann for manufacture at a validated milling center.
Medentika abutments for the Nobel Biocare Nobel Active®* 3.0mm, Dentsply Sirona Astra Tech OsseoSpeed EV®* 3.0mm and TX®* 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Medentika PreFace CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Medentika Preface is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. The final patient matched form is a MedentiCAD abutment.
Medentika abutments for the Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Multi-unit abutments are indicated for use with dental implants as a support for multi-unit screw retained bridges and bars in the maxilla or mandible of a partially or fully edentulous patient.
The Medentika abutments include abutments, abutment screws, caps, and bases which are labelled under a specific Medentika series and are compatible with a specified dental implant system. The abutments include sinqle-unit abutments intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. The abutments also include multi-unit abutments indicated for use with dental implants as a support for multi-unit screw retained bridges and bars in the maxilla or mandible of a partially or fully edentulous patient.
The purpose of this premarket notification is to add additional abutments. The subject abutments include abutments compatible with additional dental implant systems forming a new Medentika series (the OT series). The subject abutments also include abutments compatible with new implant diameters in existing Medentika series (E, EV, F, and S). Lastly, the subject abutments include new abutment designs compatible with existing implant diameters in existing Medentika series (R).
This looks like a 510(k) Summary for a medical device (dental abutments), which means the document is about proving "substantial equivalence" to a predicate device, not about proving clinical effectiveness or performance against pre-defined acceptance criteria in the way one might for a novel AI/software medical device.
Therefore, the information requested in your bullet points (e.g., acceptance criteria table, sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) is not applicable to this type of regulatory submission because the device is a mechanical one, not an AI/software device. The data presented here is focused on demonstrating physical and mechanical compatibility and equivalence to previously cleared devices.
Here's why each point is not applicable and what information is provided:
-
A table of acceptance criteria and the reported device performance: This document doesn't provide a typical "acceptance criteria" table as would be seen for an AI/software device measuring diagnostic performance (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating that the new abutments perform similarly to existing, cleared abutments through "dynamic fatigue testing" and "dimensional analysis and reverse engineering." The performance is implicitly "accepted" if these tests show equivalence to the predicate.
-
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): No "test set" in the context of diagnostic performance is mentioned. The "testing" refers to non-clinical, physical testing (fatigue, dimensional analysis). There is no patient data involved in this type of submission for a mechanical device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth, in the AI/software sense, is not established for this device. The "truth" is based on engineered specifications and physical testing.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No human adjudication of diagnostic output is relevant here.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" is engineering specifications and physical measurements.
-
The sample size for the training set: Not applicable. There is no AI/machine learning component to "train."
-
How the ground truth for the training set was established: Not applicable for the same reason as above.
What the document does provide regarding device performance and testing:
-
Type of Testing:
- Dynamic fatigue testing according to FDA guidance and ISO 14801 (Dentistry - implants dynamic loading test for endosseous dental implants).
- Dimensional analysis and reverse engineering of the implant-to-abutment connection platform.
- Sterilization validation (steam and gamma irradiation) referenced from K191123, ISO 17665-1, ISO/TS 17665-2, ISO 11137-1, ISO 11137-2.
- Sterile packaging validation referenced from K191123, ISO 11607-1, ISO 11607-2.
- Biocompatibility evaluations referenced from K142167, K170838, K191123, K150203, K061804 in accordance with ISO 10993-1.
- MR testing referenced from K180564 in accordance with ASTM F2052-15, ASTM F2213-06 (2011), ASTM F2182-11a, and ASTM F2119-13.
-
Conclusion: The tests demonstrated "implant to abutment compatibility" and "established substantial equivalency of the proposed device with predicate devices." This is the "proof" that the device meets the (implicit) acceptance of being substantially equivalent to existing, legally marketed devices.
In summary, this document is for a traditional mechanical medical device, and the regulatory pathway does not involve performance studies in the way you've outlined for AI/software-based devices.