Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213859
    Date Cleared
    2022-01-13

    (34 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stronghold Group Nitrile Examination Gloves, Powder Free is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    The subject device in this 510(k) Notification is Stronghold Group Nitrile Examination Glove, Powder Free. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250. Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    The Stronghold Group Nitrile Examination Gloves, Powder Free, device performance was evaluated through non-clinical testing to demonstrate substantial equivalence to a legally marketed predicate device (K203191).

    1. Acceptance Criteria and Reported Device Performance

    The device performance was evaluated against the standards outlined in ASTM D6319-19 and ISO 10993 for various physical properties and biocompatibility. The table below summarizes the acceptance criteria and the reported results:

    Test MethodAcceptance CriteriaReported Device PerformanceResult
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions Test (Width)Width: 91 mm (Mean) (for medium size)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions Test (Length)Length: 241 mm (Mean) (for medium size)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions Test (Thickness: Finger)Thickness: Finger – 0.13 mm (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions Test (Thickness: Palm)Thickness: Palm – 0.09 mm (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements Test (Tensile Strength Before Aging)Before Acceleration Aging: Tensile Strength (MPa): 34 (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements Test (Ultimate Elongation Before Aging)Before Acceleration Aging: Ultimate Elongation (%): 601 (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements Test (Tensile Strength After Aging)After Acceleration Aging: Tensile Strength (MPa): 34 (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements Test (Ultimate Elongation After Aging)After Acceleration Aging: Ultimate Elongation (%): 571 (Mean)Passed (The specific measured value is not provided, but it met the criteria to be termed "Passed")Passed
    ASTM D6319-19 (ASTM D5151-11) Standard Test Method for Detection of Holes in Medical GlovesAQL 2.5PassedPassed
    ASMT D6319-19 (ASTM D6124-11) Standard Test Method for Residual Powder on Medical Gloves≤ 2.0 mg/glove0.12 mg/glovePassed
    ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin irritation and skin sensitization (Irritation)Under the condition of study not an irritantUnder the condition of study not an irritantPassed
    ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin irritation and skin sensitization (Sensitization)Under the conditions of the study not a sensitizer.Under the conditions of the study not a sensitizer.Passed
    ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityUnder the conditions of study non cytotoxicNon-cytotoxic reactivity at 100%, 50%, 25%, 12.5% and 6.25% extract concentration. (This indicates it was non-cytotoxic across all tested concentrations, meeting or exceeding the "non-cytotoxic" acceptance criteria).Passed
    ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for acute systemic toxicityUnder the conditions of the study no systemic toxicityUnder the conditions of the study no systemic toxicityPassed
    ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for skin irritationUnder the conditions of the study, no skin irritationUnder the conditions of the study, no skin irritationPassed

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each non-clinical test conducted for the Stronghold Group Nitrile Examination Gloves. However, tests were conducted according to established ASTM and ISO standards, which typically specify minimum sample sizes for valid results.

    The data provenance is from non-clinical testing (laboratory-based testing) of the Stronghold Group Nitrile Examination Gloves. The country of origin for the data is not specified directly in the provided text, but the testing was for a device manufactured by Stronghold Group. The testing is prospective in the sense that it evaluates the performance of the manufactured device directly.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation involved non-clinical laboratory testing against established physical and chemical standards (ASTM, ISO), rather than a clinical study requiring expert interpretation of results or ground truth establishment by medical experts. The "ground truth" for these tests is defined by the objective measurement criteria of the standards themselves.


    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation involved non-clinical laboratory testing with objective pass/fail criteria from recognized standards, not a process that would require expert adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The device is a "Nitrile Examination Glove, Powder Free" and does not involve AI or human reader interpretation. The testing focuses on the physical and biocompatibility properties of the glove.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable, as the device is a physical medical glove and does not involve any algorithms or AI.


    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is based on established standard specifications and methodologies from ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). For example, for dimensions, the ground truth is the specified range in ASTM D6319-19; for cytotoxicity, it's the qualitative assessment of "non-cytotoxic" based on ISO 10993-5.


    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical glove, and its performance evaluation did not involve machine learning models that require a training set.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as above (no training set for a physical device).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1